2198947-100
Revision 0
do not duplicate
TABLE OF CONTENTS
REVISION HISTORY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
LIST OF EFFECTIVE PAGES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
i
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
Blank page
ii
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
iii
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
iv
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
REVISION HISTORY
Title page 0
i thru vi 0
v
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
Blank page
vi
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
CHAPTER CONVENTIONS
The following diagrams will be use in this chapter to illustrate jumper settings.
1 2 3 1
2
2 AND 3 ARE COVERED
3
1 IS SET TO ON
SECTION 2
ACCESSING THE VDP BOARDS
Use a Phillips screwdriver to remove the 10 screws (with tooth lock and flat washers) (1) (2)
(3) from the VDP front panel.
8–1
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–1
DLX VDP BACKPLANE DIAGRAM MONO–PLAN
GE B1 OP. GREY
GE B2 OP. GREY
GE B1 OP. BLUE
GE B2 OP. BLUE
PANEL
VDP
(BIP)
(BIP)
(BIP)
(BIP)
8–2
REV 0
ILLUSTRATION 8–2
GE Medical Systems
VDP
MVME 167
+ IVP I/O
PANEL
W525
J24
J23
J3 W507
DLX VDP BACKPLANE DIAGRAM BI–PLAN
J2 W511 J2
W504
W521
J7 W522
J1 W514 J27 W517 W408
Y J1
J1
W459 J2 W520
J1 W515 J28 J2
W527
J3 W460 J3 W528
J4 W482 VD W495
J1 W479
W496
8–3
J2 W480
W481 TRIG W424
W534 GE B1 OP. BLUE
GE–B1 BLUE
GE–B1 GREY
W533 GE B1 OP. GREY
W536
Board
GE–B2 BLUE GE B2 OP. BLUE
Optical
J18 W487 J6 GE–B2 GREY W535 GE B2 OP. GREY
J9 W406
W445
J1
J16 W502
W503
W505
J19 W501
J15 W493 J7
W509
W510
J14 W494 J7
W419
Blank page
8–4
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
D Remove front panel on the DLX1 A1 VDP. See Job Card D/R020 A.
SECTION 5
TASK DESCRIPTION
D Refer to Illustrations 8–3 and AUCUN LIEN for the DLX VDP board configuration.
8-5
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–3
DLX1A1 VDP DRAWER CONFIGURATION (2122364)
XDG 1 (OPTION)
XDG 3 (OPTION)
XDG 2 (OPTION)
ITRL (OPTION)
DVA (OPTION)
MVME 167
VAD 3 F
XDG 0
VAPZ
GEIM
CP4
CPI
32 31 30 29 28 27 26 25 24 23 22 21 20 19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1
VDP RACK 2122364: Remove the terminator board when an option board is installed.
8-6
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
D Open the DLX cabinet. Remove the front panel on the VDP (see Job Card D/R020 A).
SECTION 5
PROCEDURE
Verify that switches and jumpers are installed as shown in Illustration 8–4 (factory installed).
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for board part number, and for the
fuse F1 part number.
8-7
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–4
IVP I/O BOARD
1
STRAP ON
1 AND 2
1
1
STRAP ON
1 AND 2
MVME 167
BOARD
1 2 3 4 5 6 7 8
B = BIPLANE
M = MONOPLANE
8
M
B
8-8
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
D Verify that switches and jumpers are installed as shown in Illustration 8–5 (factory installed).
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for board part number.
8-9
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–5
DX1A1A1 MVME 167 BOARD CONFIGURATION
A1 A32 A1 A32
B1 P1 B32 B1 P2 B32
C1 C32 C1 C32 J7
1 1
J6
3 3
39 29
40 28
1 XU4
SKT
6 18
7 17
39 29
40 28
1 XU3
6 SKT 18
7 17 MEZZANINE BOARD
39 29
40 28
1
XU2
6 18
7 17
39 29
40 28
1
XU1
6 18
7 17
2 J1 16
2
Ä Ä
J2
DS3 1
Ä Ä
DS1
DS2
DS4
1 60 J4 2 69 J5 2
19
20 J3 2 S1 S2 59 1 59 1
2
J2 J6 J7
J1 1 1
2 16
1 15 3 3
8-10
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
D Open DLX VDP drawer. See Job Cards D/R020 A and D/R021 A.
