Professional Documents
Culture Documents
Saturn Evo Operators Manual 4th Edition 2008 PDF
Saturn Evo Operators Manual 4th Edition 2008 PDF
Saturn Evo Operators Manual 4th Edition 2008 PDF
OPERATOR’S MANUAL
4th edition
JUNE 2008
SATURN EVO ANAESTHESIA SYSTEM
OPERATOR’S MANUAL
0499
4th edition
June 2008
MEDEC BENELUX NV
All rights reserved. No part of this publication may be reproduced, stored in a
retrieval system or transmitted, in any form or by any means, electronically,
mechanically, by photocopying, recording or otherwise, without the prior written
permission of Medec Benelux NV.
Medec Benelux NV
Lion D’Orweg 19
9300 Aalst
Belgium
Telephone : (32) 53 / 70.35.44
Fax : (32) 53 / 70.35.33
Website : www.medecbenelux.be
E-mail : sales@medecbenelux.be
4th edition
June 2008
Table of Contents
1. Saturn Evo system overview
1.1 General 1
1.2 User responsibility 1
1.3 Main components 1
1.1 GENERAL
Personnel should make themselves familiar with the contents of this manual before
using the anaesthesia system.
The equipment described in this manual has been built to conform with the
specifications and operating instructions stated in this manual. To ensure proper and
safe operation of the equipment, it must be checked and serviced to at least the
minimum standards laid out in this manual.
The equipment must be repaired and serviced only in accordance with written
instructions issued by Medec Benelux NV and must not be altered or modified in any
way without written approval of Medec Benelux NV. The user of this equipment shall
have the responsibility for any malfunction which results from improper use,
maintenance, repair, damage or alteration by anyone other than Medec Benelux NV or
its appointed agents.
Monitor shelf
Electronic ventilator
Writing shelf
Absorber system
Drawer unit
suction jar
• Suction (fig. 2)
The collection jar is connected to the suction input connector (1) and operated
via the ventilator front panel.
FIG 1
Schematic diagram
• Top tray
Instruments, monitoring etc. may be put on the top tray (1).
• Light (3)
Built-in light. 3 positions + off.
• Drawers (4)
The drawers may be used to store instruments, etc.
FIG 3
0499
4th edition
June 2008
MEDEC BENELUX NV
All rights reserved. No part of this publication may be reproduced, stored in a
retrieval system or transmitted, in any form or by any means, electronically,
mechanically, by photocopying, recording or otherwise, without the prior written
permission of Medec Benelux NV.
Medec Benelux NV
Lion D’Orweg 19
9300 Aalst
Belgium
Telephone : (32) 53 / 70.35.44
Fax : (32) 53 / 70.35.33
Website : www.medecbenelux.be
E-mail : sales@medecbenelux.be
4th edition
June 2008
Table of Contents
1. General information 1
1.1 General 1
1.2 User responsibility 1
2. System overview 2
2.1 Front view 3
2.2 Side view 4
2.3 Back view 5
2.4 Schematic diagram of gas supply 6
2.5 Pneumatic connections 7
2.6 Front 9
2.7 Electrical connections 10
3. Operation 11
3.1 Operator interface 11
3.2 Switch on the ventilator 12
3.3 The display screen 13
3.4 Operating instructions 20
3.4.1 Preparations before use 20
3.4.2 Changing button parameters 20
3.4.3 Standby mode 21
3.4.4 Spontaneous / manual ventilation mode (MAN) 21
3.4.5 Controlled mandatory ventilation mode (CMV) 22
3.4.6 Pressure controlled ventilation mode (PCV) 22
3.4.7 Synchronized mandatory ventilation mode (SIMV) 23
3.4.8 Tidal volume (VOLUME) (CMV /SIMV mode) 24
3.4.9 Peak pressure (PEAK P.) (PCV mode) 24
3.4.10 Frequency (FREQUENCY) (CMV / SIMV / PCV mode) 24
3.4.11 Inspiration / expiration ratio (I/E ratio) (CMV / SIMV / PCV mode) 24
3.4.12 Inspiratoire pause (INSP. PAUZE) (CMV / SIMV mode) 24
3.4.13 Positive end expiration pressure (PEEP) (CMV / SIMV /PCV mode)25
3.4.14 Trigger (SIMV / PCV mode) 25
3.4.15 Sigh (CMV/ SIMV mode) 25
3.5 Oxygen measurement 26
3.5.1 Principle of operation 26
3.5.2 Operation 26
3.5.3 Set-up the O2 measurement 27
3.5.4 O2 calibration 29
3.6 Alarms, warnings and indicators 30
3.6.1 Lower limit pressure alarm (CMV/SIMV/ PCV mode) 31
3.6.2 Apnea alarm (Man/Spont mode) 31
3.6.3 Upper limit pressure alarm (CMV / SIMV and Man/Spont mode) 31
3.6.4 Switching between the different ventilation modes 32
3.6.5 Warning window 33
3.7 Performing a leaktest 36
3.7.1 Entering the leaktest 36
3.7.2 Performing the leaktest 37
3.7.3 Leaving the leaktest menu 38
3.8 The setup menu 39
3.8.1 Entering the setup menu 39
3.8.2 The setup menu 39
3.9 Battery backup 42
6. Specifications 50
6.1 General 50
6.2 Ratings 50
1. GENERAL INFORMATION
1.1 GENERAL
Personnel should make themselves familiar with the contents of this manual before
using the anesthesia system.
