Yamada 2015

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

Article

pubs.acs.org/ac

Ultratrace Measurement of Acetone from Skin Using Zeolite: Toward


Development of a Wearable Monitor of Fat Metabolism
Yuki Yamada,*,†,‡ Satoshi Hiyama,‡ Tsuguyoshi Toyooka,‡ Shoji Takeuchi,§ Keiji Itabashi,∥
Tatsuya Okubo,∥ and Hitoshi Tabata†

Department of Bioengineering, The University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8656, Japan

Research Laboratories, NTT DOCOMO, Inc., 3-6 Hikarinooka, Yokosuka, Kanagawa 239-8536, Japan
§
Institute of Industrial Science, The University of Tokyo, 4-6-1 Komaba, Meguro-ku, Tokyo 153-8505, Japan

Department of Chemical System Engineering, The University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8656, Japan
*
S Supporting Information

ABSTRACT: Analysis of gases emitted from human skin and contained


in human breath has received increasing attention in recent years for
noninvasive clinical diagnoses and health checkups. Acetone emitted from
human skin (skin acetone) should be a good indicator of fat metabolism,
which is associated with diet and exercise. However, skin acetone is an
analytically challenging target because it is emitted in very low
concentrations. In the present study, zeolite was investigated for
concentrating skin acetone for subsequent semiconductor-based analysis.
The adsorption and desorption characteristics of five zeolites with
different structures and those hydrophobicities were compared. A
hydrophobic zeolite with relatively large pores (approximately 1.6 times
larger than the acetone molecule diameter) was the best concentrator of
skin acetone among the zeolites tested. The concentrator developed using
zeolite was applied in a semiconductor-based gas sensor in a simulated
mobile environment where the closed space was frequently collapsed to reflect the twisting and elastic movement of skin that
would be encountered in a wearable device. These results could be used to develop a wearable analyzer for skin acetone, which
would be a powerful tool for preventing and alleviating lifestyle-related diseases.

M ore than 200 compounds have been identified in human


breath, and some of these are associated with various
diseases, physiological changes, and physical conditions.
for skin acetone would provide a powerful tool for preventing
and alleviating lifestyle-related diseases. However, development
of such a device is challenging. This is because skin acetone is
Consequently, breath analysis has received increasing attention typically emitted at concentrations of only several tens of parts
in recent years for noninvasive clinical diagnoses and health per billion (ppb), which is too low for detection by small
checkups.1,2 Among the many compounds in human breath, commercially available sensors. For example, semiconductor-
acetone is expected to be a good indicator of fat metabolism, based gas sensors have limits of detection around 200 ppb.4 By
which is associated with diet,3,4 aerobic exercise,5,6 and contrast, existing high-sensitivity methods for skin acetone
diabetes.7,8 Obesity increases the risk of lifestyle-related analysis use large apparatus, such as gas chromatographs (GCs)
diseases, and enabling patient monitoring of breath acetone or liquid chromatographs combined with skin gas collection
concentrations could play a pivotal role in day-to-day bags,13 cooling concentrators,14 trapping filters,15 or solid-phase
management of dieting and diabetes control.3,9 However, microextraction cartridges,16−18 which are not suitable for
with breath analysis, it is difficult to motivate patients to blow development into a wearable device. To develop a wearable
into a measurement device on a daily basis. Therefore, a simpler device, either the size of the high-sensitivity devices or the
method that requires less user input is required. sensitivity requirements of the small devices needs to be
Many compounds, including acetone, are emitted from reduced.
human skin,10−12 and breath and skin acetone concentrations Here, to reduce the detector sensitivity requirements,
are correlated.13 Compared with breath acetone, skin acetone is concentration of the acetone emitted from skin was
advantageous for analysis because it is continuously emitted investigated (Figure 1). Zeolite was selected for concentration
and analysis requires no active action on the part of the user,
unlike the deep exhalation required for breath analysis. Skin Received: January 23, 2015
analysis is also more precise than breath analysis as it excludes Accepted: July 3, 2015
factors such as the flow rate of exhalation. A wearable analyzer

