Saturn Evo Operators Manual 4th Edition 2008 PDF

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 63

SATURN EVO

OPERATOR’S MANUAL
4th edition
JUNE 2008
SATURN EVO ANAESTHESIA SYSTEM

OPERATOR’S MANUAL

0499

4th edition
June 2008

MEDEC BENELUX NV
All rights reserved. No part of this publication may be reproduced, stored in a
retrieval system or transmitted, in any form or by any means, electronically,
mechanically, by photocopying, recording or otherwise, without the prior written
permission of Medec Benelux NV.

Medec Benelux NV reserves the right to change specifications without prior


notification. Careful attention has been paid to the compilation of this publication.

Medec Benelux NV
Lion D’Orweg 19
9300 Aalst
Belgium
Telephone : (32) 53 / 70.35.44
Fax : (32) 53 / 70.35.33
Website : www.medecbenelux.be
E-mail : sales@medecbenelux.be

4th edition
June 2008
Table of Contents
1. Saturn Evo system overview
1.1 General 1
1.2 User responsibility 1
1.3 Main components 1

2. Saturn Evo anaesthesia configuration 2


2.1 Gas supply & trolley connections 3
2.2 Mechanical functions 5
1. SATURN EVO SYSTEM OVERVIEW

1.1 GENERAL

This manual is intended to provide authorized personnel with following information:

• how to operate the Saturnus anaesthesia system


• the operating principles of the system
• maintenance checks that should be performed

Personnel should make themselves familiar with the contents of this manual before
using the anaesthesia system.

1.2 USER RESPONSIBILITY

The equipment described in this manual has been built to conform with the
specifications and operating instructions stated in this manual. To ensure proper and
safe operation of the equipment, it must be checked and serviced to at least the
minimum standards laid out in this manual.

The equipment must be repaired and serviced only in accordance with written
instructions issued by Medec Benelux NV and must not be altered or modified in any
way without written approval of Medec Benelux NV. The user of this equipment shall
have the responsibility for any malfunction which results from improper use,
maintenance, repair, damage or alteration by anyone other than Medec Benelux NV or
its appointed agents.

1.3 MAIN COMPONENTS

An overview of the system’s components is shown on the following page.


The system’s main components are:

• Saturn Evo trolley


• Saturn Evo electronic ventilator – operator’s manual
• Saturn Evo CO2 absorber system – operator’s manual
• Saturn Evo ORC rotameter

Each of these components is described in a separate section of this manual.

Saturn Evo anaesthesia system 1 Operator's manual


4th edition – June 2008
2. SATURN EVO ANAESTHESIA CONFIGURATION

Monitor shelf

Vaporizers - three stations


Rotameter with
oxygen ratio controller

Patient arm and


autoclavable circuit

Electronic ventilator

Writing shelf

Absorber system
Drawer unit

suction jar

Saturn Evo anaesthesia system 2 Operator's manual


4th edition – June 2008
2.1 GAS SUPPLY & TROLLEY CONNECTIONS

Numbers in text refer to figure 1 and figure 2.

• NIST connectors - see Saturn Evo electronic ventilator manual.

• Cylinder holders (fig. 1)


The trolley is provided with 3 cylinder holders (4) on the back of the trolley.
Cylinders and regulators are optional items.

• Fresh gas outlet - absorber (fig. 2)


When the front/absorber knob on the ventilator is switched to absorber, the
fresh gas flow is directed via the fresh gas outlet connector (2) on the trolley to
the absorber system, into the circle system.

• Suction (fig. 2)
The collection jar is connected to the suction input connector (1) and operated
via the ventilator front panel.

• Oxygen measurement (fig. 2)


The oxygen measurement set is connected to the input connector (3) on the
trolley panel.

FIG 1

Saturn Evo anaesthesia system 3 Operator's manual


4th edition – June 2008
FIG 2

Schematic diagram

See operator’s manual Saturn Evo ventilator

Saturn Evo anaesthesia system 4 Operator's manual


4th edition – June 2008
2.2 MECHANICAL FUNCTIONS

Numbers in the text refer to figure 3

• Top tray
Instruments, monitoring etc. may be put on the top tray (1).

• Writing shelf (2)


Push to open

• Light (3)
Built-in light. 3 positions + off.

• Drawers (4)
The drawers may be used to store instruments, etc.

• Braking system (5)


The braking system to block the wheels is activated by pressing pedal 1 and
pedal 2 located on the front wheels.

FIG 3

Saturn Evo anaesthesia system 5 Operator's manual


4th edition – June 2008
SATURN EVO
ELECTRONIC VENTILATOR
OPERATOR’S MANUAL

0499

4th edition
June 2008

MEDEC BENELUX NV
All rights reserved. No part of this publication may be reproduced, stored in a
retrieval system or transmitted, in any form or by any means, electronically,
mechanically, by photocopying, recording or otherwise, without the prior written
permission of Medec Benelux NV.

Medec Benelux NV reserves the right to change specifications without prior


notification. Careful attention has been paid to the compilation of this publication.

Medec Benelux NV
Lion D’Orweg 19
9300 Aalst
Belgium
Telephone : (32) 53 / 70.35.44
Fax : (32) 53 / 70.35.33
Website : www.medecbenelux.be
E-mail : sales@medecbenelux.be

4th edition
June 2008
Table of Contents
1. General information 1
1.1 General 1
1.2 User responsibility 1

2. System overview 2
2.1 Front view 3
2.2 Side view 4
2.3 Back view 5
2.4 Schematic diagram of gas supply 6
2.5 Pneumatic connections 7
2.6 Front 9
2.7 Electrical connections 10

3. Operation 11
3.1 Operator interface 11
3.2 Switch on the ventilator 12
3.3 The display screen 13
3.4 Operating instructions 20
3.4.1 Preparations before use 20
3.4.2 Changing button parameters 20
3.4.3 Standby mode 21
3.4.4 Spontaneous / manual ventilation mode (MAN) 21
3.4.5 Controlled mandatory ventilation mode (CMV) 22
3.4.6 Pressure controlled ventilation mode (PCV) 22
3.4.7 Synchronized mandatory ventilation mode (SIMV) 23
3.4.8 Tidal volume (VOLUME) (CMV /SIMV mode) 24
3.4.9 Peak pressure (PEAK P.) (PCV mode) 24
3.4.10 Frequency (FREQUENCY) (CMV / SIMV / PCV mode) 24
3.4.11 Inspiration / expiration ratio (I/E ratio) (CMV / SIMV / PCV mode) 24
3.4.12 Inspiratoire pause (INSP. PAUZE) (CMV / SIMV mode) 24
3.4.13 Positive end expiration pressure (PEEP) (CMV / SIMV /PCV mode)25
3.4.14 Trigger (SIMV / PCV mode) 25
3.4.15 Sigh (CMV/ SIMV mode) 25
3.5 Oxygen measurement 26
3.5.1 Principle of operation 26
3.5.2 Operation 26
3.5.3 Set-up the O2 measurement 27
3.5.4 O2 calibration 29
3.6 Alarms, warnings and indicators 30
3.6.1 Lower limit pressure alarm (CMV/SIMV/ PCV mode) 31
3.6.2 Apnea alarm (Man/Spont mode) 31
3.6.3 Upper limit pressure alarm (CMV / SIMV and Man/Spont mode) 31
3.6.4 Switching between the different ventilation modes 32
3.6.5 Warning window 33
3.7 Performing a leaktest 36
3.7.1 Entering the leaktest 36
3.7.2 Performing the leaktest 37
3.7.3 Leaving the leaktest menu 38
3.8 The setup menu 39
3.8.1 Entering the setup menu 39
3.8.2 The setup menu 39
3.9 Battery backup 42