The equipment described in this manual has been built to conform with the
specifications and operating instructions stated in this manual. To ensure proper and
safe operation of the equipment, it must be checked and serviced to at least the
minimum standards laid out in this manual.
The equipment must be repaired and serviced only in accordance with written
instructions issued by Medec Benelux NV and must not be altered or modified in any
way without written approval of Medec Benelux NV. The user of this equipment shall
have the responsibility for any malfunction which results from improper use,
maintenance, repair, damage or alteration by anyone other than Medec Benelux NV or
its appointed agents.
The Saturn Evo combines some never seen abilities, high technological research and a
high level of security. Clear monitoring gives you full control at a single glance. Our
well-studied platform is easy to handle and gives you many exceptional features. The
platform is based on the well-known horizontal bag-in-bottle system.
The Saturn Evo is an anesthesia ventilator that can handle spontaneous breathing as
well. The system incorporates a semi-closed breathing circuit, to which a fresh gas
flow is continuously directed.
The patient inspires and expires either in the hand balloon (in spontaneous/manual
mode) or in the bag (in CMV,SIMV or PCV mode).
CO2 accumulation is prevented by means of an absorber.
This manual will describe all control buttons and functions of the electronic
ventilator. Both the alarm system and the cleaning procedures will be attended to as
well.
Oxygen, nitrous oxide and air from the wall supply flows through the rotameter and
vaporizer-section to the selected fresh gas outlet (absorber or front outlet). Via a
bypass, oxygen can also flow directly to the fresh gas outlet. Air is used as pneumatic
supply to the ventilator. If no wall supply is available, oxygen and nitrous oxide can
be supplied by high-pressure cylinders on the back of the trolley.
6 5
3 4
7 8 9 10
2. N2O NIST connection: two N2O NIST connectors for wall and/or cylinder
supply.
3. Air NIST connection: air NIST connection for wall or cylinder supply.
7. Patient Evacuation.
1 2 3
4 5 6
1. O2 input pressure indicator.
2 3
1. The mains power inlet: connect the power cord with the mains power inlet.
Turn switch ON. The ventilator is working on mains supply and the battery is
charging. The green mains indicator at the front will be lit. The battery will not
be charged with the main switch in the OFF position. The ventilator will work
on battery and the green mains indicator is off.
3. Mains outlet: outlet for connection with 4 mains outlets located at upper back
panel.
5
6
3. Silent button: all the alarms are mute for one minute when this button is
pressed.
4. ON / OFF button: press this button for at least 1 second to switch the
ventilator on/off.
As soon as the power cord is connected to mains supply and the main switch is turned
on, the mains indicator will be illuminated.
When a failure is detected, the display screen will show an error code and a short
description of the failure. Refer to chapter ‘TROUBLESHOOTING’ for more
information about the error codes.
After turning the ventilator on, it will always start up in the standby window.
3 Standby window
1. Setup: the setup button in the standby window is used to access the setup
menu.
2. Leaktest: the leaktest button in the standby window is used to access the leak
test menu.
3. Mode: with the mode button, you can switch between the different ventilator
modes: CMV,SIMV, PCV, Man/Spont and standby mode. Press the mode
button, change the value by rotating the control knob and execute by pressing
the control knob. You can cancel by pressing the mode button once again.
The settings of the ventilator are presented on a number of large screen push buttons.
13 12 11 10 9
1
8
4 CMV window
5 6 7
1. Limit pressure: sets upper limit and lower limit pressure.
2. PEEP: the PEEP can be adjusted by using this button. The PEEP range is
between 0 - 20 hPa / 0 – 20 mbar.
