Professional Documents
Culture Documents
Diseño y Desarrollo de Un Exoesqueleto Ligero de Tobillo para El Aumento de La Marcha Humana.
Diseño y Desarrollo de Un Exoesqueleto Ligero de Tobillo para El Aumento de La Marcha Humana.
Mechatronics
journal homepage: www.elsevier.com/locate/mechatronics
a r t i c l e i n f o a b s t r a c t
Keywords: Most of powered ankle exoskeletons add considerable distal mass to the user which limit their capacity in reduc-
Lower limb exoskeleton ing the metabolic energy of walking. The objective of the work presented in this paper is to develop an ankle
Human augmentation exoskeleton with a minimum added distal mass compared to existing autonomous powered ankle exoskeletons
Ankle exoskeleton
while it can provide at least 30 Nm of assistive plantarflexion torque. The proposed exoskeleton uses Bowden
Wearable robot
cables to transmit the mechanical force from the actuation unit attached to the waist to the carbon fiber struts
fixed on the boot. As the struts are pulled, they create an assistive ankle plantarflexion torque. A 3d-printed brace
was attached to the shin to adjust the direction of the cables. A design optimization study was performed to min-
imize the mass of the struts, thereby limiting the total added distal mass, attached to the shin and foot, to only
348 g. The main result obtained from walking tests was the reduction of the soleus and gastrocnemius muscles
activity by a maximum of 37% and 44% respectively when walking with the exoskeleton compared to normal
walking. This result shows the potential of the proposed exoskeleton to reduce the metabolic cost of walking and
emphasizes the importance of minimizing the distal mass of ankle exoskeletons. Tests with more subjects will be
carried in the future to confirm this result.
1. Introduction largest compared to other joints of the leg [18]. Therefore, adding posi-
tive power during this phase on the ankle can have a beneficial effect on
Walking while carrying heavy loads is a fundamental skill in daily the walking energetics. There are different design approaches for ankle
activities of some professions. For example, soldiers are often required exoskeletons in the literature; one of the most distinctive design features
to walk while carrying heavy backpacks that weigh more than 50% of is the actuation form as either passive or active, as described in the two
their lean body mass [1]. Firefighters also wear protective equipment next paragraphs, respectively.
and breathing apparatus that negatively impact their physical conditions Passive exoskeletons are suitable for augmentation on the ankle joint
[2]. Several exoskeletons have been developed to reduce the effect of as the power on the ankle has a negative period where energy can be
heavy payloads during walking [3,4]. The design of these exoskeletons stored in an elastic element and a positive period where the stored en-
varies from full leg exoskeletons [5–9] to single joint exoskeletons that ergy can be released. Using this approach, Collins et al. developed a
assist the user at the hip [10,11] or at the ankle [12–14]. It is important lightweight device equipped with a passive clutch mechanism attached
to note that the most commonly used metric to evaluate these exoskele- to a series spring to augment the ankle torque during push off [19]. Tests
tons is the metabolic cost of walking which is the energy consumed by performed on healthy typically developed subjects walking with this de-
the human body during walking [4,15]. Another common metric is the vice showed a reduction in metabolic cost by approximately 7% [13].
analysis of electromyography (EMG) signals to examine changes in mus- The advantage of passive ankle exoskeletons is that they are lightweight
cle activation [15]. Both of these metrics will be referred to in the rest and simple; however, the positive power they provide is limited com-
of the paper. pared to active exoskeletons, due to the lack of actuators.
Among single joint exoskeletons, ankle exoskeletons have shown Pneumatic actuation was used in many ankle foot orthoses for both
great potential in reducing the metabolic cost of walking [13,16,17]. rehabilitation and gait training purposes [20–23]. Pneumatic actua-
The power generated on the ankle joint during the push off phase is the tors have a high specific power [24], and thus they represent a good
Abbreviations: EMG, Electromyography; FSR, Force sensitive resistor; WAXO, Walking augmentation ankle exoskeleton; PID, Proportional integrator derivative;
PC, Personal computer; BLDC, Brushless direct current.
✩
This paper was recommended for publication by Associate Editor Prof. Mahdi Tavakoli.
∗
Corresponding author.
E-mail addresses: yacine.bougrinat@polymtl.ca (Y. Bougrinat), sofiane.achiche@polymtl.ca (S. Achiche), maxime.raison@polymtl.ca (M. Raison).
https://doi.org/10.1016/j.mechatronics.2019.102297
Received 8 March 2018; Received in revised form 24 August 2019; Accepted 27 October 2019
0957-4158/© 2019 Elsevier Ltd. All rights reserved.
Y. Bougrinat, S. Achiche and M. Raison Mechatronics 64 (2019) 102297
Table 1
Mass of selected active ankle exoskeletons.
