Professional Documents
Culture Documents
Shining OptimScan-5M User Manual8.4.11
Shining OptimScan-5M User Manual8.4.11
Shining OptimScan-5M
User Manual
-I-
OptimScan-5M User Manual
Dear customers,
⚫ Thank you very much for buying Shining3D scanning system.
⚫ Please read this manual carefully to ensure correct operations. Read the safety
regulation carefully in case any damage to you or the equipment caused by
improper operations.
⚫ Please refer to this user manual for troubleshoot possible problems. And please
contact us if you have any other question.
⚫ Please do not connect the computer which using our equipment to the Internet, in
case any damage caused by computer virus.
⚫ Please contact us for a new version manual if the product is updated, and you can
also download the electronic version manual on our official website.
II
--
OptimScan-5M User Manual
CONTENTS
CONTENTS............................................................................................ III
Chapter 1. Introduction............................................................................ 5
1.1 About Shining3D scanner .............................................................................. 5
1.2 System accessories ............................................................................................ 5
1.3 About the instruction ......................................................................................... 6
Chapter 2 Safety ........................................................................................ 7
2.1 Safety identification and introduction ............................................................... 7
2.2 Safety identification .......................................................................................... 7
2.3 General safety information ............................................................................. 11
Chapter 3 Basic concept .........................................................................13
3.1 3D scanner ...................................................................................................... 13
3.2 Image collection .............................................................................................. 13
3.3 Camera lens parameter adjustment ................................................................. 14
3.4 Camera calibration .......................................................................................... 15
3.5 Software usage mode ...................................................................................... 16
3.6 File format for 3D data ............................................................................... 17
Chapter 4 Software installation .............................................................18
4.1 Software requirement ...................................................................................... 18
4.2 Scanning software installation ........................................................................ 24
4.3Firmware Upgrade ........................................................................................... 30
Chapter 5 Software Interface ................................................................32
5.1 Introduction ..................................................................................................... 32
5.2Calibration mode .............................................................................................. 33
5.2.1 The command in the toolbar......................................................错误!未定义书签。
5.3 Scanning mode ................................................................................................ 36
5.3.1 Toolbar command introduction .............................................................................. 36
5.4 Work panel ...................................................................................................... 47
5.5 Window ........................................................................................................... 54
Chapter 6 System operation under scan mode ....................................56
6.1 System startup ................................................................................................. 56
III
--
OptimScan-5M User Manual
IV
--
天远三维扫描系统 用户使用指南
Chapter 1. Introduction
5
天远三维扫描系统 用户使用指南
The instruction describes how to use Shining3D scanner in details, thus can be
used as reference material for training and studying.
6
天远三维扫描系统 用户使用指南
Chapter 2 Safety
This chapter explains certain safety issues related to the equipment. Please read
the safety indications carefully before operation.
7
天远三维扫描系统 用户使用指南
Figure 2-1
8
天远三维扫描系统 用户使用指南
Figure 2-2
○1 Threaded rod used to control cloud deck, can be used to rotate the cloud deck
up and down
○2 Threaded rod used to control cloud deck, can be used to rotate the cloud deck
horizontally.
○3 Threaded rod used to control cloud deck, can be used to rotate the cloud deck
left and right
Three threaded rods and two screws must be tightened before operation,
otherwise the equipment can drop and break.
2.2.3 Tripod
9
天远三维扫描系统 用户使用指南
Figure 2-3
○1 Rocker of tripod, can be used to control the device to rise and fall
Tighten the jackscrew of tripod each time after adjusting the device
height, otherwise the device can shack during measurement and affects axis
rigidity
10
天远三维扫描系统 用户使用指南
Warning
⚫ Power supply for the device should be 220V 50HZ.
⚫ To ensure safety, operation can only be done after professional training.
11
天远三维扫描系统 用户使用指南
Caution
⚫ The device should be operated in clean workshop. Do not operate in dusty
environment.
⚫ Do not unplug data transmission cable during software operation.
⚫ Do not touch lens of laser transmitter or lens of cameras with hands.
⚫ Do not strike camera or laser transmitter with edge tools.
⚫ Do not strike the surface of calibration system with edge tools.
Notice
⚫ Lock the calibration system box after calibration is done.