4. Mode: with the mode button, you can change between CMV,PCV,SIMV,
Man/Spont and standby mode.
5. Tidal volume: the tidal volume can be set from 10 ml to 1600 ml.
6. Frequency: the frequency range can be set in the range from 4 - 80 breaths
per minute. Limitations on the frequency depending on the I/E ratio setting.
8. Measurement window: this window contains the measured values from the
last inspiration / expiration cycle and the barograph. These values are updated
at the end of each expiration. The measured values are: expired tidal volume,
minute volume (MIN.VOL.), peak inspiration pressure (PIP), plateau pressure
(PLATEAU), positive end expiration pressure (PEEP), mean pressure
(MEAN) and compliance (COMPL.). The barograph is updated continuously.
9. Silent indicator: the silent symbol is flashing after the silent button is pressed,
and no apnea alarm is active.
A down counter (starting from 60 seconds) gives an indication how long the
Alarm disable function will be active.
12. Mains / battery indicator: when working on mains supply, the following
symbol is shown . When working on batteries, a battery symbol
indicates the battery status .
13. Barograph: the barograph indicates airway pressure in the range from
-10 hPa / -10mbar to 100 hPa / 100mbar.
SIMV window
1.Trigger: sets the trigger on/off and adjusts the trigger level –2 to –20 hPa
-2 to -20mbar.
Trigger is a trigger below Peep and Max adjusted trigger level that can be set is
equal to peep setting.
All the other settings and buttons are similar as described in CMV mode.
PCV window
1
All the other settings and buttons are similar as described in CMV and SIMV mode.
Man/Spont window
8
2
7
3
4
6
Graph window
5
1. Pressure scale: changes the graph pressure scale between 25 hPa/25 mbar,
50 hPa/50 mbar and 100 hPa/100 mbar.
2. Time scale: changes the graph time scale between 5 sec, 10 sec and
20 sec.
3. Freeze / Run: freeze the graph by pressing this button. Pressing this button a
second time, will set the graph in run mode.
8. Slider: the slider will synchronize with the start of a new inspiration. The
slider waits at the start of the graph until a new inspiration takes place. In
MAN mode, the slider is continuously running independent of the inspiration /
expiration cycle.
- connect manual system with the hand balloon and patient system.
Rotating the control knob changes the value of the setting. Pushing the control knob
will save the changed value. Pressing the selected button again will cancel the
changed parameter value and the value restores to the previous value.
The ventilator will always start in "standby mode". From the standby window you can
perform following actions:
The spontaneous / manual mode (MAN mode) allows spontaneous breathing of the
patient and manually controlled ventilation with a hand balloon. The ventilator will be
in this mode if:
In MAN mode, the Saturn Evo can even be used without electrical and/or compressed
air connections. When no electrical connection is available, the readouts of the
measured values, alarms and warnings will of course not be displayed.
Switch the ventilator to the pressure controlled ventilation mode (PCV mode) with the
use of the mode button.
The following screen pops up:
Push the control knob to confirm and switch over to PCV mode or rotate selector
knob to select cancel.
In CMV / PCV/ SIMV mode, several controls are available. If a control is changed,
the new value will be used after a complete expiration cycle. The lower limit pressure
and upper limit pressure values will be used immediately.
When switching over from MAN mode to CMV / PCV/ SIMV mode, the Saturn Evo
will start with an inspiration, when patient pressure is lower than 2mbar.
The ventilator will wait to go to CMV / PCV/ SIMV mode due the following reasons:
The MODE text will be blinking during waiting changing mode from MAN to
CMV/PCV/SIMV or CMV/PCV/SIMV to MAN.
The tidal volume is set in ml. It may take a few breathing cycles before the set volume
is reached.
The Saturn Evo electronics will automatically correct any deviation to the preset
volume. This correction is being calculated by a microprocessor, based on a number
of parameters (compliance, leakage, fresh gas flow, etc.).
This setting is applicable in CMV & SIMV mode.
Set up the peak pressure in mbar with this button. The ventilator will regulate the flow
during inspiration to reach the peak pressure as fast as possible and keep it at this
level for the whole duration of inspiration. The peak pressure can be set from 5 - 60
hPa.
The peak pressure is visualized on the barograph (upper dotted line).
This setting is applicable in PCV mode only.
