EXPERIMENTAL SET-UP
Collimation system adjusted to 30 x 30 cm2 field at 1 m from the focus of the x-ray tube;
Exposure parameters: 100 kVp and 1 mA; values normalized to 1 min of scopy;
Measures performed at two different heights (“x-ray tube height” and “phantom height”), in the horizontal,
Air kerma measures performed with an environmental ionization chamber (Victoreen 451P, serial No. 1420)
calibrated by Global Calibration Laboratory (Cleveland, OH, USA) at the energy of the Cs-137 (E = 662 keV)
(calibration report No. 1470-12-15-2006) (Tolerance Statement of measure during calibration: inaccuracy
±10%);
[File: 035_07_Eng_Rev0]
TECHNIX TECHNIX S.p.A. – TCA6
TOP VIEW
Air kerma measures, at the “phantom height”, normalized to 1 min of scopy (μGy/min) for single exposure;
Measures performed at 1 and 2 m perpendicularly to the beam axis.
0,9 1,4
4,2 3,4
5,1 3,8
6,4
2,0 2,0
1,5
[File: 035_07_Eng_Rev0]
TECHNIX TECHNIX S.p.A. – TCA6
TOP VIEW
Air kerma measures, at the “X-ray tube height”, normalized to 1 min of scopy (μGy/min) for single exposure;
Measures performed at 1 and 2 m perpendicularly to the beam axis.
1,6 1,8
6,6 5,1
6,5 5,3
6,5
1,5 1,5
1,8
[File: 035_07_Eng_Rev0]
TECHNIX TECHNIX S.p.A. – TCA6
LATERAL VIEW
Air kerma measures, normalized to 1 min of scopy (μGy/min) for single exposure;
Measures performed at 1 and 2 m perpendicularly to the beam axis.
6,4 1,5
6,5 1,8
[File: 035_07_Eng_Rev0]
TECHNIX TECHNIX S.p.A. – TCA6
FRONT VIEW
Air kerma measures, normalized to 1 min of scopy (μGy/min) for single exposure;
Measures performed at 1 and 2 m perpendicularly to the beam axis.