© XXXX American Chemical Society A DOI: 10.1021/acs.analchem.5b00296


Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article

Figure 1. Schematic diagram of our proposed method using zeolite to concentrate the very low concentrations of acetone emitted from human skin.
(a) Acetone is continuously emitted from human skin. (b) Emitted acetone is adsorbed into zeolite pores. (c) Adsorbed acetone is desorbed by
heating the zeolite, and then the acetone is detected by a semiconductor-based gas sensor.

of skin acetone, which was then detected using a semi- Showa K.K. (Tokyo, Japan). This zeolite has a MFI structure,
conductor-based gas sensor. Zeolites are microporous materials and its pores are slightly larger than those of FER-156. The
that adsorb gaseous molecules, and are widely used as zeolites 390HUA, 385HUA, and 350HUA were purchased from
adsorbents, ion exchangers, and catalysts.19 It is commonly Tosoh Corp. (Tokyo, Japan). These zeolites have FAU
used in automotive exhaust treatment 20 and chemical structures and their pores are larger than those of
industries,21 and has been used to coat a quartz crystal HISIV3000. The SiO2/Al2O3 ratio of the zeolite was used as
microbalance for gas sensing.22 More than 200 structure types an indicator of the zeolite hydrophobicity, with a high ratio
of zeolites have been recognized, and it is expected that a corresponding to high hydrophobicity. On the basis of this
specific zeolite could be targeted to efficient acetone grading, 390HUA has higher hydrophobicity than 385HUA and
adsorption/desorption agent by the selection of appropriate 350HUA (see Table 1).
structure type and its hydrophobicity. Adsorption/Desorption of Pure Acetone Gas into/
The aims of this work were: (1) to investigate and compare from the Zeolites. An aliquot (5 mg) of zeolite powder was
the adsorption/desorption properties of acetone gas with placed in a 16.9 mL glass vial (CV-140, Osaka Chemical Co.,
various zeolites with different structures and those hydro- Ltd., Osaka, Japan) and the open vial was heated at 200 °C for
phobicities, and to identify the best zeolite for concentration of 20 min on a hot plate to remove any volatile impurities from
skin acetone, (2) to verify whether the concentrated skin the zeolite. The vial was then sealed and pure acetone gas
acetones can be measured using a semiconductor-based gas diluted with nitrogen gas was injected into the vial using a
sensor, and (3) to confirm the feasibility of the developed syringe so that the final concentration of acetone in the vial was
method for analysis of acetone emitted from the skin of human 800 ppb. This initial acetone concentration was used as a
volunteers. To the best of our knowledge, this is the first study reference to evaluate the acetone adsorption/desorption
to demonstrate the concentration of skin acetones using zeolite characteristics of each zeolite, because the concentration in
and their subsequent detection using a semiconductor-based the vial without zeolite samples will stay the same during the
gas sensor. These results could help us in developing a wearable experiments. After 15 min, the acetone concentration in the vial
analyzer for skin acetone that could be used for self-monitoring was measured by GC (SGEA-P2, FIS Inc., Hyogo, Japan) to
of fat metabolism in daily life. determine the acetone adsorption capacity of each zeolite.