3.4.11 Inspiration / expiration ratio (I/E ratio) (CMV / PCV / SIMV mode)
The I/E ratio can be set 1:1, 1:1.5 , 1:2 , 1:3, 1:4 , 1:5 , 1:6 , 2:1 ,3:1 and 4:1. This
value reflects the relation between inspiration and expiration and is therefore
applicable in CMV , PCV and SIMVmode.
For example: the I/E ratio is equal to 1:2, i.e. the duration of expiration is twice as
long as the duration of inspiration.
The insp. pause can be set 0 - 50% in steps of 5%. This value is equal to the pause
used during inspiration expressed by percents of the total inspiration time.
This setting is applicable in CMV and SIMV mode only.
The PEEP is the Postive End Expiration Pressure. The PEEP range is 0 - 20 mbar.
Note: When setting a (higher) PEEP, one has to take into consideration that
"building" a PEEP requires extra volume. This extra volume will be
delivered to the system through the fresh gas flow. Because the Saturn
Evo is suitable for very low fresh gas flows, the PEEP may be built
very slowly.
A trigger event is visualized in the trigger button and on the airway pressure graph
with the symbol .
Note: You must select a PEEP of minimum 2 mbar to use the trigger
function. If you select a PEEP lower than 2 mbar, the trigger function
is automatically switched off.
The sigh button is shown when the graph window is selected . Press the sigh button
once to set sigh on. A sigh is generated every 64 breaths. During a sigh, a volume of
approximately 1,5 x tidal volume is delivered to the patient.
Press the sigh button once again for sigh off. This setting is applicable in CMV and
SIMV mode.
Note: The sigh function is standard off after switching the ventilator on.
The oxygen measurement is done with an O2 fuel cell. The oxygen sensor function is
similar to a typical battery. The sensor consists of an anode, a cathode and a unique
weak acid electrolyte. Oxygen molecules diffuse through a non-porous membrane and
are reduced at the cathode. The current generated between the electrodes is directly
proportional to the concentration of oxygen in the measured gas.
Just like a battery, the output voltage of the sensor drops at the end of its operating life
and the sensor will fail to measure oxygen concentrations.
3.5.2 Operation
Oxygen measurement is available in manual mode, CMV ,PCV and SIMV mode.
1 2 3
2. Low O2 alarm: lower O2 alarm can be set in the range from 18% to 99% O2.
If the O2 concentration is lower than the set value, the lower O2 alarm is
activated, which will be visual and audible.
3. High O2 alarm: high O2 alarm can be set in the range from 19% to 99% O2
and OFF. If the O2 concentration exceeds the upper O2 alarm, the O2 high
alarm is activated, which will be visual and audible.
The lower O2 alarm will be activated. Change the value by rotating the control knob.
Press the control knob once to confirm the new value.
The high O2 alarm will not be selected. Change the high O2 alarm by rotating the
control knob. Press the control knob once to confirm the high alarm.
alarm indicator
Set the ventilator to standby mode and press the setup button to enter the setup menu.
Always disconnect the patient before entering the setup menu.
After entering the setup menu, swap to the SETUP O2 window by pressing the
button.
4
2 5
O2 setup menu 3
1. O2 alarm: sets up the low and high O2 alarm for use after turning the
ventilator on.
5. Status line: the status of the O2 module is shown over here. The following
error codes are available:
Oxygen measurement is done by means of a "fuel cell" type sensor. These sensors do
have a certain lifecycle expressed by "O2 % hours".
Perform a 21% O2 calibration if the O2 sensor doesn’t show 21.0% when the sensor is
kept in ambient air. Perform a calibration each time the sensor is replaced.
• 1-points calibration
Put the O2 sensor in ambient air and wait for a stable reading.
The message "O2 calibration failed" is shown when the calibration fails.
The O2 sensor must be kept in ambient air. Also check the electrical
connections of the sensor. Replace the sensor if the problem still remains.
• 2-points calibration
After a successful 21% O2 calibration, put the O2 sensor in pure oxygen and
wait for a stable O2 reading.
The message "O2 calibration failed" is shown when the calibration fails.
The O2 sensor must kept in pure oxygen. Also check the electrical
connections of the sensor. Replace the sensor if the problem still remains.
Note: 100% O2 calibration is not necessary. You can reduce the sensor
linearity error with a 2-points calibration.
Press the button to store the calibration factors. Press the button to ignore
the new calibration factor. The old calibration factors are restored.
1 11 10 12 9 8 7
2
6
3
3
5
1. Upper limit pressure alarm indicator: text is flashing when an upper limit
pressure alarm is active.
2. Lower limit pressure alarm indicator: text is flashing when a lower limit
pressure alarm is active.
