Beli ed BELIMED MST-V 606-VS2


DOUBLE DOOR STEAM STERILIZER
Dwg. No. :
CUT-606-VS2
Rev. :
F
Infection Control Typical installation of a Belimed double door Drawn By :
A. JONES
Date :
05/11/18
1535 Hobby Street, Charleston, SC 29405 MST-V 606-VS2 Sterilizer Scale : Page :
Tel: (800) 457-4117 Web: www.belimed.com N.T.S. 1 of 6
LEGEND
SERVICE SHUT-OFF VALVE, BY OTHERS

STEAM PARTICLE STRAINER, BY OTHERS

STEAM TRAP, BY OTHERS

35 1/2" AREA TO BE LEVEL


FOR DOCKING OF CART
STERILE SIDE
FINISHED WALL OPENING 40 1/2"
LINE TYPES
20 1/4"
B
4
16 3/4"
WALL
FLOOR LEVEL OUTLINE
MAXIMUM TOLERANCE FOR FINISHED FLOOR LEVEL 1/4"

14 3/4"
(BY OTHERS)

SRV2
SRV1
1"
4"
HATCH TYPES

DR
STAINLESS STEEL TRIM, BY BELIMED

11"
EL

DP

20"
DR
STERILIZER 44 1/8"

DP

26"
6" STAINLESS STEEL ENCLOSURES, BY BELIMED
115" OVERALL

31"
SRV2
SRV1
SRV2

SRV1 GENERAL NOTES


1. SEE SHEETS 5 & 6 FOR TECHNICAL SPECIFICATIONS.
CW

CR
CA

CS
SS

2. SEE TYPICAL INSTALLATION DETAILS ON SHEET 4.


12 3/4" 4" 4" 4" 4"
35 1/2" AREA TO BE LEVEL
FOR DOCKING OF CART

STERILIZER 39"

CLEAN SIDE

40 1/2" AREA TO BE LEVEL


FOR DOCKING OF CART

PLAN VIEW
SEE DETAIL D/4 FOR TYPICAL
OPTIONAL CHILLED WATER
BYPASS.

WALL
(BY OTHERS)

EL, DP

CA, CW, SS, CS, CR


SS
EL

CONDENSATE
REMOVAL,
BY OTHERS
DP
CW

CS

CR
CA

SS

F
4

CLEAN SIDE
STERILE SIDE
CEILING HEIGHT

CEILING HEIGHT
CEILING HEIGHT

UTILITY CONNECTION POINT 84"


UTILITY CONNECTION POINT 84"

CHAMBER
WIDTH 26"
STERILIZER 77 1/2"
STERILIZER 77 1/2"
HEIGHT 26"
CHAMBER

1"
DR
DR
A
STERILIZER 39" 4 STERILIZER 44 1/8"

SEE DETAIL C/4 FOR TYPICAL


UTILITY LAYOUT.

CLEAN SIDE ELEVATION SIDE ELEVATION

Beli ed BELIMED MST-V 606-VS2


DOUBLE DOOR STEAM STERILIZER
TECHNICAL DRAWING
Dwg. No. :
CUT-606-VS2
Rev. :
F
Infection Control Typical installation of an inset Belimed Drawn By :
A. JONES
Date :
05/11/18
1535 Hobby Street, Charleston, SC 29405 MST-V 606-VS2 Sterilizer. Scale : Page :
Tel: (800) 457-4117 Web: www.belimed.com 1/4" = 1' 2 of 6
LEGEND
SERVICE SHUT-OFF VALVE, BY OTHERS

STEAM PARTICLE STRAINER, BY OTHERS

STEAM TRAP, BY OTHERS

35 1/2" AREA TO BE LEVEL


FOR DOCKING OF CART
LINE TYPES
STERILE SIDE
FINISHED WALL OPENING 40 1/2"
20 1/4"
FLOOR LEVEL OUTLINE
16 3/4"
14 3/4" WALL HATCH TYPES
MAXIMUM TOLERANCE FOR FINISHED FLOOR LEVEL 1/4"

(BY OTHERS)

SRV1
SRV2

1"
DR
STAINLESS STEEL TRIM, BY BELIMED

11"
20"
DP
STERILIZER 44 1/8"

DR

GENERAL NOTES

26"
115" OVERALL

31"
SRV2
SRV1
SRV2

SRV1
SRV 1. SEE SHEETS 5 & 6 FOR TECHNICAL SPECIFICATIONS.
2. SEE TYPICAL INSTALLATION DETAILS ON SHEET 4.

DP
EL

WALL
CW

CR
CA

CS
SS

(BY OTHERS)
1"

14" 4" 4" 4" 4" 4"


35 1/2" AREA TO BE LEVEL

B
FOR DOCKING OF CART

3/4" STERILIZER 39" 3/4" 4


FINISHED WALL OPENING 40 1/2"

CLEAN SIDE

40 1/2" AREA TO BE LEVEL


FOR DOCKING OF CART

PLAN VIEW
SEE DETAIL D/4 FOR TYPICAL
OPTIONAL CHILLED WATER
BYPASS.

