Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

3 4058

page 1 of 1

Technical Data Bulletin


#150, September, 2001 — Reusable Respirators
Issue Date 11/01/01
Inspection, Cleaning and schedule for individually assigned 5. Respirator components must be
respirators, but they must not be the inspected prior to each use. A
Storage Procedures for only method in place. During fit- respirator with any damaged or
3M Reusable Respirators testing, wipes may also be used deteriorated components must be
between employees being tested. repaired or discarded.
Inspection However, these respirators must be 6. The cleaned respirator should
3MTM 6000 and 7000 Series thoroughly cleaned at the end of be stored away from contaminated
Respirators must be inspected each day, using procedures in areas when not in use.
before each use to ensure good appendix B-2 of 29 CFR 1910.134.
operating condition. The facepiece NOTE: The above information
must be repaired or replaced if WARNING is also outlined in the users
there are damaged or defective instructions. Additionally, see
Do not clean with solvents.
parts. The following inspection Cleaning with solvents may specific product user instructions
procedure is suggested: degrade some respirator packaged with each respirator
components and reduce facepiece for additional
1. Check facepiece for cracks, tears
respirator effectiveness. information.
and dirt. Be certain facepiece,
especially faceseal area, is not Inspect all respirator
components before each For more Information:
distorted.
use to ensure good
2. Examine inhalation valves for operating condition. Failure In the U.S., contact:
signs of distortion, cracking or to do so may result Technical Assistance
tearing. in sickness or death. 1-800-243-4630
Sales Assistance/Local Distributor
3. Make sure that head straps are 1-800-328-1667
1. Remove cartridges and filters.
intact and have good elasticity. Fax On Demand
2. Clean facepiece (excluding 1-800-646-1655
4. Examine all plastic parts for
filters and cartridges) by Internet
signs of cracking or fatiguing.
immersing in warm cleaning www.3M.com/occsafety
Make sure filter gaskets or seal
solution, water temperature not to E-mail
areas are in good condition. occsafety@mmm.com
exceed 120º F (49º C), and scrub
5. Remove exhalation valve cover with soft brush until clean. Add For other 3M products
and examine exhalation valve and 1-888-3M HELPS
neutral detergent if necessary. Do
valve seat for signs of dirt, not use cleaners containing lanolin In Canada, contact:
distortion, cracking or tearing. or other oils. 3M Canada Company, OH&ESD
Replace exhalation valve cover. P.O. Box 5757
3. Disinfect facepiece by soaking London, Ontario N6A 4T1
6. Inspect lens of full facepiece for in a solution of quaternary Technical Assistance (Canada only)
any damage that may impair ammonia disinfectant or sodium 1-800-267-4414
respirator performance or vision. hypochlorite (1 oz [30 ML] Sales Assistance
1-800-265-1840, ext. 6137
household bleach in 2 gallons [7.5
Cleaning and Storage L] of water), or other disinfectant. Internet
Cleaning is recommended after www.3M.com/CA/occsafety
4. Rinse in fresh, warm water and 3M Canada E-mail
each use. 3MTM Respirator
air dry in noncontaminated ohes@ca.mmm.com
Cleaning Wipes 504 may be used
atmosphere. Technical Assistance In Mexico
as an interim method in the cleaning 01-800-712-0646
Technical Assistance In Brazil
0800-550705
3M Occupational Health and Environmental Safety Division
For all other OUS locations:
3M Center, Building 235-2W-70 1-651-732-6530
St. Paul, MN 55144-1000

You might also like