Then, the vial was opened and the gas it contained was
EXPERIMENTAL SECTION exchanged with clean air. After resealing the vial, the vial was
heated at 200 °C for 5 min on the hot plate. The acetone
Zeolites Used in This Study. Five representative zeolites concentration in the sealed vial was then measured using the
with different structures and hydrophobicities were selected GC, and this was used to determine how much acetone
(Table 1). The zeolite FER-156 was synthesized by a research desorbed from each zeolite.
group of one of the authors (T. Okubo) according to a Development of the Concentrator Using Zeolite.
published method.23 This zeolite has a FER structure, and its Zeolite powder (4.42 g) was mixed with 4.42 g of silica sol
pore size is comparable with the acetone molecule diameter (SNOWTEX N30G, Nissan Chemical Industries, Ltd., Tokyo,
(4.6 Å). The zeolite HISIV3000 was purchased from Union Japan), 0.35 g of carboxymethyl cellulose ammonium salt
(Wako Pure Chemical Industries, Ltd., Osaka, Japan), and 20
Table 1. Characteristics of Zeolites Used in This Studya mL of distilled water. A platinum coil heater (300 × 300 × 500
Structure type Pore size [Å] SiO2/Al2O3 [mol/mol] μm) was then dipped into the mixture. Any of the mixture that
FER-156 FER 3.5−5.4 156
adhered to the coil was air-dried on the coil surface at room
HISIV3000 MFI 5.1−5.6 800
temperature. This dip and dry process was repeated several
390HUA FAU 7.4 500
times, and then the coil heater was used to calcine the air-dried
385HUA FAU 7.4 100
mixture at 500 °C for 1 h. The obtained calcined mixture
350HUA FAU 7.4 10 weighed 0.6 mg, and was used to concentrate either pure
acetone gas or acetone emitted from the skin of human
a
Codes consisting of three capital letters, such as FER, MFI, and FAU, volunteers.
indicate a topology of zeolite frameworks and are assigned by the Adsorption/Desorption of Pure Acetone Gas into/
structure commission of the International Zeolite Association. from the Concentrator in Repeated Measurements. The
B DOI: 10.1021/acs.analchem.5b00296
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article

concentrator was heated at 330 °C for 30 s with the coil heater portion of the concentrator were placed far enough away from
to remove any volatile impurities from the concentrator and the skin surface. In addition, the heated portions were limited
then moved into a 6.4 mL glass vial. The vial was sealed and to the micrometer scale and thus their radiant heat did not
pure acetone gas diluted with nitrogen gas was injected into the affect the skin surface. The acetone concentration in the vial
vial using a syringe so that the final concentration of acetone in was measured using the acetone sensor. For comparison,
the vial was 800 ppb. After 15 min, the acetone concentration acetone samples emitted from the skin were collected over 10
in the vial was measured using the GC to determine the min using the same glass vial without the concentrator and
acetone adsorption capacity of the concentrator. The acetone sensor. The acetone concentration in the vial was
concentrator was then moved into another 6.4 mL glass vial measured using the GC.
and the vial was sealed. The concentrator was heated at 330 °C Skin Acetone Sensing under Real Conditions. A 6.4 mL
for 30 s and the acetone concentration in the sealed vial was bottomless glass vial containing the concentrator and the
measured using the GC to determine how much acetone acetone sensor was attached to the skin of the left forearm of a
desorbed from the concentrator. The above experiments were human volunteer to create a closed space. Skin acetone was
repeated nine times. collected over 20 min. The closed space was intentionally
Characterization of the Concentrator Using Acetone collapsed in a cyclic manner during the collection. This was
Emitted from Skin. Skin acetone samples were collected from performed by manually detaching the vial to open the closed
six healthy male volunteers aged in their 20s and 30s. A capped space, and then attaching it again to close the space. The open/
glass vial with the bottom removed was attached on the skin of close process was repeated every 30 s during the 20 min
the left forearm of each volunteer to create a closed space. This collection period. This process replicates the conditions that
vial contained the concentrator. Acetone samples emitted from could occur in a wearable device during daily activities, where
the skin were collected and adsorbed into the concentrator over twisting and elastic movement of the skin surface could
3, 6, 9, 12, or 15 min. After sample adsorption, the concentrator temporarily collapse the closed space of the sensor. After the
was transferred into a 16.9 mL glass vial, which was then sealed collection time was complete, the concentrator was heated at
and heated at 330 °C for 30 s with the coil heater. The acetone 330 °C for 30 s using the coil heater to desorb the adsorbed
concentration in the vial was measured using the GC. skin acetone.
Accuracy of Our Proposed Method in Skin Acetone These processes were repeated three times in succession, and
Measurements. Skin acetone samples were collected twice changes in the sensitivity of the acetone sensor were
from six healthy volunteers. A capped 16.9 mL glass vial with continuously monitored during the experiments. For compar-
the bottom removed was attached on the skin of the left ison, the same experiments were conducted without equipping
forearm of each volunteer to create a closed space (Figure 2). the vial with the concentrator.

