Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Cd

The Chrom
Doctor

Impact of GC Parameters on The Separation


Part I: Choice of the Stationary Phase
Jaap de Zeeuw, Restek Corporation, Middelburg, The Netherlands.

In this series, we will discuss the impact of different parameters on the actual separation. We will do this by discussing how they impact and
then explain the impact by some practical examples. In GC there are seven important parameters that we can choose to get a separation. All
parameters have an impact and selection of a certain parameter has an impact on the separation.

Strategic column choices can improve based on retention factor and 2 shows the seven major factors that Figure 1

lab productivity by assuring that efficiency required. Understanding affect chromatographic separation.
speed and resolution are optimized. how the separation factor, retention
While the number of choices available factors and efficiency influence Selecting The Stationary Phase
can be challenging to choose from, separations allows analysts to make This is the most important decision
consideration of the resolution effective choices and quickly select as the stationary phase defines the
equation variables, Figure 1, being the best column dimensions for selectivity for your components. Figure 1: Resolution equation.
the selectivity factor, retention specific separations and techniques. The ‘selectivity’ of stationary phases
(capacity) factor and efficiency, After the column has been selected depends on interactions between components that you would like to
simplifies the decision. The most there are several non-column related analyte and stationary phase. These separate’. This will enhance interaction
important separation factor to select factors that can be adjusted to tune interactions can be based on dipole, with target analytes and results in
is the stationary phase. Once the the resolution of the compounds of hydrogen bridge, van der Waals and better separations. For example it
stationary phase has been chosen, interest. These non-column factors steric interactions. means that for alcohol separation, a
physical dimensions (inner diameter, are oven temperature, carrier gas As a rule of thumb, you should use ‘Stabilwax’ phase is used, and for a
film thickness, length) can be selected type and carrier gas velocity. Figure a stationary phase that ‘looks like the hydrocarbon type, the Rtx-1 is used,

2 Impact of GC parameters on the Separation - Part 1 separation science — volume 6 issue 1 www.sepscience.com
Figure 2 Figure 4

Figure 2: Seven major factors that affect chromatographic resolution. Note: ID = inner diameter,
µ = film thickness.
Figure 3

Figure 4: Separation of benzofluoranthenes on Rxi17Sil MS; Full details, see ref. 6.

being 100% polydimethyl siloxane. PEG type phase the ethyl benzene,
So, in what cases will phase meta- and para xylene are not
selectivity play a critical role? resolved. If a high resolution capillary
is used with 100,000 theoretical
1. If there are only a limited number plates, all components are resolved
of plates available on a similar type phase, (Figure 3B).
If efficiency is low, a separation is only Choosing a more selective stationary
obtained if the phase used, is very phase in the packed column, a
selective. This is the reason why there full base line separation is also
are so many stationary phases used obtained (see Figure 3C). Despite
with packed column applications. developments in capillary GC a
With a packed column, only 5–10,000 large number of applications are still
plates are available resulting in performed with packed columns. If
relative broad peaks. Figure 3 shows these columns do the analysis in an
Figure 3: Separation of BTX; A: 2 m x ¼” packed column with 3% PEG; B: 30 m x 0.25 mm capillary, the separation of benzene, toluene, acceptable time, show stability and
df =0.25 um PEG; C: 2 m, 1/8”, 2 mm ID 5% Rt®-1200, 5% Bentone® 34; Peak identification: 3 = ethyl benzene and the three xylenes deliver the required detection limits,
benzene, 4 = toluene, 5 = ethylbenzene, 6 = p-xylene, 7 = m-xylene, 8 = o-xylene. on different columns. Using a polar there is no reason to use a capillary