WALL WALL
SS (BY OTHERS) (BY OTHERS)

EL, DP
CA, CW, SS, CS, CR
EL

CONDENSATE
REMOVAL,
BY OTHERS

DP
CR
CW

CS
CA

1"
SS

1"

E
4
STERILE SIDE
CEILING HEIGHT

CEILING HEIGHT

CEILING HEIGHT
CONNECTION POINT FOR UTILITIES 84"

CLEAN SIDE
UTILITY CONNECTION POINT 84"

FINISHED WALL OPENING HEIGHT 79"

FINISHED WALL OPENING HEIGHT 79"

CHAMBER
WIDTH 26"
STERILIZER 77 1/2"

STERILIZER 77 1/2"

STERILIZER 77 1/2"
HEIGHT 26"
CHAMBER

DR DR

A
STERILIZER 39" 4
STERILIZER 44 1/8"

SEE DETAIL C/4 FOR TYPICAL


UTILITY LAYOUT.

CLEAN SIDE ELEVATION SIDE ELEVATION

Beli ed BELIMED MST-V 606-VS2


DOUBLE DOOR STEAM STERILIZER
TECHNICAL DRAWING
Dwg. No. :
CUT-606-VS2
Rev. :
F
Infection Control Typical installation of a recessed Belimed Drawn By :
A. JONES
Date :
05/11/18
1535 Hobby Street, Charleston, SC 29405 MST-V 606-VS2 Sterilizer. Scale : Page :
Tel: (800) 457-4117 Web: www.belimed.com 1/4" = 1' 3 of 6
5/8" DIA. HILTI KB-TZ EXPANSION
ANCHORS (MIN. EMBED. = 4").
(x4 LOCATIONS).
WALL

3/4"

STERILIZER

#8 1-1/2" SHEET METAL


SCREW INTO STUD.

STAINLESS STEEL
TRIM, BY BELIMED.

DETAIL FOR FOR REGIONS


REQUIRING SEISMIC ANCHORAGE.

TYP. SEISMIC ANCHORING DETAIL TYP. TRIM ATTACHMENT DETAIL


VERTICAL SERIES STERILIZER VERTICAL STERILIZER
A N.T.S. B N.T.S.

UTILITY ROUGH-IN, BY OTHERS OPTIONAL CHILLED


WATER CONNECTIONS
CW

CR
CA

CS
SS
EL

CW

CR
CS
SS
PIPE
INSULATION,
TYPICAL
DP

PIPE
UTILITY CONNECTIONS INSULATION,

84" A.F.F . TYPICAL

UTILITY CONNECTIONS

84" A.F.F .

TO DRAIN
MANUAL SERVICE VALVES, BY OTHERS.

1. REFER TO EQUIPMENT PLUMBING,


MECHANICAL AND ELECTRICAL
EL - ELECTRICAL PLANS FOR UTILITY LOCATIONS.
CA - COMPRESSED AIR MANUAL SERVICE VALVES, BY OTHERS.
CW - COLD WATER 2. SEE TECHNICAL SPECIFICATIONS
SS - STEAM FOR UTILITY SIZING, INPUT
CS - CHILLED WATER SUPPLY (OPTIONAL) REQUIREMENTS, AND FINAL CW - COLD WATER
CR - CHILLED WATER RETURN (OPTIONAL) CONNECTIONS. CS - CHILLED WATER SUPPLY
CR - CHILLED WATER RETURN

TYPICAL UTILITY LAYOUT TYP. (OPTIONAL) BY PASS DETAIL


VERTICAL STERILIZER - (STEAM) CHILLED WATER CONNECTION
C N.T.S. D N.T.S.

WALL, BY OTHERS.
CHANNEL SECTION
ATTACHMENT ANCHOR,
BY OTHERS

CEILING, BY OTHERS. UNISTRUT CHANNEL


SECTIONS, BY OTHERS
UNISTRUT CONDUIT PIPE
CLAMP, (TYPICAL) BY OTHERS.
CHANNEL SECTION THREADED ROD ,
ATTACHMENT ANCHOR, BY OTHERS.
UTILITIES, BY BY OTHERS
OTHERS. UNISTRUT CONDUIT PIPE
CLAMP, (TYPICAL) BY OTHERS.
UNISTRUT CHANNEL
SECTIONS, BY OTHERS

UTILITIES, BY
OTHERS.
EQUIPMENT
WALL OPENING
1. UTILITY ATTACHMENT ASSEMBLY
1. UTILITY ATTACHMENT ASSEMBLY SHALL FALL WITH IN THE FOOTPRINT
SHALL FALL WITH IN THE FOOTPRINT OF THE SPECIFIED EQUIPMENT.
OF THE SPECIFIED EQUIPMENT.
2. REFER TO EQUIPMENT PLUMBING,
2. REFER TO EQUIPMENT PLUMBING, MECHANICAL AND ELECTRICAL
MECHANICAL AND ELECTRICAL NOTE: REFER TO LOCAL BUILDING PLANS FOR UTILITY LOCATIONS.
NOTE: REFER TO LOCAL BUILDING PLANS FOR UTILITY LOCATIONS. CODES AND/OR SEISMIC
CODES AND/OR SEISMIC ANCHORAGE REQUIREMENTS FOR 3. SEE TECHNICAL SPECIFICATIONS
ANCHORAGE REQUIREMENTS FOR 3. SEE TECHNICAL SPECIFICATIONS FINAL ATTACHMENT METHOD. FOR UTILITY SIZING, INPUT
FINAL ATTACHMENT METHOD. FOR UTILITY SIZING, INPUT REQUIREMENTS, AND FINAL
REQUIREMENTS, AND FINAL CONNECTIONS.
CONNECTIONS.