12
天远三维扫描系统 用户使用指南
3.1 3D scanner
The device captures 3D surface data of the object with certain methods. The
scanning methods can be divided into two categories, contact and non-contact
scanning. Our device is optical non-contact 3D scanner.
13
天远三维扫描系统 用户使用指南
:Click the button, grab the current camera to display the image and save as;
Camera lens has two parameters which need adjust, clarity and aperture size,
14
天远三维扫描系统 用户使用指南
Refer to figure 3-4, loosen “1”, rotate outer ring “A” to adjust camera clarity, loosen
“2”, rotate inner ring “B” to adjust camera aperture size.
tighten rotary knobs “1” and “2” after adjustment, otherwise measurement
accuracy will be affected. Adjustment should be done under technician guidance
of our company.
15
天远三维扫描系统 用户使用指南
The software usage mode includes scan mode and light pen mode.
Scan mode is used for 3D Scan system. When building a new project, three
modes can be selected.
⚫ Scan and registration mode
16
天远三维扫描系统 用户使用指南
For big objects that complete data cannot be collected once, reference points can
be pasted on the object to complete the object data with scan and registration mode.
Attentions should be paid to the following issues during reference point alignment:
(1)Reference point should be pasted on a plane surface or curved surface with
small curvature change.
(2)Do not line the reference points.
(3)Between successive two scan, communal reference points should be at least
4, Due to image quality, shooting angles etc., some reference points can't be
identified correctly. Thus it is suggested to use more reference points, normally
6-8.
⚫ Targets mode
For some big objects, measuring error will be enlarged by accumulated
errors. In order to control the integrated error, frame should be established by
using targets mode before measurement. And then select the scan and registration
mode for scanning.
Common 3D data formats are ASC, STL, DXF,IGES, OBJ etc. In this system,
we output ASC,STL and DGM files. All of these can be used by other software to get
what user requires.
17
天远三维扫描系统 用户使用指南
The software should run in the operating system Microsoft Windows 7/10
64bit(Ultimate). The computer hardware configuration requirements are:
CPU: Intel i7 and higher;
Memory: 32G and higher;
GPU: Graphics memory 2G and higher, the GPU accelerate function only support
NVIDIA Graphics card .
Figure 4-1
18
天远三维扫描系统 用户使用指南
Click the ‘Next’ button enter the License Agreement Interface shown as figure
4-2, select the ‘I accept the terms in the license agreement’ option and then the ‘Next’
button becomes available.
Figure 4-2
Click the ‘Next’ button again and enter the Customer Information Interface, use
the default ‘User Name’ and ‘Organization’, select ‘Anyone who uses this
computer(all users)’option, shown as figure 4-3.
Figure 4-3
19
天远三维扫描系统 用户使用指南
Click the ‘Next’ button again and enter the Options to be installed Interface, set
the ‘Install JAI GigE Vision Filter Driver’ option checked, set the ‘Enable built-in
debug features’ option checked, shown as figure 4-4.
Figure4-4
Click the ‘Next’ button again and set the ‘Install JAI USB 3.0 Driver’ option
checked, shown as figure 4-5.
20
天远三维扫描系统 用户使用指南
Figure4-5
Click the ‘Next’ button once again and enter the Destination Folder Interface,
clicked the ‘Change’ button can change the folder path which the program will be
installed, shown as figure 4-6.
Figure 4-6
Click the ‘Next’ button enter the Setup Type Interface, select the ‘Complete’
option to install the program completely, shown as figure 4-7.
21
天远三维扫描系统 用户使用指南
Figure 4-7
Click the ‘Next’ button and enter the Ready to Install the Program Interface,
shown as figure 4-8, click the ‘Install’ button if there is no doubt about these settings
above.
Figure 4-8
Then the Install Shield Wizard will cost a few minutes to install the program,
shown as figure 4-9.
Figure 4-9
22
天远三维扫描系统 用户使用指南
When the install is finished, the Install Shield Wizard is shown as figure 4-10.
Figure 4-10
Click the ‘Finish’ button, the program may remind you restart your computer,
shown as figure 4-11, you can select ‘Yes’ to restart right now or select ‘No’ to restart
later, please note to save your file which is using on your computer carefully.