5. Warning and alarm window: window contains all kind of information about
the status of the ventilator.
6. Sigh status: symbol is shown when sigh is active during the complete
inspiration/expiration cycle.
7. Silence indicator: indicates that alarm sounds are suppressed by pressing the
silent button and a down counter (seconds starting at 60) is visual to give the
user an idea how long the alarm is suppressed.
10. Low battery indicator: the battery symbol is blinking when the battery is
almost empty. A plug is drawn when working on mains supply.
The lower limit pressure alarm is functional in CMV,PCV and SIMV mode. The
lower limit pressure alarm is visualized digitally, in the barograph and on the airway
pressure graph. The alarm will sound and the indicator will start blinking if the airway
pressure does not cross the preset level during the inspiration or when during
expiration the airway pressure stays below preset level. The alarm will automatically
be disabled as soon as the patient pressure exceeds the lower limit pressure during
inspiration.
You can silence the alarm for 1 minute by pressing the button. The silence
indicator will then start blinking ,the down counter is set to 60.
You can silence the alarm for 1 minute by pressing the button. The silence
indicator will then start blinking ,the down counter is set to 60.
The Apnea alarm can be switched OFF but when the ventilator is switched OFF and
ON the apnea alarm is set ON.
The ventilator will automatically evacuate gas to prevent unlimited pressure rise if
working in CMV or SIMV mode.
You can silence the alarm for 1 minute by pressing the button. The silence
indicator will then start blinking ,the down counter is set to 60.
The alarm will automatically be disabled as soon as the patient pressure no longer
exceeds the upper limit pressure.
Note: At the time of an upper limit pressure alarm, the ventilator may not be
able to deliver the preset tidal volume. This is because the safety valve
is opened and gas evacuation is taking place.
When changing to another ventilator mode, the ventilator will first end his inspiration
/ expiration cycle and then switch over to the new selected mode.
While waiting to switch over, the mode text is blinking.
When switching to CMV, PCV or SIMV mode, several warning windows can be
displayed and the ventilator is waiting to go to the new selected mode due the
following reasons:
If working in CMV / PCV / SIMV mode, the ventilator can be forced to MAN mode
because a malfunction is detected. The following warning window is shown and the
mode text is blinking:
One or more warning windows display the reason why the ventilator is forced to
MAN mode. These warning windows are cleared when the ventilator is switched to
standby or MAN mode.
The warning windows are projected above the read-outs of the measured values. Only
the expired tidal volume is visible. At the same time, a warning window is drawn.
A warning window indicates a malfunction, an alarm, a wrong setting, etc. of the
ventilator. It is a helpful item for the operator using the ventilator in a correct way.
When several warning windows have to be displayed, it will swap every 3 seconds to
another warning window.
The display will automatically switch over from the graph window to the main
window when warnings or alarms are active.
Set the ventilator to standby mode and press the leaktest button. Always disconnect
the patient before entering the leaktest menu.
Note: It is not possible to enter the leaktest menu when the airway pressure is
greater than 2 hpa/2 mbar. Disconnect patient so that pressure can
decrease to 0 hpa/0 mbar.
2
5
3 4
1. Patient circuit + internal bag button: press this button to select the patient
breathing circuit and the internal balloon leaktest.
2. Patient circuit + handset button: press this button to select the patient
breathing circuit and the external handset.
4. PEEP valve leakage detected: this message appears when the PEEP valve
leakage test fails.
To perform a leaktest on the patient breathing circuit and the external handset,
proceed as follows:
- Press button.
- Make all the connections in the patient breathing unit; see the
operating instructions of the CO2 absorber.
- Close safety valve on handset.
- Close / disconnect sample tube of gas monitor (optional).
- Close the output of the Y-piece.
- Set a fresh gas flow.
- Let the airway pressure increase to 30 hPa.
- Close the fresh gas flow.
- The time in which the pressure decreases to 20 hPa must be
15 seconds or longer.
If the leak test is not within the specifications, check all connections in the patient
breathing circuit. If no leaks can be found, contact your service department.
After successful performing the previously described leaktest, you can perform a
leaktest on the patient breathing circuit and the internal balloon.
Proceed as follows:
- Press button.
- Close the output of the Y-piece.
- Set a fresh gas flow.
- Let the airway pressure increase to 30 hPa.
- Close the fresh gas flow.
- The time in which the pressure decreases to 20 hPa must be
15 seconds or longer.
If the leak test is not within the specifications, check all connections in the patient
breathing circuit. Also check the internal balloon for leaks. If no leaks can be found,
contact your service department.
