RS-3000 Specifications

OCT scanning
Anterior segment module (optional)
Scanning pattern Cornea scan
Advanced OCT/ SLO System
Technology Spectral domain OCT Software analysis Angle measurement
OCT resolution Optical Z: 7 µm, XY: 20 µm Corneal thickness measurement
Digital Z: 4 µm, XY: 3 µm Corneal thickness map
Scanning range Z: 2.1 mm
XY: 3 to 9 mm
OCT light source SLD, 880 nm
Scanning speed 53,000 A-scans / s
Acquisition time of 3D image 1.6 s Footprint (mm)
Internal fixation lamp / Wavelength Cross shape (normal or large) / 635 nm
External fixation lamp Red / Green
Wide
Auto alignment Z direction 983
Area
Minimum pupil diameter
Focus adjustment range
ø 2.5 mm
-15 to +10 D (VD=12 mm)
632 Scan
Working distance 35.5 mm (from the objective lens to the pupil) OCT
Scanning pattern Macula line (scan angle changeable by 15º)

452
Macula cross
Macula map
Macula multi (X - Y: 5 x 5)
Disc circle
Disc map

592
Radial scan
Software analysis Segmentation of 6 retinal layers
Macular thickness map
RNFL thickness map
GCC analysis
Optic nerve analysis 472
Follow-up examination of pathological progress
SLO imaging
Technology Confocal scanning laser ophthalmoscope
SLO light source 785 nm
Field of view 40º x 30º (zoom: 20º x 15º)
Focusing method Auto focus
PC networking Available
Display Tiltable 8.4-inch color LCD
Power supply AC 100, 120, 230 V ±10%
50 / 60 Hz
Power consumption 300 VA
Maximum power output (transformer) 1000 VA
Dimensions / Mass 380 (W) x 524 (D) x 499 to 531 (H) mm / 34 kg
15.0 (W) x 20.6 (D) x 19.7 to 20.9 (H)“ / 75.0 lbs.

Motorized optical table (optional)


Dimensions / Mass 592 (W) x 472 (D) x 596 to 794 (H) mm / 27 kg
23.3 (W) x 18.6 (D) x 23.5 to 31.3 (H)“ / 59.5 lbs.
Power supply AC 100 V
50 / 60 Hz
Power consumption 150 W

PC rack (optional)
Dimensions / Mass 632 (W) x 452 (D) x 703 (H) mm / 34 kg
24.9 (W) x 17.8 (D) x 27.7 (H)“ / 75.0 lbs.

FDA 510(K) pending


Specifications and design are subject to change without notice.

HEAD OFFICE TOKYO OFFICE NIDEK INC. NIDEK S.A. NIDEK TECHNOLOGIES Srl
34-14 Maehama, Hiroishi (International Div.) 47651 Westinghouse Drive Europarc Via dell'Artigianato, 6 / A
Gamagori, Aichi 443-0038, Japan 3F Sumitomo Fudosan Hongo Bldg., Fremont, CA 94539, U.S.A. 13, rue Auguste Perret 35020 Albignasego (Padova), Italy
Telephone : +81-533-67-6611 3-22-5 Hongo, Bunkyo-ku, Tokyo Telephone : +1-510-226-5700 94042 Creteil, France Telephone : +39 049 8629200 / 8626399
Facsimile : +81-533-67-6610 113-0033, Japan : +1-800-223-9044 (US only) Telephone : +33-1-49 80 97 97 Facsimile : +39 049 8626824
URL : http://www.nidek.co.jp Telephone : +81-3-5844-2641 Facsimile : +1-510-226-5750 Facsimile : +33-1-49 80 32 08 URL : http://www.nidektechnologies.it
[ Manufacturer ] Facsimile : +81-3-5844-2642 URL : http://usa.nidek.com URL : http://www.nidek.fr
URL : http://www.nidek.com

CNIDEK 2009 Printed in Japan RS-3000 OIDNMK5


Wide
Area High-Speed Scan leads to High-Quality image
Wide Area Scan Scan
OCT

The RS-3000 provides an overall examination report simultaneously indicating macula and [NFL+GCL+IPL]
analyses of right and left eyes. It is useful to assist clinician’s decision in his or her diagnostic flow.

Macula thickness map


Color-coded thickness map of all
retinal layers at and around the
macula (ILM to RPE / BM)
Map of layers from ILM to IS / OS
is also settable.

