Operating Instructions
Micropilot NMR81
Tank Gauging
Order code: XXXXX-XXXXXX
1. Ser. no.: XXXXXXXXXXXX
Ext. ord. cd.: XXX.XXXX.XX
Serial number
3.
A0023555
2 Endress+Hauser
Table of contents
Table of contents
13 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
6 Electrical connection . . . . . . . . . . . . . . 19
13.1 General information on repairs . . . . . . . . . . . . 85
6.1 Terminal assignment . . . . . . . . . . . . . . . . . . . 19 13.2 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . 85
6.2 Connecting requirements . . . . . . . . . . . . . . . . 31 13.3 Endress+Hauser services . . . . . . . . . . . . . . . . 86
6.3 Ensuring the degree of protection . . . . . . . . . . 32 13.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
6.4 Post-connection check . . . . . . . . . . . . . . . . . . 32 13.5 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
7 Operability . . . . . . . . . . . . . . . . . . . . . . . . 33 14 Accessories . . . . . . . . . . . . . . . . . . . . . . . 87
7.1 Overview of the operation options . . . . . . . . . 33 14.1 Device-specific accessories . . . . . . . . . . . . . . . 87
7.2 Structure and function of the operating 14.2 Communication-specific accessories . . . . . . . . 89
menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 14.3 Service-specific accessories . . . . . . . . . . . . . . . 89
7.3 Access to the operating menu via the local 14.4 System components . . . . . . . . . . . . . . . . . . . . 90
display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
7.4 Access to the operating menu via the service
interface and FieldCare . . . . . . . . . . . . . . . . . 47 15 Operating menu . . . . . . . . . . . . . . . . . . . 91
7.5 Access to the operating menu via Tankvision 15.1 Overview of the operating menu . . . . . . . . . . . 91
Tank Scanner NXA820 and FieldCare . . . . . . . 48 15.2 "Operation" menu . . . . . . . . . . . . . . . . . . . . . 100
15.3 "Setup" menu . . . . . . . . . . . . . . . . . . . . . . . . 109
8 System integration . . . . . . . . . . . . . . . . 51 15.4 "Diagnostics" menu . . . . . . . . . . . . . . . . . . . . 226
9 Commissioning . . . . . . . . . . . . . . . . . . . . 52
9.1 Terms related to tank measurement . . . . . . . . 52
Endress+Hauser 3
About this document Micropilot NMR81
1.2 Symbols
Symbol Meaning
DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.
WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.
CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.
NOTE!
NOTICE This symbol contains information on procedures and other facts which do not result in
personal injury.
Symbol Meaning
Torx screwdriver
A0013442
A0011220
4 Endress+Hauser
Micropilot NMR81 About this document
Symbol Meaning
Cross-head screwdriver
A0011219
Allen key
A0011221
Hexagon wrench
A0011222
Symbol Meaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation
Reference to page
Reference to graphic
, , … Series of steps
Visual inspection
Symbol Meaning
A, B, C, ... Views
Hazardous area
- Indicates a hazardous area.
Endress+Hauser 5
About this document Micropilot NMR81
Symbol Meaning
Safety instructions
Observe the safety instructions contained in the associated Operating Instructions.
6 Endress+Hauser
Micropilot NMR81 About this document
1.3 Documentation
For an overview of the scope of the associated Technical Documentation, refer to the
following:
• The W@M Device Viewer : Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate
or scan the 2-D matrix code (QR code) on the nameplate.
Endress+Hauser 7
About this document Micropilot NMR81
FE FM C/US XP-AIS Cl.I Div.1 Gr.A-D, AEx d[ia] IIC T4...T1 XA01436G (in preparation)
8 Endress+Hauser
Micropilot NMR81 About this document
Endress+Hauser 9
Basic safety instructions Micropilot NMR81
Residual risk
During operation the sensor may assume a temperature near the temperature of the
measured material.
Danger of burns due to heated surfaces!
‣ For high process temperatures: Install protection against contact in order to prevent
burns.
10 Endress+Hauser
Micropilot NMR81 Basic safety instructions
2.5.1 CE mark
The measuring system meets the legal requirements of the applicable EC guidelines. These
are listed in the corresponding EC Declaration of Conformity together with the standards
applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
Endress+Hauser 11
Product description Micropilot NMR81
3 Product description
1 1 1
2 2 2
3
4
5 5 5
6 7 8
A0027765
12 Endress+Hauser
Micropilot NMR81 Incoming acceptance and product identification
Endress+Hauser 13
Incoming acceptance and product identification Micropilot NMR81
4.2.1 Nameplate
1
Micropilot
Order code: 2 18
Ser. no.: 3
Ext. ord. cd.: 19
4
5
MWP: 6
Tp max.: 7 Ta > 60 °C: t >85°C
20
8
Mat.: 9
DeviceID: 10
FW: 11 Dev.Rev.: 12 ex works
13
Tank ID:
Tank ref.height: 14 FCC ID: LCG NMR8X
Zero point:
Dual seal (MWP: 370bar)
Ta: 15 21
22
if modification
X = see sep. label
16 17 24
Date: 23
A0027791
14 Endress+Hauser
Micropilot NMR81 Incoming acceptance and product identification
4.3.2 Transport
NOTICE
Housing or antenna may be damaged or break away.
Risk of injury
‣ Transport the measuring device to the measuring point in its original packaging or at
the process connection.
‣ Do not fasten lifting devices (hoisting slings, lifting eyes etc.) at the housing or the
antenna but at the process connection. Take into account the mass center of the device
in order to avoid unintended tilting.
‣ Comply with the safety instructions, transport conditions for devices over 18kg
(39.6lbs) (IEC61010).
Endress+Hauser 15
Installation Micropilot NMR81
5 Installation
5 m (16 ft) 0.3 m (0.98 ft) 0.17 m (0.55 ft) 0.13 m (0.44 ft)
10 m (33 ft) 0.6 m (1.9 ft) 0.33 m (1.1 ft) 0.27 m (0.87 ft)
15 m (49 ft) 0.9 m (2.9 ft) 0.5 m (1.6 ft) 0.4 m (1.3 ft)
20 m (66 ft) 1.2 m (3.9 ft) 0.67 m (2.2 ft) 0.53 m (1.7 ft)
25 m (82 ft) 1.5 m (4.9 ft) 0.83 m (2.7 ft) 0.67 m (2.2 ft)
30 m (98 ft) 1.8 m (5.9 ft) 1.0 m (3.3 ft) 0.8 m (2.6 ft)
16 Endress+Hauser
Micropilot NMR81 Installation
Adjustable seal
4°
ød
øD
1
2
h
±8 °
hmax
3
hmin
A0027787
PS PT PU
2)
Order code 71285499 71285501 71285503
Length of screws 100 mm (3.9 in) 100 mm (3.9 in) 100 mm (3.9 in)
1) With this ordering feature the adjustable seal is supplied together with the device.
2) This order code must be used if the adjustable seal is ordered separately.
Endress+Hauser 17
Installation Micropilot NMR81
α
°
90
A0027776
m Are the measuring point identification and labeling correct (visual inspection)?
18 Endress+Hauser
Micropilot NMR81 Electrical connection
6 Electrical connection
D 1 2 3 4
1
E
D 1 2 3 4
1
E 2
CDI
1
C 1 2 3 4 5 6 7 8 2
i
F 21 3
1
4
B 1 2 3 4 5 6 7 8
1
G2
3
HR
WP
on
A 1 2 3 4
CDI
POWER SIM
C 1 2 3 4 5 6 7 8
F 12 3
4
B 1 2 3 4 5 6 7 8
1
G2 HR
3 on
WP A 1 2 3 4
POWER SIM
G1 N AC 85...264 V
G3 L
A0026372
Endress+Hauser 19
Electrical connection Micropilot NMR81
Spare parts
for:
Ser.-no.:8A21A Proservo NMS81
C098AF4
Spare part
1 Cover Spare no./stru
2 Display cture
3 Display 71023451
asm. 71023451
4 Main electr.
ewer
71023451
5 IOM-V1 XPF0002-AAB
AAEAEBEFLA ICR+
.com/devicevi
information:
Slot A
6 IOM-A/RTD XPF0002-AAB LC76
Slot B XPF0002-AAB EFEG+
7 IOM-D XPF0002-AAB EFEG+
XPF0002-AAB EFEG+
Additional
www.endress
Slot C
8 IO Mod XPF0002-AAB EFEG+
FF XPF0002-AAB EFEG+
Slot D XPF0002-AAB EFEG+
9 SMS electr.
10 Detector XPF0002-AAB EFEG+
71273689 EFEG+
11 Displac
er 7122xxxx
12 Wire Drum XPF0002-AAB
AAAAACDEFEEFEG+
XPF0002-AAA G+
AAHAHCHRIJJCABADJ+
AJBKP
1 12
2,3 4-8 9 10
11
D
1
B
A0030070
20 Endress+Hauser
Micropilot NMR81 Electrical connection
D
Output IO Analog IO Digital
Ex d/XP
1 2 3 4
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8
1 2 3 4
A0023888
A1 X0 X0 Modbus - - -
A1 X0 A1 Modbus - - Digital
A1 X0 B1 Modbus Modbus - -
Endress+Hauser 21
Electrical connection Micropilot NMR81
D
Output IO Analog IO Digital
Ex d/XP
1 2 3 4
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8
1 2 3 4
A0023888
B1 X0 X0 V1 - - -
B1 X0 A1 V1 - - Digital
B1 X0 A2 V1 - Digital Digital
B1 A1 X0 V1 Analog Ex d/XP - -
B1 B1 X0 V1 Analog Ex i/IS - -
22 Endress+Hauser
Micropilot NMR81 Electrical connection
D
Output IO Analog IO Digital
Ex d/XP
1 2 3 4
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8
1 2 3 4
A0023888
E1 X0 X0 - Analog Ex d/XP - -
Endress+Hauser 23
Electrical connection Micropilot NMR81
D
Output IO Analog IO Digital
Ex d/XP
1 2 3 4
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8
1 2 3 4
A0023888
H1 X0 X0 - Analog Ex i/IS - -
24 Endress+Hauser
Micropilot NMR81 Electrical connection
D 1 2 3 4
1
E 2 D 1 2 3 4
1
CDI
C 1 2 3 4 5 6 7 8
i D1-4
F 12 3
4
B 1 2 3 4 5 6 7 8
A 1 2 3 4
HR
on
POWER
WP A 1 2 3 4
SIM
A1-4
A0031200
6 Designation of the "Modbus" or "V1" modules (examples); depending on the device version these modules
may also be in slot B or C.
Depending on the device version, the "Modbus" and/or "V1" module may be in different
slots of the terminal compartment. In the operating menu the "Modbus" and "V1" interfaces
are designated by the respective slot and the terminals within this slot: A1-4, B1-4, C1-4,
D1-4.
X2 0V Common reference
1) In this column, "X" stands for one of the slots "A", "B", "C", or "D".
X2 not connected
1) In this column, "X" stands for one of the slots "A", "B", "C", or "D".
Endress+Hauser 25
Electrical connection Micropilot NMR81
D 1 2 3 4
1
E 2
CDI
1 C 1 2 3 4 5 6 7 8
C 1 2 3 4 5 6 7 8
i
C1-3 C4-8
F 12 3
4
B 1 2 3 4 5 6 7 8
B 1 2 3 4 5 6 7 8
HR
on
POWER
WP A 1 2 3 4
SIM
B1-3 B4-8
A0031168
B1-3 Analog input or output • Passive usage: → 26 Analog I/O B1-3 (→ 128)
(configurable) • Active usage: → 28
C1-3 Analog I/O C1-3 (→ 128)
D 1 2 3 4
1
E 2
1
CDI
D 1 2 3 4
C
1
CDI
E
1
2
1 2 3 4 5 6 7 8
C 1 2 3 4 5 6 7 8
F 21 3
4
B 1 2 3 4 5 6 7 8
i
1
G2
F 12 3
HR
3 on
WP A 1 2 3 4
POWER SIM
4
B 1 2 3 4 5 6 7 8
1
G2
G HR
3 on
WP A 1 2 3 4
a
POWER SIM c
-
+
!
b
A0027931
26 Endress+Hauser
Micropilot NMR81 Electrical connection
D 1 2 3 4
1
E 2
1
CDI
D 1 2 3 4
C
1
CDI
E
1
2
1 2 3 4 5 6 7 8
C 1 2 3 4 5 6 7 8
F 12 3
4
B 1 2 3 4 5 6 7 8
i
1
G2
F 12 3
HR
3 on
WP A 1 2 3 4
POWER SIM
4
B 1 2 3 4 5 6 7 8
1
G2
G HR
3 on
WP A 1 2 3 4
a
POWER SIM
–
+ b
!