D Remove DLX terminator board.
SECTION 5
PROCEDURE
D The DLX “dummy” terminator board, which has no jumpers, is shown in Illustration 8–5.
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for board part number.
8-11
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–6
DLX TERMINATOR BOARD
8-12
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
D Verify that switches and jumpers are installed as shown in Illustration 8–7 (factory installed).
Note: There are two types of GEIM board, one for Monoplane and one for Biplane, that can only be
distinguished by their part number (physically they look exactly the same). Refer to the illustrated parts
list, in the Section “Renewal Parts”, for these board numbers.
8-13
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–7
DLX GEIM BOARD CONFIGURATION (ADVANTX–E)
4 FIBER OPTIC
CABLE CONNECTIONS
1 2 3 1 2 3
8-14
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
D Verify that switches and jumpers are installed as shown in Illustration 8–8 (factory installed).
Note: Refer to the illustrated parts list, in the Section “Renewal Parts”, for board part number.
8-15
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–8
DLX1A1A15 XDG1 BOARD CONFIGURATION
8-16
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
D Verify that switches and jumpers are installed as shown in Illustration 8–9 (factory installed).
Note: Refer to the illustrated parts list, in the Section “Renewal Parts”, for board number.
8-17
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–9
DLX1A1A17 XDG0 BOARD CONFIGURATION
8-18
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
D Verify that switches and jumpers are installed as shown in Illustration 8–10 (factory installed).
Note: Refer to the illustrated parts list, in the Section “Renewal Parts”, for board part number.
8-19
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–10
DLX1A1A19 VAPZ BOARD CONFIGURATION
NO JUMPER
8-20
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
D Open DLX1A1 VDP drawer. See Job Cards D/R020 A and D/R023–39 A.
D Pull to remove DLX1A1A20 ARRAY board.
SECTION 5
PROCEDURE
D Verify that switches and jumpers are installed as shown in Illustration 8–11 (factory installed).
Note: Refer to the illustrated parts list, in the Section “Renewal Parts”, for board part number.
8-21
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–11
DLX1A1A20 ARRAY BOARD CONFIGURATION
8-22
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
D Verify that switches and jumpers are installed as shown in Illustration 8–12 (factory installed).
Note: Refer to the illustrated parts list, in the Section “Renewal Parts”, for board part number.
8-23
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–12
DLX IMAGE MEM A BOARD CONFIGURATION
FACTORY SET
X6 (1024)
X5 (512)
ILLUSTRATION 8–13
DLX IMAGE MEM B BOARD CONFIGURATION
FACTORY SET
X1 (1024)
X2 (512)
* For option Bolus chasing , spin and stepping you must have Memb with Jumper in 1024.
8-24
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
D Verify that switches and jumpers are installed as shown in Illustration 8–14 (factory installed).
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for board part number.
8-25
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–14
DLX1A1A23 CPI BOARD CONFIGURATION
1
2
3
8-26
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
Time: 10 mn Personnel: 1
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for board part number.
8-27
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–15
DLX1A1A25 CP4 BOARD CONFIGURATION
8-28
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
D Verify that switches and jumpers are installed as shown in Illustration 8–16 (factory installed).
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for board part number.
8-29
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–16
DLX1A1A26 ITRL BOARD CONFIGURATION
1–2 = RS232
2–3 = RS422
STROBE PERIOD
CONFIGURATION
OSCILLATOR
VALIDATION 800ns
1Ms
1.5Ms
2Ms
2.5Ms
8-30
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS REQUIRED
Not applicable.
