Professional Documents
Culture Documents
1297 01
1297 01
CALIBRATION
Plain Rings (0.125 to 0.25) in (30 + 1.6L) in Master ring and standard
(0.25 to 11) in (20 + 1.6L) in measuring machine
(11 to 16) in (27 + 1.6L) in
Plain Plugs (0.125 to 16) in (8.0 + 1.6L) in Master plug and standard
measuring machine
Threaded Plugs/Rings –
Plug Pitch Diameter (0.1 to 10) in (54 + 2.2D) in Gage blocks, standard
measuring machine,
master thread wires
Plug Major Diameter (0.1 to 10) in 22 in Gage blocks and standard
measuring machine.
Ring Major Diameter (0.06 to 2) in 190 µin Master setting plugs
Micrometers3 –
Digital Up to 6 in 34 in Gage blocks
Vernier (1 to 36) in (62 + 2.5L) in
Up to 3 in 35 in
AC Voltage – Measure
and Generate
22 mV
(10, 20) mV 10 Hz 0.038 % AC/DC transfer
20 Hz to 20 kHz 0.033 % standard 792A with
50 kHz 0.036 % Datron 1271 DMM and
Datron 4808 calibrator
220 mV
20 mV 10 Hz 0.042 %
20 Hz 0.028 %
40 Hz to 20 kHz 0.024 %
50 kHz 0.027 %
60 mV 10 Hz 0.025 %
20 Hz 0.015 %
40 Hz to 20 kHz 0.010%
50 kHz 0.013 %
700 mV
(200, 600) mV 10 Hz 0.026 %
20 Hz 0.010 %
40 Hz to 20 kHz 0.008 %
(50 to 100) kHz 0.019 %
300 kHz to 1 MHz 0.072 %
2.2 V
(0.6, 1, 2) V 10 Hz 0.025 %
20 Hz 0.010 %
40 Hz to 20 kHz 0.065 %
(50 to 100) kHz 0.008 %
300 kHz to 1 MHz 0.064 %
7V
(2, 6) V 10 Hz 0.025 %
20 Hz 0.010 %
40 Hz to 20 kHz 0.006 %
(50 to 100) kHz 0.009 %
300 kHz to 1 MHz 0.055 %
AC Voltage – Measure
and Generate (cont)
22 V
(6, 10) V 10 Hz 0.025 % AC/DC transfer
20 Hz 0.010% standard
40 Hz to 20 kHz 0.007 % 792A with Datron 1271
(50 to 100) kHz 0.010 % DMM and Datron 4808
300 kHz to 1 MHz 0.056 % calibrator
20 V 10 Hz 0.025 %
20 Hz 0.010 %
40 Hz to 50 kHz 0.070 %
70 V
(20, 60) V 10 Hz 0.025 %
20 Hz 0.010 %
40 Hz to 20 kHz 0.007 %
50 kHz 0.010 %
220 V
(60, 100, 200) V 10 Hz 0.025 %
20 Hz 0.012 %
40 Hz to 20 kHz 0.008 %
50 kHz 0.010 %
1000 V
(200, 600, 1000) V 10 Hz 0.025 %
20 Hz 0.012 %
40 Hz to 20 kHz 0.008 %
Measure Only
Inductance – Generate
Fixed Point
200 mH 100 Hz, 200 Hz, 0.15 % GenRad 1482M std
400 Hz, 1 kHz, inductor
and 10 kHz
Inductance3 – Measure
100 μH to 10 H 100 Hz, 200 Hz, 0.20 % GenRad 1689 RLC
400 Hz, 1 kHz, Digibridge
and 10 kHz
Electrical Simulation of
Thermocouple
Indicators –
Type E (-250 to -100) °C 0.12 °C Fluke 5500A/ ice
(-100 to -25) °C 0.16 °C reference
(-25 to 350) °C 0.14 °C
(350 to 650) °C 0.16 °C
(650 to 1000) °C 0.21 °C
Type J (-210 to -100) °C 0.27 °C
(-100 to -30) °C 0.16 °C
(-30 to 150) °C 0.14 °C
(150 to 760) °C 0.17 °C
(760 to 1200) °C 0.23 °C
Type K (-200 to 100) °C 0.33 °C
(-100 to -25) °C 0.18 °C
(-25 to 120) °C 0.16 °C
(120 to 1000) °C 0.26 °C
(1000 to 1372) °C 0.4 °C
Type N (-200 to -100) °C 0.4 °C
(-100 to -25) °C 0.22 °C
(-25 to 120) °C 0.19 °C
(120 to 410) °C 0.18 °C
(410 to 1300) °C 0.27 °C
Electrical Simulation of
Thermocouple
Indicators – (cont)
Type R (0 to 250) °C 0.57 °C Fluke 5500A/ ice
(250 to 400) °C 0.35 °C reference
(400 to 1000) °C 0.33 °C
(1000 to 1767) °C 0.40 °C
Electrical Simulation of
Thermocouple
Indicators3 –
Electrical Simulation of
RTD Indicating Systems