A0027933
D 1 2 3 4
1
E 2
1
CDI
D 1 2 3 4
C
1
CDI
E
1
2
1 2 3 4 5 6 7 8
C 1 2 3 4 5 6 7 8
F 21 3
4
B 1 2 3 4 5 6 7 8
i
1
G2
F 12 3
HR
3 on
WP A 1 2 3 4
POWER SIM
4
B 1 2 3 4 5 6 7 8
1
G2
G HR
3 on
WP A 1 2 3 4
a
POWER SIM
–
+ b
A0027934
9 Passive usage of the Analog I/O module in the HART master mode
a Power supply
b Up to 6 external devices with HART signal output
Endress+Hauser 27
Electrical connection Micropilot NMR81
D 1 2 3 4
1
E 2
1
CDI
D 1 2 3 4
C
1
CDI
E
1
2
1 2 3 4 5 6 7 8
C 1 2 3 4 5 6 7 8
F 12 3
4
B 1 2 3 4 5 6 7 8
i
1
G2
F 12 3
HR
3 on
WP A 1 2 3 4
POWER SIM
4
B 1 2 3 4 5 6 7 8
1
G2
G HR
3 on
WP A 1- 2 3 +4
POWER SIM b
a
A0027932
D 1 2 3 4
1
E 2
1
CDI
D 1 2 3 4
C
1
CDI
E
1
2
1 2 3 4 5 6 7 8
C 1 2 3 4 5 6 7 8
F 21 3
4
B 1 2 3 4 5 6 7 8
i
1
G2
F 12 3
HR
3 on
WP A 1 2 3 4
POWER SIM
4
B 1 2 3 4 5 6 7 8
1
G2
G HR
3 on
POWER
WP A 1- 2 3 +4
SIM
A0027935
28 Endress+Hauser
Micropilot NMR81 Electrical connection
D 1 2 3 4
1
E 2
1
CDI
D 1 2 3 4
C
1
CDI
E
1
2
1 2 3 4 5 6 7 8
C 1 2 3 4 5 6 7 8
F 12 3
4
B 1 2 3 4 5 6 7 8
i
1
G2
F 12 3
HR
3 on
WP A 1 2 3 4
POWER SIM
4
B 1 2 3 4 5 6 7 8
1
G2
G HR
3 on
POWER
WP A 1 2 3 4
SIM
- +
A0027936
12 Active usage of the Analog I/O module in the HART master mode
a Up to 6 external devices with HART signal output
The maximum current consumption for the connected HART devices is 24 mA (i.e.
4 mA per device if 6 devices are connected).
A B C
1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8
A0026371
Endress+Hauser 29
Electrical connection Micropilot NMR81
D 1 2 3 4
1
E 2 C 1 2 3 4 5 6 7
1
CDI
C 1 2 3 4 5 6 7 8
i C1-2 C3-4
F 12 3
4
B 1 2 3 4 5 6 7 8
A 1 2 3 4
HR
on
POWER
WP A 1 2 3 4
SIM
A1-2 A3-4
A0026424
30 Endress+Hauser
Micropilot NMR81 Electrical connection
Terminals
Ground terminal in the terminal compartment max. 2.5 mm2 (13 AWG)
V1 communication line
• Two wire (twisted pair) screened or un-screened cable
• Resistance in one cable: ≤ 120 Ω
• Capacitance between lines: ≤ 0.3 µF
Endress+Hauser 31
Electrical connection Micropilot NMR81
A0013960
5. Insert blind plugs appropriate for the safety rating of the device (e.g. Ex d/XP).
m Are all cable glands installed, firmly tightened and correctly sealed?
m Does the supply voltage match the specifications on the transmitter nameplate?
If supply voltage is present: Is the device ready for operation and do values appear on the display
m
module?
32 Endress+Hauser
Micropilot NMR81 Operability
7 Operability
Endress+Hauser 33
Operability Micropilot NMR81
1) On entering the "Expert" menu, an access code is always requested. If a customer specific access code has
not been defined, "0000" has to be entered.
34 Endress+Hauser
Micropilot NMR81 Operability
XX
XX
XX
XX
X
mm
1
2
A0028345
Endress+Hauser 35
Operability Micropilot NMR81
2 XXXXXXXXX F 3
6
4 4841.00
5 mm
A0028317
Status symbols
Symbol Meaning
"Failure"
A device error is present. The measured value is no longer valid.
A0013956
"Function check"
The device is in service mode (e.g. during a simulation).
A0013959
"Out of specification"
The device is operated:
• Outside of its technical specifications (e.g. during startup or a cleaning)
A0013958
• Outside of the configuration carried out by the user (e.g. level outside configured span)
"Maintenance required"
Maintenance is required. The measured value is still valid.
A0013957
• Tank level
• Measured level
A0028148
• Tank level %
Water level
A0028149
Liquid temperature
A0028528
Vapor temperature
A0028528 A0027990
Air temperature
A0028528 A0027991
• Tank ullage
• Tank ullage %
A0027993
P1 (bottom)
1
A0028151 A0028141
36 Endress+Hauser
Micropilot NMR81 Operability
P2 (middle)
2
A0028151 A0028142
P3 (top)
3
A0028151 A0028146
GP 1 value
1 This is used for an external device.
A0027992 A0028141
GP 2 value
2
A0027992 A0028142
This is used for an external device.
GP 3 value
3
A0027992 A0028146
This is used for an external device.
GP 4 value
4 This is used for an external device.
A0027992 A0028147
Status "Alarm"
The measurment is interrupted. The output assumes the defined alarm value. A diagnostic
message is generated.
A0012102
Status "Warning"
The device continues measuring. A diagnostic message is generated.
A0012103
Display parameter
Marks display-only parameters which cannot be edited.
A0011978
Device locked
• In front of a parameter name: The device is locked via software and/or hardware.
A0011979 • In the header of the measured value screen: The device is locked via hardware.
Enter key
• Pressing the key briefly opens the operating menu.
A0028326 • Pressing the key for 2 s opens the context menu.
Endress+Hauser 37
Operability Micropilot NMR81
1 /../Setup 00215-1 2
Tank level
5500.00 mm
Mapping
Advanced Setup 3
A0028346-EN
16 Navigation view
1 Current submenu or wizard
2 Quick access code
3 Display area for navigation
Navigation symbols
Symbol Meaning
Operation
Is displayed:
A0011975
• in the main menu next to the selection Operation
• in the header, if you are in the Operation menu.
Setup
Is displayed:
A0011974
• in the main menu next to the selection Setup
• in the header, if you are in the Setup menu
Expert
Is displayed:
• in the main menu next to the selection Expert
A0011976
• in the header, if you are in the Expert menu
Diagnostics
Is displayed:
• in the main menu next to the selection Diagnostics
A0011977
• in the header, if you are in the Diagnostics menu
Submenu
A0013967
Wizard
A0013968
Parameter locked
When displayed in front of a parameter name, indicates that the parameter is locked.
A0013963
38 Endress+Hauser
Micropilot NMR81 Operability
Minus key
Moves the selection bar upwards in a picklist.
A0028324
Plus key
Moves the selection bar downwards in a picklist.
A0028325
Enter key
• Pressing the key briefly opens the selected menu, submenu or
parameter.
A0028326 • For parameters: Pressing the key for 2 s opens the help text for the
function of the parameter (if present).
Endress+Hauser 39
Operability Micropilot NMR81
1 /../Mapping
Confirm distance
Distance unknown
Distance 4756.5 mm
2
A0028349-EN
A0013972
A0013978
A0013976
A0013977
In the wizard view the meaning of the keys is indicated by the navigation symbol
directly above the respective key (softkey functionality).
40 Endress+Hauser
Micropilot NMR81 Operability
1 20
0 1 2 3 4
5 6 7 8 9
2
A0028341
Symbol Meaning
A0016619
Confirms selection.
A0013985
A0016621
A0013986
A0014040
Minus key
In the input mask, moves the selection bar to the left (backwards).
A0028324
Plus key
In the input mask, moves the selection bar to the right (forwards).
A0028325
Enter key
• Pressing the key briefly
– Opens the selected group.
A0028326 – Carries out the selected action.
• Pressing the key for 2 s confirms the edited parameter value.
Endress+Hauser 41
Operability Micropilot NMR81
A0028342
Toggle
Aa1 • Between upper-case and lower-case letters
• For entering numbers
A0013981
• For entering special characters
Confirms selection.
A0013985
A0013987
A0013986
A0014040
A0013989
A0013991
A0013990
A0013988
42 Endress+Hauser
Micropilot NMR81 Operability
Minus key
In the input mask, moves the selection bar to the left (backwards).
A0028324
Plus key
In the input mask, moves the selection bar to the right (forwards).
A0028325
Enter key
• Pressing the key briefly
– Opens the selected group.
A0028326 – Carries out the selected action.
• Pressing the key for 2 s confirms the edited parameter value.
Endress+Hauser 43
Operability Micropilot NMR81
Key Meaning
44 Endress+Hauser
Micropilot NMR81 Operability
• The description of parameters states which role is needed at least for read and write
access to each parameter → 100.
• The current user role is indicated by the Access status display parameter.
• If the access code is "0000", every user is in the Maintenance role. This is the
default setting on delivery of the device.
Endress+Hauser 45
Operability Micropilot NMR81
3
4
B
HR
on
WP A
SIM
3 mm
2.5 Nm (1.84 lbf ft)
A0028363
The display module can be attached to the edge of the electronics compartment. This
makes it easier to access the lock switch.
A0028381
46 Endress+Hauser
Micropilot NMR81 Operability
XXXXXXXXX
20.50
XX XX
A0015870
A0023737
Endress+Hauser 47
Operability Micropilot NMR81
5
6
NXA820
1 2 3
A0025621
22 Connection of Tank Gauging devices to FieldCare via the Tankvision Tank Scanner NXA820
1 Proservo NMS8x
2 Tankside Monitor NRF81
3 Micropilot NMR8x
4 Field protocol (e.g. Modbus, V1)
5 Tankvision Tank Scanner NXA820
6 Ethernet
7 Computer with FieldCare installed
48 Endress+Hauser
Micropilot NMR81 Operability
A0028515
A0028516
Open the configuration of the DTM and enter the required data (IP address of the
NXA820; "Password" = "hart"; "Tank identification" only with NXA V1.05 or higher)
Endress+Hauser 49
Operability Micropilot NMR81
5.
A0028517
A0028519
50 Endress+Hauser
Micropilot NMR81 System integration
8 System integration
Manufacturer ID 0x11
Endress+Hauser 51
Commissioning Micropilot NMR81
9 Commissioning
1
8
7
2
6
5
A0029794
52 Endress+Hauser
Micropilot NMR81 Commissioning
This setting only affects the language on the display module. To set the language in
the operating tool use the language setting functionality of FieldCare or DeviceCare,
respecitvely.
Use the following parameters to set the date and time: Year, Month, Day, Hour,
Minutes.
Endress+Hauser 53
Commissioning Micropilot NMR81
4.
Go to the Set date parameter and select the Confirm time option.
The real-time clock is set to the current date and time.
54 Endress+Hauser
Micropilot NMR81 Commissioning
Basic settings
Navigation path: Setup
Device tag Define a name to identify the measuring point within the plant.
Units preset Select a set of units for length, pressure and temperature. → 109
Empty Enter the distance from the lower edge of the device flange to the datum plate. → 110
Set level Can be used to correct a constant shift of the measured level. → 111
If the indicated level does not match the actual level: Enter the actual level into this
parameter.