Figure4-11
When the computer is restarted, open the Device Manager and find the device
shown with the red rectangle on the figure 4-12, if the device is not found, try to
refresh or re-install the JAI camera driver. You can use the ‘Add / Remove Programs’
function on the ‘Control Panel’ or the JAI camera driver folder’s ‘Remove’ option to
uninstall the JAI camera driver.
23
天远三维扫描系统 用户使用指南
Figure 4-12
24
天远三维扫描系统 用户使用指南
Figure 4-14
Input User information, click “Next”,and show as figure 4-15
25
天远三维扫描系统 用户使用指南
Figure 4-15
Confirm installation path,click”Next”,and show as figure 4-16
Figure 4-16
Shortcut Folder,click”Next”,and show as figure 4-17
26
天远三维扫描系统 用户使用指南
Figure 4-17
Check the information ,and click “Next”,and show as figure 4-18
Figure 4-18
Software installation process is shown as figure 4-19
27
天远三维扫描系统 用户使用指南
Figure 4-19
Figure 4-20 shows installation completed.
Figure 4-21
When you run the program on Windows 10, you need to do the
following:
First, verify that Windows OS is at 15063 and above (figure as 4-18),right-click on the
shortcut,choose properties,click compatibility,check the Override high DPI scaling
behavior,and choose the system from Scaling performed by,check the Run this
program as an administrator,click apply(figure as 4-19).
If you don't, the program will display an exception
Figure 4-18
29
天远三维扫描系统 用户使用指南
Figure 4-19
4.3Firmware Upgrade
When the program version is updated, the firmware needs to be detected and
upgraded.
30
天远三维扫描系统 用户使用指南
Figure as 4-20
click ,If prompted, "this version is the latest version and does
not need to be upgraded",click the close.
Figure as 4-21
31
天远三维扫描系统 用户使用指南
5.1 Introduction
32
天远三维扫描系统 用户使用指南
5.2Calibration mode
Select the calibration mode in the mode selection menu ,enter calibration mode,the
inside buttons can help you complete the calibration.
Command instrument:
If you hover your mouse over a command for a few seconds, it will show a simple
explanation of the command.
:The rough adjustment button can help the user to complete the initial
33
天远三维扫描系统 用户使用指南
adjustment of the equipment, determine the scope of work, and select the correct
equipment model as required.
:Click the camera focal length button, which can guide the user to
adjust the camera focus to the appropriate state.
:Click the aperture adjustment button, which guides the user to adjust
the camera aperture to the appropriate state.
:Click the raster adjustment button, which can instruct the user to
adjust the light machine clearly.
:After adjusting the lens and grating, click the calibration button to enter
the camera calibration interface.(The calibration instructions for the equipment are
detailed in 6.2)
:Click this button to reconnect the device after an abnormal power failure
and the cameras disconnected.
Scanner Pos Show: Select this function,the model and position of the scanning head
will be displayed
34
天远三维扫描系统 用户使用指南
Figure 5-5
Measure Size:Selecting this feature will show the actual scan range of the scan head.
Figure 5-7
35
OptimScan-5M User Manual
New construction:Use, the first step must be new, click on "new project", in the
pop-up dialog, select the need to model, to set up the project directory and type in the
name of engineering, engineering support directory and name in Chinese.
Click "new project" ,pop up the following dialog box, figure5-5
Select the project mode needed. Under project configuration, click (browse
folder), select “project directory” and click. Dialog shown as 5-5 will pop up. Enter
project name after “ Project name”. All the data for this scanning project will be saved
in the file named after “Project name” under “Project directory”.
36
OptimScan-5M User Manual
37
OptimScan-5M User Manual
Open project: open a scanning project already exists. Note: select project
scanning mode first during operation, and then select the corresponding project file.
Click open project, below interface pops up, figure 5-10.
38
OptimScan-5M User Manual
The “.spc” file marked in above figure is the file for scanning point cloud project.
“1111111” is the name for the new established project. To open a frame scanning
Grating: Project grating on the object, and adjust the focal distance of
grating emitter to reveal the grating on measured object to its clearest. Observe the
grating clarity to judge whether the distance between measured object and scanner is
suitable or not. Too close or too far can not obtain clear grating.