■ RESULTS AND DISCUSSION


Five representative zeolites were studied to determine how
pore size and hydrophobicity affected the concentration of skin
acetone. The adsorption and desorption characteristics of FER-
156, HISIV3000, and 390HUA zeolites for pure acetone gas
were compared (Figure 3). With an acetone concentration of
880 ppb in a sealed glass vial, more than 800 ppb acetone was
adsorbed by all three zeolites after a 15 min incubation period
(Figure 3a). These results indicate that approximately 92% of
the acetone was adsorbed into the zeolites, and their adsorption
capacities were similar. In contrast, desorption of acetone by
heating was different for the three zeolites. The zeolite
390HUA showed much better desorption performance than
the other zeolites, and approximately 85% of the adsorbed
acetone was desorbed from this zeolite under the present
Figure 2. Photograph of skin acetone collection and measurement conditions (Figure 3b). The differences in desorption are
from human skin on the left forearm. A closed space is created using a presumably caused by the structural differences of the zeolites.
glass vial containing the concentrator and acetone sensor. The inset Acetone would desorb from zeolite 390HUA more easily than
shows a photograph of the concentrator, which is composed of
the other zeolites because its pores are approximately 1.6 times
390HUA zeolite. The scale bar corresponds to 500 μm.
larger than the acetone molecule diameter. If the zeolite pores
are too large, acetone and the zeolite will not strongly interact.
This vial contained the concentrator and a semiconductor- This indicates that there is a trade-off between adsorption
based acetone sensor4 made of platinum-doped tungsten oxide capacity and desorption kinetics, and this could be used to
(FIS Inc., Hyogo, Japan). We used a sensor that has high select a zeolite to act as the best concentrator for a specific
sensitivity for acetone and low sensitivity for the other gases molecule.
emitted from human skin, such as acetaldehyde and ammonia To investigate the effect of the differences in zeolite
(Figure S1, Supporting Information). Acetone samples emitted hydrophobicity, the adsorption and desorption characteristics
from the skin were collected and adsorbed into the of pure acetone gas with the FAU type zeolites (390HUA,
concentrator over 10 min. After sample adsorption, the 385HUA, and 350HUA) were compared (Figure 4). When an
concentrator was heated at 330 °C for 30 s with the coil acetone concentration of 780 ppb was generated in the sealed
heater equipped inside the concentrator. The volunteers did glass vial, more than 740 ppb acetone was adsorbed by all three
not feel heat because the heater and the corresponding heated of the zeolites after a 15 min incubation period (Figure 4a).
C DOI: 10.1021/acs.analchem.5b00296
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article

Figure 3. (a) Acetone (pure gas) adsorption characteristics of three


zeolites with different structures. The error bars represent the standard
error from three measurements. (b) Desorption characteristics of the
three zeolites. The error bars represent the standard error from three Figure 4. (a) Acetone (pure gas) adsorption characteristics of three
measurements. zeolites with different hydrophobicities. The error bars represent the
standard error from three measurements. (b) Desorption character-
istics of the three zeolites. The error bars represent the standard error
These results indicate that approximately 95% of the acetone from three measurements.
was adsorbed into the zeolites, and their adsorption capacities
were similar. Desorption of acetone from the three zeolites by
heating was quite different. Zeolite 390HUA showed much
better desorption performance than the other zeolites (Figure
4b). Approximately 81% of the adsorbed acetone desorbed
from zeolite 390HUA, whereas only around 3% of the adsorbed
acetone desorbed from zeolite 350HUA. These results indicate
that hydrophilic zeolites, such as 350HUA, strongly adsorb
polar molecules such as acetone. These results show that a
hydrophobic zeolite with relatively large pores will be most
suitable for use as a concentrator of skin acetone. Among the
five zeolites examined in this study, 390HUA is the best choice
as a concentrator of acetone.
Next, the acetone adsorption capacity and the reproducibility
of acetone adsorption/desorption of the calcined 390HUA
concentrator were investigated (Figure 5). The concentrator
used in this study (0.6 mg) was able to adsorb up to 688 ppb
pure acetone gas in an average incubation time of 15 min for
843 ppb of generated pure acetone gas. This adsorption Figure 5. Adsorption and desorption characteristics of the
capacity is feasible because the skin acetone emission rate per concentrator for pure acetone in repeated measurements.
unit time and area will be 2 to 3 ppb, and it will thus take 3.8 to
5.7 h to reach 688 ppb collected skin acetone gas. Assuming from the concentrator for the nine distinct measurements.
that our proposed method will be implemented in wearable These results indicate that our proposed method has high
devices to monitor fat metabolism, a 15 min to 1 h collection reproducibility.
time will be sufficient for this application and saturation of the We then verified adsorption/desorption of acetone with the
zeolite pores will not occur in such a collection time. We also concentrator made of 390HUA zeolite in human volunteers.
found that 93% ± 3% of the adsorbed pure acetone desorbed Gas samples were collected and adsorbed into 390HUA over 3,
D DOI: 10.1021/acs.analchem.5b00296
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article

