MDS-101
Periodic replacement manual
6000h
MDS-101
Table of contents
Table of contents
Introduction
This system is a single patient dialysis system intended for the artificial hemodialysis of ESRD patients and
should not be used for other purposes.
This periodic replacement manual describes the required items for the maintenance of the system.
・ Make sure that maintenance is conducted by personnel who took the system maintenance lecture and
is fully versed in the system.
・ In order to keep performance and safety during the lifetime of this system and to avoid a breakdown, we
highly recommend customers to conduct a periodic preventive maintenance for this system. Especially,
concerning possible fatigue, replacement of the “periodic replacement parts” within a specified period is
recommended.
・ Remove the top cover and open the front panel. Then open the both sides of the side cover.
Securing enough working space is recommended for the work.
・ Periodic replacement includes such works as decomposition of parts and removal of tubes containing
liquid.
Therefore, please prepare tray and waste cloths to treat the drain from the parts and tubes.
・ Regarding specific parts, adjustment and operation check of the machine are required after periodic
replacement.
Please conduct them, consulting the Service manual.
・ Confirm if the magnetic shield of the control board is attached firmly and there’s no tools left in the
machine before closing the front cover and side covers.
・ Be sure to conduct chemical and water rinse after the replacement of parts.
Part composition figure /
Part replacement procedure
MDS-101
Part composition figure / Part replacement procedure
2
MDS-101
Part composition figure / Part replacement procedure
Pressure relief
valve spring (12000h)
3
MDS-101
Part composition figure / Part replacement procedure
Body
Partition
Body
4
MDS-101
Part composition figure / Part replacement procedure
5
MDS-101
Part composition figure / Part replacement procedure
【Assembling (12000h)】
1) Replace with the new air separator. Insert
the screw into the notch on the air separator
and fasten the screw.
2) Connect the tubes and remove the forceps.
To V10
To the water relief valve
6
MDS-101
Part composition figure / Part replacement procedure
MEMO
7
MDS-101
Part composition figure / Part replacement procedure
Adjusting screw
Base
Mounting plate
8
MDS-101
Part composition figure / Part replacement procedure
9
MDS-101
Part composition figure / Part replacement procedure
10
MDS-101
Part composition figure / Part replacement procedure
MEMO
11
MDS-101
Part composition figure / Part replacement procedure
V10
V11
V12
Body
Diaphragm unit
SVB-1A plunger
(6000h) (12000h)
Solenoid ASSY
SVB-1A diaphragm
(6000h) (12000h)
Retainer
Spring
12
MDS-101
Part composition figure / Part replacement procedure
1.4.4 Confirmation
1) Refer to the separate [Service Manual 5.42
Solenoid Valve].
13
MDS-101
Part composition figure / Part replacement procedure
Differential
pressure relief valve spring (12000h)
14
MDS-101
Part composition figure / Part replacement procedure
15
MDS-101
Part composition figure / Part replacement procedure
Concentrate pump
Concentrate bottom BF
Concentrate head BF (6000h) (12000h)
(6000h) (12000h)
Retainer bottom
Piston Cylinder
Retainer
Cylinder head
16
MDS-101
Part composition figure / Part replacement procedure
17
MDS-101
Part composition figure / Part replacement procedure
【Grease application】
15) Apply Stabrags Grease on to the ball screws.
Apply an ear-pick full of grease onto the each
ball nut part of upper and lower nuts.
Note
18
MDS-101
Part composition figure / Part replacement procedure
11) Attach the BF diaphragm onto the cylinder 16) Fix cylinder to the motor with 4 screws,
case. Then fix and straighten the BF having the motor connect side face to the
diaphragm concentrate stroke pin side.
as the 17) Attach the interrupter PCB, with the “UP”
picture on mark on it up.
the right. Set as this 18) Attach the concentrate pump in the reverse
order indicated in the disassembling section
2) - 4).
1.6.3 Adjustment
・ Try to avoid any sagging of the BF diaphragm. 1) There’s no need for adjustment after the
parts replacement.
12) Attach the concentrate stroke pin to the
Note
retainer bottom.
13) Using 4 screws, attach the cylinder head and ・ There’s no need for adjustment of concentration.
cylinder base to the cylinder. Align the faces
of the concentrate stroke pin and hose 1.6.4 Confirmation
opening on the cylinder horizontally. Align the 1) Conduct the inspection of the concentrate
cylinder head hose and cylinder hose pump according to the Service manual
vertically. 5.37.1.
2) Operate the machine under the rinse mode
to check the following points.
・ Be cautious about the direction of each part ・If there’s any abnormal sound or vibration.
when assemble.
・If there’s any leakage.
・If there’s any needless alarming.
14) With the surface having the smaller hole on it
up, mount the thrust bearing on the motor.
19
MDS-101
Part composition figure / Part replacement procedure
V20-1 V20-2
20
MDS-101
Part composition figure / Part replacement procedure
1.7.3 Adjustment
1) No need for adjustment after replacing the
parts.
C328-H416A diaphragm boss
1.7.4 Confirmation
C328-H416A spring 1) Refer to the separate [Service Manual 5.42
C328-H416 Valve seats Solenoid Valve].