4. Cleaning and disinfecting 43


4.1 Ventilator housing 43
4.2 Patient breathing circuit 43

5. Maintenance, malfunction and troubleshooting 45


5.1 Maintenance 45
5.2 Malfunction - error codes 46
5.3 Troubleshooting 48
5.4 Fuse replacement 49

6. Specifications 50
6.1 General 50
6.2 Ratings 50
1. GENERAL INFORMATION

1.1 GENERAL

This manual is intended to provide authorized personnel with following information:

• how to operate the Saturn Evo ventilator


• the operating principles of the system
• maintenance checks that should be performed

Personnel should make themselves familiar with the contents of this manual before
using the anesthesia system.

1.2 USER RESPONSIBILITY

The equipment described in this manual has been built to conform with the
specifications and operating instructions stated in this manual. To ensure proper and
safe operation of the equipment, it must be checked and serviced to at least the
minimum standards laid out in this manual.

The equipment must be repaired and serviced only in accordance with written
instructions issued by Medec Benelux NV and must not be altered or modified in any
way without written approval of Medec Benelux NV. The user of this equipment shall
have the responsibility for any malfunction which results from improper use,
maintenance, repair, damage or alteration by anyone other than Medec Benelux NV or
its appointed agents.

Saturn Evo ventilator 1 Operator’s manual


4th edition – June 2008
2. SYSTEM OVERVIEW

The Saturn Evo combines some never seen abilities, high technological research and a
high level of security. Clear monitoring gives you full control at a single glance. Our
well-studied platform is easy to handle and gives you many exceptional features. The
platform is based on the well-known horizontal bag-in-bottle system.

The Saturn Evo is an anesthesia ventilator that can handle spontaneous breathing as
well. The system incorporates a semi-closed breathing circuit, to which a fresh gas
flow is continuously directed.
The patient inspires and expires either in the hand balloon (in spontaneous/manual
mode) or in the bag (in CMV,SIMV or PCV mode).
CO2 accumulation is prevented by means of an absorber.

This manual will describe all control buttons and functions of the electronic
ventilator. Both the alarm system and the cleaning procedures will be attended to as
well.

Saturn Evo ventilator 2 Operator’s manual


4th edition – June 2008
2.1 FRONT VIEW

Saturn Evo ventilator 3 Operator’s manual


4th edition – June 2008
2.2 SIDE VIEW

Saturn Evo ventilator 4 Operator’s manual


4th edition – June 2008
2.3 BACK VIEW

Saturn Evo ventilator 5 Operator’s manual


4th edition – June 2008
2.4 SCHEMATIC DIAGRAM OF GAS SUPPLY

Oxygen, nitrous oxide and air from the wall supply flows through the rotameter and
vaporizer-section to the selected fresh gas outlet (absorber or front outlet). Via a
bypass, oxygen can also flow directly to the fresh gas outlet. Air is used as pneumatic
supply to the ventilator. If no wall supply is available, oxygen and nitrous oxide can
be supplied by high-pressure cylinders on the back of the trolley.

Saturn Evo ventilator 6 Operator’s manual


4th edition – June 2008
2.5 PNEUMATIC CONNECTIONS

6 5

3 4

7 8 9 10

Saturn Evo ventilator 7 Operator’s manual


4th edition – June 2008
1. O2 NIST connection: two O2 NIST connectors for wall and/or for cylinder
supply.

2. N2O NIST connection: two N2O NIST connectors for wall and/or cylinder
supply.

3. Air NIST connection: air NIST connection for wall or cylinder supply.

4. Vacuum NIST connection.

5. Air, N2O and O2 output: for connection with rotameter.

6. Fresh gas input: delivered by rotameter.

7. Patient Evacuation.

8. Connection for GASOVAC.

9. Suction jar connection.

10. Fresh gas output: to absorber.

Saturn Evo ventilator 8 Operator’s manual


4th edition – June 2008
2.6 FRONT

1 2 3

4 5 6
1. O2 input pressure indicator.

2. N2O input pressure indicator.

3. Air input pressure indicator.

4. Fresh gas front outlet: for connection with a mask.

5. Fresh gas absorber-front selection knob: select absorber to connect the


fresh gas flow to the absorber. Select front to use the front outlet as fresh gas
output.

Saturn Evo ventilator 9 Operator’s manual


4th edition – June 2008
6. O2 flush button: bypasses the rotameter and vaporizer and directs a flow of
35 l/min O2 directly to the selected fresh gas output.
There is a detection in CMV&SIMV when an oxygen flush is generated to
prevent high pressure if the upper limit was set to high.
The Max pressure allowed during an O2 flush is last peak pressure + 20 hPa /
20 mbar and this level must be below upper limit setting (last pip +20 is
below upper limit setting).
When the upper limit alarm is lower ,the upper limit alarm level is used.
The min level is 30 hPa / 30 mbar if upper limit is higher.
When the upper limit alarm level is lower then the calculated safety, the upper
limit alarm level has priority.

7. Suction On/Off knob: for use with suction jar.

8. Vacuum pressure indicator.

2.7 ELECTRICAL CONNECTIONS

2 3

1. The mains power inlet: connect the power cord with the mains power inlet.
Turn switch ON. The ventilator is working on mains supply and the battery is
charging. The green mains indicator at the front will be lit. The battery will not
be charged with the main switch in the OFF position. The ventilator will work
on battery and the green mains indicator is off.

2. +12V DC output for lighting: power for top shelf lighting.

3. Mains outlet: outlet for connection with 4 mains outlets located at upper back
panel.

4. O2 sensor input: connection for O2 measurement.

Saturn Evo ventilator 10 Operator’s manual


4th edition – June 2008
3. OPERATION

3.1 OPERATOR INTERFACE

5
6

1. Buttons: to select and change ventilator settings.

2. Control knob: for changing the settings.

3. Silent button: all the alarms are mute for one minute when this button is
pressed.

4. ON / OFF button: press this button for at least 1 second to switch the
ventilator on/off.

5. Display screen: displays measured data and settings.

6. Mains indicator: this green indicator lights up when the ventilator is


connected to the mains power and the power switch on the back panel is
switched on.