6, 9, 12, or 15 min. Figure 6 shows that three representative


patterns of the emission flux of skin acetone observed from

Figure 7. Scatter plots of skin acetone concentrations in 10 min


collection time detected by our proposed system (concentrator and
acetone sensor) and by GC.
Figure 6. Relationship between acetone collection time and
concentration of skin acetone desorbed from the concentrator. The proposed method tended to be slightly lower than those
experiments were conducted in the order of decreasing collection time, obtained from the GC. This is presumably because some of the
and thus each of the lapse times from the start of the experiment adsorbed skin acetone molecules remain in the concentrator, as
corresponds to the sum of the collection times. shown in Figure 5.
To investigate the feasibility of our proposed method, skin
acetone was measured using the semiconductor-based acetone
three volunteers among the six. Results obtained from the rest sensor with and without the concentrator (390HUA zeolite)
of three volunteers were similar to either Subject A or Subject under simulated real life conditions. To achieve this, the closed
C, and thus these results were not shown in this figure for space was intentionally collapsed in a cyclic manner during the
clarity. Assuming that the emission rate of skin acetone per unit 20 min gas collection (Figure 8). The sensitivity of the acetone
time and area is constant, the relationship between desorbed
acetone and collection time should theoretically be linear. The
emission rate, however, fluctuates over time according to the
lifestyle behavior, such as the diet and exercise of each subject.
Considering that the experiments were conducted in the order
of decreasing collection time, the emission rates for the longer
collection times (e.g., 12 and 15 min, which correspond to 30
and 45 min in lapse time from the start of the experiment) may
have changed from the rate at the start of the experiment. Thus,
we assumed that the emission rate was constant during the first
9 min in lapse time from the start of the experiment, and found
that the skin acetone concentration desorbed from the
390HUA zeolite linearly increased with increasing collection
time. These results show that 390HUA zeolite can adsorb/
desorb skin acetone as well as pure acetone gas. The obtained
results that the skin acetone concentration desorbed from the
390HUA zeolite did not linearly increase after the first 9 min in
lapse time indicate that the emission rates significantly
Figure 8. Continuous monitoring of the sensitivities of the
increased in Subject B and decreased in Subject C. The slopes
semiconductor-based acetone sensor with and without the concen-
of the curves in Figure 6 show the differences in skin acetone trator under simulated real conditions. The closed space was
emission among the volunteers. Among the six volunteers, intentionally collapsed in a cyclic manner during each 20 min
Subject B emitted acetone at the highest rate per unit time and collection time. Acetone detection is possible when the sensitivity is
area during the experiments. less than 0.8.
To investigate the accuracy of our proposed method, the
concentrations of skin acetone in 10 min collection time
detected by our proposed system (with the concentrator and sensor without the concentrator drastically decreased from 1.00
acetone sensor) and those determined by GC were compared to 0.83 within the first 5 min from the start of the measurement
(Figure 7). The plots show that there is a strong correlation because of the increased concentration of skin acetone in the
between the concentrations of skin acetone obtained from our closed space. It then plateaued at around 0.83 because of the
proposed method and those obtained by the GC (R2 = 0.90, P intentional and cyclic collapsing of the closed space. Such
< 0.001), confirming that our proposed method is practical with collapsing is very likely in daily life, where some activities are
a reasonable range of measurement deviation. It should be accompanied by twisting and elastic movement of the skin
noted that the skin acetone concentrations obtained from our surface. According to the results of the six volunteers in Figure
E DOI: 10.1021/acs.analchem.5b00296
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article