SECTION 3
SAFETY, PRECAUTIONS TO TAKE
Set the DLX1A51A2 AC POWER ENABLE switch on DLX1 cabinet to OFF (position ”0”).
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE – SEE ILLUSTRATION 1
5.1 Verify that switches and jumpers are installed as shown in Illustration 8–17 (factory installed)
Note: Refer to the illustrated parts list, in the Section “Renewal Parts”, for board number.
8-31
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–17
DLX1A1A11 XDG2 BOARD CONFIGURATION
8-32
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS REQUIRED
Not applicable.
SECTION 3
SAFETY, PRECAUTIONS TO TAKE
Set the DLX1A51A2 AC POWER ENABLE switch on DLX1 cabinet to OFF (position ”0”).
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
5.1 Verify that switches and jumpers are installed as shown in Illustration 8–18 (factory installed)
Note: Refer to the illustrated parts list, in the Section “Renewal Parts”, for board number.
8-33
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–18
DLX1A1A13 XDG3 BOARD CONFIGURATION
8-34
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
D Verify that switches and jumpers are installed as shown in Illustration 8–16 (factory installed).
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for board number.
8-35
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–19
DLX1A1A31 VAD3 BOARD CONFIGURATION (FRONTAL)
X1
X2
X6
Module MA3
J2
Select 2.56V
8-36
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS REQUIRED
Not applicable.
SECTION 3
SAFETY, PRECAUTIONS TO TAKE
Set the DX1A51A2 AC POWER ENABLE switch on the DLX1 cabinet to OFF (position ”0”).
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
5.1 Verify that switches and jumpers are installed as shown in Illustration 8–20 (factory installed)
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for board number.
8-37
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–20
DX1A1A29 VAD3 BOARD (LATERAL) CONFIGURATION
X1
X2
X6
Module MA3
J2
Select 2.56V
8-38
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS REQUIRED
Not applicable.
SECTION 3
SAFETY, PRECAUTIONS TO TAKE
Set DLX1A51A2 AC POWER ENABLE switch on DLX1 cabinet to OFF (position ”0”).
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
5.1 There are no jumpers or switches to configure on the DVA board. See Illustration 8–21.
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for board number.
8-39
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–21
DLX1A1A27 DVA BOARD CONFIGURATION
Identification number
8-40
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Not applicable.
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
D The SWITCH should be in the ON position when the cabinet is powered on. The VDP switch powers
ON/OFF the VDP.
Note: The Sun and flat panel power are supplied by the transformer (or UPS when installed) and are not
managed by this switch.
8-41
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–22
DLX VDP POWER SWITCH
ON OFF
8-42
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Not applicable.
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
D The SWITCH should be in the ON position when the cabinet is powered on.
Note: The Sun and flat panel power are supplied by the transformer (or UPS when installed) and are not
managed by this switch.
8-43
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–23
DLX POWER STRIP BREAKER
OUTLET STRIP
BREAKER POSITIONS
0 0
| |
ON OFF
8-44
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
D Remove the VDP front cover to access the VDP disk drive rack (see Job Card D/R020 A).
SECTION 5
DETERMINE DISK MODEL
There are two different models of Image Disk Drive for DLX:
8-45
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 6
SEAGATE : BARRACUDA ST32550N JUMPER CONFIGURATION
DC POWER
4 3 2 1
DC POWER
PIN
1 +12V
2 +12V Return
3 +5V Return
4 +5V
D Verify that switches and jumpers are installed as shown in the Illustrations 8–25, 8–26, 8–27.
Note: If you replace Disk 0 be sure to install the jumpers on J1 (use the shunts from the removed Disk 0 and
install it on the new Disk 0 if necessary).