An offset for the measured level is then automatically defined.
The Set level parameter can only be used to compensate for a constant level error. To
eliminate errors resulting from interference echos, use the interference echo
suppression (map).
Distance Shows the measured distance from the lower edge of the device flange to the → 102
product surface. Check whether this value is correct.
Confirm distance Specify whether the measured distance matches the actual distance. → 111
Present mapping Shows up to which distance a mapping has already been recorded. → 112
Mapping end point Determines up to which distance the new mapping will be recorded. → 113
Depending on the selection in Confirm distance a suitable value is preset in
this parameter. Usually, there is no need to change this value.
Record map Select Record map. This starts the recording of the new map. → 113
Dip table
The dip table is used to correct the level readings using independently taken hand dips.
The dip table is used in particular to adapt the level gauge to the specific application
conditions such as a mechanical offset and the tank or stilling well design.
The dip table is managed in the Dip-table submenu → 193.
Endress+Hauser 55
Commissioning Micropilot NMR81
4-20mA inputs → 59
RTD input → 60
Digital inputs → 62
Limit evaluation → 68
4-20mA output → 70
Modbus → 69
V1 → 70
56 Endress+Hauser
Micropilot NMR81 Commissioning
If HART devices are connected to an Analog I/O module (slot B or C in the terminal
compartment), this module must be configured as follows 2):
1. Make sure the HART devices are connected as defined by the terminal assignment
→ 26.
2. Navigate to the submenu of the respective Analog I/O module: Setup → Advanced
setup → Input/output → Analog I/O
3. Go to the Operating mode parameter (→ 128).
4. If only one HART device is connected to this loop:
Select the HART master+4..20mA input option. In this case the 4-20mA signal can
be used in addition to the HART signal.
5. If up to 6 HART devices are connected to this loop:
Select the HART master option.
1) The current software does not support HART devices with adress 0 (zero).
2) This procedure is not required for HART devices connected to the HART Ex ia input (terminal area E in the terminal compartment).
Endress+Hauser 57
Commissioning Micropilot NMR81
A0030098
3) In this expression, "X" is the HART address of the NMT. Its default value is "2".
58 Endress+Hauser
Micropilot NMR81 Commissioning
I [mA]
20mA
4mA
1 4 2 A
A0029264
The Analog I/O submenu contains additional parameters for a more detailed
configuration of the Analog Input. For a description refer to : → 128
Endress+Hauser 59
Commissioning Micropilot NMR81
3
1
A0029269
1 Datum plate
2 RTD
3 Probe position (→ 126)
Go to the Probe position parameter and enter the mounting position of the RTD
(measured from the datum plate).
This parameter, in conjunction with the measured level, determines whether the
measured temperature refers to the product or to the gas phase.
60 Endress+Hauser
Micropilot NMR81 Commissioning
R[Ω] R[Ω]
2
1
T T
A0029265
1 Ohms offset
2 Temperature offset after conversion
Endress+Hauser 61
Commissioning Micropilot NMR81
A B
DIO DIO
A0029262
The internal state of the Digital Input can then be transferred to a Digital Output or can be
used to control the measurement.
The Digital Xx-x submenu contains additional parameters for a more detailed
configuration of the Digital Input. For a description refer to → 137.
62 Endress+Hauser
Micropilot NMR81 Commissioning
Bottom water level Setup → Advanced setup → Application → Tank configuration → Level
→ Water level source
Temperature of the air surrounding Setup → Advanced setup → Application → Tank configuration
the tank → Temperature → Air temperature source
Temperature of the vapor above the Setup → Advanced setup → Tank configuration → Temperature → Vapor
product temp source
Bottom pressure (P1) Setup → Advanced setup → Application → Tank configuration → Pressure
→ P1 (bottom) source
Top pressure (P3) Setup → Advanced setup → Application → Tank configuration → Pressure
→ P3 (top) source
Depending on the application not all these parameters will be relevant in a given
situation.
Endress+Hauser 63
Commissioning Micropilot NMR81
A B
1 3 1
2 2
A0029255
1. Navigate to: "Setup → Level source" and specify from which device the level is
obtained.
2. If a temperature transmitter is connected:
Navigate to: "Setup → Liquid temp source" and specify from which device the
temperature is obtained.
64 Endress+Hauser
Micropilot NMR81 Commissioning
A B
1 4 1
2 2
D3
D1
3 3
A0029256
Endress+Hauser 65
Commissioning Micropilot NMR81
C
D
A
A0028722
66 Endress+Hauser
Micropilot NMR81 Commissioning
Endress+Hauser 67
Commissioning Micropilot NMR81
A B
10 10
1 1
2 2
3 3
4 4
5 5
6 6
7 7
8 8
A0029539
68 Endress+Hauser
Micropilot NMR81 Commissioning
HART output
Endress+Hauser 69
Commissioning Micropilot NMR81
4-20mA output
Each Analog I/O module of the device can be configured as a 4...20mA analog output. To
do so, proceed as follows:
1. Navigate to: Setup → Advanced setup → Input/output → Analog I/O
2. Go to the Operating mode parameter and select the 4..20mA output option.
3. Go to the Analog input source parameter and select the tank variable which is to be
transmitted via the 4...20mA output.
4. Go to the 0 % value parameter and enter the value of the selected tank variable
which will be mapped to 4 mA
5. Go to the 100 % value parameter and enter the value of the selected tank variable
which will be mapped to 20 mA.
The Analog I/O submenu contains more parameters which can be used for a more
detailed configuration of the analog output. For a description see → 128
Modbus
The Micropilot NMR8x acts as a Modbus slave. Measured or calculated tank values are
stored in registers which can be requested by a Modbus master.
The following submenu is used to configure the communication between the device and
the Modbus master:
Setup → Advanced setup → Communication → Modbus X1-4 →Configuration (→ 144)
V1
The following submenus are used to configure the V1 communication between the device
and the control system:
• Setup → Advanced setup → Communication → V1 X1-4 → Configuration (→ 147)
• Setup → Advanced setup → Communication → V1 X1-4 → V1 input selector (→ 150)
70 Endress+Hauser
Micropilot NMR81 Commissioning
9.6 Simulation
To check the correct configuration of the device and of the control system, it is possible to
simulate different situations (measured values, diagnostic messages etc.). See the
Simulation submenu (→ 231) for details.
Endress+Hauser 71
Operation Micropilot NMR81
10 Operation
Hardware locked The device is locked by the write-protection switch in the terminal → 46
compartment. This is indicated by the write protection symbol in the
header of the display:
XXXXXXXXX
20.50
XX XX
A0015870
SIL locked (in The device is in SIL-locked mode. See the SIL
preparation) Safety manual
(in preparation)
Temporarily Write access to the parameters is temporarily lock due to device-internal Wait for
locked processing (e.g. data upload/download, reset). Once the internal completion of
processing has been completed, the parameters can be changed once the device-
again. internal
processing.
72 Endress+Hauser
Micropilot NMR81 Diagnostics and troubleshooting
Device does not respond. Supply voltage not connected. Connect the correct voltage.
The cables do not contact the Ensure electrical contact between the
terminals properly. cable and the terminal.
Values on the display The plug of the display cable is not Connect the plug correctly.
invisible connected correctly.
CDI communication does not Wrong setting of the COM port on Check the setting of the COM port on
work. the computer. the computer and change it if necessary.
Endress+Hauser 73
Diagnostics and troubleshooting Micropilot NMR81
2 1
XXXXXXXXX S XXXXXXXXX S
20.50 S801
Supply voltage
3
4
X XX i Menu
5
A0013939-EN
1 Status signal
2 Status symbol (symbol for event level)
3 Status symbol with diagnostics event
4 Event text
5 Operating elements
Status signals
"Failure"
A device error is present. The measured value is no longer valid.
A0013956
"Function check"
The device is in service mode (e.g. during a simulation).
A0013959
"Out of specification"
The device is operated:
• Outside of its technical specifications (e.g. during startup or a cleaning)
A0013958
• Outside of the configuration carried out by the user (e.g. level outside configured span)
"Maintenance required"
Maintenance is required. The measured value is still valid.
A0013957
"Alarm" status
The measurement is interrupted. The signal outputs take on the defined alarm condition.
A0013961 A diagnostic message is generated.
"Warning" status
The device continues to measure. A diagnostic message is generated.
A0013962
74 Endress+Hauser
Micropilot NMR81 Diagnostics and troubleshooting
Diagnostics event
↓ ↓ ↓
3-digit number
If two or more diagnostic messages are pending simultaneously, only the message with the
highest priority is shown. Additional pending diagnostic messages can be shown in
Diagnostic list submenu (→ 228).
Operating elements
Plus key
Opens the message about the remedial measures.
A0013970
Enter key
Opens the operating menu.
A0013952
Endress+Hauser 75
Diagnostics and troubleshooting Micropilot NMR81
XXXXXXXXX S XXXXXXXXX S
20.50 S801
Supply voltage
X XX i Menu
Diagnostic list S
Diagnostics 1
1 S801 Supply voltage
Diagnostics 2
Diagnostics 3
A0013940-EN
The user is in the Diagnostics menu at an entry for a diagnostics event, e.g. in the
Diagnostic list submenu or in the Previous diagnostics.
1. Press .
The message for the remedial measures for the selected diagnostic event opens.
2. Press + simultaneously.
The message for the remedial measures closes.
76 Endress+Hauser
Micropilot NMR81 Diagnostics and troubleshooting
Xxxxxx/…/…/
Device name: Xxxxxxx Mass flow: 12.34 kg/h
Device tag: Xxxxxxx Volume flow: 12.34 m³/h
Status signal: Function check (C)
Xxxxxx
P Diagnostics 1: C485 Simu...
P Remedy information: Deactivate...
P Access status tooling: Mainenance Failure (F)
A0021799-EN
Furthermore, diagnostic events that have occurred can be viewed in the Diagnostic
list submenu.
Symbol Meaning
Failure
A device error has occurred. The measured value is no longer valid.
A0017271
Function check
The device is in service mode (e.g. during a simulation).
A0017278
Out of specification
The device is operated outside its technical specification limits (e.g. outside the process
A0017277 temperature range)
Maintenance required
Maintenance is required. The measured value is still valid.
A0017276
The status signals are categorized in accordance with VDI/VDE 2650 and NAMUR
Recommendation NE 107.
Endress+Hauser 77
Diagnostics and troubleshooting Micropilot NMR81
Diagnostic information
Diagnostic code
Diagnostic Diagnostic
Status signal Short text
behavior number
↓ ↓ ↓
Electronic
Example 261
modules
A0013962 A0013956
3-digit number
78 Endress+Hauser
Micropilot NMR81 Diagnostics and troubleshooting
Diagnostic of sensor
Diagnostic of electronic
Endress+Hauser 79
Diagnostics and troubleshooting Micropilot NMR81
Diagnostic of configuration
443 AIO 1 to 2 Input not HART AIO select compatible HART input. C Warning
compatible
80 Endress+Hauser
Micropilot NMR81 Diagnostics and troubleshooting
Diagnostic of process
Endress+Hauser 81
Diagnostics and troubleshooting Micropilot NMR81
82 Endress+Hauser
Micropilot NMR81 Diagnostics and troubleshooting
Endress+Hauser 83
Maintenance Micropilot NMR81
12 Maintenance
For a list of some of the measuring and test equipment, refer to the "Accessories"
chapter of the "Technical Information" document for the device.
84 Endress+Hauser
Micropilot NMR81 Repair
13 Repair
Endress+Hauser 85
Repair Micropilot NMR81
13.4 Return
The measuring device must be returned if repairs or a factory calibration are required, or if
the wrong measuring device has been ordered or delivered. According to legal regulations,
Endress+Hauser, as an ISO-certified company, is required to follow certain procedures
when handling returned products that are in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures
and conditions on the Endress+Hauser website at
www.services.endress.com/return-material
13.5 Disposal
Observe the following notes during disposal:
• Observe valid federal/national regulations.
• Ensure proper separation and reuse of the device components.