6, the skin acetone emission rate per unit time and area was 2
to 3 ppb in average, and it will thus take 7 to 10 min to reach 20

*
ASSOCIATED CONTENT
S Supporting Information
ppb of collected gas. This indicates that it will be difficult to
Additional information as noted in the text. The Supporting
measure skin acetone using only a sensor even if state-of-the-art
Information is available free of charge on the ACS Publications
acetone sensors that can measure 20 ppb of breath acetone24
website at DOI: 10.1021/acs.analchem.5b00296.


are used, because such collapsing will take place in real life
during the gas collection time of 7 to 10 min.
After the 20 min collection time, the sensitivity increased AUTHOR INFORMATION
back to 1.00 because the closed space was opened and the Corresponding Author
enclosed gas was completely exchanged with clean air. The *Y. Yamada. E-mail: yuuki.yamada.vt@nttdocomo.com.
sensitivity is indicated by the ratio of the electrical resistance of Notes
the gas sensor in air (Rair) to the electrical resistance of the gas
The authors declare no competing financial interest.


sensor in the target gas (R). Thus, it will be difficult to clearly
distinguish the signal from the noise as R/Rair → 1. We set the
limit of detection to R/Rair = 0.8 to avoid false detection and to ACKNOWLEDGMENTS
conduct reliable gas sensing. Considering that the sensitivities The authors thank Mariko Hanada, Kazuo Onaga, Junko
of the acetone sensor without the concentrator were always R/ Yanagitani, Aki Uesaka, Hitoshi Katayama, Dr. Katsuyuki
Rair > 0.8, we concluded that the acetone sensor alone would Tanaka, and Takeo Tsunemi of FIS Inc. for their help in
not be able to detect skin acetone under the simulated developing the concentrators.
conditions. In comparison, the sensitivities of the acetone
sensor with the concentrator were all much less than 0.8 when
the adsorbed skin acetone was desorbed every 20 min.
■ REFERENCES
(1) Manolis, A. Clin. Chem. 1983, 29, 5−15.
Therefore, skin acetone could be easily detected under the (2) Cao, W.; Duan, Y. Clin. Chem. 2006, 52, 800−811.
simulated conditions with the concentrator made from zeolite. (3) Kundu, S. K.; Bruzek, J. A.; Nair, R.; Judilla, A. M. Clin. Chem.
We also found that almost the same peak sensitivities of the 1993, 39, 87−92.
acetone sensor with the concentrator (0.53 ± 0.01) were (4) Toyooka, T.; Hiyama, S.; Yamada, Y. J. Breath Res. 2013, 7,
obtained from the three measurements. These results indicate 036005.
that our proposed method has high reproducibility even if the (5) King, J.; Kupferthaler, A.; Koc, H.; Teschi, S.; Teschi, G.;
closed space for skin acetone collection is in high humidity Mekisch, W.; Schubert, J.; Hinterhuber, H.; Amann, A. J. Breath Res.
because of the water vapor emitted from the skin surface. 2009, 3, 027006.
(6) Sakai, H.; Ishikawa, S.; Ueda, H.; Kimura, Y. Adv. Exerc. Sports
It should be noted that the sensitivities of the acetone sensor Physiol. 2011, 16, 97−100.
with the concentrator also plateaued after the first 5 min, and (7) Wang, C.; Mbi, A.; Sphepherd, M. IEEE Sensors J. 2010, 10, 54−
were higher than those of the acetone sensor without the 63.
concentrator. This is because skin-emitted acetone adsorbed (8) Storer, M.; Dummer, J.; Lunt, H.; Scotter, J.; McCartin, F.; Cook,
into the concentrator and the adsorbed acetone was retained J.; Swanney, M.; Kendall, D.; Logan, F.; Epton, M. J. Breath Res. 2011,
until the concentrator was heated for desorption. 5, 046011.
The ratio of skin acetone measured by the acetone sensor (9) Righettoni, M.; Tricoli, A.; Pratsinis, S. E. Anal. Chem. 2010, 82,
with the concentrator to skin acetone measured by the acetone 3581−3587.
sensor without the concentrator was 7.5. This magnitude of (10) Mochalski, P.; King, J.; Unterkofler, K.; Hinterhuber, H.;
concentrated skin acetone could help to reduce the detector Amann, A. J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2014, 959,
62−70.
sensitivity requirement.
(11) Bernier, R. U.; Kline, L. D.; Bamard, R. D.; Schreck, E. C.; Yost,
These results indicate that our proposed method is feasible A. R. Anal. Chem. 2000, 72, 747−756.
and shows potential for implementation in wearable devices.