[NFL+GCL+IPL] thickness map


Color-coded thickness map of
layers composed of
Nerve Fiber Layer
Ganglion Cell Layer
Inner Plexiform Layer
(ILM to IPL / INL)

Analysis charts (ETDRS, GChart,


Superior / Inferior pole)

SLO image

[NFL+GCL+IPL] thickness analysis, RNFL thickness analysis, Disc analysis, TSNIT analysis, Angle measurement
(optional)

The operation of RS-3000 is as easy as an Auto-Refractometer. The focus of SLO fundus image and the 9 mm x 9 mm wide area scan 1.6 s high-speed 3D scan
alignment of OCT depth are adjusted automatically by pressing the optimization button.

Steps 1 2 3
2
1
3

53,000 A-scans / second greatly helps to reduce the measurement time and minimize artifacts.
The advanced speckle-noise-reduction system by averaging images provides 4µm OCT digital resolution.

1
High-resolution image shows the discrete retinal layers.
Start Scanning
53,000

Optimize 25,000~27,000 Conventional SD - OCT

2 The focus of SLO fundus image


and the alignment of OCT depth
are adjusted automatically.
512
A-scans / s
Time Domain - OCT

3 Release
Capturing images of SLO and
OCT with one shot.
Wide
Area High-Speed Scan leads to High-Quality image
Wide Area Scan Scan
OCT

The RS-3000 provides an overall examination report simultaneously indicating macula and [NFL+GCL+IPL]
analyses of right and left eyes. It is useful to assist clinician’s decision in his or her diagnostic flow.

Macula thickness map


Color-coded thickness map of all
retinal layers at and around the
macula (ILM to RPE / BM)
Map of layers from ILM to IS / OS
is also settable.

[NFL+GCL+IPL] thickness map


Color-coded thickness map of
layers composed of Nerve Fiber
Layer, Ganglion Cell Layer, and
Inner Plexiform Layer
(ILM to IPL / INL)

Analysis charts (ETDRS, GChart,


Superior / Inferior pole)

SLO image

[NFL+GCL+IPL] thickness analysis, RNFL thickness analysis, Disc analysis, TSNIT analysis, Angle measurement
(optional)

The operation of RS-3000 is as easy as an Auto-Refractometer. The focus of SLO fundus image and the 9 mm x 9 mm wide area scan 1.6 s high-speed 3D scan
alignment of OCT depth are adjusted automatically by pressing the optimization button.

Steps 1 2 3
2
1
3

53,000 A-scans / second greatly helps to reduce the measurement time and minimize artifacts.
The advanced speckle-noise-reduction system by averaging images provides 4µm OCT digital resolution.

1
High-resolution image shows the discrete retinal layers.
Start Scanning
53,000

Optimize 25,000~27,000 Conventional SD - OCT

2 The focus of SLO fundus image


and the alignment of OCT depth
are adjusted automatically.
512
A-scans / s
Time Domain - OCT

3 Release
Capturing images of SLO and
OCT with one shot.
Glaucoma analysis

 Wide area scan of [NFL+GCL+IPL]  RNFL thickness map


The [NFL+GCL+IPL] analysis supplements clinical work-up for the early detection of optic nerve fiber Early detection of glaucoma and glaucoma follow-up are enhanced with the color coded RNFL
layer defects and 9 mm x 9 mm wide area map enables analysis of [NFL+GCL+IPL] status even in the thickness maps and quantitative analysis of the nerve fiber layer defect.
peripheral retina.

[NFL+GCL+IPL] thickness map Analysis charts (Superior / Inferior pole, GChart) RNFL thickness map Normative database
Color-coded thickness map (9 mm x 9 mm) Analysis charts of average thickness of each sector Color-coded thickness map of Color-coded map indicating distribution range
of [NFL+GCL+IPL] layers (ILM to IPL / INL) surrounding the macula with color code based on RNFL layer (ILM to NFL / GCL) of the patient’s RNFL thickness in a population
overlaid on SLO image comparison to a normative database of normative eyes

SLO image
SLO image showing optic disc

TSNIT graph
Graph showing thickness from
ILM to NFL / GCL on disc circle
with comparison to a normative
OCT image database
OCT image at any
position in X and
Y direction
OCT image of disc circle