86 Endress+Hauser
Micropilot NMR81 Accessories
14 Accessories
361.6 (14.2)
319 (12.6)
32.5 (1.28)
56.1 (2.21)
A0028019
Materials
Part Material
• The weather protection cover can be ordered together with the device:
Ordering feature 620 "Accessory Enclosed", option PA "Weather Protection Cover")
• It can also be ordered as an accessory:
Order code: 71292751 (for NMR8x and NRF8)
Endress+Hauser 87
Accessories Micropilot NMR81
4°
ød
øD
1
2
h
±8 °
hmax
3
hmin
A0027787
PS PT PU
2)
Order code 71285499 71285501 71285503
Length of screws 100 mm (3.9 in) 100 mm (3.9 in) 100 mm (3.9 in)
1) With this ordering feature the adjustable seal is supplied together with the device.
2) This order code must be used if the adjustable seal is ordered separately.
88 Endress+Hauser
Micropilot NMR81 Accessories
Accessory Description
Fieldgate FXA320 Gateway for remote monitoring of connected 4-20mA measuring devices via web
browser.
Accessory Description
Fieldgate FXA520 Gateway for remote diagnosis and parametrization of connected HART measuring
devices via web browser.
Commubox FXA195 For intrinsically safe HART communication with FieldCare via the USB interface.
HART
For details refer to Technical Information TI00404F
Accessory Description
Commubox FXA291 Connects Endress+Hauser field devices with CDI interface (= Endress+Hauser
Common Data Interface) to the USB interface of a computer.
Accessory Description
Endress+Hauser 89
Accessories Micropilot NMR81
RIA15 Compact process display unit with very low voltage drop for universal use to display
4 to 20 mA/HART signals
90 Endress+Hauser
Micropilot NMR81 Operating menu
15 Operating menu
• : Navigation path for operating module at the device
• : Navigation path for operating tool (e.g. FieldCare)
• : Parameter can be locked via software locking
Navigation Bedientool
Operation → 100
‣ Level → 100
Distance → 102
‣ Temperature → 102
Endress+Hauser 91
Operating menu Micropilot NMR81
‣ Density → 104
‣ Pressure → 106
P1 (bottom) → 106
P3 (top) → 106
‣ GP values → 107
GP 1 to 4 name → 107
GP Value 1 → 107
GP Value 2 → 107
GP Value 3 → 107
GP Value 4 → 108
Setup → 109
92 Endress+Hauser
Micropilot NMR81 Operating menu
Empty → 110
Distance → 102
‣ Input/output → 116
‣ Analog IP → 123
Endress+Hauser 93
Operating menu Micropilot NMR81
0 % value → 125
0 % value → 132
94 Endress+Hauser
Micropilot NMR81 Operating menu
‣ Communication → 142
‣ Communication interface 1 to 2
‣ Configuration → 144
‣ Configuration → 147
‣ Configuration → 153
‣ Information → 161
Endress+Hauser 95
Operating menu Micropilot NMR81
‣ Application → 163
‣ Level → 163
‣ Temperature → 166
‣ Density → 169
‣ Pressure → 171
‣ HyTD → 178
‣ CTSh → 184
‣ HTMS → 189
‣ Dip-table → 196
‣ Alarm
‣ Alarm 1 to 4 → 198
‣ Information → 209
96 Endress+Hauser
Micropilot NMR81 Operating menu
Distance → 102
‣ Display → 212
Language → 212
Separator → 214
Header → 215
Backlight → 216
Endress+Hauser 97
Operating menu Micropilot NMR81
Date/time → 221
Year → 222
Month → 222
Day → 222
Hour → 223
Minute → 223
‣ Administration → 225
Diagnostics → 226
Timestamp → 226
Timestamp → 227
Date/time → 221
Diagnostics 1 to 5 → 228
Timestamp 1 to 5 → 228
98 Endress+Hauser
Micropilot NMR81 Operating menu
‣ Simulation → 231
Endress+Hauser 99
Operating menu Micropilot NMR81
Navigation Operation
Tank level
Description Shows the distance from the zero position (tank bottom or datum plate) to the product
surface.
Write access -
Tank Level %
Write access -
Tank ullage
Write access -
100 Endress+Hauser
Micropilot NMR81 Operating menu
Tank ullage %
Description Shows the remaining empty space in the tank as a percentage of the full tank space.
Write access -
Description Shows measured interface level from zero position (tank bottom or datum plate). Value is
updated when device generates a valid Interface measurement.
Write access -
Description Shows measured interface level from zero position (tank bottom or datum plate). Value is
updated when device generates a valid interface measurement.
Write access -
Bottom level
Write access -
Endress+Hauser 101
Operating menu Micropilot NMR81
Water level
Write access -
Measured level
Description Shows the measured level without any correction from the tank calculations.
Write access -
Distance
Write access -
Air temperature
Write access -
102 Endress+Hauser
Micropilot NMR81 Operating menu
Liquid temperature
Write access -
Vapor temperature
Write access -
Element temperature 0 to 23
Navigation Operation → Temperature → NMT elem. values → Element temp. → Element temp 0
to 23
Write access -
Endress+Hauser 103
Operating menu Micropilot NMR81
Element position 0 to 23
Navigation Operation → Temperature → NMT elem. values → Element position → Element pos.
0 to 23
Write access -
Observed density
Write access -
This value is calculated from different measured variables depending on the selected
calculation method → 176.
Vapor density
104 Endress+Hauser
Micropilot NMR81 Operating menu
Air density
Write access -
Write access -
Endress+Hauser 105
Operating menu Micropilot NMR81
Write access -
P1 (bottom)
Write access -
P3 (top)
Write access -
106 Endress+Hauser
Micropilot NMR81 Operating menu
GP 1 to 4 name
GP Value 1
Write access -
GP Value 2
Write access -
GP Value 3
Endress+Hauser 107
Operating menu Micropilot NMR81
Write access -
GP Value 4
Write access -
108 Endress+Hauser
Micropilot NMR81 Operating menu
Device tag
Description Enter a unique name for the measuring point to identify the device quickly within the
plant.
Units preset
If the Customer value option is selected, the units are defined in the following
parameters:
• Distance unit (→ 218)
• Pressure unit (→ 219)
• Temperature unit (→ 219)
• Mass unit (→ 219)
• Density unit (→ 220)
In any other case these are read-only parameters used to indicate the respective unit.
Endress+Hauser 109
Operating menu Micropilot NMR81
Empty
Description Distance from reference point to zero position (tank bottom or datum plate).
Description Defines the distance from the dipping reference point to the zero position (tank bottom or
datum plate).
Tank level
Description Shows the distance from the zero position (tank bottom or datum plate) to the product
surface.
Write access -
110 Endress+Hauser
Micropilot NMR81 Operating menu
Set level
Description If the level measured by the device does not match the actual level obtained by a manual
dip, enter the correct level into this parameter.
Factory setting 0 mm
The device adjusts the Empty parameter (→ 110) according to the entered value, such
that the measured level will match the actual level.
Confirm distance
Description Specify, whether the measured distance matches the real distance. Depending on the
selection the device automatically sets the range of mapping.
Selection • Distance ok
• Distance unknown
• Distance too small *
• Distance too big *
• Tank empty
• Manual map
• Factory map
Endress+Hauser 111
Operating menu Micropilot NMR81
Present mapping
Write access -
4) Only available for " Evaluation mode (→ 211)" = "Short time history"
112 Endress+Hauser
Micropilot NMR81 Operating menu
Record map
Selection • No
• Record map
• Recalculate map
• Overlay map
• Factory map
• Delete partial map
• Stop overlay
Factory setting No
Endress+Hauser 113
Operating menu Micropilot NMR81
Distance
Write access -
114 Endress+Hauser
Micropilot NMR81 Operating menu
Locking status
Description Indicates the write protection with the highest priority that is currently active.
Write access -
Description Shows the access authorization to the parameters via the operating tool.
Write access -
Endress+Hauser 115
Operating menu Micropilot NMR81
"Input/output" submenu
Number of devices
Write access -
116 Endress+Hauser
Micropilot NMR81 Operating menu
Device name
Write access -
Polling address
Write access -
Device tag
Write access -
Endress+Hauser 117
Operating menu Micropilot NMR81
Operating mode
Selection • PV only
• PV,SV,TV & QV
• Level
• Measured level
Communication status
Write access -
Write access -
118 Endress+Hauser
Micropilot NMR81 Operating menu
Write access -
Write access -
Write access -
Output pressure
Prerequisite Not available for Micropilot S FMR5xx and Prothermo 53x. (In these cases the measured
variables are allocated automatically).
Endress+Hauser 119
Operating menu Micropilot NMR81
Selection • No value
• Primary variable (PV)
• Secondary variable (SV)
• Tertiary variable (TV)
• Quaternary variable (QV)
Output density
Prerequisite Not available for Micropilot S FMR5xx and Prothermo 53x. (In these cases the measured
variables are allocated automatically).
Selection • No value
• Primary variable (PV)
• Secondary variable (SV)
• Tertiary variable (TV)
• Quaternary variable (QV)
Output temperature
Prerequisite Not available for Micropilot S FMR5xx and Prothermo 53x. (In these cases the measured
variables are allocated automatically).
Description Defines which variable of the HART device is the product temperature.
Selection • No value
• Primary variable (PV)
• Secondary variable (SV)
• Tertiary variable (TV)
• Quaternary variable (QV)
120 Endress+Hauser
Micropilot NMR81 Operating menu
Prerequisite Not available for Micropilot S FMR5xx and Prothermo 53x. (In these cases the measured
variables are allocated automatically).
Description Defines which variable of the HART device is the vapor temperature.
Selection • No value
• Primary variable (PV)
• Secondary variable (SV)
• Tertiary variable (TV)
• Quaternary variable (QV)
Output level
Prerequisite Not available for Micropilot S FMR5xx and Prothermo 53x. (In these cases the measured
variables are allocated automatically).
Selection • No value
• Primary variable (PV)
• Secondary variable (SV)
• Tertiary variable (TV)
• Quaternary variable (QV)
Endress+Hauser 121
Operating menu Micropilot NMR81
Forget device
Navigation Setup → Advanced setup → Input/output → HART devices → Forget device → Forget
device
122 Endress+Hauser
Micropilot NMR81 Operating menu
Operating mode
Selection • Disabled
• RTD temperature input
• Gauge power supply
RTD type
Endress+Hauser 123
Operating menu Micropilot NMR81
Process value
Description Shows the measured value received via the analog input.
Write access -
Process variable
124 Endress+Hauser
Micropilot NMR81 Operating menu
0 % value
Factory setting 0 mm
100 % value
Factory setting 0 mm
Description Minimum approved temperature of the connected probe. If the temperature falls below
this value, the W&M status will be 'invalid'.
Endress+Hauser 125
Operating menu Micropilot NMR81
If the temperature rises above the value entered in this parameter, the W&M status will be
"invalid".
Probe position
Description The position of the temperature probe (measured from the datum plate).
This parameter, in conjunction with the measured level, determines whether the
temperature probe is still covered by the product. If this is no longer the case, the status of
the temperature value will be "invalid".
126 Endress+Hauser
Micropilot NMR81 Operating menu
Damping factor
Factory setting 0s
Gauge current
Description Shows the current on the power supply line for the connected Micropilot S.
Write access -
Endress+Hauser 127
Operating menu Micropilot NMR81
Operating mode
Selection • Disabled
• 4..20mA input
• HART master+4..20mA input
• HART master
• 4..20mA output
• HART slave +4..20mA output
Disabled - -
Depending on the terminals used, the Analog I/O module is used in the passive or active
mode.
1 2 3
128 Endress+Hauser
Micropilot NMR81 Operating menu
Current span
Prerequisite Operating mode parameter (→ 128) ≠ Disabled option or HART master option
Description Defines the current range for the measured value transmission.
Fixed current Constant current, defined in the Fixed current parameter (→ 129).
In the case of an error, the output current assumes the value defined in the Failure
mode parameter (→ 131).
Fixed current
Factory setting 4 mA
Endress+Hauser 129
Operating menu Micropilot NMR81
Prerequisite Operating mode (→ 128) = 4..20mA output or HART slave +4..20mA output
Description Defines the value which will be transmitted via the Analog I/O module.