Saturn Evo ventilator 11 Operator’s manual


4th edition – June 2008
3.2 SWITCH ON THE VENTILATOR

As soon as the power cord is connected to mains supply and the main switch is turned
on, the mains indicator will be illuminated.

Pushing the button for at least 1 second turns on the ventilator.


A short beep indicates that the ventilator is turned on. The display remains dark for a
few seconds during start-up.
During initialization, the display will show a logo and all main functions - including
the microprocessor functions, external component tests, etc. - will be tested.
After succeeding the self-test, the ventilator enters standby mode.

When a failure is detected, the display screen will show an error code and a short
description of the failure. Refer to chapter ‘TROUBLESHOOTING’ for more
information about the error codes.

Saturn Evo ventilator 12 Operator’s manual


4th edition – June 2008
3.3 THE DISPLAY SCREEN

• The standby window

After turning the ventilator on, it will always start up in the standby window.

3 Standby window

1. Setup: the setup button in the standby window is used to access the setup
menu.

2. Leaktest: the leaktest button in the standby window is used to access the leak
test menu.

3. Mode: with the mode button, you can switch between the different ventilator
modes: CMV,SIMV, PCV, Man/Spont and standby mode. Press the mode
button, change the value by rotating the control knob and execute by pressing
the control knob. You can cancel by pressing the mode button once again.

Saturn Evo ventilator 13 Operator’s manual


4th edition – June 2008
• The CMV window

The settings of the ventilator are presented on a number of large screen push buttons.

13 12 11 10 9

1
8

4 CMV window
5 6 7
1. Limit pressure: sets upper limit and lower limit pressure.

2. PEEP: the PEEP can be adjusted by using this button. The PEEP range is
between 0 - 20 hPa / 0 – 20 mbar.

3. Inspiratoire pause: the INSP. PAUSE can be changed in steps of 5% from


0% to 50%.

4. Mode: with the mode button, you can change between CMV,PCV,SIMV,
Man/Spont and standby mode.

5. Tidal volume: the tidal volume can be set from 10 ml to 1600 ml.

6. Frequency: the frequency range can be set in the range from 4 - 80 breaths
per minute. Limitations on the frequency depending on the I/E ratio setting.

I/E ratio Frequency range


1:1 4 – 80
1:1.5 4 – 80
1:2 4 – 80
1:3 4 – 40
1:4 4 – 50
1:5 4 – 60
1:6 4 – 70
2:1 4 – 80
3:1 4 – 80
4:1 4 – 80

Saturn Evo ventilator 14 Operator’s manual


4th edition – June 2008
7. I/E ratio: this button sets the I/E ratio. The possible values are 1:1, 1:1.5, 1:2,
1:3, 1:4, 1:5,1:6, 2:1, 3:1, 4:1. Limitations on the I/E ratio depending on the
frequency setting.

Frequency range I/E setting


4 - 40 1:1, 1:1.5, 1:2, 1:3, 1:4, 1:5,1:6, 2:1, 3:1, 4:1
41 – 50 1:1, 1:1.5, 1:2, 1:3, 1:4, 1:5, 2:1, 3:1, 4:1
51 – 60 1:1, 1:1.5, 1:2, 1:3, 1:4, 2:1, 3:1, 4:1
61 – 70 1:1, 1:1.5, 1:2, 1:3, 2:1, 3:1, 4:1
71 – 80 1:1, 1:1.5, 1:2, 2:1, 3:1, 4:1

8. Measurement window: this window contains the measured values from the
last inspiration / expiration cycle and the barograph. These values are updated
at the end of each expiration. The measured values are: expired tidal volume,
minute volume (MIN.VOL.), peak inspiration pressure (PIP), plateau pressure
(PLATEAU), positive end expiration pressure (PEEP), mean pressure
(MEAN) and compliance (COMPL.). The barograph is updated continuously.

9. Silent indicator: the silent symbol is flashing after the silent button is pressed,
and no apnea alarm is active.
A down counter (starting from 60 seconds) gives an indication how long the
Alarm disable function will be active.

10. Time: actual time is displayed over here.

11. Date: actual date is displayed over here.

12. Mains / battery indicator: when working on mains supply, the following
symbol is shown . When working on batteries, a battery symbol
indicates the battery status .

13. Barograph: the barograph indicates airway pressure in the range from
-10 hPa / -10mbar to 100 hPa / 100mbar.

Saturn Evo ventilator 15 Operator’s manual


4th edition – June 2008
• The SIMV window

SIMV window

1.Trigger: sets the trigger on/off and adjusts the trigger level –2 to –20 hPa
-2 to -20mbar.
Trigger is a trigger below Peep and Max adjusted trigger level that can be set is
equal to peep setting.
All the other settings and buttons are similar as described in CMV mode.

Saturn Evo ventilator 16 Operator’s manual


4th edition – June 2008
• The PCV window

PCV window
1

1. Peak pressure: sets the peak pressure.

All the other settings and buttons are similar as described in CMV and SIMV mode.

Saturn Evo ventilator 17 Operator’s manual


4th edition – June 2008
• The Man/Spont window

Man/Spont window

In MAN mode, the upper limit pressure alarm can be set.


Apnea alarm can set ON or OFF.

Saturn Evo ventilator 18 Operator’s manual


4th edition – June 2008
• The graph window

Select the graph window by rotating the control knob.

Note: It is not possible to change to the graph window when alarms or


warnings are still active.

8
2

7
3

4
6

Graph window
5
1. Pressure scale: changes the graph pressure scale between 25 hPa/25 mbar,
50 hPa/50 mbar and 100 hPa/100 mbar.

2. Time scale: changes the graph time scale between 5 sec, 10 sec and
20 sec.

3. Freeze / Run: freeze the graph by pressing this button. Pressing this button a
second time, will set the graph in run mode.

4. Sigh: sets sigh on/off.


Can only be set in CMV or SIMV mode.

5. Mode indicator: indicates the actual mode.

6. Measurement window: contains the measured values from the last


inspiration/expiration cycle.

7. Graph: airway pressure (hPa/mbar) versus time (seconds) is displayed.

8. Slider: the slider will synchronize with the start of a new inspiration. The
slider waits at the start of the graph until a new inspiration takes place. In
MAN mode, the slider is continuously running independent of the inspiration /
expiration cycle.

Saturn Evo ventilator 19 Operator’s manual


4th edition – June 2008
3.4 OPERATING INSTRUCTIONS

3.4.1 Preparations before use

Before use, make the following connections:

- check supply of air, oxygen and nitrous oxide.

- check the main supply indicator.

- connect patient circuit through absorber system.

- connect manual system with the hand balloon and patient system.

- connect the patient tubing to absorber outlets.

- switch the ventilator on by pressing the button for at least 1 second.

- check the battery backup by pulling the mains plug (optional).

- perform a leak test (optional).

After these preparations the ventilator is ready for use.

3.4.2 Changing button parameters

A button press inverts the selected setting.