Note: The peripheral order should always be: 0, 1, 2, 3 with the disk’s configuration and SCSI number
corresponding to the disk number. When removing a disk, reconfigure the remaining disks so that the
numbers are consecutive. If you remove Disk 0, Disk 1 should be configured as Disk 0, Disk 2 as Disk
1, and so on. Change the SCSI number, if necessary, so that it matches the new disk configuration (this
is modified on jumper J4).
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for disk drive part numbers.
8-46
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–25
J4 CONFIGURATION
J4
FRONT VIEW WHEN DISK IS INSTALLED
J4 SCSI ID
DISK 0 Pin 1
DISK 1 Pin 1
DISK 2 Pin 1
DISK 3 Pin 1
8-47
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–26
J1 CONFIGURATION
J1 CONFIG
DISK 0
DISK 1
For Disk 0:
D Enable SCSI Term DISK 2
D Term. Power from drive
DISK 3
ILLUSTRATION 8–27
J2 CONFIGURATION
NO STRAP ON J2
NO STRAP = DISABLED
NO STRAP = DISABLED
NO STRAP = DISABLED
8-48
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 7
FUJITSU : M2952SYU JUMPER CONFIGURATION
DC POWER
1 2 3 4
DC POWER
PIN
1 +12V
2 +12V Return
3 +5V Return
4 +5V
D Verify that switches and jumpers are installed as shown in the Illustrations 8–29, 8–30, 8–31
Note: The peripheral order should always be: 0, 1, 2, 3 with the disk’s configuration and SCSI number
corresponding to the disk number. When removing a disk, reconfigure the remaining disks so that the
numbers are consecutive. If you remove Disk 0, Disk 1 should be configured as Disk 0, Disk 2 as Disk
1, and so on.
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for disk drive part numbers.
8-49
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–29
CNF7 CONFIGURATION
CN7
FRONT VIEW WHEN DISK IS INSTALLED
8-50
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–30
CNF7 PIN ASSIGNMENT
CN7
8-51
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–31
CNF6 CONFIGURATION
CN6
8-52
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS TO TAKE
When the OK message appears, turn off the power to the station.
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
D Disconnect the Sun station, and remove it from the cabinet. See Job Card D/R 51 A.
D Open the Sun station to access the Ethernet board.
8-53
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 5
PROCEDURE
There are no jumpers to be set on the Ethernet board. The board is shown in Illustration 8–32.
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for board part number.
ILLUSTRATION 8–32
ETHERNET BOARD
PROM
810P01160
Connector P2
F1
Fuse
8-54
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
D Open the DLX cabinet, and remove the VDP front panel. See Job Cards D/R 020A and D/R 75 A.
SECTION 5
PROCEDURE
D Verify that the switches and jumpers are installed as shown in Illustration 8–33 (factory installed).
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for board part number.
8-55
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–33
DLX1A3A1 CSS BOARD CONFIGURATION
J1 Power
Connector
Transfo CSS
J4 Power
Connector
J3 Linelock
Connector
8-56
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
SECTION 1
SUPPLIES
D Paralan board.
SECTION 2
TOOLS
D Not applicable.
SECTION 3
SAFETY PRECAUTIONS
D Set the DLX VDP/options AC power switch on the DLX cabinet to OFF.
ON / OFF
SECTION 4
PREREQUISITES
SECTION 5
PROCEDURE
Note: Refer to the illustrated parts list, in the chapter “Renewal Parts”, for board part number.
8-57
GE Medical Systems ADVANTX DLX CENTRAL LISTING
REV 0 sm 2198947-100
ILLUSTRATION 8–34
PARALAN BOARD CONFIGURATION
RESISTORS RN1, RN2, AND RN3 ARE FACTORY INSTALLED – DO NOT REMOVE.
MOLEX
YELLOW LED (SIGNAL)
RN3
JUMPERS ARE FACTORY SET.
SINGLE RN2
CONNECTOR
RN1
MOLEX
DIFFERENTIAL CONNECTOR
POWER CONNECTOR
8-58