39
OptimScan-5M User Manual
Manual Register: Align several scanned data by manual select features, this
function could be used only when the relative option in the system setting is selected.
Clicked "Manual Register" button and the interface shown as figure 5-11:
40
OptimScan-5M User Manual
Figure 5-12
The display window is divided into 3 regions, after select the fixed data and
floating data on the left "Manual Registration" panel, the fixed data will shown on the
up-left region and the floating data will shown on the up-right region. Users should
pick some feature points on the fixed data and then pick the same points on the
floating data, click "Register" button and register result will shown on the bottom
region, click "Clear" button to re-pick feature points and, click "Ok" or "Cancel"
button to close the "Manual Registration" panel.
Register: Align several scanned data into one coordinate system and remove
common parts.
Display control:
Frame cloud point: Display frame point cloud (reference point information).
Click 'Turntable Measure' button, the Turntable Control dialog will pop up, shown as
figure 5-17.
:Click the delete points button to remove the point clouds below the
fitting plane.the selected point cloud is shown as red,the fitting plane data will be
automatically saved.
43
OptimScan-5M User Manual
also directly click the desired location on the arc, and the program will automatically
determine the current Angle value.
Manual Scan Mode:You can change the Angle of the turntable by entering the values
of X and Y,X represents the rotation Angle of the rotary table, range: 0~359 degree;Y
represents the rotation Angle of the rocker arm of the turntable, range: 0~90 degrees
Auto Scan model:Load the turntable path that has been set,the scanner will scan
automatically after rotating to the specified position.
44
OptimScan-5M User Manual
fit windows
Select the point cloud for the current page using the rectangular lasso tool.
Select the point cloud of the current page using the polygon.
Figure 5-15View
45
OptimScan-5M User Manual
(2)Language
Figure 5-16Language
Figure 5-16shows the language setting dialog.
Language:Set the language of program, this program support 2 languages: Chinese
and English, please restart program to make the language setting effect.
(3)Calib
46
OptimScan-5M User Manual
Figure 5-17Calib
47
OptimScan-5M User Manual
Remove boundry:If this option is checked, the boundary of scan data will be deleted
a certain extent.
Minimum:Set the minimum number of the vertex of effective point cloud, the point
cloud which the number of vertex is less than the minimum number is determined to
be invalid and deleted automatically.
48
OptimScan-5M User Manual
Gray Capture:Select this option to collect the gray information of the object surface
during the scanning process. Using the ‘Gray View’ button on the menu bar can
display or hide the gray color of the scan data.
Auto Delete Target:Check this option to remove the point cloud surrounding the
mark point during the scanning process, and input the diameter range of the removed
mark point.
The environmental monitoring:This function can monitor the vibration frequency
around the device, and if the vibration has an effect on the scan results, it will prompt
the user to give up the current data.
(5)Register
Work Mode:There are ‘Single Regist Mode’ and ‘Global Regist Mode’.
Figure 5-24Register
If ‘Single Regist mode’ is selected, the dialog interface is shown as Figure 5-24. Scan
system output point cloud data and display point cloud model. After each scan process,
the ‘Register’ button need to be clicked manually to merge two point clouds into
one point cloud. If the "Manual Registration" option is selected, after each scan
process, the "Manual Register" button on the ribbon bar could be use, please read
relative part to learn how to use this function.
49
OptimScan-5M User Manual
Figure 5-25
If ‘Global Regist Mode’ is selected, the dialog interface is shown as Figure 5-25. Scan
system output point cloud data and mesh data, display point cloud model. After each
scan process, the new point cloud data will automatically match the previous point
cloud data. When all the scans are completed, click ‘Register’ button to merge
total scan data, output final point cloud data and mesh data, display final mesh model.
The ‘Feature Registration’ option below could make different scan data of measure
object merge together just based on its own features without reference points. But the
precision of feature registration is not as good as targets registration, so generally not
recommended. If the "Manual Registration" option is selected, after each scan process,
the "Manual Register" button on the ribbon bar could be use, please read relative part
to learn how to use this function.
Registration:
If the ‘Single Regist Mode’ is selected, the registration group-box displays two items,
"Only by targets" and "And by Point Cloud". The "Only by targets" means the scan
system will merge single scan data together based on frame points and then remove
redundancy, the "And by Point Cloud" means after the merger of single scan data, the
merge result will be optimized based on total point cloud, and then remove
redundancy.