Example of a button with one parameter:

Example of a button with two or more parameters:

Rotating the control knob changes the value of the setting. Pushing the control knob
will save the changed value. Pressing the selected button again will cancel the
changed parameter value and the value restores to the previous value.

Saturn Evo ventilator 20 Operator’s manual


4th edition – June 2008
Note: When no button is pressed or when the control knob is rotated within
10 seconds, the selected setting will automatically restore to the
previous value.
3.4.3 Standby mode

The ventilator will always start in "standby mode". From the standby window you can
perform following actions:

- Open the setup menu.


- Open the leaktest menu.
- Start the ventilator in manual mode.
- Start the ventilator in CMV mode.
- Start the ventilator in PCV mode.
- Start the ventilator in SIMV mode.

3.4.4 Spontaneous / manual ventilation mode (MAN)

The spontaneous / manual mode (MAN mode) allows spontaneous breathing of the
patient and manually controlled ventilation with a hand balloon. The ventilator will be
in this mode if:

- it is not switched on.


- it is switched on and in the standby window.
- it is switched on and in MAN mode.
- in case of a technical failure.

In MAN mode, the Saturn Evo can even be used without electrical and/or compressed
air connections. When no electrical connection is available, the readouts of the
measured values, alarms and warnings will of course not be displayed.

Saturn Evo ventilator 21 Operator’s manual


4th edition – June 2008
3.4.5 Controlled mandatory ventilation mode (CMV)

If the ventilator is activated, it can be switched to the controlled mandatory ventilation


mode (CMV mode) with the use of the mode button.
The following screen pops up:

Set up the parameters for use during CMV mode.


Push the control knob to confirm and switch over to CMV mode or rotate selector
knob to select cancel.

3.4.6 Pressure controlled ventilation mode (PCV)

Switch the ventilator to the pressure controlled ventilation mode (PCV mode) with the
use of the mode button.
The following screen pops up:

Push the control knob to confirm and switch over to PCV mode or rotate selector
knob to select cancel.

Saturn Evo ventilator 22 Operator’s manual


4th edition – June 2008
3.4.7 Synchronized intermitted mandatory ventilation mode (SIMV)

If the ventilator is activated, it can be switched to the Synchronized intermitted


mandatory ventilation mode (SIMV mode) with the use of the mode button.
The following screen pops up:

Set up the parameters for use during SIMV mode.


Push the control knob to confirm and switch over to SIMV mode or rotate selector
knob to select cancel.

In CMV / PCV/ SIMV mode, several controls are available. If a control is changed,
the new value will be used after a complete expiration cycle. The lower limit pressure
and upper limit pressure values will be used immediately.
When switching over from MAN mode to CMV / PCV/ SIMV mode, the Saturn Evo
will start with an inspiration, when patient pressure is lower than 2mbar.
The ventilator will wait to go to CMV / PCV/ SIMV mode due the following reasons:

- low oxygen input pressure.


- low nitrous oxide input pressure.
- low air input pressure.
- fresh gas volume higher than set volume.

The MODE text will be blinking during waiting changing mode from MAN to
CMV/PCV/SIMV or CMV/PCV/SIMV to MAN.

Saturn Evo ventilator 23 Operator’s manual


4th edition – June 2008
3.4.8 Tidal volume (VOLUME) (CMV & SIMV mode)

The tidal volume is set in ml. It may take a few breathing cycles before the set volume
is reached.
The Saturn Evo electronics will automatically correct any deviation to the preset
volume. This correction is being calculated by a microprocessor, based on a number
of parameters (compliance, leakage, fresh gas flow, etc.).
This setting is applicable in CMV & SIMV mode.

3.4.9 Peak pressure (PEAK P.) (PCV mode)

Set up the peak pressure in mbar with this button. The ventilator will regulate the flow
during inspiration to reach the peak pressure as fast as possible and keep it at this
level for the whole duration of inspiration. The peak pressure can be set from 5 - 60
hPa.
The peak pressure is visualized on the barograph (upper dotted line).
This setting is applicable in PCV mode only.

3.4.10 Frequency (FREQUENCY) (CMV / PCV/ SIMV mode)

The breathing frequency can be set from 4 - 80 breaths per minute.


This setting is applicable in CMV , PCV and SIMV mode.

3.4.11 Inspiration / expiration ratio (I/E ratio) (CMV / PCV / SIMV mode)

The I/E ratio can be set 1:1, 1:1.5 , 1:2 , 1:3, 1:4 , 1:5 , 1:6 , 2:1 ,3:1 and 4:1. This
value reflects the relation between inspiration and expiration and is therefore
applicable in CMV , PCV and SIMVmode.
For example: the I/E ratio is equal to 1:2, i.e. the duration of expiration is twice as
long as the duration of inspiration.

3.4.12 Inspiratoire pause (INSP. PAUSE) (CMV & SIMV mode)

The insp. pause can be set 0 - 50% in steps of 5%. This value is equal to the pause
used during inspiration expressed by percents of the total inspiration time.
This setting is applicable in CMV and SIMV mode only.

Saturn Evo ventilator 24 Operator’s manual


4th edition – June 2008
3.4.13 Positive end expiration pressure (PEEP) (CMV / PCV / SIMV mode)

The PEEP is the Postive End Expiration Pressure. The PEEP range is 0 - 20 mbar.

Note: When setting a (higher) PEEP, one has to take into consideration that
"building" a PEEP requires extra volume. This extra volume will be
delivered to the system through the fresh gas flow. Because the Saturn
Evo is suitable for very low fresh gas flows, the PEEP may be built
very slowly.

3.4.14 Trigger (SIMV / PCV mode)


Trigger below peep.
With this button the trigger function can be enabled or disabled.
The trigger value can be set from - 2 to - 20 mbar. The Max trigger level that can be
set is equal to the peep setting. Press the trigger button once; rotate control knob and
press control knob to enable/disable the trigger function. Press the trigger button
twice, rotate control knob and press control knob to change trigger value.

A trigger event is visualized in the trigger button and on the airway pressure graph
with the symbol .

Note: You must select a PEEP of minimum 2 mbar to use the trigger
function. If you select a PEEP lower than 2 mbar, the trigger function
is automatically switched off.

3.4.15 Sigh (CMV / SIMV mode)

The sigh button is shown when the graph window is selected . Press the sigh button
once to set sigh on. A sigh is generated every 64 breaths. During a sigh, a volume of
approximately 1,5 x tidal volume is delivered to the patient.

Press the sigh button once again for sigh off. This setting is applicable in CMV and
SIMV mode.

Note: The sigh function is standard off after switching the ventilator on.

Saturn Evo ventilator 25 Operator’s manual


4th edition – June 2008
3.5 OXYGEN MEASUREMENT

3.5.1 Principle of operation

The oxygen measurement is done with an O2 fuel cell. The oxygen sensor function is
similar to a typical battery. The sensor consists of an anode, a cathode and a unique
weak acid electrolyte. Oxygen molecules diffuse through a non-porous membrane and
are reduced at the cathode. The current generated between the electrodes is directly
proportional to the concentration of oxygen in the measured gas.
Just like a battery, the output voltage of the sensor drops at the end of its operating life
and the sensor will fail to measure oxygen concentrations.

