Professional Documents
Culture Documents
Introduction To PET/CT: Jon A. Anderson
Introduction To PET/CT: Jon A. Anderson
Jon A. Anderson
Department of Radiology
The University of Texas Southwestern Medical Center at Dallas
γ2
Detector Ring e+
γ1
Angle (θ)
θ
Position (t)
Count on an LOR gets mapped to corresponding place in sinogram
J. A. Anderson 2/18/04
PET Fundamentals: Real Case
RANDOMS
with false
γ2
LOR γ2
suppress
with small
coincidence
γ1 time and
collimation γ1
TRUES
with correct
LOR SCATTERS
γ2
γ1 ’ with misplaced
scatter
γ1 LOR
LOR = line of response suppress with energy
resolution and collimation
J. A. Anderson 2/18/04
What Influences These
Contributions?
Trues
↑ as radio-nuclide concentration ↑ (the good stuff)
↓ as patient size ↑ (absorption and scatter effects)
Randoms
↑↑ as count rate ↑ (varies as square of count rate)
Effects dominate image noise at high injected activities
Reduce by faster electronics, faster crystals
Scatters
↓ with collimation
(about 15% for 2D PET and 50% for 3D PET)
Reduce with collimation
(energy selection not efficient in PET)
J. A. Anderson 2/18/04
2D and 3D Flavors of PET (Similar
to single-row vs multi-detector CT)
2D 3D
collimators
2D
Sensitivity
100
80
60
40
20 d = 35 cm
0 → ACF ≈ 28
0 10 20 30 40 50 60
Path Length [cm]
J. A. Anderson 2/18/04
One Last Concept: Specific
Uptake Value
Measured Activity Concentration [Bq/ml]
SUV =
Injected Activity (Bq)
Pt. Mass/ (1 g/ml)
concentration if activity
uniformly distributed in
the body
SUV is used as an index to determine if a hotspot is significant
Its use depends on
STANDARDIZED UPTAKE TIMES
CALIBRATED SCANNER
J. A. Anderson 2/18/04 NO ARTIFACTS THAT CHANGE COUNTS
PET and PET/CT Imaging
- uses rotating isotopic sources for
acquiring transmission image
- 70 % E/30% T time split
- typically, 15 cm axial FOV
- 2D or 3D acquisitions
- newer crystals allow lower noise,
shorter acquisition times
Conventional PET
J. A. Anderson 2/18/04
PET Alone
Esophageal CA
constricting the esophagus
J. A. Anderson 2/18/04
Current Implementations
CPS
(marketed by CTI and Siemens)
Reveal RT/Reveal XVI
Biograph LSO/Biograph Sensation 16
GE
Discovery LS
Discovery ST
J. A. Anderson 2/18/04
PET Detector Crystal Performance
Compared to NaI(Tl)
Detector Performance
Attenuation Length
120%
* Light Output
NaI(Tl)
100%
80% Energy Resolution
60% *
40% * = a winner
20% *
0% *
BGO LSO GSO
Detector Material
J. A. Anderson 2/18/04
Issues Specific to PET/CT
• Proper mapping of attenuation coefficients
from 70 keV to 511 keV
• Recognition of PET artifacts due to
– Attenuation correction model failures
– Patient motion
– CT artifacts
– CT truncation (mismatch between PET,CT
FOV
• QA and test procedures
J. A. Anderson 2/18/04
Converting µ70keV →µ511keV
10.00
Soft Tissue Total
Soft Tissue PE
Adipose Tissue Total
Adipose Tissue PE
1.00 CT Energy
70 keV Bone Total
µ/ρ [cm /g]
Bone PE
2
PET Energy
511 keV
0.10
Photoelectric
Component
Compton Component Accounts
of Attenuation
for Almost All of Attenuation
Coefficient
Coefficient
0.01
10 100 1000
Energy [keV]
J. A. Anderson 2/18/04
Conclusions Regarding the
Relation of µ70keV to µ511keV
1) The attenuation coefficient at 511 keV can
be obtained from that at 70 keV by simple
multiplication by a constant, independent
of tissue type, for all tissues having
attenuation dominated by Compton scatter
at CT energies (≈ 70 keV).
0.16
plot
0.14
0.12 Hybrid
Segmentation
0.10 threshold=200 HU
0.08
PET/CT<200 = 0.5
PET/CT>200 = 0.41
0.06
Mixture
0.04
threshold = 0 HU
0.02 µH2O,80 keV = 0.184 cm-1
µbone,80 keV = 0.428 cm-1
0.00 µH2O,511 = 0.096 cm-1
-1000 -500 0 500 1000 µbone,511 keV = 0.172 cm-1
CT Number (HU)
J. A. Anderson 2/18/04
Principal Artifacts Specific to
PET/CT
• Artifacts due to fundamental failure of the
attenuation correction model
• Artifacts due to patient motion between the
CT and PET scans
• Truncation artifacts due to patient extending
outside the CT field of view
J. A. Anderson 2/18/04
Failure of the Attenuation
Correction Model
• There is no perfect way to tell from CT number
what material we’ve got or how much, but
only what the total attenuation is at CT energies
• Materials not in the model (metal, iodine, barium)
can scale differently (CT → PET) and give the
wrong attenuation correction
• Incorrect PET attenuation values →
wrong corrections →
incorrect activities, too high or too low
The size of the ACF (factors of 20, 50, more) means an error
here can cause a big artifact!