and Measure3 –
Pt 385, 100 (-200 to 0) °C 0.05 °C Decade resistors and
(0 to 100) °C 0.07 °C Datron 1271, Wavetek
(100 to 300) °C 0.09 °C 1281, Fluke 744
(300 to 400) °C 0.10 °C
(400 to 630) °C 0.12 °C
(630 to 800) °C 0.23 °C
Pt 3926, 100 (-200 to 0) °C 0.05 °C
(0 to 100) °C 0.07 °C
(100 to 300) °C 0.09 °C
(300 to 400) °C 0.10 °C
(400 to 630) °C 0.12 °C
Pt 3916, 100 (-200 to -190) °C 0.25 °C
(-190 to -80) °C 0.04 °C
(-80 to 0) °C 0.05 °C
(0 to 100) °C 0.06 °C
(100 to 260) °C 0.07 °C
(260 to 300) °C 0.08 °C
(300 to 400) °C 0.09 °C
(400 to 600) °C 0.10 °C
(600 to 630) °C 0.23 °C
Pt 385, 200 (-200 to 100) °C 0.04 °C
(100 to 260) °C 0.05 °C
(260 to 300) °C 0.12 °C
(300 to 400) °C 0.13 °C
(400 to 600) °C 0.14 °C
(600 to 630) °C 0.16 °C
Pt 385, 500 (-200 to -80) °C 0.04 °C
(-80 to 100) °C 0.05 °C
(100 to 260) °C 0.06 °C
(260 to 400) °C 0.08 °C
(400 to 600) °C 0.09 °C
(600 to 630) °C 0.11 °C
Pt 385, 1000 (-200 to 0) °C 0.03 °C
(0 to 100) °C 0.04 °C
(100 to 260) °C 0.05 °C
(260 to 300) °C 0.06 °C
(300 to 600) °C 0.07 °C
(600 to 630) °C 0.23 °C
Oscilloscopes3 –
Squarewave Signal –
50 @ 1 kHz 1 mV to 6.6 V p-p 0.25 % + 40 µV Fluke 5500A/SC600
RF Power –
(1 to 10) mW (10 to 20) MHz 1.6 % Tegam F1109 coaxial
(20 to 50) MHz 1.5 % power standard and
50 MHz to 8 GHz 1.4 % Agilent E4418B power
(8 to 18) GHz 1.5 % meter
RF Attenuation –
Volumetric Ware –
(To Contain or (1.0 to 10) mL 0.0054 mL NISTIR 7383 SOP #14 by
Deliver) (> 10 to 25) mL 0.0063 mL gravimetric determination
(> 25 to 50) mL 0.0085 mL
(> 50 to 100) mL 0.016 mL
(> 100 to 200) mL 0.026 mL
(> 200 to 250) mL 0.032 mL
(> 250 to 500) mL 0.064 mL
(> 500 to 1000) mL 0.13 mL
(> 1000 to 2000) mL 0.19 mL
VI. Mechanical
Torque –
Wrenches3, (0.4 to 1000) ft·lbf 0.47 % Torque transducers
Screwdrivers indicator
Watches (0.6 to 100) in·ozf 0.33 % Torque arms with ASTM
Class 4 dead weight
Torque Analyzers, (1 to 100) in·ozf 0.053 % Torque arms with ASTM
Transducers (5 to 50) in·lbf 0.049 % Class 4 dead weight
(50 to 1200) in·lbf 0.046 %
(100 to 600) ft·lbf 0.049 %
(600 to 1000) ft·lbf 0.068 %
Pressure – Measuring
Equipment
Gas Medium (0.2 to 25) psi 17 x 10-6 psi Dead weight tester
(25 to 100) psi 19 x 10-6 psi
(100 to 1000) psi 21 x 10-6 psi
Oil Medium (1000 to 20 000) psi 49 x 10-6 psi
Frequency Dissemination 10 MHz 2.5 parts in 1011 Hz/Hz Fluke 910 GPS controlled
frequency standard
Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005
General requirements for the competence of testing and calibration laboratories. This laboratory also meets R205 – Specific
Requirements: Calibration Laboratory Accreditation Program. This accreditation demonstrates technical competence for a
defined scope and the operation of a laboratory quality management system
(refer to joint ISO-ILAC-IAF Communiqué dated April 2017).
_______________________
President and CEO
For the Accreditation Council
Certificate Number 1297.01
Valid to May 31, 2020
For the calibrations to which this accreditation applies, please refer to the laboratory’s Calibration Scope of Accreditation.