50
OptimScan-5M User Manual
If the ‘Global regist Mode’ is selected, the registration group-box displays two items,
Only by targets register and Global point cloud register. The ‘Only by targets register’
means the scan system will merge single scan data together based on frame points, the
‘Global point cloud register’ means after the merger of single scan data, the merge
result will be optimized based on total point cloud.
Output Reduction Setting : When working mode selects "Global Regist
Mode","Output Settings" appear below the dialog,you can select the output file type.
Register reduce sampling:Before registering, the registering value of each piece of
cloud will be sampled at most,a reduction of up to one in 16.
Min duster points reserved:This parameter means the smallest unit of all the grid
collections that can be preserved when the output data is printed,the grid data that is
less than the required number is automatically deleted.
Only keep the Biggest Component:This parameter is only the largest single piece of
data, and the scattered small data is deleted.
Edge sharp:The edge sharpening can be sharpened to the right Angle, not checked
this function, the Angle of the scan may be shown as a circular arc.
Smooth Setting:If the work mode is ‘Global Regist Mode’, ‘Smooth Setting’
group-box will appear below, a scroll-bar named smooth degree in this group-box will
control the smoothness of the generated mesh model. The larger the parameter is, the
higher the smoothness is.
Output point num :This parameter is the number of model point cloud generated
after global regist,check this function, then the number of point clouds output allows
customization.
(6)Hardware
51
OptimScan-5M User Manual
In the board display on the tool bar, click check box in front of the “work panel”,
“work panel” control window will pop up on the left.
Work panel shown as figure 5-22
52
OptimScan-5M User Manual
“All”, “Part”, “Reverse”, “Delete” are all operation for point cloud
Reference point error will show up in this bar during reference point scanning and
frame scanning. Figure 5-23.
53
OptimScan-5M User Manual
5.5 Window
54
OptimScan-5M User Manual
1. Make sure hardware is right connected first and then switch on the power.
2. Power on computer and launch raster radiator
3. Open the program, the main software interface and two real-time camera
image will be displayed
4. Adjusting cameras parameter accordingly until getting satisfied photo quality
Figure6-1
Adjust the scanning head position according to the graphic and explanatory text,figure
56
OptimScan-5M User Manual
as 6-2.
Figure6-2
Adjust the camera parameters according to the instructions, and move the camera to
the appropriate hole position. After the adjustment is completed, click finish,figure as
6-3
57
OptimScan-5M User Manual
Figure6-3
Figure6-4
Adjust the focal length of the left camera: click the left camera focal distance button,
adjust the camera parameters according to the guide text.
58
OptimScan-5M User Manual
After the adjustment of the left and right cameras is completed, click the aperture
adjustment,adjust the device position according to the picture and adjust the camera
aperture size according to the text description,when four green checkmarks appear at
the same time, the position qualifies
59
OptimScan-5M User Manual
Set the Calib board according to the picture, adjust the lens aperture value, so that the
aperture indicator bar changes from red to green, tighten the screw, and click finish.
Click ,adjust the lens according to the text description of the picture.
60
OptimScan-5M User Manual
After the adjustment of the camera and project is completed, click the calibration
1. Place calibration block. Pick the block and place it on the black fabric. Notice:
the fabric needs to lie on a flat surface, and one of big dots of the calibration board
should face upward .All the four big dots must be seen and try to see more small
points
2.Check the information on the Calib Info Confirm, and click OK,if the Use Calib
Plane function is selected, the plane calibration menu bar is displayed
62
OptimScan-5M User Manual
3.Make sure that two cameras are in shoot models, the raster radiator projects blue
light
Adjust the position of the scanning head according to the screen prompt until the red
symbol turns to green checkmark. Then click position 1 to complete the collection of
the image at the first calibration position.
63
OptimScan-5M User Manual
After block calibration, continue to white panel calibration with the reverse side
of calibration board.
64
OptimScan-5M User Manual
Make sure the reverse side of calibration board is clear and scratch free during
white panel calibration, otherwise accuracy of the scanned data will be affected. Take
as much white panel as possible into the camera during calibration ,otherwise no data
will show in place not calibrated.