130 Endress+Hauser
Micropilot NMR81 Operating menu
Failure mode
Prerequisite Operating mode (→ 128) = 4..20mA output or HART slave +4..20mA output
Selection • Min.
• Max.
• Last valid value
• Actual value
• Defined value
Error value
Factory setting 22 mA
Input value
Prerequisite Operating mode (→ 128) = 4..20mA output or HART slave +4..20mA output
Write access -
Endress+Hauser 131
Operating menu Micropilot NMR81
0 % value
Prerequisite Operating mode (→ 128) = 4..20mA output or HART slave +4..20mA output
100 % value
Prerequisite Operating mode (→ 128) = 4..20mA output or HART slave +4..20mA output
Input value %
Prerequisite Operating mode (→ 128) = 4..20mA output or HART slave +4..20mA output
Description Shows the output value as a percentage of the complete 4...20mA range.
Write access -
132 Endress+Hauser
Micropilot NMR81 Operating menu
Output value
Prerequisite Operating mode (→ 128) = 4..20mA output or HART slave +4..20mA output
Write access -
Process variable
Factory setting 0 mm
Endress+Hauser 133
Operating menu Micropilot NMR81
Factory setting 0 mm
Navigation Setup → Advanced setup → Input/output → Analog I/O → Error event type
Description Defines the type of event in case of an error in the analog I/O module.
Selection • None
• Warning
• Alarm
Process value
Write access -
134 Endress+Hauser
Micropilot NMR81 Operating menu
Input value in mA
Write access -
Navigation Setup → Advanced setup → Input/output → Analog I/O → Input value [%]
Description Shows the input value as a percentage of the complete 4...20mA current range.
Write access -
Damping factor
Factory setting 0s
Navigation Setup → Advanced setup → Input/output → Analog I/O → Used for SIL
Endress+Hauser 135
Operating menu Micropilot NMR81
Description in preparation.
Selection • Enabled
• Disabled
Navigation Setup → Advanced setup → Input/output → Analog I/O → Expec. SIL chain
Description in preparation.
Write access -
136 Endress+Hauser
Micropilot NMR81 Operating menu
D 1 2 3 4
1
E 2 C 1 2 3 4 5 6 7
1
CDI
C 1 2 3 4 5 6 7 8
i C1-2 C3-4
F 12 3
4
B 1 2 3 4 5 6 7 8
A 1 2 3 4
HR
on
POWER
WP A 1 2 3 4
SIM
A1-2 A3-4
A0026424
Operating mode
Selection • Disabled
• Output passive
• Input passive
• Input active
Endress+Hauser 137
Operating menu Micropilot NMR81
Selection • None
• Alarm x any
• Alarm x High
• Alarm x HighHigh
• Alarm x High or HighHigh
• Alarm x Low
• Alarm x LowLow
• Alarm x Low or LowLow
• Digital Xx-x
• Primary Modbus x
• Secondary Modbus x
Input value
Prerequisite Operating mode (→ 137) = "Input passive" option or "Input active" option
Write access -
5) Only present if "Operating mode (→ 137)" = "Input passive" or "Input active" for the respective Digital I/O module.
138 Endress+Hauser
Micropilot NMR81 Operating menu
Contact type
Output simulation
Selection • Disable
• Simulating active
• Simulating inactive
• Fault 1
• Fault 2
3 1
A0028602
Endress+Hauser 139
Operating menu Micropilot NMR81
The switching state of these relays is defined by the Output simulation parameter as
follows:
The Fault 1 and Fault 2 options can be used to check the correct switching behavior
of the two relays.
Output value
Write access -
Readback value
Write access -
Navigation Setup → Advanced setup → Input/output → Digital Xx-x → Used for SIL
140 Endress+Hauser
Micropilot NMR81 Operating menu
Selection • Enabled
• Disabled
Endress+Hauser 141
Operating menu Micropilot NMR81
"Communication" submenu
This menu contains a submenu for each digital communication interface of the device. The
communication interfaces are designated by "X1-4" where "X" specifies the slot in the
terminal compartmen and "1-4" the terminals within this slot.
D 1 2 3 4
1
E 2 D 1 2 3 4
1
CDI
C 1 2 3 4 5 6 7 8
i D1-4
F 12 3
4
B 1 2 3 4 5 6 7 8
A 1 2 3 4
HR
on
POWER
WP A 1 2 3 4
SIM
A1-4
A0031200
33 Designation of the "Modbus" or "V1" modules (examples); depending on the device version these modules
may also be in slot B or C.
Navigation Setup → Advanced setup → Communication → Modbus X1-4 → Commu I/F protoc
Write access -
Description in preparation.
142 Endress+Hauser
Micropilot NMR81 Operating menu
Write access -
Communications info
Description in preparation.
Write access -
Endress+Hauser 143
Operating menu Micropilot NMR81
"Configuration" submenu
Baudrate
Parity
Selection • Odd
• Even
• None / 1 stop bit
• None / 2 stop bits
144 Endress+Hauser
Micropilot NMR81 Operating menu
Modbus address
Factory setting 1
Bus termination
Description Activates or deactivates the bus termination at the device. Should only be activated on the
last device in a loop.
Selection • Off
• On
Endress+Hauser 145
Operating menu Micropilot NMR81
146 Endress+Hauser
Micropilot NMR81 Operating menu
"Configuration" submenu
This submenu is only present for devices with a V1 communication interface.
Selection • None
• V1
V1 address
User entry 0 to 99
Factory setting 1
V1 address
Endress+Hauser 147
Operating menu Micropilot NMR81
Factory setting 1
Level mapping
Description Determines the transmittable range of levels. '+ve' Only positive values can be transmitted
(0.0 mmm to 99999.9 mm) '+ve & -ve' Positive and negative numbers can be transmitted
(-49999.9 mm to +50000.0 mm).
Selection • +ve
• +ve & -ve
In V1, the level is always represented by a number in the range from 0 to 999 999. This
number corresponds to a level as follows:
0 0.0 mm
0 0.0 mm
148 Endress+Hauser
Micropilot NMR81 Operating menu
Line impedance
Description Adjusts the impedance of the communication line. Note: The line impedance affects the
voltage difference between a logical 0 and a logical 1 on the message of the device to the
bus. The default setting is suitable for most applications.
User entry 0 to 15
Factory setting 15
The line impedance affects the voltage difference between a logical 0 and a logical 1
on the message of the device to the bus. The default setting is suitable for most
applications.
Service relay
Description Activates the service relay and disconnects the device from the bus.
Selection • Off
• On
Endress+Hauser 149
Operating menu Micropilot NMR81
Selection • None
• Alarm 1 any
• Alarm 1 High
• Alarm 1 HighHigh
• Alarm 1 High or HighHigh
• Alarm 1 Low
• Alarm 1 LowLow
• Alarm 1 Low or LowLow
• Alarm 2 any
• Alarm 2 High
• Alarm 2 HighHigh
• Alarm 2 High or HighHigh
• Alarm 2 Low
• Alarm 2 LowLow
• Alarm 2 Low or LowLow
• Alarm 3 any
• Alarm 3 High
• Alarm 3 HighHigh
• Alarm 3 High or HighHigh
• Alarm 3 Low
• Alarm 3 LowLow
• Alarm 3 Low or LowLow
• Alarm 4 any
• Alarm 4 High
• Alarm 4 HighHigh
• Alarm 4 High or HighHigh
• Alarm 4 Low
• Alarm 4 LowLow
• Alarm 4 Low or LowLow
150 Endress+Hauser
Micropilot NMR81 Operating menu
Selection • None
• Alarm 1 any
• Alarm 1 High
• Alarm 1 HighHigh
• Alarm 1 High or HighHigh
• Alarm 1 Low
• Alarm 1 LowLow
• Alarm 1 Low or LowLow
• Alarm 2 any
• Alarm 2 High
• Alarm 2 HighHigh
• Alarm 2 High or HighHigh
• Alarm 2 Low
• Alarm 2 LowLow
• Alarm 2 Low or LowLow
• Alarm 3 any
• Alarm 3 High
• Alarm 3 HighHigh
• Alarm 3 High or HighHigh
• Alarm 3 Low
• Alarm 3 LowLow
• Alarm 3 Low or LowLow
• Alarm 4 any
• Alarm 4 High
• Alarm 4 HighHigh
• Alarm 4 High or HighHigh
• Alarm 4 Low
• Alarm 4 LowLow
• Alarm 4 Low or LowLow
Description in preparation.
Endress+Hauser 151
Operating menu Micropilot NMR81
Selection • None
• Tank level %
• Tank ullage %
• AIO B1-3 value %
• AIO C1-3 value %
152 Endress+Hauser
Micropilot NMR81 Operating menu
"Configuration" submenu
User entry 0 to 63
Factory setting 15
No. of preambles
User entry 5 to 20
Factory setting 5
Assign PV
Endress+Hauser 153
Operating menu Micropilot NMR81
Write access -
0 % value
Factory setting 0 mm
154 Endress+Hauser
Micropilot NMR81 Operating menu
100 % value
Factory setting 0 mm
Percent of range
Description Shows the value of the primary variable as a percentage of the defined 0% to 100% range.
Write access -
Assign SV
Endress+Hauser 155
Operating menu Micropilot NMR81
• Flow velocity
• Volume flow
• Vapor density
• Middle density
• Upper density
• Correction
• Tank level %
• GP 1 ... 4 value
• Measured level
• P3 position
• Tank reference height
• Local gravity
• P1 position
• Manual density
• Tank ullage
• Average profile density
• Lower density
• Upper interface level
• Lower interface level
• Bottom level
• Displacer position
• HART device 1 ... 15 PV
• HART device 1 ... 15 SV
• HART device 1 ... 15 TV
• HART device 1 ... 15 QV
• HART device 1 ... 15 PV mA
• HART device 1 ... 15 PV %
• Element temperature 1 ... 24
• AIO B1-3 value
• AIO C1-3 value
• AIP B4-8 value
• AIP C4-8 value
• None
Write access -
156 Endress+Hauser
Micropilot NMR81 Operating menu
Assign TV
Endress+Hauser 157
Operating menu Micropilot NMR81
Write access -
Assign QV
158 Endress+Hauser
Micropilot NMR81 Operating menu
Write access -
PV mA selector
Description Defines which parameter will be returned as the primary value (PV) current.
Selection • None
• AIO B1-3 value mA
• AIO C1-3 value mA
• AIP B4-8 value mA
• AIP C4-8 value mA
Endress+Hauser 159
Operating menu Micropilot NMR81
PV source
160 Endress+Hauser
Micropilot NMR81 Operating menu
"Information" submenu
User entry Max. 8 characters: A-Z, 0-9 and certain special characters (e.g. punctuation marks, @, %).
Device tag
Description Enter a unique name for the measuring point to identify the device quickly within the
plant.
HART descriptor
Endress+Hauser 161
Operating menu Micropilot NMR81
HART message
Description HART message which is sent via the HART protocol if requested by the master.
162 Endress+Hauser
Micropilot NMR81 Operating menu
"Application" submenu
"Level" submenu
Empty
Description Distance from reference point to zero position (tank bottom or datum plate).
Navigation Setup → Advanced setup → Application → Tank config → Level → Tank ref height
Description Defines the distance from the dipping reference point to the zero position (tank bottom or
datum plate).
Endress+Hauser 163
Operating menu Micropilot NMR81
Tank level
Navigation Setup → Advanced setup → Application → Tank config → Level → Tank level
Description Shows the distance from the zero position (tank bottom or datum plate) to the product
surface.
Write access -
Set level
Navigation Setup → Advanced setup → Application → Tank config → Level → Set level
Description If the level measured by the device does not match the actual level obtained by a manual
dip, enter the correct level into this parameter.
Factory setting 0 mm
The device adjusts the Empty parameter (→ 110) according to the entered value, such
that the measured level will match the actual level.
Navigation Setup → Advanced setup → Application → Tank config → Level → Water level src
164 Endress+Hauser
Micropilot NMR81 Operating menu
Navigation Setup → Advanced setup → Application → Tank config → Level → Man. water level
Factory setting 0 mm
Water level
Navigation Setup → Advanced setup → Application → Tank config → Level → Water level
Write access -
Blocking distance
Navigation Setup → Advanced setup → Application → Tank config → Level → Blocking dist.