Device Overall mass (g) Distal mass (shank and foot) on one leg (g)
3. Mechanical design
Fig. 3. (A) The cable routing brace attached to the shin was
used to change the direction of the cables (B) a zoomed figure
of the cable routing brace.
be placed near the center of mass of the body. Although using Bowden point O coincides with the metatarsophalangeal joint representing the
cables results in loss of mechanical power due to friction, the advan- origin of the coordinate system. One end of the strut is fixed on the point
tage of this choice is that it reduces the effect of the actuation unit on O and the other end is represented by the point C. C’ is the projection
the metabolic energy because the actuation unit has the largest weight of the point C on the x axis. The point B is the location of the posterior
compared to other main parts of the device. extremity of the boot. The point A represents the location of the ankle
Two Bowden cables were used to pull the two struts. Each cable was joint. The point P is the lower tip of the V brake noodle, thus, the line
attached to the spool on one end and to one of the two struts on the (CP) defines the direction of the force exerted by the cable. The point
other end. When the spool rotates, the cables rotate simultaneously in D in the line (CP) defines the perpendicular distance 𝐴𝐷 between point
the same direction pulling the two struts forward. A brace was custom- A and the line (CP). P’ is the intersection of the horizontal line passing
designed, and 3D printed, in such a way as to change the direction of through the point C and the vertical line passing through the point P.
the cables from vertical to the pulling direction (see Fig. 3A). Two “V
brake” noodles were inserted on the left and right sides of the brace to 3.4.1.1. Design variables. Three design variables were chosen for this
guide the cables (see Fig. 3B). optimization study as shown in Table 2. These variables control the ge-
ometrical configuration of the struts and the position of the cable routing
3.3.3. Struts brace on the shank.
The struts set, which constitutes an exoskeletal foot rigidly attached
to the boot, is the physical interface between the exoskeleton and the 3.4.1.2. Design parameters. Design parameters are different from design
user and guarantees the transfer of power to the biological joint. The variables. Unlike the design variables that are changed by the optimiza-
struts were optimized to provide a large moment arm while keeping tion algorithm at each iteration. The design parameters can vary in gen-
their mass to a minimum. Details of the optimization are given in the eral but are fixed for each optimization run. Design parameters can be
next section. The struts were fixed by bolts to the boot insole on the used to examine how their variation affects the final design solution.
metatarsophalangeal joint and under the ankle joint. This configuration The design parameters chosen in our optimization problem are given in
was chosen to reduce undesired forces and moments on these joints that Table 3.
may alter the natural gait. We denote the design variables and parameters by x and p respec-
tively, that is:
3.4. Struts optimization [ ]𝑇
𝒙 = 𝑙 𝑠 , 𝜃𝑠 , 𝜃𝑐
The struts are the main distal mass added around the ankle joint by
the WAXO exoskeleton. Therefore, as our main objective is to reduce the [ ]𝑇
𝒑 = 𝐷1𝑚𝑎𝑥 , 𝐷2𝑚𝑖𝑛 , 𝐻𝑚𝑎𝑥 , 𝐿𝑚𝑎𝑥 , 𝜃𝑚𝑎𝑥 , 𝑥𝐴 , 𝑦𝐴 , 𝑥𝑃 , 𝑑𝐵
distal mass while providing a high torque, an optimization study of the
struts assembly was performed. This optimization was done through two 3.4.1.3. Objective function formulation. The main goal of the layout op-
distinct optimization problems. First, the geometrical layout of the struts timization is to maximize the artificial lever arm, 𝐴𝐷[mm], of the aug-
was optimized to result in a maximum torque lever arm; thus, reducing mentation torque when the foot is on the ground with no plantarflexion.
force requirements of the motor. Then, the thickness and width of the Eq. (3) represents the objective function written in terms of design vari-
struts were optimized to minimize the mass. In the next subsections, we ables and parameters. The details of the development of this equation
give details of these two optimization problems. are given in Appendix A.
( ( ( ) ))
3.4.1. Struts geometrical layout optimization 𝑙s cos 𝜃𝑠 − 𝑥𝐴
Optimizing the struts layout is central to reducing the added distal 𝑓 (𝒙, 𝒑) = 𝐴𝐶 𝑠𝑖𝑛 𝜃𝑐 + 𝑎𝑟𝑐𝑜𝑠 (3)
𝐴𝐶
mass while providing high torque comparable to the biological torque.
From each side of the foot, the structure consists of two struts attached to Where:
√
each other near the ankle joint (Fig. 3A). The longer strut is bolted to the ( ( ) )2 ( ( ) )2
𝐴𝐶 = 𝑙s cos 𝜃𝑠 − 𝑥𝐴 + 𝑙s 𝑠𝑖𝑛 𝜃𝑠 − 𝑦𝐴
boot insole on the metatarsophalangeal joint on one end and attached
to the Bowden cable on the opposite end. The shorter strut is fixed to
3.4.1.4. Constraints. The constraints associated with the optimization
the previous strut from one end and to the back side of the insole on
are as follows:
the other end. The shorter struts help reinforce the bolt in holding the
longer struts fixed under the applied cable force, thereby, maximizing 1. When the foot is on the ground (no plantarflexion), the part of
the transfer of power from the actuator to the user joint. the struts behind the foot should not exceed a maximum value
Fig. 4 shows the design variables as well as the different points defin- D1max [mm] on the horizontal axis. A large strut part behind the
ing the lines and distances used in the optimization problem below. The foot might make the device impractical or difficult to use in a real
Y. Bougrinat, S. Achiche and M. Raison Mechatronics 64 (2019) 102297
Table 2
Design variables.