Detailed calibration methods: The down most button is for panel calibration. Five
times calibration is required. The distance between measuring head to while panel
should be from near to far. Only data in these five panel positions is accurate.
For the first panel calibration position, align the cross to the center, move the
device downward for 3 circles, shoot the first calibration panel.
For the second panel calibration position, move the device upward for 1.5 circles,
shoot the second calibration panel.
For the third panel calibration position, move the device upward for another1.5 circles
of Position 2, shoot the third calibration panel(the cross is aligned to the center)
For the forth panel calibration position, move the device upward for another1.5 circles
of Position 3, shoot the forth calibration panel.
For the fifth panel calibration position, move the device upward for 1.5 circles of
Position 4, shoot the second calibration panel.(Upward for 3 circles with cross aligned
to the center)
To sum up, the first and the fifth position, the second and the forth position are
all aligned to the center based on the third position( best position with cross aligned to
the center). After taking pictures of all five positions, scanning can be performed.
Only data in these positions is correct during scanning. Incorrect will be deleted
automatically. That is to say, no data will be read in places beyond ±3 circles of
calibration area . For small calibration panel with ±2 circles calibrated(one circle
means 1 interval), no data will be read in places beyond ±2 circles of calibration area.
Click “Return” on work pane, the following dialog will pop up.
After checking the plane calibration option, the plane calibration process
must be completed, otherwise the device cannot be scanned.
If calibration in some positions is not satisfactory enough or calibration block is
not rotated, click “cancel” to go back and recalibrate it again.
6.4Turntable Calibration
Do not move the relative position between the turntable and the
measuring head after the turntable is calibrated.
General process:
1 , normal boot system, from the menu bar "turntable calibration" (Figure 6-17 )
option to enter the turntable calibration interface (Figure 6-18 )
Figure 6-17
66
OptimScan-5M User Manual
Figure 6-18
3, Click , where the turntable rotates for a week, shoot the mark point
trajectory
4, in the scene, select a complete point trajectory with the mouse (do not omit the
location) and make a circle .
67
OptimScan-5M User Manual
Figure 6-19
Reference point alignment scan uses information on common points between two
shootings to align the data of two shootings. Before using reference points, analyze
the measured object first, and paste the reference points on appropriate positions. Data
needed should be captured after several times of scanning and alignment.
4. Put the object needs to be scanned in the vision center and adjust
exposing time and blue background of the two cameras.
5. Adjust the raster radiator’s focus accord to Chapter 6.2 till clear raster
images are obtained.
6. Click new files in toolbar, and change the project scanning mode to
“reference point alignment scan”, select “project directory”, enter “project name”.
If the “real-time recognition” is in high light , reference points in two cameras are
real-time displayed
and click again . After scanning for the second position, the error of two
reference points will be shown in the left working panel. Repeat scanning until
scanning for the whole object is completed.
if the mutual area of every two scanning has less than four reference points or
align error is over set, a dialog will come out as figure 6-26
69
OptimScan-5M User Manual
Move the position of object to be scanned or the position of raster and perform
scanning again.
10. Scan finished
Click “New project”, and reset “Project directory”“Project name” to complete
current reference point alignment and start scanning for next object.
1. Make sure that two cameras are in shoot model and the raster radiator is open
and blue light projected
2. confirm that the camera calibration is finished which means “camera.bin” files
can be found under“.\CalibData”
3. Put the object needs to be scanned in the vision center and adjust exposing
time and Gain strength of the two cameras based on the color of scanned object and
environment light.
4. Adjust the raster radiator’s focus accord to Chapter 6.2 till clear raster images
are obtained.
5.Click new files in toolbar, and change the project scanning mode to
“single ply alignment scan”, select “project directory”, enter “project name”. A series
of raster will be projected on objects during scanning, and real-time image will show
in the camera vision.
6.Save results
After raster projection finished, a dialog will come out as figure 6-21, Click
“OK” to finish scanning.
70
OptimScan-5M User Manual
7. Result display
After scanning finished, result display window will show up.