3 Impact of GC parameters on the Separation - Part 1 separation science — volume 6 issue 1 www.sepscience.com
Table 1
column. Also packed columns have groups of components for which
benefitted from developments in unique selective stationary phases Permanent gases Freons/CFC
capillary technology. Because 1/4/, have been developed Separation of CO and CO2 Blood alcohols
1/8 and 1/16” metal tubing could be
C1-C6 hydrocarbons and isomers Optical Isomers
deactivated using the Siltek processes 3. Using mass spectrometric
[1], there is no need anymore for methods and target components C1-C20 hydrocarbons and isomers Cis/trans FAME
‘glass’ packed columns. show similar fragmentation C5-C120 hydrocarbons USP-467 solvents
There are also 0.53 mm capillary As mass spectrometric detection Xylenes/substituted benzenes Biodiesel
columns that are operated at higher has become a much more used
flow-rate. Under these conditions, technique, the selectivity of GC- Bio-ethanol Volatile Organic Compounds
such columns do not generate a lot phases is not always required. By Chlorinated pesticides Sulfur gases
of plates, so selectivity becomes even using selective scanning for certain Dioxins, dibenzofurans Mass spectroscopy
more important. fragments, a mass spectrometer can
Selection of the most selective also be used to separate peaks that Brominated flame retardants Oxygenates
stationary phase is also very co-elute from a GC column. PCBs VOC and semi-VOC acc to EPA
important If the analysis time has Even with mass spectroscopy, PAHs
to be reduced. This will allow a selectivity is required for separating
short column and/or high carrier isobaric compounds. Figure 4 Table 1: Example of applications for which optimized stationary phases are used.
gas velocities to be used, which will shows the separation of the 3 benzo
generate the shortest analysis times. fluoranthenes, using the Rxi-17Sil
MS phase. Only this mid-polar phase 4. When a method needs to be to different selectivity, [8].
2. If components are to be detected can separate the benzofluoranthenes confirmed by chromatography
by a non-selective detector with full baseline resolution. Special Besides using mass spectroscopy, 5. If selective detection is not
A non-selective detector means selective low-bleed phases have chromatography can also be used ‘selective’ enough
it detects every component and been developed for separating PAH for a positive identification. The same Selective detection is perfect if
chromatographic separation is [2], Dioxins [3] and PCB [4]. Also if sample is analysed on two columns the signal is not impacted by other
essential. Non-selective detectors components have similar degradation with different selectivities. If the components that may elute together
such as FID and TCD are very popular products, a bias can occur. For retention times of the target analyte with the target analyte. This effect
as they have linearity of 6–7 decades, instance If PCBs, dioxins or even on the two columns match with is called ‘quenching’. The signal of a
are sensitive and relatively low in cost. hydrocarbons are measured, the the standard, there is a high level selective detector is always impacted
Components that elute at the same fragmentation products in a MS, can of confidence that the component if another component elutes. It
retention time, cannot be quantified be very similar. identification is correct. This is done depends on the selectivity of the
individually. Here the stationary For optimal use of mass in chlorinated pesticide analysis (EPA detector if it’s a problem or not. For
phase has to do the separation spectroscopic detection, it is 8081), where special selective phases instance, if selectivity of a sulfur
and selecting the stationary phase important to use low-bleed and are developed [7]; Another example specific detector is six magnitudes, it
becomes very important. The inert GC columns [5]. Despite is the analysis of blood alcohols. means it will start to show response if
challenge increases when more the separation power of a mass Figure 5 shows the separation of the co-eluting peak is a million times
components are present in the spectrometer, it is always better to the components of interest on two higher concentration than the target
sample. Here it is essential to use the have a chromatographic separation columns, with optimized selectivities analyte. This may seem a lot, but in the
existing information on separations. first. for this type of separation. The elution case of sulfur, amounts of 5–50 ppb of
Table 1 shows some examples of order of both columns is different due H2S/COS must be quantified in typical

4 Impact of GC parameters on the Separation - Part 1 separation science — volume 6 issue 1 www.sepscience.com
Figure 5 Figure 6

Figure 5: For blood alcohol analysis, conformation is done by analysing the same sample on 2 Figure 6: Analysis of ppb levels of COS and H2S in ethylene and propylene, using the Rt XL Sulfur.
columns of different selectivity; Full details see ref. 8. Even with relative low efficiency, this separation was sufficient to avoid the quenching. Using the
PLOT column as referred in [9], even lower sensitivities can be expected.
hydrocarbon streams of ethylene, very powerful but two different
propylene and propane, see [9]. It is selectivities need to be chosen. columns have low plate numbers but injected on the second column every
very important that a selective phase Usually a non-polar and a polar phase because two selectivities are used, a 8–15 seconds. To make this work
is chosen where the matrix peak is not are used. Co-eluting components lot of separations can be done. The in the most optimal way, the two
interfering with the trace components on the first column, can be easily packed columns are often preferred phases used must have a selectivity
to be measured. Figure 6 shows an separated on a second other column because of their robustness. that ideally must be orthogonal. As
example of such a selective phase. The which has a different selectivity this is not possible, usually in the first
H2S and COS are well separated from phase. The transfer is often done with 7. Multi dimensional separations: dimension a separation on vapour
the main ethylene and propylene valves but most of the instrument GCxGC pressure on a non-polar column
peaks. Despite the detector being vendors also supply solutions In comprehensive GC, every is performed and on the second a
sulfur-selective, there is still a large based on Deans/flow switching. The component is separated in a first long polar phase is used with different
signal of the matrix. multidimensional separations are still high resolution column, usually 30 m interactions and sometimes also with
done a lot in process applications and x 0.25 mm, followed in series by a a shape-selectivity. See reference 10
6. Multi dimensional separations: also many analyser configurations second column, which is about 1–2m for some typical combinations. This
column switching are often equipped with packed long and has a different selectivity. set-up can also be reversed, using a
Multi dimensional separations are columns/switching devices. The Eluting fractions are focused and re- polar first dimension and a non-polar