Table 3
Design parameters.
D1max Maximum distance behind the strut tip and the back of the insole [mm]
D2min Minimum distance between the strut tip and the vertical line passing through the ankle joint [mm]
Hmax Maximum height of the cable routing brace from the insole [mm]
Lmax Maximum length of the strut [mm]
𝜃 max Maximum value for the cable angle 𝜃 c [rad]
xA x coordinate of the ankle joint (point A) [mm]
yA y coordinate of the ankle joint (point A) [mm]
xP x coordinate of the point P [mm]
dB Distance between the strut attachment on the insole (point O) and the back side of the insole (point B). [mm]
setting as the strut or the cable can get stuck to objects from the security of the user.
surrounding environment. This constraint could be expressed in ( )
terms of design variables and parameters as follows: 𝑔2 (𝒙, 𝒑) = 𝐷2𝑚𝑖𝑛 − 𝑙𝑠 𝑐𝑜𝑠 0.26 + 𝜃𝑠 + 𝑂𝐴 cos (0.26 + ∠𝐴𝑂𝐵) < 0
(5)
( )
𝑔1 (𝒙, 𝒑) = 𝑙𝑠 𝑐𝑜𝑠 𝜃𝑠 − 𝑑𝐵 − 𝐷1𝑚𝑎𝑥 < 0 (4) Replacing the angle ∠AOB and distance 𝑂𝐴 by their explicit form
in terms of design parameters, we obtain:
2. When maximum plantar flexion is reached (around 15°) (Fig. 4B), ( ( ))
( ) √( 2 ) 𝑦
the distance on the horizontal axis between the tip of the longer 𝑔2 (𝒙, 𝒑) = 𝐷2𝑚𝑖𝑛 − 𝑙𝑠 𝑐𝑜𝑠 0.26 + 𝜃𝑠 + 𝑥𝐴 + 𝑦2𝐴 cos 0.26 + atan 𝐴 <0
𝑥𝐴
strut (point C in Fig. 4) and the vertical line passing through the
(6)
ankle joint A must be bigger that D2min . This constraint is neces-
sary for the proper functioning of the cable routing brace and the Here, atan returns values in the first quadrant.
Y. Bougrinat, S. Achiche and M. Raison Mechatronics 64 (2019) 102297
3. When the foot is on the ground (no plantarflexion), the height Table 4
of the cable routing brace must be lower than Hmax .This ensures Numerical values for parame-
that the brace be placed on the shank area, at a given distance ters.
from the knee, making its attachment easy. Parameter Value Unit
The height of the brace represented by the point P is given by:
D1max 30 [mm]
( ) ( ( ) ) ( )
𝑔3 (𝒙, 𝒑) = 𝑙𝑠 𝑠𝑖𝑛 𝜃𝑠 + 𝑙𝑠 cos 𝜃𝑠 − 𝑥𝑃 tan 𝜃𝑐 − 𝐻𝑚𝑎𝑥 < 0 (7) D2min 0 [mm]
Hmax 300 [mm]
Where xP is the x coordinate of the point P. Lmax 330 [mm]
4. To prevent excessive deflection of the strut when the force is ap- 𝜃 max 0.52 [rad]
xA 130 [mm]
plied to it, the strut length, 𝑂𝐶 [mm], was limited to an upper
yA 105 [mm]
bound, Lmax [mm]. However, this limit was considered as a de- xP 70 [mm]
sign parameter to examine how its variation affects the solution. dB 195 [mm]
This bound constraint is given by:
𝑙𝑠 < 𝐿𝑚𝑎𝑥 (8) Table 5
Obtained results.
5. The cable angle, 𝜃 c [rad], was bounded by zero on the lower
bound. A negative angle means that the cable end attached to Variable Optimal value Unit
the strut tip is pulled downward, which could cause high deflec- ls 330 [mm]
tion of the struts as the body is propelled upward during push 𝜃s 1.0 [rad]
off. Although this deflection will not necessarily cause a damage, 𝜃c 0.2 [rad]
( ( ))
( ) √( 2 ) 𝑦
𝑔2 (𝒙, 𝒑) = 𝐷2𝑚𝑖𝑛 − 𝑙𝑠 𝑐𝑜𝑠 0.26 + 𝜃𝑠 + 𝑥𝐴 + 𝑦2𝐴 cos 0.26 + atan 𝐴 <0
𝑥𝐴
( ) ( ( ) ) ( )
𝑔3 (𝒙, 𝒑) = 𝑙𝑠 𝑠𝑖𝑛 𝜃𝑠 + 𝑙𝑠 cos 𝜃𝑠 − 𝑥𝑃 tan 𝜃𝑐 − 𝐻𝑚𝑎𝑥 < 0
𝑙𝑠 < 𝐿𝑚𝑎𝑥 Fig. 5. Objective function and isocontours of the constraints. The contraints
“distance to shank” (red), “cable brace height” (green) and “Max. strut length”
(yellow) are active. Changing those contraints will change the solution.