Description No echos are evaluated within the blocking distance BD. Therefore, BD can be used to
suppress interference echos in the vicinity of the antenna.
No echos are evaluated within the blocking distance BD. Therefore, BD can be used to
suppress interference echos in the vicinity of the antenna.
Endress+Hauser 165
Operating menu Micropilot NMR81
"Temperature" submenu
Navigation Setup → Advanced setup → Application → Tank config → Temperature → Liq temp
source
Liquid temperature
Write access -
166 Endress+Hauser
Micropilot NMR81 Operating menu
Navigation Setup → Advanced setup → Application → Tank config → Temperature → Air temp.
source
Navigation Setup → Advanced setup → Application → Tank config → Temperature → Manual air
temp.
Factory setting 25 °C
Air temperature
Navigation Setup → Advanced setup → Application → Tank config → Temperature → Air temp.
Write access -
Endress+Hauser 167
Operating menu Micropilot NMR81
Description Defines the source from which the vapor temperature is obtained.
Factory setting 25 °C
Vapor temperature
Write access -
168 Endress+Hauser
Micropilot NMR81 Operating menu
"Density" submenu
Navigation Setup → Advanced setup → Application → Tank config → Density → Density source
Selection • HTG
• HTMS
• Average profile density *
• Upper density
• Middle density
• Lower density
Density value
Navigation Setup → Advanced setup → Application → Tank config → Density → Density value
Write access -
Air density
Navigation Setup → Advanced setup → Application → Tank config → Density → Air density
Endress+Hauser 169
Operating menu Micropilot NMR81
Vapor density
Navigation Setup → Advanced setup → Application → Tank config → Density → Vapor density
170 Endress+Hauser
Micropilot NMR81 Operating menu
"Pressure" submenu
P1 (bottom) source
Navigation Setup → Advanced setup → Application → Tank config → Pressure → P1 (bot) source
P1 (bottom)
Write access -
Endress+Hauser 171
Operating menu Micropilot NMR81
P1 position
Description Defines the position of the bottom pressure transmitter (P1), measured from zero position
(tank bottom or datum plate).
P1 offset
Description Defines the offset for the bottom pressure (P1). The offset is added to the measured
pressure prior to any tank calculation.
The offset is added to the measured pressure prior to any tank calculation.
P1 absolute / gauge
Description Defines whether the connected pressure transmitter measures an absolute or a gauge
pressure.
Selection • Absolute
• Gauge
172 Endress+Hauser
Micropilot NMR81 Operating menu
P3 (top) source
Navigation Setup → Advanced setup → Application → Tank config → Pressure → P3 (top) source
P3 (top)
Write access -
Endress+Hauser 173
Operating menu Micropilot NMR81
P3 position
Description Defines the position of the top pressure transmitter (P3), measured from zero position
(tank bottom or datum plate).
P3 offset
Description Defines the offset for the top pressure (P3). The offset is added to the measured pressure
prior to any tank calculation.
The offset is added to the measured pressure prior to any tank calculation.
P3 absolute / gauge
Description Defines whether the connected pressure transmitter measures an absolute or a gauge
pressure.
Selection • Absolute
• Gauge
174 Endress+Hauser
Micropilot NMR81 Operating menu
Ambient pressure
Endress+Hauser 175
Operating menu Micropilot NMR81
"HyTD" submenu
Overview
Hydrostatic Tank Deformation can be used to compensate the vertical movement of the
Gauge Reference Height (GRH) due to bulging of the tank shell caused by the hydrostatic
pressure exerted by the liquid stored in the tank. The compensation is based on a linear
approximation obtained from manual hand dips at several levels divided over the full
range of the tank.
C
D
A
A0028722
176 Endress+Hauser
Micropilot NMR81 Operating menu
H 1
3 L
L0
A0028724
L < L0 = CHyTD = 0
L > L0 = CHyTD = - (L - L0) x D
A0028715
L Measured level
L0 Starting level
D Deformation factor
Endress+Hauser 177
Operating menu Micropilot NMR81
Description of parameters
Navigation Setup → Advanced setup → Application → Tank calculation → HyTD → HyTD corr.
value
Description Shows the correction value from the Hydrostatic Tank Deformation.
Write access -
HyTD mode
Navigation Setup → Advanced setup → Application → Tank calculation → HyTD → HyTD mode
Selection • No
• Yes
Factory setting No
Starting level
Navigation Setup → Advanced setup → Application → Tank calculation → HyTD → Starting level
Description Defines the starting level for the Hydrostatic Tank Deformation.
178 Endress+Hauser
Micropilot NMR81 Operating menu
Deformation factor
Navigation Setup → Advanced setup → Application → Tank calculation → HyTD → Deform factor
Description Defines the deformation factor of the Hydrostatic Tank Deformation (change of gauge
position per change of level).
Endress+Hauser 179
Operating menu Micropilot NMR81
"CTSh" submenu
Overview
CTSh (correction of the thermal tank shell expansion) compensates for effects on the
Gauge Reference Height (GRH) due to temperature effects on the tank shell or stilling well.
The temperature effects are separated into two parts, respectively effecting the 'dry' and
'wetted' part of the tank shell or stilling well. The calculation is based on thermal expansion
coefficients of steel and insulation factors for both the 'dry' and 'wet' shell. The assessed
temperatures are based on manual or measured values and the temperature of the shell
when the tank was calibrated (for details refer to API MPMS Chapter 12.1).
This correction is recommended for the following situations:
• if the operating temperature deviates considederably from the temperature during
calibration (ΔT > 10 °C (18 °F))
• for extremely high tanks
• for refrigerated, cryogenic or heated applications
As the use of this correction will influence the innage level reading, it is
recommended to review the manual hand dip and level verification procedures prior
to enabling this correction method.
This mode should not be used in conjunction with HTG as with HTG the level is not
measured relative to the gauge reference height.
180 Endress+Hauser
Micropilot NMR81 Operating menu
TA
TV
TD
A
TW
TP
A0028712
TP Product temperature
Yes TP TA
Open top
No (7/8) TP + (1/8) TA TA
1) This option is also valid for insulated tanks without a stilling welll. This is due to the temperature inside
and outside of the tank shell being the same due to the insulation of the tank.
Endress+Hauser 181
Operating menu Micropilot NMR81
1 2 3 4 5
A0030382
182 Endress+Hauser
Micropilot NMR81 Operating menu
L Measured level
TD Temperature of the dry part of the tank shell (calculated from TP, TV and TA)
TW Temperature of the wetted part of the tank shell (calculated from TP, TV and TA)
Endress+Hauser 183
Operating menu Micropilot NMR81
Description of parameters
Navigation Setup → Advanced setup → Application → Tank calculation → CTSh → CTSh corr
value
Write access -
CTSh mode
Navigation Setup → Advanced setup → Application → Tank calculation → CTSh → CTSh mode
Selection • No
• Yes
Factory setting No
Covered tank
Navigation Setup → Advanced setup → Application → Tank calculation → CTSh → Covered tank
184 Endress+Hauser
Micropilot NMR81 Operating menu
The Covered option is only valid for fixed tank roofs. For a floating roof select Open
top.
Stilling well
Navigation Setup → Advanced setup → Application → Tank calculation → CTSh → Stilling well
Selection • No
• Yes
Factory setting No
Calibration temperature
Description Enter the temperature at which the measurement has been calibrated.
Factory setting 25 °C
Navigation Setup → Advanced setup → Application → Tank calculation → CTSh → Linear exp
coeff
Description Enter the linear expansion coefficient of the tank shell material.
Endress+Hauser 185
Operating menu Micropilot NMR81
186 Endress+Hauser
Micropilot NMR81 Operating menu
"HTMS" submenu
Overview
The Hybrid Tank Measurement System (HTMS) is a method to calculate the density of a
product in a tank based on both a (top mounted) level and at least one (bottom mounted)
pressure measurement. An additional pressure sensor can be installed at the top of the
tank to provide information about the vapor pressure and to make the density calculation
more accurate. The calculation method also takes into account a possible level of water at
the bottom of the tank to make density calculations as accurate as possible.
HTMS parameters
P3
ρA
ρV
g
A
HP3
ρP ΔP
LP
B
P1 V
LW
HP1
A0028847
37 HTMS parameters
A Product
B Water
HP1 (Position of P1 transmitter) Setup → Advanced setup → Tank configuration → Pressure → P1 position
HP3 (Position of P3 transmitter) Setup → Advanced setup → Tank configuration → Pressure → P3 position
1)
ρP (Density of the product ) • Measured value: Setup → Advanced setup → Calculation → HTMS → Density value
• User-defined value: Setup → Advanced setup → Calculation → HTMS → Manual upper density
ρA (Ambient air temperature) Setup → Advanced setup → Tank configuration → Density → Air density
V = LW - HP1
ΔP = LP - LW = LP - V - HP1
Endress+Hauser 187
Operating menu Micropilot NMR81
HTMS modes
Two HTMS modes can be selected in the HTMS mode parameter (→ 189). The mode
determines whether one or two pressure values are used. Depending on the selected mode
a number of additional parameters are required for the calculation of the product density.
The HTMS P1+P3 option must be used in pressurized tanks in order to compensate
for the pressure of the vapor phase.
HTMS P1 • P1 • g ρP
• LP • HP1
• LW (optional)
HTMS P1+P3 • P1 • ρV ρP
• P3 • ρA (more precise calculation
• LP • g for pressurized tanks)
• HP1
• HP3
• LW (optional)
Minimum level
The density of the product can only be calculated if the product has a minimum thickness :
ΔP ≥ ΔP, min
A0028864
LP - V ³ D P ,min + H P1 = Lmin
A0028863
Lmin is defined in the Minimum level parameter (→ 190). As can be seen from the
formula it always must be bigger than HP1.
If LP - V falls below this limit, the density is calculated as follows:
• If a previous calculated value is available, this value will be kept as long as no new
calculation is possible.
• If no value was previously calculated, the manual value (defined in the Manual upper
density parameter) will be used.
Hysteresis
The level of the product in a tank is not constant but slightly varies, due for example to
filling disturbances. If the level oscillates around the changeover level (Minimum level
(→ 190)), the algorithm will constantly switch between calculating the value and
holding the previous result. To avoid this effect a positional hysteresis is defined around
the changeover point.
188 Endress+Hauser
Micropilot NMR81 Operating menu
Hr
Lmin
2
A0029148
38 HTMS hysteresis
1 Value calculated
2 Value held/manual
Lmin Minimum level (→ 190)
Hr Hysteresis (→ 191)
Description of parameters
HTMS mode
Navigation Setup → Advanced setup → Application → Tank calculation → HTMS → HTMS mode
Description Defines the HTMS mode. Depending on the mode one or two pressure transmitters are
used.
Selection • HTMS P1
• HTMS P1+P3
Manual density
Endress+Hauser 189
Operating menu Micropilot NMR81
Density value
Navigation Setup → Advanced setup → Application → Tank calculation → HTMS → Density value
Write access -
Minimum level
Navigation Setup → Advanced setup → Application → Tank calculation → HTMS → Min. level
If LP - V falls below the limit defined in this parameter, the density retains its last value or
the manual value is used instead.
Minimum pressure
190 Endress+Hauser
Micropilot NMR81 Operating menu
If the pressure P1 falls below the limit defined in this parameter, the density retains its last
value or the manual value is used instead.
Safety distance
Description Defines the minimum level which must be present above the bottom pressure sensor
before its signal is used for the calculation.
Hysteresis
Description Defines the hysteresis for the HTMS calculation. Prevents constant switching if the level is
near the switch-over point.
Factory setting 50 mm
Water density
Endress+Hauser 191
Operating menu Micropilot NMR81
192 Endress+Hauser
Micropilot NMR81 Operating menu
"Dip-table" submenu
Dip table
The dip table is used to correct the level readings using independently taken hand dips.