3.5.2 Operation

Oxygen measurement is available in manual mode, CMV ,PCV and SIMV mode.

1 2 3

O2 measurement in CMV mode

1. O2 concentration: the actual measured O2 concentration expressed in % O2.

2. Low O2 alarm: lower O2 alarm can be set in the range from 18% to 99% O2.
If the O2 concentration is lower than the set value, the lower O2 alarm is
activated, which will be visual and audible.

3. High O2 alarm: high O2 alarm can be set in the range from 19% to 99% O2
and OFF. If the O2 concentration exceeds the upper O2 alarm, the O2 high
alarm is activated, which will be visual and audible.

Saturn Evo ventilator 26 Operator’s manual


4th edition – June 2008
To change the lower O2 alarm, make sure there is no other setting selected and press
the control knob once.

The lower O2 alarm will be activated. Change the value by rotating the control knob.
Press the control knob once to confirm the new value.

The high O2 alarm will not be selected. Change the high O2 alarm by rotating the
control knob. Press the control knob once to confirm the high alarm.

An alarm indicator is drawn on the screen after an alarm occurred.

alarm indicator

Press the control knob once to reset the alarm indicators.

3.5.3 Set-up the O2 measurement

Set the ventilator to standby mode and press the setup button to enter the setup menu.
Always disconnect the patient before entering the setup menu.

After entering the setup menu, swap to the SETUP O2 window by pressing the
button.

Saturn Evo ventilator 27 Operator’s manual


4th edition – June 2008
1

4
2 5

O2 setup menu 3

1. O2 alarm: sets up the low and high O2 alarm for use after turning the
ventilator on.

2. Calibration buttons: use these buttons to perform a 21% or 100% O2


calibration.

3. O2 sensor on / off: you can disable O2 measurement if no O2 sensor is


available.

4. O2 concentration: the actual O2 concentration is represented to check if a


calibration is necessary.

5. Status line: the status of the O2 module is shown over here. The following
error codes are available:

O2 module is not working properly.


Please contact your local service
engineer.

O2 module is working properly.

See section "Calibration" later


described in this manual.

See section "Calibration" later


described in this manual.

Saturn Evo ventilator 28 Operator’s manual


4th edition – June 2008
3.5.4 O2 calibration

Oxygen measurement is done by means of a "fuel cell" type sensor. These sensors do
have a certain lifecycle expressed by "O2 % hours".

Note: The lifecycle of the sensor is strongly reduced by high O2


concentrations.

Perform a 21% O2 calibration if the O2 sensor doesn’t show 21.0% when the sensor is
kept in ambient air. Perform a calibration each time the sensor is replaced.

You can perform two types of calibration:

• 1-points calibration

„ Put the O2 sensor in ambient air and wait for a stable reading.

„ Press the button.

„ The message "O2 calibration successful" is shown after a successful


calibration. The display reading will be equal to 21.0%.

„ The message "O2 calibration failed" is shown when the calibration fails.
The O2 sensor must be kept in ambient air. Also check the electrical
connections of the sensor. Replace the sensor if the problem still remains.

• 2-points calibration

„ After a successful 21% O2 calibration, put the O2 sensor in pure oxygen and
wait for a stable O2 reading.

„ Push the button.

„ The message "O2 calibration successful" is shown after a successful


calibration. The display reading will be equal to 100.0%.

„ The message "O2 calibration failed" is shown when the calibration fails.
The O2 sensor must kept in pure oxygen. Also check the electrical
connections of the sensor. Replace the sensor if the problem still remains.

Note: 100% O2 calibration is not necessary. You can reduce the sensor
linearity error with a 2-points calibration.

Press the button to store the calibration factors. Press the button to ignore
the new calibration factor. The old calibration factors are restored.

Saturn Evo ventilator 29 Operator’s manual


4th edition – June 2008
3.6 ALARMS, WARNINGS AND INDICATORS

1 11 10 12 9 8 7

2
6
3
3
5

1. Upper limit pressure alarm indicator: text is flashing when an upper limit
pressure alarm is active.

2. Lower limit pressure alarm indicator: text is flashing when a lower limit
pressure alarm is active.

3. Trigger status: symbol is active during a triggered inspiration/expiration


cycle.

4. MODE indicator: indicator is blinking during switching between the different


ventilator modes.

5. Warning and alarm window: window contains all kind of information about
the status of the ventilator.

6. Sigh status: symbol is shown when sigh is active during the complete
inspiration/expiration cycle.

7. Silence indicator: indicates that alarm sounds are suppressed by pressing the
silent button and a down counter (seconds starting at 60) is visual to give the
user an idea how long the alarm is suppressed.

8. High oxygen alarm

9. Low oxygen alarm

10. Low battery indicator: the battery symbol is blinking when the battery is
almost empty. A plug is drawn when working on mains supply.

Saturn Evo ventilator 30 Operator’s manual


4th edition – June 2008
11. Limit pressure alarm indicators: these indicators are blinking when a limit
pressure alarm is active or has happened. Cleared by pressing the pressure
limit button.

12. O2 SENSOR alarm : This indicator is blinking when the O2 sensor is


disconnected.

3.6.1 Lower limit pressure alarm (CMV / PCV / SIMV mode)

The lower limit pressure alarm is functional in CMV,PCV and SIMV mode. The
lower limit pressure alarm is visualized digitally, in the barograph and on the airway
pressure graph. The alarm will sound and the indicator will start blinking if the airway
pressure does not cross the preset level during the inspiration or when during
expiration the airway pressure stays below preset level. The alarm will automatically
be disabled as soon as the patient pressure exceeds the lower limit pressure during
inspiration.
You can silence the alarm for 1 minute by pressing the button. The silence
indicator will then start blinking ,the down counter is set to 60.

3.6.2 Apnea alarm (Man/Spont mode)

The apnea alarm is functional in the Man/Spont mode. If there is no


inspiration/expiration cycle detected within 20 seconds, the alarm will sound
(continuous beep) and a warning screen is displayed:

You can silence the alarm for 1 minute by pressing the button. The silence
indicator will then start blinking ,the down counter is set to 60.
The Apnea alarm can be switched OFF but when the ventilator is switched OFF and
ON the apnea alarm is set ON.

3.6.3 Upper limit pressure alarm (CMV,SIMV and Man/Spont mode)

The upper limit pressure alarm is functional in CMV,SIMVand Man/Spont mode.


When, during a full respiratory cycle, the pressure exceeds the upper pressure limit,
the alarm will be active.
The indicator will start blinking and an audible alarm will sound.

The ventilator will automatically evacuate gas to prevent unlimited pressure rise if
working in CMV or SIMV mode.