J. A. Anderson 2/18/04
The Origin of Contrast and Metal
Artifacts in PET/CT
Mass Attenuation Coefficients for Tissue and Contrast Materials
10.00
CT Energy Soft Tissue
70 keV Bone
Barium Sulfate
Sodium Iodide
1.00
µ/ρ [cm /g]
2
PET Energy
511 keV
0.10
J. A. Anderson 2/18/04
Artifacts from Oral Contrast
CT AC
Non-
AC
J. A. Anderson 2/18/04
Artifacts from Metal: Orthopedic
Hardware
Intense activity shown on
PET/CT (SUV = 6) is associated
with metallic hardware having
CT# > 3000HU
J. A. Anderson 2/18/04
Artifacts from Metal: Orthopedic
Hardware, continued
J. A. Anderson 2/18/04
Artifacts from Patient Motion
J. A. Anderson 2/18/04
Truncation Artifacts
Arise when objects extend outside the field
of view of the CT, but are in the PET field of
view
Rim artifact due CT FOV
to out-of-field
object on CT
will generate rim
artifact on PET;
no attenuation
correction for
out-of-field
PET FOV
anatomy.
J. A. Anderson 2/18/04
Quality Assurance Program:
• QA for PET --
– Daily check per mfg with line or volume source to assure
that system performance is not drifting
– Calibration check to assure proper SUV reports
– Periodic Physics check per standard PET practice
• QA for CT --
– Daily air cals and image quality check per mfg
– Period Physics check per standard CT practice
• Registration check -- check for proper registration of
PET and CT images on periodic basis
J. A. Anderson 2/18/04
Examples: QA Procedures
CPS (Siemens-CTI):
Obtained with volume
phantom (1.5 mCi
68Ge); summary report
presented and logged;
sinogram viewer for
details.
J. A. Anderson 2/18/04
Gantry Offset Correction/Check:
Phantom Geometry
J. A. Anderson 2/18/04
Registration Test
CT FUSION PET
-0.5
(multiple replacement of
-1 Z-900 mm the phantom) was better
than σ = 0.4 mm.
-1.5 X
-2 Y The stability of the
system over the first 45
-2.5
days was characterized by
-3 σ = 0.5 mm or better for
2/23 3/5 3/15 3/25 4/4 4/14 4/24 all 3 axes.
Date
J. A. Anderson 2/18/04
What Tests are Important?
(You may get different answers from different folks!)
NEMA NU-2 1994 NEMA NU-2 2001
Spatial resolution (xverse,axial) Spatial resolution
Scatter fraction Scatter fraction, count losses, randoms
Sensitivity Sensitivity
Deadtime,count-rate losses {now part of scatter fraction}
Uniformity Uniformity
Scatter correction accuracy {now part of image quality test}
Count-rate correction accuracy Count rate correction accuracy
Attenuation correction accuracy {now part of image quality test}
Image quality (scatter, AC accuracy)
94 Scatter fraction test
94 Count rate tests
These tests were developed to compare classes of machines rather than to
provide practical acceptance test or QA procedures. NU2-94 may be more
appropriate for brain scanners and NU2-01 for WB oncology scans.
J. A. Anderson 2/18/04
Workflow at the PET Center
(FDG Whole Body Scans)
Arrival of patient Receive doses
Transport Pt to scanner *
10 min Position Pt *
* steps with highest
5-30 min Scan technologist exposure
Read study
15
Patient Number
13
11
9
7 Phase
5 Uptake
3
Scanning
8:00 9:00 10:00 11:00 12:00 13:00 14:00 15:00 16:00 17:00
Time of Day
J. A. Anderson 2/18/04
Hot Lab Details: Dose Assay
and Preparation Area Notes:
1) Calibrator
convenient to
dose storage
2) L Block
close to
calibrator
3) Note use of
carrier for
syringe
4) Note L
Block: thick
window, 2"
lead, 2" lead
wrap-around
J. A. Anderson 2/18/04
Hot Lab Details
Notes:
1) All this lead requires solid support -- have a heart-to-
heart talk with the cabinet maker
2) Counter mount of calibrator decreases tech exposure
3) Extra shielding required on well counter to shield
from sources in scanner, calibration sources, patient in
scanner, etc.
4) Use tungsten syringe shields for dose reduction to
fingers.
J. A. Anderson 2/18/04
Injection Room Details
Notes:
1) Injection room
Hot lab
PET/CT bay
are most likely to need
shielding
2) To minimize
anomalous uptake
-minimize external
stimuli
-keep patient quiet and
still on gurney or in
injection chair
The minimum number of injection areas required
3) Need adjacent hot
per scanner at a full patient load is given by
toilet for patients to use
(uptake time)/(scan room time)
after uptake period.
J. A. Anderson 2/18/04
Shielding and the PET Center
Common Viewpoint: Nuclear medicine departments
don't need shielding
0.100
(mrem/mCi)
0.080
µ Sv/MBq
0.060
0.040
0.020
0.000
Benetar Chisea Chisea McElroy UTSW Average
J. A. Anderson 2/18/04
Miscellaneous Operating
Suggestions to Reduce Tech Dose
Complete patient instruction, interaction before
injection. Minimize time near patient after injection.