0 < 𝜃𝑐 < 𝜃𝑚𝑎𝑥
where:
[ ]𝑇 Given these values, the maximum obtained lever arm is:
𝒙 = 𝑙𝑠 , 𝜃𝑠 , 𝜃𝑐
𝐴𝐷 = 179.5 𝑚𝑚
[ ]𝑇
𝒑 = 𝐷1𝑚𝑎𝑥 , 𝐷2𝑚𝑖𝑛 , 𝐻𝑚𝑎𝑥 , 𝐿𝑚𝑎𝑥 , 𝜃𝑚𝑎𝑥 , 𝑥𝐴 , 𝑦𝐴 , 𝑥𝑃 , 𝑑𝐵 The results show that the second, third and fourth constraints are
active. Therefore, a variation in the limit values of these constraints will
3.4.1.5. Results of the geometrical layout optimization. The parameters change the solution, as can be seen in Fig. 5, which presents isocontours
(xA , yA , xP ,dB ) were measured for a typically developed adult wearing of the objective function and the constraints.
the boot on which the struts were attached. Table 4 shows the numerical The graph presented in Fig. 6 shows how the solution varies when
values for the previously mentioned parameters and the chosen values the maximum strut length parameter, Lmax [mm], is varied. Up to a cer-
for the other design parameter. tain limit, the lever arm length increases linearly with the increase in
The fmincon Matlab function (Mathworks, Natrick, MA, US) was used Lmax [mm], such that an increase of 1 mm in Lmax [mm] results in an in-
to solve the optimization problem using the sequential quadratic poly- crease of 1.06 mm in the length of the lever arm. However, because of
nomial algorithm. The obtained results are shown in Table 5. other constraints, in particular, the maximum height of the cable rout-
Y. Bougrinat, S. Achiche and M. Raison Mechatronics 64 (2019) 102297
Fig. 8. (A) The design variables for the mass optimization problem were the
thickness, t[mm] and the width, w[mm] of the struts (B) The struts were fixed
on the points O and F. A force of 200 N was applied at the point C at an 11°
angle with the horizontal.
ingly, this relationship is of particular interest given that Hmax [mm] can
be increased for taller people who might need higher push-off torque.
The obtained results presented in the table (5) were used in our de-
sign. The obtained lever arm length was sufficient to deliver the required
augmentation torque. In fact, with a cable transmission efficiency 𝜂 c
Fig. 6. Effect of the maximum strut length parameter on the maximum (discussed further on Section 5.1), to deliver a supportive plantarflexion
lever arm. In this case, the maximum strut length varied while the maximum torque, T[Nm], of 30 Nm, the force, F[N], that needs to be transmitted
height is fixed at 300 mm. by the two cables is:
𝑇 30
𝐹 = = = 283.2 𝑁 (10)
𝜂𝑐 𝐴𝐷 0.59 × 0.1795
This force is lower than the 360 N maximum force that the gearbox
shaft can support.
3.4.2.2. Finite element model. The Finite element analysis tool of Solid-
works software (Dassault Systèmes, France) was used to calculate
the Von-Mises stress for each combination of the design variables.
Fig. 7. Effect of the maximum brace height parameter on the maximum Fig. 8 shows the load and the constraints applied to the struts. A 200 N
lever arm. In this case, the maximum brace height parameter is varied while
force was applied to the strut tip (point C) at an angle of 11° with the
the strut length is fixed to 360 mm in this case.
horizontal. The two struts were considered rigidly connected and fixed
at points O and F. This configuration is conservative, since in reality,
the struts structure can move with the foot under a given load.
ing brace, Hmax [mm], after a certain value of Lmax [mm] the resulting
length of the lever arm remains constant. Specifically, when Lmax [mm] 3.4.2.3. Mass optimization results. The minimization problem was
is above 350 mm, the solution remains near 185 mm. solved using the simulation module of Solidworks. The obtained solu-
The Hmax [mm] parameter exhibits the same general behavior as tion corresponds to 3-mm thick and 10-mm wide struts. The proportion
Lmax [mm] (see Fig. 7), except that the relationship between the increase between the width and the thickness of the struts allows to reduce de-
in lever arm length and the increase in Hmax [mm] is nonlinear. Interest- flection on the direction of the cable pulling force.