5) Replace the two C328-H416 valve seats
(6000h, 12000h).
6) Replace the C328-H416A spring and
C328-H416A diaphragm boss (12000h).
21
MDS-101
Part composition figure / Part replacement procedure
Fixing plate
Chambers I and II
Chamber T joint
Chamber screws
(6000h)
Seat packing
(6000h) (12000h)
Chamber diaphragm
Nuts (6000h) (12000h)
Scale chamber
Chamber I side (12000h)
22
MDS-101
Part composition figure / Part replacement procedure
23
MDS-101
Part composition figure / Part replacement procedure
5) Using chamber fixing screws and nuts, fix 【I n the case of Hot rinse machine】
the chamber I and II. Use the torque driver 6) Replace with the new Chamber nipple
for tightening, setting the power to 30kgf・cm. (12000h).
To avoid leakage, tighten all the screws in Replace the seat packing,
the order indicated with the picture 1 below. O-ring/P-10Si/white and attach them to the
Then re-tighten them all according to the Chamber nipple (6000h, 12000h).
picture 2. 7) Fix the nipple cover plate with 4 screws.
8) Paying utmost attention to the direction of
Chamber nipple, attach it. Connect the
silicon tube having the size of φ5×φ11,
connecting the Chamber nipple with
picture 1 picture 2
Solenoid valve, to the Chamber nipple.
kit】
6) Attach the seat packing and
O-ring/P-10Si/white to the originally attached
Chamber nipple.
・ Chamber nipple’s shape becomes T-shape and ・ Direction of Chamber nipple when attaching
the shape of Nipple cover plate also changes. (For both of Chamber T joint replacement kit
・ When replacing 1 set of Chamber unit, 4 sets of and Hot rinse machine)
Chamber T joint replacement kit are required.
V2 V3 V7 V6
7) Fix the nipple cover plate with 4 screws.
8) Paying utmost attention to the direction of
Chamber nipple, attach it. Then connect the
originally attached silicon tube having the
size of φ5×φ11 to the Chamber nipple.
Note
24
MDS-101
Part composition figure / Part replacement procedure
25
MDS-101
Part composition figure / Part replacement procedure
V1-V8
Body
Diaphragm unit
SVB-1A plunger
(6000h) (12000h)
Solenoid ASSY
SVB-1A diaphragm
(6000h) (12000h)
Retainer
Spring
26
MDS-101
Part composition figure / Part replacement procedure
1.9.1 Tools required piece 4) Install the chamber unit on the system and
・ Phillips driver (No.2)...........................................1 connect the solenoid valve connectors and
the tubes.
1.9.2 Replacement procedure
【Disassembling】
1) Remove the chamber unit. ・ Ensure that the tubes are correctly connected.
2) Remove the tube and connector from the
solenoid valve to be replaced. 5) Be sure to conduct air purge and operation
3) Loosen the two screws from the solenoid check after replacement. Also, confirm if
valve set plate. Slide the solenoid valve and there’s no leakage in the connecting part of
remove it from the T-shaped tube. solenoid valve.
4) From the rear side of the solenoid valve set 6) On the Reset Work Time screen, reset the
plate, loosen the two screws fixing the work time of the replaced solenoid valve.
solenoid valve to the solenoid valve set plate
and remove the solenoid valve. 1.9.3 Adjustment
1) No need for adjustment after replacing the
parts.
・ Be careful not to let the intimal spring pop out.
1.9.4 Confirmation
5) Remove the diaphragm unit and 1) Refer to the separate [Service Manual 5.41
disassemble the unit into the diaphragm, Chamber I and II] and [Service Manual 5.42
retainer and plunger. Solenoid Valve].
【Assembling】
1) Replace the SVB-1A diaphragm and
SVB-1A plunger. Assemble the unit with the
retainer inserted between the diaphragm and
plunger.
2) Assemble the solenoid valves and attach the
solenoid valve set plate to the solenoid valve.
3) Insert the solenoid valve set plate assembled
in step 2) into the T-shaped tube and fasten
the screws.
T-shaped
tube Solenoid valve
Set plate
27
MDS-101
Part composition figure / Part replacement procedure
O-ring/JASO60/Si White
Body (6000h)
Mixing plate
Body
28
MDS-101
Part composition figure / Part replacement procedure
29
MDS-101
Part composition figure / Part replacement procedure
【Disassembling (12000h)】
1) Refer to the aforementioned step 1) - 3). 1.10.3 Adjustment
2) Place a tray under the mixing chamber. 1) No need for adjustment after replacing the
Close the tubes connected to the mixing parts.
chamber with forcipes and disconnect the
tubes over the tray. 1.10.4 Confirmation
1) Refer to the separate [Service Manual 5.34
【Assembling (12000h)】 Mixing Chamber].
1) Replace the in to mixing chamber. Insert the
screw into the notch on the mixing chamber
and fasten the screw.
2) Connect the tubes and remove the forceps.
