Professional Documents
Culture Documents
Fibre-Grating Sensors For The Measurement of Physi
Fibre-Grating Sensors For The Measurement of Physi
This content has been downloaded from IOPscience. Please scroll down to see the full text.
(http://iopscience.iop.org/1402-4896/2013/T157/014022)
View the table of contents for this issue, or go to the journal homepage for more
Download details:
IP Address: 130.132.173.241
This content was downloaded on 13/03/2016 at 17:17
Abstract
Mechanical physiological pulsations are movements of a body surface incited by the
movements of muscles in organs inside the body. Here we demonstrate the use of long-period
grating sensors in the detection of cardio-vascular pulsations (CVP), in particular apex and
carotid pulsations. To calibrate the sensors, we use a mechanical tool designed specifically to
measure the sensor response to a localized perturbation at different grating curvatures as
working points. From the data we infer the amplitude of the CVP. Together with the
electrophysiological signals, the CVP signals obtained from the sensors can provide
significant information on heart function which is inaccessible to the electrocardiogram. The
low cost and easy handling of the fibre sensors increase their prospects to become the sensors
of choice for novel diagnostic devices.
T[db]
t
−4 2
P
illnesses [3, 9].
The paper is structured as follows. The working principle −6 1.5
of LPG sensors and the particulars of the interrogation scheme −8 1
used in this work are explained in section 2. Application
of these sensors to measuring carotid and apex pulsations −10 0.5
0 2 4 6 8
−1
is described and the results of the measurements are given curvature [m ]
in section 3. An independent calibration procedure used
to enable quantification of the pulsation amplitudes is also Figure 1. (a) Spectral sensitivity of LPG to bending, measured by
OSA. During bending the resonance experiences a red shift. (b)
described therein. Finally, in section 4 the advantages and Power sensitivity of LPG to bending at 1470.4 nm measured by
limitations of the LPG sensors with respect to the current OSA (blue line) and a photodiode (green line). The OSA
standards are reviewed and the prospects for their clinical use measurements in (b) are obtained as the points of intersection of the
are outlined. black line and grating spectra in (a). The observed difference
between the OSA and photodiode measurements is probably due to
the large room temperature variation during the experimental
2. Long-period fibre grating sensors campaign.
The LPG consists of a one-dimensional periodic perturbation interrogation scheme that relies on the measurement of the
of the refractive index profile or the fibre geometry power transmitted through an LPG sensor, see figure 2(a).
along the fibre axis. It has a period of several tens to Such a scheme requires tuning of the grating to resonance with
several hundreds of micrometres which enables coupling the laser. As the sensors are applied to patients with different
of light from the fundamental core mode to the resonant body curvatures and as the position of a grating spectral
co-propagating cladding modes of the fibre at discrete resonance depends on fibre curvature, a series of gratings with
wavelengths. Due to coupling to the lossy cladding modes, slightly different resonant wavelengths was fabricated to suit
the resulting transmission spectrum consists of a series of body shapes of a wide range of patients.
attenuation bands [10]. There are three major characteristics Special care was taken to solve the problem of the
of attenuation bands: resonant wavelength, magnitude and LPG cross-sensitivity to other parameters. Namely, a single
width. The resonant wavelength can be determined from the attenuation band may be simultaneously sensitive to several
phase-matching condition parameters, which makes it impossible to distinguish the
ν eff contribution of a wanted measurand [11]. Here, the grating
λ = n eff
co − n cl 3,
cross-sensitivity to the refractive index has been avoided
where n eff ν eff by using the progressive three-layered fibre whose second
co is the effective index of the core mode, n cl is the
effective index of the νth cladding mode, 3 is the period of cladding served as an isolator [12], while the effects of rapid
the grating and λ is the resonant wavelength. The magnitude temperature fluctuations have been avoided by encapsulating
depends on the mode-coupling strength (and hence on the the sensor in a low-temperature curing silicone rubber, as
index modulation of the grating and the overlap between described in [13].
the resonant modes) and the grating length, while the width
depends also on the fibre and the grating dispersion. 3. CVP measurements
All the above characteristics are sensitive to the
forces applied to the fibre (strain, load, bend) and local We used LPGs to measure small displacements and pressures
environment (temperature, refractive index of surrounding induced by cardiac muscle function and arterial blood flow.
medium) [11]. However, the high cost of the optical spectrum We assume that the apex pulsations can be approximated as
analysers needed for the full grating characterization makes small changes in the chest curvature. On the other hand, the
a straightforward application of such sensing schemes in neck pulsations caused by the blood flow through the carotid
primary care unlikely. Here, we use a simple monochromatic artery act as local pressures. It is interesting to note that this is,
2
Phys. Scr. T157 (2013) 014022 M D Petrović et al
(a) 1.55
1.54
1.53
Pt [a. u.]