8. Save result
After scanning , the scanning data is automatically saved in the file for new
projects. If you want to save a copy scanning file to appointed position, right click on
the mouse and select “exporting all the cloud point”, choose the route you want and
save. Show as figure 6-23
71
OptimScan-5M User Manual
Figure 6-24
Figure 6-25
When scanning large objects, the final measuring error will be large due to
accumulated error. To control the overall error, build a frame first and then use
72
OptimScan-5M User Manual
6. Click new files in toolbar, and change the project scanning mode to
“frame scan”, select “project directory”, enter “project name”. If the “real-time
recognition” is in high light , reference points in two cameras are real-time
displayed
7. Result display
8. After the first scan of reference points, move the scanner or object for next
scanning. Click . The green points are points in last scanning and the red ones
73
OptimScan-5M User Manual
are new scanned points. Error for each scanning will be displayed in the left work
panel. Show as figure 6-27
If the mutual area of every two scanning has less than four reference points, a
dialog will come out as figure 6-28. Readjust scanning position or scanning angle to
take images again.
74
OptimScan-5M User Manual
Repeat step 8, and scan all the reference points on the object. Hence the frame is
built as figure 6-29.
9. The frame data will be automatically saved as “**.dgm” in “Result” file under
project directory. You can save frame file to appointed position. Click right of
mouse and select “exporting frame points”, choose the route you want and save
as Figure 6-30,6-31.
75
OptimScan-5M User Manual
Figure 6-30
Figure 6-31
10. After the frame is set up, choosing “reference point align” and choose
“open frame point”, and import the scan frame. And then start scanning.
11.Shown as figure 6-32, Frame scan is used to measure the whole object. In
the measurement process frame can help in align two parts without mutual area,
show as following figure:
76
OptimScan-5M User Manual
Programs support multiple project merges,enter the Adjust Mode,click the project
manual merge option in the splicing option in the upper menu bar,as shown figure in
6-38, select point cloud as a reference point.
There are four Windows in the engineering manual merge interface,Fixed
Project,Float Project,Whole Project,Select Project Register,click the Import Project
button,import of the merged project as shown in figure 6-33,
77
OptimScan-5M User Manual
78
OptimScan-5M User Manual
Figure 6-36
Instruction:
The floating project is spliced to a fixed project,the Whole Project shows all the data
that is currently imported,the Selected Project Register shows the result of Fixed
Project and Floating Project.
Select three or more points in Fixed and Floating Project as a Align point,the Selected
Project Register in the lower right shows the results,the Register error shows the
current error, after confirming the result, click the Register button.
Set the result path and result name in the Register Operate, then click the Global
Merge button.
When using this mode, only green points can be selected as Align points,other
operation reference chapter 6.8
79
OptimScan-5M User Manual
80
OptimScan-5M User Manual
7.1Function introduction
The inspect mode can be used to characterize the scanning data for parametric
measurement.
Toolbar commands:
Figure7-1
:The open file function can import the scan data into the detection
module,which supported data is*.dgm、*.asc、*.stl、*.obj,figure as 7-2;
Figure 7-2
Figure 7-3
Figure 7-4
82
OptimScan-5M User Manual
shown in the figure below, click on the next one to perform the fitting of the second
line.
:Select a set of points to fit into a surface feature,Select the point cloud in
the same plane,click apply, and complete the fitting of plane 1,add graphic features
and click next.
83
OptimScan-5M User Manual
84
OptimScan-5M User Manual
:Select some point cloud, click apply, calculate a ball feature according to
the selected point cloud, as shown in figure 7-10;
Figure 7-10
7-11.
85
OptimScan-5M User Manual
Figure7-11
:Click into the point spacing measurement mode and select the
two point clouds that need to be measured. The left menu bar shows the coordinates
and distance of the two points, as shown in the figure 7-11
86
OptimScan-5M User Manual
87
OptimScan-5M User Manual
① Probe nib: direct contact with the surface of the measured object;
② Probe nib fastening bolts;
③ Probe frame point: used to determine the position of the probe nib;
④ Power switch;
⑤ Next: the next step in calibration, detailed introduction in the following;
⑥ Undo: undo the last step;
⑦ Trigger: used to select points in probe calibration and measure, detailed
introduction in the following;
⑧ Probe main body: measuring personnel hand-held position.
88
OptimScan-5M User Manual
89
OptimScan-5M User Manual
In figure 8-4, the standard ball, the fixed point and the calibration frame point are
used for probe calibration.