Y. Bougrinat, S. Achiche and M. Raison Mechatronics 64 (2019) 102297
While performing these tests, the safety of the device and the user
were insured by a set of hardware and software measures. For the soft-
ware, the applied torque was limited by selecting a torque profile with a
maximum value lower than 50 Nm. In addition, at any moment, the de-
vice can be stopped immediately by pressing a stop button in the graph-
ical user interface.
For the hardware safety measures, mechanical stops were added to
the spool and the spool cover to limit the travel distance of the cables.
Table 6
Human walking trials performed for the exoskeleton evaluation.
Trial Maximum torque applied by WAXO on the ankle joint (Nm) Duration (ms) Onset (%)
T1 14 200 40
T2 14 200 38
T3 11 200 38
T4 11 200 43
Table 7 avoided in WAXO design in order to reduce the mass of the actuation
WAXO parts mass. unit.
Component Location Mass (g) The priority of minimizing the exoskeleton weight is important be-
cause it reduces the effort required by the user to move with the device
Actuation Unit Waist 1050
attached to his body, and thus, the power benefit of the exoskeleton
Electronics box Waist 647
Cable routing brace Shin 290 is more significant in this case compared to a heavier exoskeleton that
Struts (including bolts) Foot 58 provides the same mechanical power. However, in the current design of
Total mass (g) 2045 WAXO, the actuation unit was located on the waist and cables were used
to transmit the force. The cables reduce the efficiency of the actuation
system and thus, WAXO will draw more electrical energy compared to
5.2.3. Data recording another exoskeleton with a more efficient transmission system. Still, we
The EMG signals of the soleus and gastrocnemius (lateralis) muscles adopted this design because increasing the exoskeleton power benefit to
were recorded using wireless surface EMG FreeEMG300 system (BTS, the user and reducing its weight are objectives of higher priority in the
Milan, Italy). These muscles were selected because they are directly current state of research in walking augmentation exoskeletons. Addi-
involved in generating the ankle plantarflexion torque. Self-adhesive tionally, more efficient transmission cables could be used to increase the
pairs of disposable Ag/AgCl surface electrodes with a 10 mm record- efficiency of the transmission system. WAXO did not include a portable
ing diameter (Covidien, Mansfield, MA, USA) were placed following power source, such as a battery to allow for a fully-autonomous system.
the SENIAM (Surface Electromyography for Non-Invasive Assessment of Adding a battery will increase the overall mass of the device, however,
Muscles) guidelines [37]. The recorded EMG signals in each trial were it will not affect the added distal mass because the battery can easily be
band-pass filtered (10–450 Hz, zero-lag 4th order), full-wave rectified, attached to the back of the user, and therefore reduce its effect on the
and smoothed with a low-pass filter (3 Hz, Butterworth zero-lag 2nd or- metabolic cost.
der). The resulting signals were then normalized by the maximum value
obtained in the NO EXO condition.
6.2. Human walking trials
Although oxygen consumption was not measured to determine the
metabolic cost of walking, it is still accepted that muscle activity in-
During the powered plantarflexion phase, the exoskeleton added to
duces a large consumption of metabolic energy [13,38]. Therefore, a
the subject ankle joint positive mechanical power of 15 ± 3 W during
significant reduction of muscle activity is an indication of the potential
trials T1 and T2 , 11 ± 2 W during trial T3 and 11 ± 1 W during trial T4
of WAXO to reduce the walking metabolic cost.
(see Fig. 10). Fig. 11 shows the average parabolic mechanical power
The current and the velocity of the motor were recorded using the
profile applied on trial T2 .Table 8 shows the results of muscle activity
motor sensors and driver. The recorded data were used to calculate the
obtained from each trial in both conditions, NO EXO and EXO ON. The
positive mechanical power delivered by the exoskeleton during walking
activity of the gastrocnemius and soleus muscles was reduced on the
trials on the EXO ON condition.
EXO ON condition compared to the NO EXO condition in all the trials.
Table 8
Results of the human walking trials. Rela-
tive reduction in EMG activity on EXO ON con-
dition compared to EXO OFF condition. Neg-
ative values indicate an increase in the EMG
activity.
T1 16 18
T2 37 44
T3 6 43
T4 11 19
Fig. 12. Results of trial T2 . The activity of the gastrocnemius and soleus muscles was reduced by 44% and 37% respectively on the EXO ON condition compared
to the NO EXO condition. The signals were normalized by the maximum EMG value obtained during the NO EXO condition of each trial. The Blue bars show the
average values over the gait cycles; error bars are standard deviations.