Saturn Evo ventilator 31 Operator’s manual


4th edition – June 2008
The ventilator will not evacuate gas during an upper limit pressure alarm in
Man/Spont mode. You have to check the setting of the safety valve on the handset
with the upper limit pressure value.

You can silence the alarm for 1 minute by pressing the button. The silence
indicator will then start blinking ,the down counter is set to 60.

The alarm will automatically be disabled as soon as the patient pressure no longer
exceeds the upper limit pressure.

Note: At the time of an upper limit pressure alarm, the ventilator may not be
able to deliver the preset tidal volume. This is because the safety valve
is opened and gas evacuation is taking place.

3.6.4 Switching between the different ventilation modes

When changing to another ventilator mode, the ventilator will first end his inspiration
/ expiration cycle and then switch over to the new selected mode.
While waiting to switch over, the mode text is blinking.

When switching to CMV, PCV or SIMV mode, several warning windows can be
displayed and the ventilator is waiting to go to the new selected mode due the
following reasons:

- Patient pressure more than 2 mbar


- Low oxygen input pressure
- Low nitrous oxide input pressure
- Low air input pressure
- Fresh gas volume higher than set volume

Look at the warning windows section (5.6.4) to solve these problems.

If working in CMV / PCV / SIMV mode, the ventilator can be forced to MAN mode
because a malfunction is detected. The following warning window is shown and the
mode text is blinking:

One or more warning windows display the reason why the ventilator is forced to
MAN mode. These warning windows are cleared when the ventilator is switched to
standby or MAN mode.

Saturn Evo ventilator 32 Operator’s manual


4th edition – June 2008
3.6.5 Warning window

The warning windows are projected above the read-outs of the measured values. Only
the expired tidal volume is visible. At the same time, a warning window is drawn.
A warning window indicates a malfunction, an alarm, a wrong setting, etc. of the
ventilator. It is a helpful item for the operator using the ventilator in a correct way.
When several warning windows have to be displayed, it will swap every 3 seconds to
another warning window.

The display will automatically switch over from the graph window to the main
window when warnings or alarms are active.

An overview of the possible warnings is given in the following table:

Warning window Mode Description

Apnea alarm occurred. A continuous beep is


MAN active during an apnea alarm.

The air input pressure is too low or the


ventilator fresh gas front output is selected.
MAN When working in CMV / PCV/SIMV mode, the
CMV ventilator will be forced to MAN mode. An
PCV audible alarm is active that can be suppressed
SIMV by the silence button.
Check absorber-front knob, air input
connection, connection tube and wall supply.

This warning is displayed when switching to


CMV CMV / PCV / SIMV mode. The airway pressure
PCV cannot be decreased. In order to prevent a
SIMV double inspiration, the ventilator will not switch
over until pressure is less than 2 mbar within
10sec.

The fresh gas transducer needs to be calibrated.


MAN Contact an authorized service engineer.
CMV Detailed maintenance procedures are covered in
PCV the technical manual.
SIMV

The fresh gas flow is too large for proper


CMV working of the ventilator.
PCV Decrease fresh gas flow by adjusting the
SIMV rotameter.

Saturn Evo ventilator 33 Operator’s manual


4th edition – June 2008
Warning window Mode Description

Unable to deliver set tidal volume. The internal


CMV balloon is completely empty after inspiration.
PCV Decrease frequency and/or tidal volume setting
SIMV or increase fresh gas flow.

The oxygen input pressure is too low. When


working in CMV / PCV / SIMV mode, the
MAN ventilator will be forced to MAN mode. An
CMV audible alarm is active that can be suppressed
PCV by the silence button.
SIMV Check oxygen input connection, connection
tube and wall supply.
The nitrous oxide input pressure is too low.
When working in CMV / PCV / SIMV mode,
MAN the ventilator will be forced to MAN mode. An
CMV audible alarm is active that can be suppressed
PCV by the silence button.
SIMV Check nitrous oxide input connection,
connection tube and wall supply.

If working in CMV / PCV / SIMV mode, the


ventilator can be forced to MAN mode because
CMV a malfunction is detected. One or more warning
PCV windows display the reason why the ventilator
SIMV is forced to MAN mode.

A major leak is detected or the patient is


CMV disconnected. Check patient, patient circuit
PCV and/or perform a leaktest. An audible alarm is
SIMV active that can be suppressed by the silent
button.

Check for proper connection of the patient


breathing circuit to the ventilator bottle.
CMV Perform a leaktest. Contact an authorized
PCV service engineer if the warning remains active.
SIMV Detailed maintenance procedures are covered in
the technical manual.

The airway pressure is higher than the upper


limit pressure and the alarm is activated. Adjust
MAN the safety valve on the handset, or check upper
limit pressure value to prevent this alarm.

Saturn Evo ventilator 34 Operator’s manual


4th edition – June 2008
Warning window Mode Description

CMV Ventilator is not able to work with the actual


PCV settings. Adjust ventilator settings.
SIMV

Maximum tank volume reached. Ventilator is


CMV unable to deliver set volume. Decrease
PCV frequency or tidal volume setting.
SIMV

Saturn Evo ventilator 35 Operator’s manual


4th edition – June 2008
3.7 PERFORMING A LEAKTEST

3.7.1 Entering the leaktest menu

Set the ventilator to standby mode and press the leaktest button. Always disconnect
the patient before entering the leaktest menu.

Note: It is not possible to enter the leaktest menu when the airway pressure is
greater than 2 hpa/2 mbar. Disconnect patient so that pressure can
decrease to 0 hpa/0 mbar.

The leaktest menu is drawn in accordance with the next drawing.


6
1

2
5

3 4

1. Patient circuit + internal bag button: press this button to select the patient
breathing circuit and the internal balloon leaktest.

2. Patient circuit + handset button: press this button to select the patient
breathing circuit and the external handset.

3. EXIT: press this button to leave the leaktest menu.

4. PEEP valve leakage detected: this message appears when the PEEP valve
leakage test fails.

5. Information window: follow these instructions to perform a leaktest.

6. Barograph: indicates airway pressure.

Saturn Evo ventilator 36 Operator’s manual


4th edition – June 2008
3.7.2 Performing the leaktest

You can perform a leaktest on:

- the patient breathing circuit and the external handset.


- the patient breathing circuit and the internal balloon.

To perform a leaktest on the patient breathing circuit and the external handset,
proceed as follows:

- Press button.
- Make all the connections in the patient breathing unit; see the
operating instructions of the CO2 absorber.
- Close safety valve on handset.
- Close / disconnect sample tube of gas monitor (optional).
- Close the output of the Y-piece.
- Set a fresh gas flow.
- Let the airway pressure increase to 30 hPa.
- Close the fresh gas flow.
- The time in which the pressure decreases to 20 hPa must be
15 seconds or longer.

If the leak test is not within the specifications, check all connections in the patient
breathing circuit. If no leaks can be found, contact your service department.