SLO image Normative database Deviation map SLO image Analysis charts Analysis table
SLO image showing scanned Color-coded map indicating Map indicating the deviation, SLO image showing scanned Analysis charts indicating average Table of optic disc analysis:
section with color-coded distribution range of the patient’s including early variation even section with color-coded thickness with color code based on C / D ratio (horizontal)
thickness map of user selected [NFL+GCL+IPL] thickness in a within normal range, from thickness map of user selected comparison to a normative database: C / D ratio (vertical)
layers population of normative eyes average thickness in a layers Whole R / D ratio (minimum)
More than 95% normative database S / I (2-sector) R / D ratio (angle)
More than 5 to 95% TSNIT (4-sector) Disc area (mm2)
More than 1 to 5% Clock Hour (12-sector) Cup area (mm2)
0 to 1%

What the [NFL+GCL+IPL] are.  TSNIT analysis  Radial scan analysis (disc radial)
The [NFL+GCL+IPL] are layers composed of The [NFL+GCL+IPL] thickness map shows objective Graph showing RNFL thickness on disc circle High resolution image of radially scanned disc
Nerve Fiber Layer (NFL), Ganglion Cell Layer status in region of nerve fiber layer defect and with comparison to a normative database cross-section. Selectable among 6 lines
(GCL), and Inner Plexiform Layer (IPL). visual field abnormality caused by glaucoma. (30˚ interval) and 12 lines (15˚ interval)

NFL
GCL
IPL

Table of each RNFL thickness with color Symmetry graph Layer thickness map
The color fundus photography is taken with another device. code based on normative database : Graph showing Color-coded RNFL thickness
Overall average symmetry / asymmetry map with thickness value
Superior pole average of the thickness of user at any location where the
Inferior pole average selected scan lines cursor is placed
Glaucoma analysis

 RGI wide analysis  RNFL thickness map


The RGI wide analysis supplements clinical work-up for the early detection of optic nerve fiber layer Early detection of glaucoma and glaucoma follow-up are enhanced with the color coded RNFL
defects and 9 mm x 9 mm wide area map enables analysis of RGI status even in the peripheral retina. thickness maps and quantitative analysis of the nerve fiber layer defect.

RGI thickness map Analysis charts (Superior / Inferior pole, RGI) RNFL thickness map Normative database
Color-coded thickness map (9 mm x 9mm) Analysis charts of average thickness of each Color-coded thickness map of Color-coded map indicating distribution range
of RGI layers (ILM to IPL / INL) overlaid on sector surrounding the macula with color code RNFL layer (ILM to NFL / GCL) of the patient’s RNFL thickness in a population
SLO image based on comparison to a normative database of normative eyes

SLO image
SLO image showing optic disc

TSNIT graph
Graph showing thickness from
ILM to NFL / GCL on disc circle
with comparison to a normative
OCT image database
OCT image at any
position in X and
Y direction
OCT image of disc circle

SLO image Normative database Deviation map SLO image Analysis charts Analysis table
SLO image showing scanned Color-coded map indicating Map indicating the deviation, SLO image showing scanned Analysis charts indicating average Table of optic disc analysis:
section with color-coded distribution range of the including early variation even section with color-coded thickness with color code based on C / D ratio (horizontal)
thickness map of user selected patient’s RGI thickness in a within normal range, from thickness map of user selected comparison to a normative database: C / D ratio (vertical)
layers population of normative eyes average thickness in a layers Whole R / D ratio (minimum)
95 to 100% normative database S / I (2-sector) R / D ratio (angle)
5 to 95% TSNIT (4-sector) Disc area (mm2)
Less than 5% Clock Hour (12-sector) Cup area (mm2)
Less than 1%

What the RGI is.  TSNIT analysis  Radial scan analysis (disc radial)
The RGI is an acronym for Retinal Nerve Fiber The RGI thickness map shows objective status in Graph showing RNFL thickness on disc circle High resolution image of radially scanned disc
Layer (RNFL), Ganglion Cell Layer (GCL), and region of nerve fiber layer defect and visual field with comparison to a normative database cross-section. Selectable among 6 lines
Inner Plexiform Layer (IPL). abnormality caused by glaucoma. (30˚ interval) and 12 lines (15˚ interval)

RNFL
GCL
IPL

Table of each RNFL thickness with color Symmetry graph Layer thickness map
The color fundus photography is taken with another device. code based on normative database : Graph showing Color-coded RNFL thickness
Overall average symmetry / asymmetry map with thickness value
Superior pole average of the thickness of user at any location where the
Inferior pole average selected scan lines cursor is placed
Retina analysis