1.52
1.51
(b)
1.5
2.3
2.2 1.8 2
1.6
t
δP
t
1
P
2.1
1.4
2 1.2 0
5.5 6 0 1 2
time [s] δ [mm]
1.9
Pt [a. u.]
1.8
Figure 2. (a) Sensing scheme. Light from a diode laser stabilized at
1.7
1470.4 nm is coupled into a fibre with an LPG sensor. The
transmitted power signal is measured by a photodiode and the thus 1.6
obtained electronic signal is digitalized for further processing. (b) 1.5
Sensor calibration gauge.
1.4
to the best of our knowledge, the first study of the response of 1.3
3
Phys. Scr. T157 (2013) 014022 M D Petrović et al
pronounced, transversally to the carotid. The measurements the long-term goal is to exploit the sensors in extending the
were performed on five healthy subjects asked not to breathe diagnostic range of standard cardio-diagnostic devices.
for a short time. A typical measured pulsation signal along
with a zoom-in on one pulsation is shown in figure 4. It Acknowledgments
agrees well with the signals reported in the literature [3, 14].
In addition, using the measured sensitivity we estimate the This work was supported by the Ministry of Education
amplitude of the neck pulsation caused by the carotid blood and Science, Republic of Serbia (Project III45010). NM
flow to be about 0.5 mm. acknowledges support from the Ministry of Education and
When the above measurements were repeated by different Science, Republic of Serbia (Project OS175016) and JP
operators, the signal amplitude varied slightly but all the support from L’OREAL-UNESCO through the Women in
characteristic features of the signals were preserved. Science National Fellowship in Serbia.
4. Conclusion References
[1] Gains M J, Grodecki K M, Jacobs R M, Dyson D and
CVP carry useful diagnostic information that has not been Foster R A 1995 Can. J. Vet. Res. 59 238
fully exploited in modern diagnostics of heart disease. Hence, [2] Avolio A P, Butlin M and Walsh A 2010 Physiol. Meas. 31 R1
there is a strong incentive to develop more advanced and [3] Hurst J W et al 1978 Hurst’s The Heart 4th edn (New York:
operator-independent tools for their recording and analysis. McGraw-Hill) p 422
In this paper we presented a novel fibre-grating CVP sensing [4] Johnson J M, Siegel W and Blomqvist G 1971 J. Appl.
Physiol. 31 796
scheme that is inexpensive, portable and only marginally [5] Willems J L H, Denef B, Kesteloot H and De Geest H 1979 Br.
operator dependent. We showed that these sensors have a Heart J. 41 716
linear response to movements of up to 2 mm, a measurement [6] Denef B, Popeye R, De Geest H and Kesteloot H 1975
resolution of 10 µm and a fast response, which makes them Circulation 51 541
suitable for measuring carotid and apex pulsations. We also [7] Salvi P, Lio G, Labat C, Ricci E, Pannier B and Benetos A
2004 J. Hypertens.S 22 2285
showed how one can eliminate sensors’ cross-sensitivity [8] Allen J 2007 Physiol. Meas. 28 R1
to environmental parameters, characteristic of long-period [9] Tafur E, Cohen L S and Levine H D 1964 Circulation 30 381
gratings. As it was noticed that the signal depends on the [10] Erdogan T 1997 J. Opt. Soc. Am. A 14 1760
way in which the sensor is attached to the body (tape, sticker [11] James S W and Tatam R P 2003 Meas. Sci. Technol. 14 R49
or just a bandage), immediate future work will be directed [12] Allsop T, Webb D J and Bennion I 2003 Opt. Fiber Technol.
9 210
towards finding an optimal measurement configuration. The [13] Allsop T, Carroll K, Lloyd G and Webb D J 2007 J. Biomed.
technique will be further advanced by improving the data Opt. 12 064003
analyses to enable more comfortable measurements that will [14] Siebenhofer A, Kemp C R W, Sutton A J and Williams B 1999
allow for normal breathing during data acquisition. Finally, J. Hum. Hypertens. 13 625