The dip table is used in particular to adapt the level gauge to the specific application
conditions such as a mechanical offset and the tank or stilling well design. Depending on
national regulations, national inspectors will dip the tank at one to three levels during a
calibration run and check the level readings.
Only one value pair must be entered into the dip table to correct the measurement offset.
If a second value pair is entered into the dip table, the device accepts the corrected
measured values identically for both value pairs. All other measured values are determined
by linear extrapolation.
If more than two value pairs are entered, the system carries out a linear interpolation
between adjacent value pairs. Outside these value pairs, extrapolation is also linear.
• The offset should not be determined and entered within the close range of the
antenna or immediately in the range of the tank bottom, because within these
ranges interferences of the radar signal may occur.
• The entries of the dip table must be arranged in an ascending order of levels. If table
values have not been entered in the correct order, they can automatically be
rearranged by selecting Table settings (→ 196) = Sort table
Endress+Hauser 193
Operating menu Micropilot NMR81
A0029166-EN
2. Use the "↑" and "↓" keys to move to the line you want to edit.
/../Edit table 13964-1
N Dip Table De. Dip Table Di.
3 1.0 1.0
4 0.0 0.0
5 0.0 0.0
A0029167-EN
A0029168-EN
194 Endress+Hauser
Micropilot NMR81 Operating menu
1.
A0029162
A0029161
2. If the device already contains a dip table: Click "Read" to load it into the editor.
3. Enter or change table values in the table on the right. A graphical representation of
the table is shown in the diagram on the right.
4. Click "Write" to write the table back to the device.
Endress+Hauser 195
Operating menu Micropilot NMR81
Description of parameters
Table settings
Selection • Manual
• Semiautomatic
• Clear table
• Sort table
Table mode
Selection • Disable
• Enable
196 Endress+Hauser
Micropilot NMR81 Operating menu
Endress+Hauser 197
Operating menu Micropilot NMR81
Alarm mode
Selection • Off
• On
• Latching
198 Endress+Hauser
Micropilot NMR81 Operating menu
A B
10 10
1 1
2 2
3 3
4 4
5 5
6 6
7 7
8 8
A0029539
Error value
Selection • No alarm
• HH+H alarm
• H alarm
• L alarm
• LL+L alarm
• All alarms
Endress+Hauser 199
Operating menu Micropilot NMR81
200 Endress+Hauser
Micropilot NMR81 Operating menu
Alarm value
Description Shows the current value of the process variable being monitored.
Write access -
HH alarm value
H alarm value
Endress+Hauser 201
Operating menu Micropilot NMR81
L alarm value
LL alarm value
HH alarm
202 Endress+Hauser
Micropilot NMR81 Operating menu
Write access -
H alarm
Write access -
HH+H alarm
Write access -
L alarm
Endress+Hauser 203
Operating menu Micropilot NMR81
Write access -
LL alarm
Write access -
LL+L alarm
Write access -
204 Endress+Hauser
Micropilot NMR81 Operating menu
Any error
Write access -
Clear alarm
Description Deletes an alarm which is still active although the alarm condition is no longer present.
Selection • No
• Yes
Factory setting No
Damping factor
Factory setting 0s
Endress+Hauser 205
Operating menu Micropilot NMR81
206 Endress+Hauser
Micropilot NMR81 Operating menu
Description Defines the delay after the occurrence of an echo loss before the device reacts as defined in
'Output echo lost'.
Safety distance
Description Defines the safety distance (measured from the reference point). A warning is issued if the
level rises into the safety distance.
Factory setting 0 mm
Endress+Hauser 207
Operating menu Micropilot NMR81
208 Endress+Hauser
Micropilot NMR81 Operating menu
"Information" submenu
Signal quality
Write access -
Navigation Setup → Advanced setup → Sensor config → Information → Abs. echo ampl.
Write access -
Description Shows the relative amplitude (i.e. the distance to the evaluation curve) of the evaluated
level signal.
Write access -
Endress+Hauser 209
Operating menu Micropilot NMR81
Distance
Write access -
210 Endress+Hauser
Micropilot NMR81 Operating menu
Evaluation mode
Navigation Setup → Advanced setup → Sensor config → Echo tracking → Evaluation mode
History reset
Navigation Setup → Advanced setup → Sensor config → Echo tracking → History reset
Endress+Hauser 211
Operating menu Micropilot NMR81
"Display" submenu
This menu is only visible if the device has a local display.
Language
Selection • English
• Deutsch *
• Français *
• Español *
• Italiano *
• Nederlands *
• Portuguesa *
• Polski *
• русский язык (Russian) *
• Svenska *
• Türkçe *
• 中文 (Chinese) *
• 日本語 (Japanese) *
• 한국어 (Korean) *
• ( الْعَرَبيّةArabic) *
• Bahasa Indonesia *
• ภาษาไทย (Thai) *
• tiếng Việt (Vietnamese) *
• čeština (Czech) *
Format display
212 Endress+Hauser
Micropilot NMR81 Operating menu
• The Value 1 to 4 display (→ 213) parameters specify which measured values are
shown on the display and in which order.
• If more measured values are specified than the current display mode permits, the values
alternate on the device display. The display time until the next change is configured in
the Display interval (→ 216) parameter.
Value 1 to 4 display
Description Select the measured value that is shown on the local display.
Selection • None 6)
• Tank level
• Measured level
• Tank level %
• Water level 6)
• Liquid temperature 6)
• Vapor temperature 6)
• Air temperature 6)
• Tank ullage
• Tank ullage %
• Observed density value 6)
• P1 (bottom) 6)
• P2 (middle) 6)
• P3 (top) 6)
• GP 1 value 6)
• GP 2 value 6)
• GP 3 value 6)
• GP 4 value 6)
• Gauge command 6)
• Gauge status 6)
• AIO B1-3 value 6)
• AIO B1-3 value mA 6)
• AIO B1-3 value % 6)
• AIO C1-3 value 6)
• AIO C1-3 value mA 6)
• AIO C1-3 value % 6)
Endress+Hauser 213
Operating menu Micropilot NMR81
Decimal places 1 to 4
Description This selection does not affect the measurement and calculation accuracy of the device.
Selection •x
• x.x
• x.xx
• x.xxx
• x.xxxx
Separator
Selection •.
•,
Factory setting .
214 Endress+Hauser
Micropilot NMR81 Operating menu
Number format
Selection • Decimal
• ft-in-1/16''
Header
Header text
Endress+Hauser 215
Operating menu Micropilot NMR81
Display interval
Description Set time measured values are shown on display if display alternates between values.
User entry 1 to 10 s
Factory setting 5s
Display damping
Backlight
Selection • Disable
• Enable
216 Endress+Hauser
Micropilot NMR81 Operating menu
Contrast display
Description Adjust local display contrast setting to ambient conditions (e.g. lighting or reading angle).
User entry 20 to 80 %
Factory setting 30 %
Endress+Hauser 217
Operating menu Micropilot NMR81
Units preset
If the Customer value option is selected, the units are defined in the following
parameters:
• Distance unit (→ 218)
• Pressure unit (→ 219)
• Temperature unit (→ 219)
• Mass unit (→ 219)
• Density unit (→ 220)
In any other case these are read-only parameters used to indicate the respective unit.
Distance unit
Factory setting mm
218 Endress+Hauser
Micropilot NMR81 Operating menu
Pressure unit
Temperature unit
Factory setting °C
Mass unit
Endress+Hauser 219
Operating menu Micropilot NMR81
Factory setting g
Density unit
220 Endress+Hauser
Micropilot NMR81 Operating menu
Date/time
Description Shows the current date and time from the real-time clock.
User interface Days (d), hours (h), minutes (m), seconds (s)
Factory setting
Write access -
Set date
Endress+Hauser 221
Operating menu Micropilot NMR81
Year
Month
User entry 1 to 12
Factory setting 1
Day
User entry 1 to 31
Factory setting 1
222 Endress+Hauser
Micropilot NMR81 Operating menu
Hour
User entry 0 to 23
Factory setting 0
Minute
User entry 0 to 59
Factory setting 0
Endress+Hauser 223
Operating menu Micropilot NMR81
224 Endress+Hauser
Micropilot NMR81 Operating menu
"Administration" submenu
Factory setting 0
Device reset
Description Reset the device configuration - either entirely or in part - to a defined state.
Selection • Cancel
• To fieldbus defaults **
• To factory defaults
• To delivery settings
• Of customer settings
• To transducer defaults
• Restart device
Endress+Hauser 225
Operating menu Micropilot NMR81
Actual diagnostics
Timestamp
User interface Days (d), hours (h), minutes (m), seconds (s)
Previous diagnostics
Description Displays the last diagnostic message which has been active before the current message.
226 Endress+Hauser
Micropilot NMR81 Operating menu
Timestamp
User interface Days (d), hours (h), minutes (m), seconds (s)
Description Displays the time the device has been in operation since the last device restart.
User interface Days (d), hours (h), minutes (m), seconds (s)
Operating time
User interface Days (d), hours (h), minutes (m), seconds (s)
Date/time
Description Shows the current date and time from the real-time clock.
User interface Days (d), hours (h), minutes (m), seconds (s)
Factory setting
Write access -
Endress+Hauser 227
Operating menu Micropilot NMR81
Diagnostics 1 to 5
Description Display the current diagnostics messages with the highest to fifth-highest priority.
Timestamp 1 to 5
User interface Days (d), hours (h), minutes (m), seconds (s)
228 Endress+Hauser
Micropilot NMR81 Operating menu
Device tag
Write access -
Serial number
Write access -
Firmware version
Factory setting
Write access -
Endress+Hauser 229
Operating menu Micropilot NMR81
Device name
Write access -
Order code
Additional information The extended order code indicates the selected option of all ordering features and thus
uniquely identifies the device.
230 Endress+Hauser
Micropilot NMR81 Operating menu
Selection • Off
• On
Description Select the diagnostic event to be simulated. Note: To terminate the simulation, select 'Off'.
Selection Off
Simulation distance on
Selection • Off
• On
Endress+Hauser 231
Operating menu Micropilot NMR81
Simulation distance
Factory setting 0 mm
Selection • Off
• On
Simulation value
232 Endress+Hauser
Micropilot NMR81 Operating menu
Endress+Hauser 233
Operating menu Micropilot NMR81
Selection • No
• Yes
Factory setting No
Write access -
Level signal
Description Shows the result of the device check for the level signal.
Write access -
234 Endress+Hauser
Micropilot NMR81 Operating menu
Near distance
Description Shows the result of the device check for the near distance area.