 Fine B-scan image & real time SLO image  Macula thickness map
Real-time, high-contrast, and wide view (40˚ x 30˚) confocal SLO imaging offers the accuracy for The macula thickness map is a color-coded thickness map of all
OCT scanning of pathology. OCT scanning position is precisely matched with SLO fundus image. retinal layers at and around the macula overlaid on SLO image. Analysis chart
Charts of average thickness (µm) and
It is useful to detect variation in thickness of the entire scanned
volume (mm3) of each sector
Macula multi scan pattern enables high-speed area and local morphological changes. surrounding the macula
scan of 5 x 5 lines (X - Y directions) at one shot.
The position, length, and interval of scanning lines
for the target are all modifiable to user preference Deviation map
Normative database
on real-time confocal SLO image. Map indicating the
Color-coded map
indicating distribution deviation, including
range of the patient's early variation even
macula thickness in a within normal range,
population of from average
normative eyes thickness in a
More than 99% normative database
More than 95 to 99%
More than 5 to 95% Surface view
More than 1 to 5% 3D surface views
0 to 1% indicating each
 High-speed averaging layer of ILM, IS / OS,
and RPE / BM
High-speed and accurate averaging achieves significant speckle-noise reduction.
Image errors are automatically detected and excluded from the averaging,
which enhances image qualities.  Radial scan analysis (macula radial)
RNFL
GCL High resolution image of radially scanned retina cross-section
IPL
Max. 50 image Selectable among 6 lines (30˚ interval) and 12 lines (15˚ interval)
INL
averaging
OPL
ONL Layer thickness map
ELM Color-coded thickness map with
thickness value at any location where
IS/OS
the cursor is placed
RPE Display of [NFL+GCL+IPL] thickness is
also settable.

Symmetry graph
Graph showing symmetry / asymmetry of the thickness (all retinal
layers, [NFL+GCL+IPL], etc.) of user selected scan lines

 3D map / Stereoview  6 layers segmentation


9 mm x 9 mm wide area map / stereoview is acquired in 1.6 seconds. Fine 3D map of whole image Layer by layer segmentation allows visual confirmation of morphological changes in each of the 6 layers.
within an arcade including macula clearly indicates layers structure and assist clinician's diagnoses.

9 mm x 9 mm wide area
6 mm x 6 mm area ILM NFL / GCL IPL / INL

OPL / ONL IS / OS RPE / BM


Retina analysis

 Fine B-scan image & real time SLO image  Macula thickness map
Real-time, high-contrast, and wide view (40˚ x 30˚) confocal SLO imaging offers the accuracy for The macula thickness map is a color-coded thickness map of all
OCT scanning of pathology. OCT scanning position is precisely matched with SLO fundus image. retinal layers at and around the macula overlaid on SLO image. Analysis chart
Charts of average thickness (µm) and
It is useful to detect variation in thickness of the entire scanned
volume (mm3) of each sector
Macula multi scan pattern enables high-speed area and local morphological changes. surrounding the macula
scan of 5 x 5 lines (X - Y directions) at one shot.
The position, length, and interval of scanning lines
for the target are all modifiable to user preference Deviation map
Normative database
on real-time confocal SLO image. Map indicating the
Color-coded map
indicating distribution deviation, including
range of the patient's early variation even
macula thickness in a within normal range,
population of from average
normative eyes thickness in a
More than 99% normative database
More than 95 to 99%
More than 5 to 95% Surface view
More than 1 to 5% 3D surface views
0 to 1% indicating each
 High-speed averaging layer of ILM, IS / OS,
and RPE / BM
High-speed and accurate averaging achieves significant speckle-noise reduction.
Image errors are automatically detected and excluded from the averaging,
which enhances image qualities.  Radial scan analysis (macula radial)
RNFL
GCL High resolution image of radially scanned retina cross-section
IPL
Max. 50 image Selectable among 6 lines (30˚ interval) and 12 lines (15˚ interval)
INL
averaging
OPL
ONL Layer thickness map
ELM Color-coded thickness map with
thickness value at any location where
IS/OS
the cursor is placed
RPE Display of [NFL+GCL+IPL] thickness is
also settable.