After successful performing the previously described leaktest, you can perform a
leaktest on the patient breathing circuit and the internal balloon.
Proceed as follows:

- Press button.
- Close the output of the Y-piece.
- Set a fresh gas flow.
- Let the airway pressure increase to 30 hPa.
- Close the fresh gas flow.
- The time in which the pressure decreases to 20 hPa must be
15 seconds or longer.

If the leak test is not within the specifications, check all connections in the patient
breathing circuit. Also check the internal balloon for leaks. If no leaks can be found,
contact your service department.

Saturn Evo ventilator 37 Operator’s manual


4th edition – June 2008
When pressing the button to perform a leaktest, at the same time the PEEP
valve located in the patient breathing unit is filled once with a pressure of ±100 mbar.
The pressure in the PEEP valve will be monitored during 15 seconds. If a pressure
drop is detected within these 15 seconds, the message "PEEP valve leakage detected"
will be displayed:

To repeat the PEEP valve test, press the button followed by a button
press.

To determine either the PEEP leakage is intern in the ventilator or extern in the
patient breathing unit, remove the patient circuit from bottle and close the PEEP
output with finger in accordance with next drawing.

Keep your finger on the PEEP output and repeat the PEEP valve test. If the PEEP
valve test succeeds, the problem is located in the patient breathing unit, otherwise the
problem is probably intern in the ventilator. Please contact your service department.

3.7.3 Leaving the leaktest menu

Leave the leaktest menu by pressing the exit button. The ventilator will return to
standby mode.

Saturn Evo ventilator 38 Operator’s manual


4th edition – June 2008
3.8 THE SETUP MENU

3.8.1 Entering the setup menu

Set the ventilator to standby mode and press the setup button to enter the setup menu.
Always disconnect the patient before entering the setup menu.

3.8.2 The setup menu

You can swap between the different setup menus by pressing the button.

• Setup window 1

Setup window 1 is equal to the CMV window. The parameters for use during
CMV mode after turning the ventilator on can be set here.
Press the button to switch to the next setup window.

• Setup window 2

Saturn Evo ventilator 39 Operator’s manual


4th edition – June 2008
Setup window 2 is equal to the PCV window. The parameters for use during PCV
mode after turning the ventilator on can be set here. Press the button to switch
to the next setup window.

• Setup window 3

Setup window is equal to the SIMV window. The parameters for use during
SIMV mode after turning the ventilator on can be set here. Press the button to
switch to the next setup window.

• Setup window 4

See paragraph ‘3.5.3 Set up the O2 measurement’ for a description of O2 setup


menu. Press the button to switch to the next setup window.

Saturn Evo ventilator 40 Operator’s manual


4th edition – June 2008
• Setup window 4

6
5

2 3 4

1. Window select: press this button to switch between the different setup
windows.

2. Language select: set up the language to be used by the ventilator. You can
chose between two different languages.

3. Cancel: press this button to cancel the changes made in the setup menu.

4. Save: store the changes made in the setup menu to the memory.

5. Sound volume: set up the speaker sound volume in the range 1-5.

6. Date: set up the day, month, century and year.

7. Time: set up hours and minutes.

8. Ventilator information window: the Saturn Evo serial number (S/N), the
software revision numbers (S/R) and the total working time in hours of the
ventilator are displayed.

Note: The 'cancel' and 'save' button have no effect on date and time. A
change of date and/or time will be stored immediately into the
memory.

Rotating the encoder knob will change the position of the selection box and adjust a
selected value by pressing the control knob. The value is inverted and a rotation of the
control knob will adjust the value. Confirm the adjusted value by pressing the control
knob.

Saturn Evo ventilator 41 Operator’s manual


4th edition – June 2008
3.9 BATTERY BACKUP

The Saturn Evo ventilator is fitted with a battery backup system that allows at least
1 hour of operation if no mains supply is available. If the mains indicator is off and
the battery symbol is drawn on the display, the battery is in use.

If the battery is almost exhausted, the low battery symbol is blinking and every
minute, an audible sound is generated.
Reconnect the mains supply or prepare for manual controlled ventilation.
A continuous beep will sound for 30 seconds after the ventilator shuts off.

The internal battery is automatically charged when the ventilator is connected to the
mains power. The charging of the battery is electronically controlled to assure
optimum battery operating conditions.

The battery does not require maintenance under normal circumstances.

Note: To check whether the battery backup is functioning properly,


disconnect the mains supply by pulling the mains plug. The ventilator
should be able to operate normally in all modes of operation.

Saturn Evo ventilator 42 Operator’s manual


4th edition – June 2008
4. CLEANING AND DISINFECTING

4.1 VENTILATOR HOUSING

The front panel of the Saturn Evo is made of a high-grade polycarbonate foil. All
common household-cleaning products can be used without harm to the front panel
surface.
Aggressive products may affect the foil.
The ventilator housing may be cleaned with a moisturized cloth.

Caution: Use cleaning solution sparingly. It may flow into the ventilator and
cause internal damage.

4.2 PATIENT BREATHING CIRCUIT

The patient breathing unit can easily be removed for disinfection by pulling the
handle (figure 4.1).

The metal part can be steam-sterilized (max. 138 °C) or gas-sterilized.

When putting back the patient breathing circuit, it is recommended to manually guide
the bag to prevent it from damage (figure 4.2).

Note: The patient breathing circuit may never be immersed in liquid. This
may result in obstruction of valves and tubes, causing the patient
breathing circuit to malfunction.
It is of utmost importance to use undamaged bags, recommended by
Medec Benelux NV. When in doubt, always replace the bag by a new
one.

It is recommended to have further maintenance performed by an authorized service


engineer. Detailed maintenance procedures are covered in the technical manual.

Saturn Evo ventilator 43 Operator’s manual


4th edition – June 2008
Figure 4.1 – Removing the patient breathing circuit

Figure 4.2 – Putting back the patient breathing circuit

Saturn Evo ventilator 44 Operator’s manual


4th edition – June 2008
5. MAINTENANCE, MALFUNCTION AND
TROUBLESHOOTING

5.1 MAINTENANCE

To assure trouble-free operation of the ventilator, the following maintenance activities


should be carried out:

• Each time the ventilator is used

- Perform a leaktest.
- Check the battery backup system.
- Check the absorber sodalime; (re-)fill if necessary.
- Check the absorber water trap.

• Every 6 months

- Preventive maintenance.

This service must be carried out by an authorized service engineer.

- Check the ventilator calibration settings and recalibrate if necessary.


- Visual inspection of the ventilator internal and external connections,
electronic boards and valves, etc.

Note: Detailed maintenance procedures are covered in the technical manual.

Saturn Evo ventilator 45 Operator’s manual


4th edition – June 2008
5.2 MALFUNCTION – ERROR CODES

If a pneumatic or electronic failure occurs during use or during power-on selftest, an


audible alarm will sound. The display will show an error code and the ventilator will
automatically switch over to spont/man mode.

