The document provides instructions for replacing the cartridge in the active inlet valve for Agilent pumps when pressure ripple is unstable or a leak test confirms an internal leak in the valve. The instructions are to use a 14 mm wrench to remove the active inlet valve, change the cartridge, and reinstall the active inlet valve, being sure to properly position the cable. Replacement cartridges and parts are also listed.
The document provides instructions for replacing the cartridge in the active inlet valve for Agilent pumps when pressure ripple is unstable or a leak test confirms an internal leak in the valve. The instructions are to use a 14 mm wrench to remove the active inlet valve, change the cartridge, and reinstall the active inlet valve, being sure to properly position the cable. Replacement cartridges and parts are also listed.
The document provides instructions for replacing the cartridge in the active inlet valve for Agilent pumps when pressure ripple is unstable or a leak test confirms an internal leak in the valve. The instructions are to use a 14 mm wrench to remove the active inlet valve, change the cartridge, and reinstall the active inlet valve, being sure to properly position the cable. Replacement cartridges and parts are also listed.
when the pressure ripple is unstable and a leak test confirms a problem with the valve due to internal leaking. The active inlet valve contains a cartridge that you can Active Inlet Valve (AIV)
exchange as follows:
1. Use a 14 mm wrench to remove the Active Inlet Valve (AIV)
active inlet valve. Description Part No. 2. Change the cartridge. 3. Reinstall the active inlet valve. Active inlet valve, without cartridge G1312-60025 Cartridge for active inlet valve, 400 bar 5062-8562 Cartridge for active inlet valve, 600 bar G1312-60020 Gold Seal for inlet valve 5001-3708
Tips & Tools
Be sure to properly position the Active Inlet Valve cable when you reinstall the valve.