30
MDS-101
Part composition figure / Part replacement procedure
MEMO
31
MDS-101
Part composition figure / Part replacement procedure
Motor
O-ring/JASO13/Si Red
(6000h) (12000h)
32
MDS-101
Part composition figure / Part replacement procedure
【Grease application】
IN-line 12) Apply specified silicone oil on the inner
O-ring bushing surface of the through-hole of the O-ring
Valve head bushing.
33
MDS-101
Part composition figure / Part replacement procedure
【Disassembling (12000h)】
1) Refer to the 【Disassembling (6000h)】 step 6) Loosen the two set screws of the feed screw
1) - 9). portion.
7) Remove the needle unit from the motor.
Note
Needle guide
4) Remove the four screws of the valve body. ・ Wipe the needle portion with clean cloth to
5) Remove the motor and the needle unit from remove the oil, the affix, etc.
the valve body.
7) Attach the interrupter mounting plate,
targeting the marks made in the prior section
Needle unit of 【Disassembling (12000h)】.
Motor 8) Refer to the 【Assembling (6000h)】 step 1)-
Valve body 6).And attach flow control valve to the unit.
1.11.3 Adjustment
1) Refer to the separate [Service Manual 5.43
Needle guide set position Flow control valve].
●
1.11.4 Confirmation
1) Refer to the separate [Service Manual 5.43
Feed screw Needle
Flow control valve].
D-shaped cut side
Set screw
Needle
Feed screw
Motor output shaft
34
MDS-101
Part composition figure / Part replacement procedure
MEMO
35
MDS-101
Part composition figure / Part replacement procedure
Base
36
MDS-101
Part composition figure / Part replacement procedure
3) Close the four tubes connected to the valve ・ After replacement, always check the dialysate
flow rate and confirm that it operates within the
with forcipes and remove the valve. criteria.
1.12.3 Adjustment
・ Be careful not to let fluid flow into the system. 1) No need for adjustment after replacing the
parts.
4) While holding the constant flow valve body
and base, loosen the six screws and 1.12.4 Confirmation
disassemble the constant flow valve. 1) Refer to the separate [Service Manual 5.44
Constant Flow Valve].
37
MDS-101
Part composition figure / Part replacement procedure
O-ring/P7/Si White
(6000h) (12000h)
Body
Washer
38
MDS-101
Part composition figure / Part replacement procedure
【Assembling】
1) Assemble the dialysis fluid conductivity
monitor and install this on the system.
39
MDS-101
Part composition figure / Part replacement procedure
Sampling port
Port cap
Port packing
(6000h) (12000h)
Port plunger
Port spring
(6000h) (12000h)
Port body
40
MDS-101
Part composition figure / Part replacement procedure
【Assembling】
1) Assemble the sampling port and install the
unit on the system.
41
MDS-101
Part composition figure / Part replacement procedure
Body
Cap
42
MDS-101
Part composition figure / Part replacement procedure
1) Attach and turn the cap clockwise to fasten it. 1.15.3 Adjustment
1) No need for adjustment after replacing the
parts.
1.15.4 Confirmation
1) Refer to the separate [Service Manual 5.48
Blood Leak Detector].
43
MDS-101
Part composition figure / Part replacement procedure
Circulation chamber
(12000h)
Body
O-ring/JASO60/Si White
(6000h) (12000h)
Body
44
MDS-101
Part composition figure / Part replacement procedure
45
MDS-101
Part composition figure / Part replacement procedure
【Assembling (12000h)】
1) Replace the in to circulation chamber. Insert
the screw into the notch on the circulation
chamber and fasten the screws.
2) Connect the tubes and remove the forceps.
46
MDS-101
Part composition figure / Part replacement procedure
MEMO
47
MDS-101
Part composition figure / Part replacement procedure
Adjusting screw
Base
Mounting plate
48
MDS-101
Part composition figure / Part replacement procedure
49
MDS-101
Part composition figure / Part replacement procedure
50
MDS-101
Part composition figure / Part replacement procedure
MEMO
51
MDS-101
Part composition figure / Part replacement procedure
52
MDS-101
Part composition figure / Part replacement procedure
1.18.1 Tools required piece 7) Assemble the solenoid valve and connect
・ Phillips driver (No.2)...........................................1 the tubes and connector.
・ Forceps...............................................................2 8) After replacement, check the solenoid valve
operation on the maintenance mode screen.
1.18.2 Replacement procedure 9) Reset the work time of the solenoid valves
【SVB - 1A】Serial Number : 80101 - 80200 replaced on the reset work time screen.
1) Remove the connector from the solenoid
valve to be replaced. 【C2KK - H803C】Serial Number : 80201-
2) Clamp the tubes with forcipes and remove 1) Refer to the above-mentioned step 1) - 5).
the tubes. 2) Replace the diaphragm and plunger.
Assemble the unit with the retainer inserted
between the diaphragm and plunger. And
・ Be careful not to let fluid flow into the system. then replace the spring.
3) Refer to the above - mentioned step 7) - 9).
3) Loosen the two screws and remove the
solenoid valve.
4) Loosen the two screws and disassemble the
solenoid valves (V16).
Spring
Tube
O-ring
Body Screw
Solenoid ASSY
Diaphragm
Plunger
Retainer
Diaphragm unit
53