Standard Calibration Procedure

Doc. No. Call/SCP/035 Rev. 00 May 01, 2015


Volumetric Measures

STANDARD CALIBRATION PROCEDURE

VOLUMETRIC MEASURES

Prepared by

Muhammad Ajlan Signature Date : 01.05.2015

Reviewed by

Muhammad Sarwar Signature Date : 01.05.2015

Approved by

Qaswar Mehmood Signature Date : 01.05.2015

Controlled Copy No. Uncontrolled Copy

Issued to :

Page 1 of 4
Standard Calibration Procedure
Doc. No. Call/SCP/035 Rev. 00 May 01, 2015
Volumetric Measures

This document is the property of Qstar Engineering and is issued only for persons or entity as
written on this document. It is not to be retransmitted, printed or copied without prior written
permission of company.

Table of Contents

Sr. No. Title Page No(s).


1.0 Scope 03
2.0 Standard Equipment 03
3.0 Preliminary Operations 03
4.0 Calibration Procedure 03
5.0 Related Documents 03

Page 2 of 4
Standard Calibration Procedure
Doc. No. Call/SCP/035 Rev. 00 May 01, 2015
Volumetric Measures

1.0 Scope

This procedure provides guidelines for calibration of Volumetric Measures.

2.0 Standard Equipment

2.1 Electronic Balance with a valid certificate of calibration traceable to international


standard in accordance with an industrial code or a regulation shall be used as a
standard for calibration.
2.2 The reference standard shall be kept under controlled environmental conditions e.g. dust
free environment and Lab. temperature 22°C ± 3°C. Lab Humidity 60%RH ± 20%RH.

3.0 Preliminary Operations

3.1 The details of Volumetric Measures received for calibration shall be entered in the
Calibration Job Form (Form: Call/001).
3.2 Inspection of the Volumetric Measures for general condition shall be carried out.
3.3 The Volumetric Measures shall be inspected and marked () in relevant block of
equipment’s primary inspection section of Calibration Data Sheet (Form: Call/013).

4.0 Calibration Procedure

4.1 The Volumetric Measures to be calibrated shall be maintained at same place of standard
Volumetric Measures / Electronic Balance for 03 hours minimum.
4.2 Volumetric Measures shall be placed on leveled and non-vibrate able Place.
4.3 Volume of Volumetric Measures shall be compared with Certified / Standard Volumetric
Measures / Electronic Balance as per graduation steps of Volumetric Measures under
Test.
4.4 After completion of calibration all the equipment shall be secured.
4.5 Calibration Certificate (Form: Call/002) shall be issued for calibrated volumetric
measures as per Calibration Data Sheet Record (Form: Call/013).
4.6 A Calibration Sticker (Form: Call/003) containing following information shall be affixed
on the calibrated volumetric measures.

Page 3 of 4
Standard Calibration Procedure
Doc. No. Call/SCP/035 Rev. 00 May 01, 2015
Volumetric Measures

a) Item code
b) Calibrated on
c) Next calibration date

5.0 Related Documents

All the latest formats are documented in Master List of Document / Record and placed in CQF
file for reference.
All the latest Standard Calibration Procedures (SCP) are controlled and issued to concerned
Calibration Engineers.