(12) Dormont, L.; Bessière, J.-M.; Cohuet, A. J. Chem. Ecol. 2013, 39,
569−578.
CONCLUSIONS (13) Turner, C.; Parekh, B.; Walton, C.; Spanel, P.; Smith, D.; Evans,
A skin acetone measurement method was developed using M. Rapid Commun. Mass Spectrom. 2008, 22, 526−532.
zeolite to concentrate the acetone gas before analysis with a (14) Naitoh, K.; Inai, Y.; Hirabayashi, T.; Tsuda, T. Anal. Chem.
2000, 72, 2797−2801.
semiconductor-based gas sensor. Among the five zeolites tested, (15) Sekine, Y.; Toyooka, S.; Watts, F. S. J. Chromatogr. B: Anal.
390HUA zeolite, a hydrophobic zeolite with relatively large Technol. Biomed. Life Sci. 2007, 859, 201−207.
pores, was the best for concentration of skin acetone. (16) Musteata, M. F.; Pawliszyn, J. J. Biochem. Biophys. Methods 2007,
Concentration of the acetone by the zeolite and subsequent 70, 181−193.
semiconductor-based gas sensor analysis was achieved in a (17) Duan, C.; Shen, Z.; Wu, D.; Guan, Y. TrAC, Trends Anal. Chem.
simulated mobile environment, where the closed space was 2011, 30, 1568−1574.
frequently collapsed to simulate the twisting and elastic (18) Dormont, L.; Bessière, J.-M.; McKey, D.; Cohuet, A. J. Exp. Biol.
movement of the skin surface that could occur with a wearable 2013, 216, 2783−2788.
device. (19) Roth, J. W.; Nachtigall, P.; Morris, E. R.; Č ejka. Chem. Rev.
These results could be used to develop a wearable analyzer 2014, 114, 4807−4837.
(20) Deka, U.; Lezcano-Gonzalez, I.; Weckhuysen, M. B.; Beal, M. A.
for skin acetone that could be used for self-monitoring of fat ACS Catal. 2013, 3, 413−427.
metabolism in daily life. In the future, we will further investigate (21) Perego, C.; Ingallina, P. Green Chem. 2004, 6, 274−279.
the characteristics of our proposed system under different (22) Huang, H.; Zhou, J.; Chen, S.; Zeng, L.; Huang, Y. Sens.
environmental conditions, such as temperature, humidity, and Actuators, B 2004, 101, 316−321.
gas composition. The next major step will be miniaturization (23) Kamimura, Y.; Kowenje, C.; Yamanaka, K.; Itabashi, K.; Endo,
and implementation in a wearable device, such as a wristwatch. A.; Okubo, T. Microporous Mesoporous Mater. 2013, 181, 154−159.

F DOI: 10.1021/acs.analchem.5b00296
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article

(24) Righettoni, M.; Tricoli, A.; Gass, S.; Schmid, A.; Amann, A.;
Pratsinis, S. E. Anal. Chim. Acta 2012, 738, 69−75.

G DOI: 10.1021/acs.analchem.5b00296
Anal. Chem. XXXX, XXX, XXX−XXX

You might also like