Y. Bougrinat, S. Achiche and M. Raison Mechatronics 64 (2019) 102297
approach was already demonstrated by Walsh and al. in their exosuit By substituting (A.3) in (A.1), we obtain the objective function as
[43]. Our design assumptions were mainly based on the results from the follows:
literature, however, modeling [44,45] and multi-criteria optimization ( ( ( ) ))
𝑙s cos 𝜃𝑠 − 𝑥𝐴
methods [46–48] can be combined to better predict and optimize the 𝑓 (𝒙, 𝒑) = 𝐴𝐶 𝑠𝑖𝑛 𝜃𝑐 + 𝑎𝑟𝑐𝑜𝑠 (A.3)
exoskeleton performance and interaction with the user. In future work, 𝐴𝐶
an extensive evaluation study for WAXO will be performed with a larger Where:
group of subjects to confirm the current results. This evaluation will also √
( ( ) )2 ( ( ) )2
include measurement of oxygen consumption to estimate the metabolic 𝐴𝐶 = 𝑙s cos 𝜃𝑠 − 𝑥𝐴 + 𝑙s 𝑠𝑖𝑛 𝜃𝑠 − 𝑦𝐴 (A.4)
expenditure as well as kinematic measurements to compare the gait of
the subjects when walking with and without WAXO. As for now, this is References
one of the limitations of our work.
Finally, the minimalist design approach we applied here will be ex- [1] Beekley MD, et al. Effects of heavy load carriage during constant-speed, simulated,
tended to other exoskeleton applications targeted to different popula- road marching. Mil Med 2007;172(6):592–5.
[2] Park K, et al. Effect of load carriage on gait due to firefighting air bottle configura-
tions such as those suffering from musculoskeletal diseases and muscu- tion. Ergonomics 2010;53(7):882–91.
lar dystrophies [49–51]. [3] Dollar AM, Herr H. Lower extremity exoskeletons and active orthoses: challenges
and state-of-the-art. IEEE Trans Robot 2008;24(1):144–58.
[4] Young AJ, Ferris DP. State of the art and future directions for lower limb robotic
7. Conclusion exoskeletons. IEEE Trans Neural Syst Rehabilit Eng 2017;25(2):171–82.
[5] Chu A, Kazerooni H, Zoss A. On the biomimetic design of the berkeley lower extrem-
The objective of this study was to develop an ankle exoskeleton with ity exoskeleton (BLEEX). In: Robotics and Automation, 2005. ICRA 2005. Proceed-
ings of the 2005 IEEE International Conference on. IEEE; 2005.
a minimum added distal mass compared to existing autonomous pow- [6] Part I Cao H, et al. Design and control of a hydraulic-actuated leg exoskeleton for
ered ankle exoskeletons while it can provide at least 30 Nm of assistive load-carrying augmentation. In: Liu H, et al., editors. Intelligent robotics and ap-
plantarflexion torque. The optimized design reduced the minimal dis- plications: third international conference, ICIRA 2010, Shanghai, China, November
10-12, 2010. Proceedings Springer Berlin Heidelberg; 2010. p. 590–9. Part I.
tal mass to 348 g. This value is the lowest one compared to existing
[7] Marcheschi S, et al. Body extender: whole body exoskeleton for human power aug-
motorized ankle exoskeletons. This shows that it is possible to design mentation Robotics and Automation (ICRA), 2011 IEEE international conference on.
powerful exoskeletons with minimal weight, which becomes a feature IEEE; 2011.
that should be considered amongst the criteria when designing augmen- [8] Yu S, et al. Design of an under-actuated exoskeleton system for walking assist while
load carrying. Adv Robot 2012;26(5–6):561–80.
tation exoskeletons. [9] Walsh CJ, Endo K, Herr H. A quasi-passive leg exoskeleton for load-carrying aug-
Tests with a typically developed human subject showed that the gas- mentation. Int J Humanoid Robot 2007;4(3):487–506.
trocnemius muscle activity was reduced by 44% when walking with the [10] Seo K, et al. Fully autonomous hip exoskeleton saves metabolic cost of walking
Robotics and Automation (ICRA), 2016 IEEE international conference on. IEEE;
exoskeleton compared to normal walking. This result shows that the pro- 2016.
posed exoskeleton has a potential in reducing the energy spent during [11] Kusuda Y. In quest of mobility–Honda to develop walking assist devices. Ind Robot
human walking. The perspective of this work is to develop a bilateral 2009;36(6):537–9.
[12] van Dijk W, Meijneke C, van der Kooij H. Evaluation of the Achilles Ankle exoskele-
exoskeleton that supports both joints of the ankle using the same design ton. IEEE Trans Neural Syst Rehabilit Eng 2016.
approach will be developed. Additionally, tests on a larger sample of [13] Collins SH, Wiggin MB, Sawicki GS. Reducing the energy cost of human walking
subjects could be performed to examine the effect of this exoskeleton using an unpowered exoskeleton. Nature 2015;522(7555):212–15.
[14] Mooney LM, Rouse EJ, Herr HM. Autonomous exoskeleton reduces metabolic cost
on the metabolic cost during walking. of human walking during load carriage. J Neuroeng Rehabil 2014;11(1):80.
[15] Ferris DP, Sawicki GS, Daley MA. A physiologist’s perspective on robotic exoskele-
Role of the funding source tons for human locomotion. Int J Humanoid Robot 2007;4(3):507–28.