90
OptimScan-5M User Manual
8.2.1 Preparation
First turn on the power switch of the probe device, the power indicator lights up.
Then to look for "human input device" in the computer "device manager", next to find
"USB input device" option in the drop-down menu and click the mouse right button to
select "properties" command. The hardware Id information can be obtained. If the ID
value is 0471, Bluetooth and computer connection is successful, and the probe link
lights lit. Open the TenYoun 3D scanning system “3DScan.exe”, switch to the probe
measuring mode, click "open Bluetooth" command, then pull the probe trigger, the
buzzer will sound "dididi" alarm, indicating that the probe and Bluetooth adapter
connection is successful.
Attentions:
The stitching error displayed in the left side of the work panel must be guaranteed
within 0.05.If the stitching error more than 0.05, indicating that the frame is not good,
must be removed and then shooting again.
91
OptimScan-5M User Manual
following shape:
Free movement of the probe at a fixed point, pull the trigger for collection. If not
satisfied with the measurement results in a position, press the "undo" key to cancel.
To ensure that the probe nib fixed, try to move the probe angle as large as possible.
When the measurement position reaches more than 70 times, you can click "Next"
button to complete the column calibration. After the column calibration is complete,
the program automatically enters the ball calibration status.
94
OptimScan-5M User Manual
2.Measurement process
95
OptimScan-5M User Manual
Click on the new construction command get the "new project" dialog box, select
project directory and enter the project name. Then open the object frame and click
"ok" button to measure .
Free movement of the probe within the camera vision field, pull the trigger to get
probe contact position. Probe collection has two modes: One is single-point collection,
and the other is continuous collection. For single-point collection, you can obtain the
nip current position data by the trigger pull once. For continuous collection, pull the
trigger once to start collecting data. Moving probe and pull the trigger again end of
collection. The data can be collected by all the nib position in this process. If not
satisfied with the points, press "undo" key to cancel. When all points have been
measured, all collection points can be found in the project directory which format for
"ASC".
●The system has the function of automatically saving data, in order to prevent
measurement data loss which caused by wrong operation or power off. The program
saves measure file real-time in every dot.
96
OptimScan-5M User Manual
3. data shows
97
OptimScan-5M User Manual
Chapter 9 Troubleshooting
Fault phenomenon
The program cannot open after installation
Solution: 1. Check if camera cable is connected.
2. Check if 2 camera drives in “device manager” is on. If not, try to
drive again.
3. Check if the softdog light is on
Fault cause
The cause can be camera or softdog is not correctly connected. Cable connection
or drive installation can also cause failure.
Fault phenomenon
Fault cause
(1) Raster emitter is not correctly connected
(2) Focus setting of raster emitter is not correctly set.
Solution
(1)Confirm if the raster emitter power cable and data cable are correctly
connected.
(2) Adjust the distance between measuring head and calibration block, adjust the
focus of raster emitter until clear raster is shown.
98
OptimScan-5M User Manual
Error 1
Error 2
99
OptimScan-5M User Manual
(1) It is better not to use the 3D scanner under strong light (such as sunlight);
(2) Temperature:+5oC ~ +40oC
(3) Humidity: 20% ~ 80%
(4) Environment: no dust, no strong vibration
(1) Do not let impurities like dust to get into the cameras during camera lens
installation;
(2) Put the cameras back to equipment box when it is not in use.
Tripod is the moving part and needsto check and maintain accordingly. When it is
difficult to swing the tripod and abnormal sound is found ,add lubricant into the tripod.
Pencil powder can also do. Place the pencil powder on the rocker gear, and shake for
several times.
After long distance transportation, there might be some loose screws, especially
the camera and screws between lens. Twist lens to see if tight or not after long
distance transportation.
Calibrate the equipment every time after long distance transportation to ensure
accuracy.
100
OptimScan-5M User Manual
Appendix 1 Others
Contact Us
Hangzhou Shining 3D Tech Co.,Ltd
Add: 18th Floor, Block A, HuaRui Center, No.66, JianShe one Road, Xiaoshan,
HangZhou, China
Tel: 086-571-82999513
Fax: 086-571-82999539
Zip Code: 311215
Skype:shining_support
Email:support@shining3d.com
Web Site: www.shining3d.com
101