To repeat the PEEP valve test, press the button followed by a button
press.
To determine either the PEEP leakage is intern in the ventilator or extern in the
patient breathing unit, remove the patient circuit from bottle and close the PEEP
output with finger in accordance with next drawing.
Keep your finger on the PEEP output and repeat the PEEP valve test. If the PEEP
valve test succeeds, the problem is located in the patient breathing unit, otherwise the
problem is probably intern in the ventilator. Please contact your service department.
Leave the leaktest menu by pressing the exit button. The ventilator will return to
standby mode.
Set the ventilator to standby mode and press the setup button to enter the setup menu.
Always disconnect the patient before entering the setup menu.
You can swap between the different setup menus by pressing the button.
• Setup window 1
Setup window 1 is equal to the CMV window. The parameters for use during
CMV mode after turning the ventilator on can be set here.
Press the button to switch to the next setup window.
• Setup window 2
• Setup window 3
Setup window is equal to the SIMV window. The parameters for use during
SIMV mode after turning the ventilator on can be set here. Press the button to
switch to the next setup window.
• Setup window 4
6
5
2 3 4
1. Window select: press this button to switch between the different setup
windows.
2. Language select: set up the language to be used by the ventilator. You can
chose between two different languages.
3. Cancel: press this button to cancel the changes made in the setup menu.
4. Save: store the changes made in the setup menu to the memory.
5. Sound volume: set up the speaker sound volume in the range 1-5.
8. Ventilator information window: the Saturn Evo serial number (S/N), the
software revision numbers (S/R) and the total working time in hours of the
ventilator are displayed.
Note: The 'cancel' and 'save' button have no effect on date and time. A
change of date and/or time will be stored immediately into the
memory.
Rotating the encoder knob will change the position of the selection box and adjust a
selected value by pressing the control knob. The value is inverted and a rotation of the
control knob will adjust the value. Confirm the adjusted value by pressing the control
knob.
The Saturn Evo ventilator is fitted with a battery backup system that allows at least
1 hour of operation if no mains supply is available. If the mains indicator is off and
the battery symbol is drawn on the display, the battery is in use.
If the battery is almost exhausted, the low battery symbol is blinking and every
minute, an audible sound is generated.
Reconnect the mains supply or prepare for manual controlled ventilation.
A continuous beep will sound for 30 seconds after the ventilator shuts off.
The internal battery is automatically charged when the ventilator is connected to the
mains power. The charging of the battery is electronically controlled to assure
optimum battery operating conditions.
The front panel of the Saturn Evo is made of a high-grade polycarbonate foil. All
common household-cleaning products can be used without harm to the front panel
surface.
Aggressive products may affect the foil.
The ventilator housing may be cleaned with a moisturized cloth.
Caution: Use cleaning solution sparingly. It may flow into the ventilator and
cause internal damage.
The patient breathing unit can easily be removed for disinfection by pulling the
handle (figure 4.1).
When putting back the patient breathing circuit, it is recommended to manually guide
the bag to prevent it from damage (figure 4.2).
Note: The patient breathing circuit may never be immersed in liquid. This
may result in obstruction of valves and tubes, causing the patient
breathing circuit to malfunction.
It is of utmost importance to use undamaged bags, recommended by
Medec Benelux NV. When in doubt, always replace the bag by a new
one.
5.1 MAINTENANCE
- Perform a leaktest.
- Check the battery backup system.
- Check the absorber sodalime; (re-)fill if necessary.
- Check the absorber water trap.
• Every 6 months
- Preventive maintenance.
During the alarm, even if electronics are turned off, the patient can be ventilated by
means of the hand balloon.
This type of failure will require the attention of an authorized service engineer, who
can easily test the basic functions by means of the incorporated software.
If the ventilator is not working properly as described in this manual, try to resolve the
problem by means of the following table.
If the malfunction cannot be corrected using the suggestions in the table below, an
authorized service engineer must perform further troubleshooting.
- Turn off the Saturn Evo and disconnect the power cord from the mains
supply.
- Insert a screwdriver in the small AC receptacle and pull out the fuse
holder block.
- Replace the blown fuses and place the fuse holder block back into the
AC receptacle.
Note: Always replace the blown fuses with the same type and ratings.
6.1 GENERAL
Operating principle: Bag-in-bottle
6.2 RATINGS
Main supply: 100 - 240V, 50 - 60Hz
Power consumption: < 100VA
IEC 601 class and type: Class I, type B