Subject
Sampler Motor

Dear Service Manager,

We would like to inform you that all new Samplers from S/N 7-0???
date - - are built with :

a new main motor-type P/N 6002-039


a new Electronic Relay board P/N 6002-035.

The reason is that the main motor P/N 3363-235 is out of production.
In the Manual the main motor is no. 85 in the drawing of the motor-
unit (see page 3 of this info).
The new motor is larger than the old one, therefor did the
Electronic Relay board require a redesign.

The new Electronic Relay board is a universal board, thus it can be


used with the old and new type of the motor.

The voltage selection on the new board has changed and the following
settings are possible :

Type 1: Old motor-type 110V (100V to 130V)


220V (200V to 240V)

Type 2: New motor-type 110V (100V to 130V)


220V (200V to 240V)

It is essential to insert the jumpers of the voltage selector for


each situation in the correct position. (see fig.)