[16] Mooney LM, Herr HM. Biomechanical walking mechanisms underlying the
metabolic reduction caused by an autonomous exoskeleton. J Neuroeng Rehabil
The funding organization was not involved in any of these: study 2016;13(1):4.
design; the collection, analysis and interpretation of data; the writing of [17] Malcolm P, et al. A simple exoskeleton that assists plantarflexion can reduce the
the report; and in the decision to submit the article for publication. metabolic cost of human walking. PLoS ONE 2013;8(2):e56137.
[18] Winter D. Biomechanics and motor control of human gait: normal, elderly and patho-
logical. Waterloo: Waterloo Biomechanics; 1991.
Declaration of Competing Interest [19] Wiggin MB, Sawicki GS, Collins SH. An exoskeleton using controlled energy storage
and release to aid ankle propulsion. Rehabilitation robotics (ICORR), 2011 IEEE
international conference on. IEEE; 2011.
None [20] Park Y-L, et al. Bio-inspired active soft orthotic device for ankle foot pathologies
Intelligent Robots and Systems (IROS), 2011 IEEE/RSJ international conference on.
Appendix A. Objective Function Formulation IEEE; 2011.
[21] Gordon KE, Sawicki GS, Ferris DP. Mechanical performance of artificial pneumatic
muscles to power an ankle–foot orthosis. J Biomech 2006;39(10):1832–41.
The goal of the layout optimization is to maximize the artificial lever [22] Sawicki GS, Ferris DP. A pneumatically powered knee-ankle-foot orthosis (KAFO)
arm, 𝐴𝐷[mm], of the augmentation torque when the foot is on the with myoelectric activation and inhibition. J Neuroeng Rehabil 2009;6(1):23.
ground with no plantarflexion. This lever arm, 𝐴𝐷 [mm], is the per- [23] Ferris DP, Czerniecki JM, Hannaford B. An ankle-foot orthosis powered by artificial
pneumatic muscles. J Appl Biomech 2005;21(2):189–97.
pendicular distance between the ankle joint, A, and the line of action of [24] Veale AJ, Xie SQ. Towards compliant and wearable robotic orthoses: a review of
the force exerted by the cable, (CP). Formally, the distance, 𝐴𝐷[mm] is current and emerging actuator technologies. Med Eng Phys 2016;38(4):317–25.
given by: [25] Meijneke C, van Dijk W, van der Kooij H. Achilles: an autonomous lightweight ankle
exoskeleton to provide push-off power. Biomedical robotics and biomechatronics
( )
𝐴𝐷 = 𝐴𝐶 𝑠𝑖𝑛∠𝐷𝐶𝐴 = 𝐴𝐶 𝑠𝑖𝑛 𝜃𝑐 + 𝜃𝑠 + ∠𝑂𝐶𝐴 (A.1) (2014 5th IEEE RAS & EMBS international conference on. IEEE; 2014.
[26] Asbeck AT, et al. Biologically-inspired soft exosuit. in rehabilitation robotics
∠OCA could be written as follows: (ICORR), 2013 IEEE international conference on. IEEE; 2013.
( ) [27] Asbeck AT, et al. A biologically inspired soft exosuit for walking assistance. Int J
𝑥 − 𝑥𝐴
∠𝑂𝐶𝐴 = 𝑎𝑟𝑐𝑜𝑠 𝐶 − 𝜃𝑠 (A.2) Rob Res 2015;34(6):744–62.
[28] Asbeck AT, et al. Stronger, smarter, softer: next-generation wearable robots. IEEE
𝐴𝐶
Robot Automat Mag 2014;21(4):22–33.
where arcos returns values in the first quadrant. [29] Panizzolo FA, et al. A biologically-inspired multi-joint soft exosuit that can reduce
That is: the energy cost of loaded walking. J Neuroeng Rehabil 2016;13(1):43.
[30] Yandell MB, Popov D, Quinlivan B, Walsh CJ, Zelik KE. Systematic evaluation of
⎛ ( ) ⎞ human-exosuit physical interfaces. Dynamic Walking 2016.
⎜ 𝑙s cos 𝜃𝑠 − 𝑥𝐴 ⎟ [31] Yandell MB, et al. Physical interface dynamics alter how robotic exosuits augment
∠𝑂𝐶𝐴 = 𝑎𝑟𝑐𝑜𝑠⎜ √ ⎟ − 𝜃𝑠 (A.3)
⎜ ( ( ) ) ( ( ) ) human movement: implications for optimizing wearable assistive devices. J Neuro-
𝑙s cos 𝜃𝑠 − 𝑥𝐴 + 𝑙s 𝑠𝑖𝑛 𝜃𝑠 − 𝑦𝐴 ⎟⎠
2 2
⎝ eng Rehabil 2017;14(1):40.