Symmetry graph
Graph showing symmetry / asymmetry of the thickness (all retinal
layers, [NFL+GCL+IPL], etc.) of user selected scan lines

 3D map / Stereoview  6 layers segmentation


9 mm x 9 mm wide area map / stereoview is acquired in 1.6 seconds. Fine 3D map of whole image Layer by layer segmentation allows visual confirmation of morphological changes in each of the 6 layers.
within an arcade including macula clearly indicates layers structure and assist clinician's diagnoses.

9 mm x 9 mm wide area
6 mm x 6 mm area ILM NFL / GCL IPL / INL

OPL / ONL IS / OS RPE / BM


Anterior segment analysis Analysis reports
With an optional anterior segment module the RS-3000 enables observation and analyses
of the anterior segment.

Angle measurement
• AOD500 (AOD700)
Distance between iris and a point 500 µm (or 700 µm)
away from scleral spur on posterior corneal surface
• TISA500 (TISA750)
Area circumscribed with AOD500 (or AOD700) line, posterior 変更未
corneal surface, line drawn from scleral spur in parallel with
AOD line, and iris surface
• ACA
Angle between posterior corneal surface and iris surface
Macula multi Macula radial 12 lines

 Cornea measurement
The cornea measurement provides thickness of corneal apex, thickness
of any two sites, and corneal thickness map.

Corneal thickness Corneal thickness map


Anterior segment adaptor

Macula map, both eyes (overall tab) Disc map, both eyes (overall tab)
Follow-up examination with auto-tracking function

High contrast SLO fundus image and auto-tracking function achieve excellent reproducibility for
follow-up examination. Auto-tracking function tracks eye movement and ensures the OCT scans are
aligned with the previous exams. Analyses of pre and post treatment of precisely aligned maps
enables quantitative assessment.

Adjustment
guide arrow

Macula map (macula tab) Macula map (3D tab)

Adjustment
target
Fundus image
deviation gauge
Large deviation Small deviation No deviation

変更未 変更未

Disc circle Disc map Macula map Angle line Angle radial
Anterior segment analysis Analysis reports
With an optional anterior segment module the RS-3000 enables observation and analyses
of the anterior segment.

Angle measurement
• AOD500 (AOD700)
Distance between iris and a point 500 µm (or 700 µm)
away from scleral spur on posterior corneal surface
• TISA500 (TISA750)
Area circumscribed with AOD500 (or AOD700) line, posterior
corneal surface, line drawn from scleral spur in parallel with
AOD line, and iris surface
• ACA
Angle between posterior corneal surface and iris surface
Macula multi Macula radial 12 lines

 Cornea measurement
The cornea measurement provides thickness of corneal apex, thickness
of any two sites, and corneal thickness map.

Corneal thickness Corneal thickness map


Anterior segment adaptor

Macula map, both eyes (overall tab) Disc map, both eyes (overall tab)
Follow-up examination with auto-tracking function

High contrast SLO fundus image and auto-tracking function achieve excellent reproducibility for
follow-up examination. Auto-tracking function tracks eye movement and ensures the OCT scans are
aligned with the previous exams. Analyses of pre and post treatment of precisely aligned maps
enables quantitative assessment.

Adjustment
guide arrow

Macula map (macula tab) Macula map (3D tab)

Adjustment
target
Fundus image
deviation gauge
Large deviation Small deviation No deviation

Disc circle Disc map Macula map Angle line Angle radial
Scan patterns Clinical images

Multiple and easy scan modes Myopic retinoschisis

Macula scan

6 x 6 mm Scan 9 mm Scan

Macula line Macula cross Macula multi

Central Serous Chorioretinopathy

Macula radial Macula map


9 x 9 mm Scan 9 mm Scan

Disc scan

Macular Hole

Disc radial Disc circle Disc map

Cornea scan (with an optional anterior segment module)


9 x 9 mm Scan 9 mm Scan

Age-related Macular Degeneration (AMD)

Cornea line Cornea cross Cornea radial Angle line

Combo release mode


9 x 9 mm Scan 9 mm Scan
The combo release mode combines scan patterns according to
preset scan pattern order and facilitates the examination which
needs several scan patterns.
The scan pattern order can be modified to user preferences.
Epiretinal membrane

9×9 mm Scan 9 mm Scan


Scan patterns Clinical images

Multiple and easy scan modes Myopic retinoschisis

Macula scan

6 x 6 mm Scan 9 mm Scan

Macula line Macula cross Macula multi

Central serous chorioretinopathy

Macula radial Macula map


9 x 9 mm Scan 9 mm Scan

Disc scan

Macular hole

Disc radial Disc circle Disc map

Cornea scan (with an optional anterior segment module)