During the alarm, even if electronics are turned off, the patient can be ventilated by
means of the hand balloon.
This type of failure will require the attention of an authorized service engineer, who
can easily test the basic functions by means of the incorporated software.

A description of the possible error codes is given in the following table:

Error code Error description


1 Display board - Internal program memory error
2 Display board - Internal SRAM error
3 Display board - Internal timer error
4 Display board - Internal EEPROM error
5 Display board - Internal watchdog error
6 Reserved
7 Reserved
8 Reserved
9 Reserved
10 Reserved
11 Display board - External flash error
12 Display board - External SRAM error
13 Display board - External display driver error
14 Reserved
15 Reserved
16 Reserved
17 Reserved
18 Reserved
19 Board communication failure
20 Reserved
21 Keyboard controller - Internal program memory error
22 Keyboard controller - Internal SRAM error
23 Keyboard controller - Internal timer error
24 Keyboard or control knob blocked
25 Keyboard controller - Internal watchdog error
26 Keyboard controller - Communication timeout
27 Keyboard controller not detected
28 Board communication error - 25 ms timeout
29 Board communication error - Parity error start-up
30 Board communication error - Parity error start-receive byte
31 Board communication error - Receive string length error
32 Board communication error - Eot counter out of range
33 Board communication error - Target slave not ready
34 Board communication error - Transmit string wrong
35 Reserved

Saturn Evo ventilator 46 Operator’s manual


4th edition – June 2008
Error code Error description
36 Reserved
37 Reserved
38 Reserved
39 Reserved
40 Reserved
41 Pneumatic board global error
42 Reserved
43 Reserved
44 Pneumatic board – Not able to start A/D converter
45 Pneumatic board – A/D converter busy time-out
46 Pneumatic board – A/D readings equal to $FFF
47 Pneumatic board – A/D readings equal to $000
48 Pneumatic board – Internal EEPROM memory error
49 Pneumatic board – 12V input voltage too low
50 Pneumatic board – 12V valves enable circuit error
51 Pneumatic board – 12V valves disable circuit error
52 Pneumatic board – Not able to fill tank
53 Pneumatic board – Not able to stop tank filling
54 Pneumatic board – Tank low flow valve close time
55 Pneumatic board – Tank high flow valve close time
56 Pneumatic board – End exp. pressure to high
57 Pneumatic board – PIP press. higher than upp.lim.
58 Pneumatic board – Not able to lower tank pressure
59 … 99 Reserved

Saturn Evo ventilator 47 Operator’s manual


4th edition – June 2008
5.3 TROUBLESHOOTING

If the ventilator is not working properly as described in this manual, try to resolve the
problem by means of the following table.

If the malfunction cannot be corrected using the suggestions in the table below, an
authorized service engineer must perform further troubleshooting.

Symptom Possible causes Action


The ventilator is off and Battery completely Check power cord and power
there is no response when exhausted switch at the back of the
pressing the ON/OFF trolley
button Blown fuses
Replace fuses at the back of
The ventilator shuts off Battery at end of life time the trolley
after a short period of time
Replace battery (performed
by an authorized service
engineer)
Bad and/or unstable Calibration factors not Contact authorized service
operation in CMV/SIMV up-to-date engineer
or PCV mode
Can not swap to the graph Alarms are still active Change the settings so that
window by rotating the no alarms are active
control knob Warnings are displayed
on the screen Resolve warnings

Control knob failure Check possible control knob


failure by changing a setting
The ventilator is working An obstruction in the Check for obstruction in the
in MAN mode, there is no patient breathing circuit patient breathing circuit
patient connected and the
airway pressure is not zero Calibration factors not If the problem can’t be
up-to-date resolved, contact authorized
service engineer
Upper limit pressure alarm An obstruction in the Check upper limit pressure
patient breathing circuit setting

Check for obstructions in the


patient breathing unit
Major leakage not able to No freshgas flow. Perform both leaktests.
deliver set volume Tubes disconnected
Leakage in circuit Disconnect tubes from
patient breathing unit and
Zero level TRX1 check level patient pressure
negative Must be zero.
If the problem can’t be
Broken bottle venting resolved, contact authorized
valve service engineer

Saturn Evo ventilator 48 Operator’s manual


4th edition – June 2008
Symptom Possible causes Action
Can not enter leaktest Airway pressure more Disconnect patient and try
menu than 2 mbar again

5.4 FUSE REPLACEMENT

Replace the power mains fuses by means of the following instructions:

- Turn off the Saturn Evo and disconnect the power cord from the mains
supply.

- Insert a screwdriver in the small AC receptacle and pull out the fuse
holder block.

- Replace the blown fuses and place the fuse holder block back into the
AC receptacle.

- Connect the power cord back to the mains supply.

Note: Always replace the blown fuses with the same type and ratings.

Saturn Evo ventilator 49 Operator’s manual


4th edition – June 2008
6. SPECIFICATIONS

6.1 GENERAL
Operating principle: Bag-in-bottle

Volume adjustment: Semi-closed loop control

Flow pattern: Decelerating (CMV)


Decelerating (SIMV)
Decelerating (PCV)

Alarms: Lower limit pressure / apnea


Upper limit pressure
Low battery
Malfunction
Low air input pressure
Low oxygen input pressure
Low nitrous oxide input pressure
PEEP valve leakage detect
Major leak or patient disconnected

Warnings: Several warnings to indicate incorrect


use of the ventilator by the operator.

System screen: EL display


Multi-language: Multiple languages available (on request)

6.2 RATINGS
Main supply: 100 - 240V, 50 - 60Hz
Power consumption: < 100VA
IEC 601 class and type: Class I, type B

Battery type: Sealed lead-acid battery


Battery run time: minimal 1 hour when fully charged
Battery life time: Approx. 3 years at 20°C

Pneumatic supply: compressed air or oxygen


Pressure: 3 - 8 bar
Connection: According to NIST
Gas evacuation: Max. – 0,5 bar
O2 flush: 35 liter/minute
External fresh gas outlet: front output and absorber output (selectable)

Ventilation modes: CMV, PCV,SIMV and Man/Spont mode


Tidal volume (CMV): 10 – 1600 ml
Peak pressure (PCV): 5 – 60 hPa / 5 – 60 mbar
Rate: 4 – 80 bpm
PEEP: 0 – 20 hPa / 0 – 20 mbar
Inspiratoire pause: 0 – 50 %
I/E ratio: 1:1,1:1.5, 1:2, 1:3, 1:4, 1:5,1:6,2:1,3:1,4:1.
Trigger: OFF; -2 to –20 hPa below PEEP
Sigh: approximately 1,5 x tidal volume (CMV/SIMV)
Airway pressure: -10 – 100 hPA / -10 – 100 mbar
Lower limit pressure: 2 – 94 hPa / 2 – 94 mbar
Upper limit pressure: 7 – 99 hPa / 7 – 99 mbar

Saturn Evo ventilator 50 Operator’s manual


4th edition – June 2008

You might also like