Y. Bougrinat, S. Achiche and M. Raison Mechatronics 64 (2019) 102297
[32] Mooney LM, Rouse EJ, Herr HM. Autonomous exoskeleton reduces metabolic cost [50] Yazji M, et al. Are the mediolateral joint forces in the lower limbs different between
of human walking. J Neuroeng Rehabil 2014;11(1):151. scoliotic and healthy subjects during gait? Scoliosis 2015;10(2):S3.
[33] Browning RC, et al. The effects of adding mass to the legs on the energetics and [51] Hart R, et al. Impact of exercise-induced fatigue on the strength, postural con-
biomechanics of walking. Med Sci Sports Exerc 2007;39(3):515. trol, and gait of children with a neuromuscular disease. Am J Phys Med Rehabilit
[34] Bovi G, et al. A multiple-task gait analysis approach: kinematic, kinetic and EMG 2014;93(8):649–55.
reference data for healthy young and adult subjects. Gait Posture 2011;33(1):6–13.
[35] Witte KA, et al. Design of two lightweight, high-bandwidth torque-controlled ankle Yacine Bougrinat is a master candidate in mechanical engi-
exoskeletons Robotics and automation (ICRA), 2015 IEEE international conference neering at the École Polytechnique de Montréal. He is also
on. IEEE; 2015. member of the Rehabilitation Engineering Chair Applied to
[36] Galle S, et al. Reducing the metabolic cost of walking with an ankle exoskeleton: Pediatrics at the CRME–Ste-Justine University Hospital Cen-
interaction between actuation timing and power. J Neuroeng Rehabil 2017;14(1):35. ter, Montreal. He holds a state engineer degree in mechanical
[37] Hermens HJ, et al. European recommendations for surface electromyography. Roess- engineering from École Nationale Polytechnique in Algeria.
ingh Res Dev 1999;8(2):13–54. His research interests include wearable robotics, human aug-
[38] Blake OM, Wakeling JM. Estimating changes in metabolic power from EMG. mentation and mechatronics systems design and integration.
Springerplus 2013;2(1):229.
[39] Zizoua C, et al. Detecting muscle contractions using strain gauges. Electron Lett
2016;52(22):1836–8.
[40] Duivenvoorden A, et al. Sensor fusion in upper limb area networks: a survey Global
information infrastructure and networking symposium (GIIS), 2017. IEEE; 2017.
[41] Geoffroy P, et al. A two-stage suboptimal approximation for variable compliance Sofiane Achiche received the M.Sc.A. and Ph.D. degrees from
and torque control Control conference (ECC), 2014 European. IEEE; 2014. the École Polytechnique de Montréal, Canada. He is currently
[42] Geoffroy P, et al.. From inverse kinematics to optimal control, in advances in robot a Full Professor at the Mechanical Engineering Department,
kinematics. Springer; 2014. p. 409–18. Design of Machinery Section, École Polytechnique de Mon-
[43] Asbeck AT, et al. Multi-joint soft exosuit for gait assistance Robotics and automation tréal. His research interests focus upon evolutionary compu-
(ICRA), 2015 IEEE international conference on. IEEE; 2015. tational intelligence applied to engineering problems, such as
[44] Raison M. On the quantification of joint and muscle efforts in the human body during condition monitoring. Furthermore, he works in the field of
motion. UCL-Université Catholique de Louvain; 2009. design support of mechatronics and applied mechatronics in
[45] Hernandez S, et al. Refinement of exoskeleton design using multibody modeling: an aerospace and biomedical engineering.
overview CCToMM mechanisms, machines, and mechatronics symposium; 2015.
[46] Leroux M, et al. Design guidelines for shoulder design of an anthropomorphic robotic
arm. In: DS 87-4 proceedings of the 21st international conference on engineering
design (ICED 17) Vol 4. Design Methods and Tools; 2017. p. 21–5. 08. 2017. Maxime Raison received the degree in electrical engineering
[47] Mohebbi A, et al. Trends in concurrent, multi-criteria and optimal design of mecha- and the Ph.D. degree in mechanical engineering from the Uni-
tronic systems: a review. In: Innovative design and manufacturing (ICIDM), Proceed- versity catholique de Louvain, Belgium. He is currently a Pro-
ings of the 2014 international conference on. IEEE; 2014. fessor in mechanical and biomedical engineering with École
[48] Mohebbi A, Achiche S, Baron L. Mechatronic multicriteria profile (MMP) for concep- Polytechnique de Montréal, Canada. He is responsible for the
tual design of a robotic visual servoing system. ASME 2014 12th biennial conference Rehabilitation Engineering Chair Applied to Pediatrics at the
on engineering systems design and analysis. American Society of Mechanical Engi- CRME–Ste-Justine University Hospital Center, Montréal. His
neers; 2014. main research interests are biomechanics, multibody dynamic
[49] Samadi B, et al. Custom sizing of lower limb exoskeleton actuators using gait dy- modeling, and assistive robotics.
namic modelling of children with cerebral palsy. Comput Methods Biomech Biomed
Engin 2016;19(14):1519–24.