5 Impact of GC parameters on the Separation - Part 1 separation science — volume 6 issue 1 www.sepscience.com
Figure 7 Figure 8

Figure 7: GCxGC separation of fire Debris using a 100m Rtx-DHA first dimension and a Stabilwax for Figure 8: Modelated chromatogram for aromatics; 1 = benzene; 2 = toluene; 3 = ethylbenzene;
the second dimension column. 4 = meta-xylene; 5 = para-xylene; 6 = ortho-xylene; Full details, see [12].

second dimension. Such set-ups have Later the development of new


huge peak capacities as thousands of generation stationary phases [1], substituted compounds but they separations. Chromatograms are
components can be separated. significantly contributed in this field. do lack temperature stability as very helpful for that. You can always
Figure 7 shows the tremendous By stabilization of siloxane polymers they cannot be heated over 260 °C. do a literature search but this takes
separation power of such a GCxGC using silphenylene as well as surface Despite this, they are still used for a time and is also expensive. It’s easier
set-up. bonding today’s phases can be lot of applications. As stated before, to use chromatogram databases,
operated from -60 °C up to 360 °C. for special groups, unique stationary which are made available by several
Selectivity of Stationary Phases Even silphenylene stabilized 50% phases are available (see Table 1). GC companies. Such databases
The siloxane-based stationary phenyl type phases have this stability can be searched on component
phases are mostly used. The polarity (Figure 4). How to Find the Most Selective name, matrix, group, method and
can be changed by incorporating Besides the siloxanes, there are Phase? even synonyms. See for example
methyl, phenyl, cyanopropyl or also polyethylene glycol type From the structure it remains difficult reference 11.
trifluoropropyl groups. Such phases phases, also known as ‘wax’ phases. to predict selectivity. It only gives Another interesting development
show ideal ‘liquid’ behaviour and have Such phases are highly selective for you a ‘general’ idea. The best way to is the availability of chromatogram
good temperature stability. alcohol, aldehydes and aromatic choose selectivity is to look at existing modeler software that is now also

A Impact of GC parameters on the Separation Part 1 separation science — volume 6 issue 1 www.sepscience.com
accessable on-line (see reference 12). inside the capillary column. He has published more
Using existing chromatographic than 100 publications in the field of GC on column
measurements as a reference, technology and application. He worked for 27 years
separations can be predicted. Figure 8 for Chrompack/Varian and for the last six years
shows an example of the separation has served as an international specialist on gas
of benzene, toluene and xylenes. chromatography for Restek in The Netherlands.
Reference 13 shows a link to a
practical guide.

References
1. http://www.americanlaboratory.com/913-
Technical-Articles/124285-Deactivation-of-Metal-
Surfaces-Applications-in-Gas-Chromatography-
GC-for-the-Past-15-Years/
2. http://www.restek.com/pdfs/GNTS1718-UNV.
pdf
3. http://www.restek.com/pdfs/pres-2004-dioxin2.
pdf
4. http://www.restek.com/pdfs/pres-2004-pcb.pdf
5. http://www.restek.com/Landing-Pages/
Content/gen_B004
6. http://www.restek.com/images/cgram/gc_
ev1160.pdf
7. http://www.restek.com/chromatogram/view/
GC_EV00740/CLPesticides2
8. http://www.restek.com/pdfs/CFTS1483-UNV.pdf
9. http://blog.restek.com/?p=9234
10. http://www.restek.com/gcxgc-combo
11. http://www.restek.com/chromatogram/search
12. http://www.restek.com/chromatogram/ezgc/
13. http://www.restek.com/pdfs/GNBR1724-UNV.
pdf

Jaap de Zeeuw studied six


years of chemistry and
graduated in 1979. Jaap has
33 years’ experience in GC
capillary technology and has
developed many PLOT columns as well as bonded-
phase columns. He is also the originator of simple
concepts for fast GC–MS using a high vacuum

A Impact of GC parameters on the Separation Part 1 separation science — volume 6 issue 1 www.sepscience.com

You might also like