9 x 9 mm Scan 9 mm Scan

Age-related Macular Degeneration (AMD)

Cornea line Cornea cross Cornea radial Angle line

Combo release mode


9 x 9 mm Scan 9 mm Scan
The combo release mode combines scan patterns according to
preset scan pattern order and facilitates the examination which
needs several scan patterns.
The scan pattern order can be modified to user preferences.
Epiretinal membrane

9×9 mm Scan 9 mm Scan


RS-3000 Specifications
OCT scanning
Anterior segment module (optional)
Scanning pattern Cornea scan
Advanced OCT/ SLO System
Technology Spectral domain OCT Software analysis Angle measurement
OCT resolution Optical Z: 7 µm, XY: 20 µm Corneal thickness measurement
Digital Z: 4 µm, XY: 3 µm Corneal thickness map
Scanning range Z: 2.1 mm
XY: 3 to 9 mm
OCT light source SLD, 880 nm
Scanning speed 53,000 A-scans / s
Acquisition time of 3D image 1.6 s Footprint (mm)
Internal fixation lamp / Wavelength Cross shape (normal or large) / 635 nm
External fixation lamp Red / Green
Wide
Auto alignment Z direction 983
Area
Minimum pupil diameter
Focus adjustment range
ø 2.5 mm
-15 to +10 D (VD=12 mm)
632 Scan
Working distance 35.5 mm (from the objective lens to the pupil) OCT
Scanning pattern Macula line (scan angle changeable by 15º)

452
Macula cross
Macula map
Macula multi (X - Y: 5 x 5)
Disc circle
Disc map

592
Radial scan
Software analysis Segmentation of 6 retinal layers
Macular thickness map
RNFL thickness map
GCC analysis
Optic nerve analysis 472
Follow-up examination of pathological progress
SLO imaging
Technology Confocal scanning laser ophthalmoscope
SLO light source 785 nm
Field of view 40º x 30º (zoom: 20º x 15º)
Focusing method Auto focus
PC networking Available
Display Tiltable 8.4-inch color LCD
Power supply AC 100, 120, 230 V ±10%
50 / 60 Hz
Power consumption 300 VA
Maximum power output (transformer) 1000 VA
Dimensions / Mass 380 (W) x 524 (D) x 499 to 531 (H) mm / 34 kg
15.0 (W) x 20.6 (D) x 19.7 to 20.9 (H)“ / 75.0 lbs.

Motorized optical table (optional)


Dimensions / Mass 592 (W) x 472 (D) x 596 to 794 (H) mm / 27 kg
23.3 (W) x 18.6 (D) x 23.5 to 31.3 (H)“ / 59.5 lbs.
Power supply AC 100 V
50 / 60 Hz
Power consumption 150 W

PC rack (optional)
Dimensions / Mass 632 (W) x 452 (D) x 703 (H) mm / 34 kg
24.9 (W) x 17.8 (D) x 27.7 (H)“ / 75.0 lbs.

FDA 510(K) pending


Specifications and design are subject to change without notice.

HEAD OFFICE TOKYO OFFICE NIDEK INC. NIDEK S.A. NIDEK TECHNOLOGIES Srl
34-14 Maehama, Hiroishi (International Div.) 47651 Westinghouse Drive Europarc Via dell'Artigianato, 6 / A
Gamagori, Aichi 443-0038, Japan 3F Sumitomo Fudosan Hongo Bldg., Fremont, CA 94539, U.S.A. 13, rue Auguste Perret 35020 Albignasego (Padova), Italy
Telephone : +81-533-67-6611 3-22-5 Hongo, Bunkyo-ku, Tokyo Telephone : +1-510-226-5700 94042 Creteil, France Telephone : +39 049 8629200 / 8626399
Facsimile : +81-533-67-6610 113-0033, Japan : +1-800-223-9044 (US only) Telephone : +33-1-49 80 97 97 Facsimile : +39 049 8626824
URL : http://www.nidek.co.jp Telephone : +81-3-5844-2641 Facsimile : +1-510-226-5750 Facsimile : +33-1-49 80 32 08 URL : http://www.nidektechnologies.it
[ Manufacturer ] Facsimile : +81-3-5844-2642 URL : http://usa.nidek.com URL : http://www.nidek.fr
URL : http://www.nidek.com

CNIDEK 2009 Printed in Japan RS-3000 OJDKOP6