Write access -
Endress+Hauser 235
Index
Index
0…9 Communications status graph (Parameter) . . . . . . . . 142
0 % value (Parameter) . . . . . . . . . . . . . . . 125, 132, 154 Configuration (Submenu) . . . . . . . . . . . . . 144, 147, 153
100 % value (Parameter) . . . . . . . . . . . . . 125, 132, 155 Confirm distance (Parameter) . . . . . . . . . . . . . . . . . 111
Contact type (Parameter) . . . . . . . . . . . . . . . . . . . . . 139
A Contrast display (Parameter) . . . . . . . . . . . . . . . . . . 217
Absolute echo amplitude (Parameter) . . . . . . . . . . . 209 Covered tank (Parameter) . . . . . . . . . . . . . . . . . . . . 184
Access status tooling (Parameter) . . . . . . . . . . . . . . 115 CTSh (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
Accessories CTSh correction value (Parameter) . . . . . . . . . . . . . . 184
Communication specific . . . . . . . . . . . . . . . . . . . . 89 CTSh mode (Parameter) . . . . . . . . . . . . . . . . . . . . . 184
Service specific . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 Current output 1 simulation (Parameter) . . . . . . . . . 232
Actual diagnostics (Parameter) . . . . . . . . . . . . . . . . 226 Current output 2 simulation (Parameter) . . . . . . . . . 232
Administration (Submenu) . . . . . . . . . . . . . . . . . . . 225 Current span (Parameter) . . . . . . . . . . . . . . . . . . . . 129
Advanced setup (Submenu) . . . . . . . . . . . . . . . . . . . 115
Air density (Parameter) . . . . . . . . . . . . . . . . . . 105, 169 D
Air temperature (Parameter) . . . . . . . . . . . . . . 102, 167 Damping factor (Parameter) . . . . . . . . . . 127, 135, 205
Air temperature source (Parameter) . . . . . . . . . . . . . 167 Date / time (Submenu) . . . . . . . . . . . . . . . . . . . . . . 221
Alarm 1 (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . . 198 Date/time (Parameter) . . . . . . . . . . . . . . . . . . 221, 227
Alarm 1 input source (Parameter) . . . . . . . . . . . . . . 150 Day (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
Alarm 2 input source (Parameter) . . . . . . . . . . . . . . 151 DD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Alarm mode (Parameter) . . . . . . . . . . . . . . . . . . . . . 198 Deactivate SIL/WHG (Wizard) . . . . . . . . . . . . . . . . . 224
Alarm value (Parameter) . . . . . . . . . . . . . . . . . . . . . 201 Decimal places 1 (Parameter) . . . . . . . . . . . . . . . . . 214
Alarm value source (Parameter) . . . . . . . . . . . . . . . . 200 Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . 11
Ambient pressure (Parameter) . . . . . . . . . . . . . . . . . 175 Define access code (Parameter) . . . . . . . . . . . . . . . . 225
Analog I/O (Submenu) . . . . . . . . . . . . . . . . . . . . . . 128 Deformation factor (Parameter) . . . . . . . . . . . . . . . . 179
Analog input 0% value (Parameter) . . . . . . . . . . . . . 133 Delay time echo lost (Parameter) . . . . . . . . . . . . . . . 207
Analog input 100% value (Parameter) . . . . . . . . . . . 134 Density (Submenu) . . . . . . . . . . . . . . . . . . . . . 104, 169
Analog input source (Parameter) . . . . . . . . . . . . . . . 130 Density unit (Parameter) . . . . . . . . . . . . . . . . . . . . . 220
Analog IP (Submenu) . . . . . . . . . . . . . . . . . . . . . . . 123 Density value (Parameter) . . . . . . . . . . . . . . . . 169, 190
Any error (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 205 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Device alarm simulation (Parameter) . . . . . . . . . . . . 231
Residual risk . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Device check (Submenu) . . . . . . . . . . . . . . . . . . . . . 234
Application (Submenu) . . . . . . . . . . . . . . . . . . . . . . 163 Device Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Assign PV (Parameter) . . . . . . . . . . . . . . . . . . . . . . 153 Device ID (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 145
Assign QV (Parameter) . . . . . . . . . . . . . . . . . . . . . . 158 Device information (Submenu) . . . . . . . . . . . . . . . . 229
Assign SV (Parameter) . . . . . . . . . . . . . . . . . . . . . . 155 Device name (Parameter) . . . . . . . . . . . . . . . . 117, 230
Assign TV (Parameter) . . . . . . . . . . . . . . . . . . . . . . 157 Device replacement . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Device reset (Parameter) . . . . . . . . . . . . . . . . . . . . . 225
B Device tag (Parameter) . . . . . . . . . . . 109, 117, 161, 229
Backlight (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 216 Diagnostic event simulation (Parameter) . . . . . . . . . 231
Baudrate (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 144 Diagnostic events . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Blocking distance (Parameter) . . . . . . . . . . . . . . . . . 165 Diagnostic information
Bottom level (Parameter) . . . . . . . . . . . . . . . . . . . . 101 Design, description . . . . . . . . . . . . . . . . . . . . . . . . 78
Bus termination (Parameter) . . . . . . . . . . . . . . . . . . 145 FieldCare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Diagnostic list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
C Diagnostic list (Submenu) . . . . . . . . . . . . . . . . . . . . 228
Calibration temperature (Parameter) . . . . . . . . . . . . 185 Diagnostic message . . . . . . . . . . . . . . . . . . . . . . . . . . 74
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Diagnostics
Cleaning Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . 84 Diagnostics (Menu) . . . . . . . . . . . . . . . . . . . . . . . . . 226
Clear alarm (Parameter) . . . . . . . . . . . . . . . . . . . . . 205 Diagnostics 1 to 5 (Parameter) . . . . . . . . . . . . . . . . . 228
Communication (Submenu) . . . . . . . . . . . . . . . . . . . 142 Diagnostics event . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Communication interface protocol (Parameter) . . . . 142 Digital input source (Parameter) . . . . . . . . . . . . . . . 137
Communication interface protocol variant Digital Xx-x (Submenu) . . . . . . . . . . . . . . . . . . . . . . 137
(Parameter) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147 DIP switch
Communication status (Parameter) . . . . . . . . . . . . . 118 see Write protection switch
Communications info (Parameter) . . . . . . . . . . . . . . 143 Dip-table (Submenu) . . . . . . . . . . . . . . . . . . . . . . . . 196
236 Endress+Hauser
Index
Endress+Hauser 237
Index
N R
Near distance (Parameter) . . . . . . . . . . . . . . . . . . . . 235 Readback value (Parameter) . . . . . . . . . . . . . . . . . . 140
NMT element values (Submenu) . . . . . . . . . . . . . . . 103 Recalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
No. of preambles (Parameter) . . . . . . . . . . . . . . . . . 153 Record map (Parameter) . . . . . . . . . . . . . . . . . . . . . 113
Number format (Parameter) . . . . . . . . . . . . . . . . . . 215 Relative echo amplitude (Parameter) . . . . . . . . . . . . 209
Number of devices (Parameter) . . . . . . . . . . . . . . . . 116 Remedial measures
Calling up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
O Closing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Observed density (Parameter) . . . . . . . . . . . . . . . . . 104 Repair concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Observed density source (Parameter) . . . . . . . . . . . . 169 Replacing a device . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Operating elements Requirements for personnel . . . . . . . . . . . . . . . . . . . . 10
Diagnostics message . . . . . . . . . . . . . . . . . . . . . . . 75 Result device check (Parameter) . . . . . . . . . . . . . . . . 234
Operating mode (Parameter) . . . . . . 118, 123, 128, 137 Returning devices . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Operating time (Parameter) . . . . . . . . . . . . . . . . . . . 227 RTD connection type (Parameter) . . . . . . . . . . . . . . 124
Operating time from restart (Parameter) . . . . . . . . . 227 RTD type (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 123
Operation (Menu) . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 S
Order code (Parameter) . . . . . . . . . . . . . . . . . . . . . . 230 Safety distance (Parameter) . . . . . . . . . . . . . . . 191, 207
Output density (Parameter) . . . . . . . . . . . . . . . . . . . 120 Safety instructions
Output echo lost (Parameter) . . . . . . . . . . . . . . . . . . 207 Basic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Output level (Parameter) . . . . . . . . . . . . . . . . . . . . . 121 Safety settings (Submenu) . . . . . . . . . . . . . . . . . . . . 207
Output pressure (Parameter) . . . . . . . . . . . . . . . . . . 119 Secondary variable (SV) (Parameter) . . . . . . . . . . . . 156
Output simulation (Parameter) . . . . . . . . . . . . . . . . 139 Sensor config (Submenu) . . . . . . . . . . . . . . . . . . . . . 209
Output temperature (Parameter) . . . . . . . . . . . . . . . 120 Separator (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 214
Output value (Parameter) . . . . . . . . . . . . . . . . 133, 140 Serial number (Parameter) . . . . . . . . . . . . . . . . . . . 229
Output values (Parameter) . . . . . . . . . . . . . . . . . . . . 140 Service relay (Parameter) . . . . . . . . . . . . . . . . . . . . . 149
Output vapor temperature (Parameter) . . . . . . . . . . 121 Set date (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . 221
Set level (Parameter) . . . . . . . . . . . . . . . . . . . . 111, 164
P Setup (Menu) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
P1 (bottom) (Parameter) . . . . . . . . . . . . . . . . . 106, 171 Signal quality (Parameter) . . . . . . . . . . . . . . . . . . . . 209
P1 (bottom) manual pressure (Parameter) . . . . . . . . 171 SIL confirmation (Wizard) . . . . . . . . . . . . . . . . . . . . 224
P1 (bottom) source (Parameter) . . . . . . . . . . . . . . . 171 Simulation (Submenu) . . . . . . . . . . . . . . . . . . . . . . . 231
P1 absolute / gauge (Parameter) . . . . . . . . . . . . . . . 172 Simulation distance (Parameter) . . . . . . . . . . . . . . . 232
P1 offset (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 172 Simulation distance on (Parameter) . . . . . . . . . . . . . 231
P1 position (Parameter) . . . . . . . . . . . . . . . . . . . . . 172 Simulation value (Parameter) . . . . . . . . . . . . . . . . . 232
P3 (top) (Parameter) . . . . . . . . . . . . . . . . . . . . 106, 173 Start device check (Parameter) . . . . . . . . . . . . . . . . . 234
P3 (top) manual pressure (Parameter) . . . . . . . . . . . 173 Starting level (Parameter) . . . . . . . . . . . . . . . . . . . . 178
P3 (top) source (Parameter) . . . . . . . . . . . . . . . . . . 173 Status signals . . . . . . . . . . . . . . . . . . . . . . . . . . . 74, 77
P3 absolute / gauge (Parameter) . . . . . . . . . . . . . . . 174 Stilling well (Parameter) . . . . . . . . . . . . . . . . . . . . . 185
P3 offset (Parameter) . . . . . . . . . . . . . . . . . . . . . . . 174 Submenu
P3 position (Parameter) . . . . . . . . . . . . . . . . . . . . . 174 Administration . . . . . . . . . . . . . . . . . . . . . . . . . . 225
Parity (Parameter) . . . . . . . . . . . . . . . . . . . . . . . . . 144 Advanced setup . . . . . . . . . . . . . . . . . . . . . . . . . 115
Percent of range (Parameter) . . . . . . . . . . . . . . . . . . 155 Alarm 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198
Polling address (Parameter) . . . . . . . . . . . . . . . . . . . 117 Analog I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128
Present mapping (Parameter) . . . . . . . . . . . . . . . . . 112 Analog IP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Pressure (Submenu) . . . . . . . . . . . . . . . . . . . . 106, 171 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Pressure unit (Parameter) . . . . . . . . . . . . . . . . . . . . 219 Communication . . . . . . . . . . . . . . . . . . . . . . . . . 142
Previous diagnostics (Parameter) . . . . . . . . . . . . . . . 226 Configuration . . . . . . . . . . . . . . . . . . . 144, 147, 153
Primary variable (PV) (Parameter) . . . . . . . . . . . . . . 154 CTSh . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184
238 Endress+Hauser
Index
T
Table mode (Parameter) . . . . . . . . . . . . . . . . . . . . . 196
Table settings (Parameter) . . . . . . . . . . . . . . . . . . . 196
Tank calculation (Submenu) . . . . . . . . . . . . . . . . . . 176
Tank configuration (Submenu) . . . . . . . . . . . . . . . . 163
Tank level (Parameter) . . . . . . . . . . . . . . 100, 110, 164
Tank Level % (Parameter) . . . . . . . . . . . . . . . . . . . . 100
Tank reference height (Parameter) . . . . . . . . . 110, 163
Tank ullage (Parameter) . . . . . . . . . . . . . . . . . . . . . 100
Tank ullage % (Parameter) . . . . . . . . . . . . . . . . . . . 101
Temperature (Submenu) . . . . . . . . . . . . . . . . . 102, 166
Temperature unit (Parameter) . . . . . . . . . . . . . . . . . 219
Tertiary variable (TV) (Parameter) . . . . . . . . . . . . . . 158
Timestamp (Parameter) . . . . . . . . . . . . . . 226, 227, 228
Trouble shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
U
Units preset (Parameter) . . . . . . . . . . . . . . . . . 109, 218
Upper interface level (Parameter) . . . . . . . . . . . . . . 101
Used for SIL (Parameter) . . . . . . . . . . . . . . . . . 135, 140
V
V1 address (Parameter) . . . . . . . . . . . . . . . . . . . . . . 147
V1 input selector (Submenu) . . . . . . . . . . . . . . . . . . 150
Value 1 display (Parameter) . . . . . . . . . . . . . . . . . . 213
Endress+Hauser 239
*71318192*
71318192
www.addresses.endress.com