RECOMMENDED TYPICAL WALL HUNG RECOMMENDED TYPICAL CEILING HUNG


UTILITY ATTACHMENT DETAIL (SIDE VIEW) UTILITY ATTACHMENT DETAIL (SIDE VIEW)
E N.T.S. F N.T.S.

Beli ed BELIMED MST-V 606-VS2


DOUBLE DOOR STEAM STERILIZER
TYPICAL DETAILS
Dwg. No. :
CUT-606-VS2
Rev. :
F
Infection Control Typical installation details for a Belimed Drawn By :
A. JONES
Date :
05/11/18
1535 Hobby Street, Charleston, SC 29405 MST-V 606-VS2 Sterilizer. Scale : Page :
Tel: (800) 457-4117 Web: www.belimed.com N.T.S. 4 of 6
MST-V 606VS2 STERILIZER - TECHNICAL SPECIFICATIONS
REQUIRED INPUTS

Symbol Description Utility Notes Connection Final Utility Conn. By:


Value
Method Others Belimed
Utility: Steam
Steam must be dry, saturated, and free from
Connection Size/Type: 3/4 inch / MNPT
all impurities such as gases, oil, dirt, or other
Pressure: Min: 40 Max: 65 psig
SS contaminants. Provide adequate steam F
exl X
Peak Consumption: 1.7 lbs/min strainer typical for sterilizers. All steam drops
*Consumption per Load: 24 lbs must be piped from top of steam supply line.
Consumption while Idle: 0.05 lbs/min
Utility: Cold Water Cdo wuslsiam
eoptrdihetcondensaeptumpandcool
Connection Size/Type: 1/2 inch / MNPT htecondensO aetm
.p eitm alpe6saurn0eitrcF
°i(om ngi
CW Pressure: Min: 30 Max: 70 psig wetam
etrpmu bse7t0eF °slo.)r TDS
ob<
t4e00ppm. F
exl X
Peak Consumption: 3 gal/min Wsaehtrodu blesotnfeo0d6
t-0ppm Boaw l.cfk
*Consumption per Load: 42 gal perve
onim tu bsetuepdibpcyluostm aopscel,oea rdrle.
Utility: Compressed Air
Connection Size/Type: 1/4 inch / MNPT
Air must be dry and oil free. Used to operate
CA Pressure: Min: 70 Max: 100 psig F
exl X
pneumatic control system.
Peak Consumption: 1 CFM
*Consumption per Load: 10 CF
Utility: Electric
Connection Type: Receptacle
Voltage (Nominal): 208 VAC
Voltage (Phase): Junction Box with Receptacle. NEMA
3 Phase
EL Voltage (Frequency): 60 Hz
15-20R to NEC local codes. Grainger part Recep
ae
tcl X
No. 5C793.
Circuit Protection: 10 Amps
Peak Consumption: 2.3 kW
*Consumption per Load: 0.9 kWh
Utility: Data Port
Connection Method: TCP/IP Ethernet to hospital network X
Connection Quantity: One (1) RJ45 Ethernet port per sterilizer X
Facility Wiring: *Cat-5, Cat-6 or Cat-6e wire Location: Ethernet port should be located X
NIC/Com Speed: 10/100 Mbps Networking Port (RJ45) within 5 ft of machine within the enclosed X
DP Broadcast Method: Unicast service area of the sterilizer. X
MTU Size: 1500 X
Belimed provided patch cables for * (By others)
Final Connection:
final connection
X
Port IP Setup: Each RJ45 Ethernet Port with
individual static IP address
X
Utility: Equipment Area Exhaust Provide sufficient air exhaust to remove neIrdcit
Connection Type: Exhaust return grille 1.5 kW heat rejection. Exhaust X
Utility (Chamber): Safety Pressure Relief Valve Provide connection to safety relief valve and pipe
SRV1 Had
P
rpe
i X
Outlet size: 1 inch / FNPT to drain, atmosphere, or as required by local code.
Utility (Jacket): Safety Pressure Relief Valve Provide connection to safety relief valve and pipe
SRV2 Had
P
rpe
i X
Outlet size: 1 inch / FNPT to drain, atmosphere, or as required by local code.
Optional Utilities for chilled water cooling (used for reduction of cold water consumption).
Utility: Chilled Water Supply
Connection Size/Type: 1/2 inch / MNPT
nIsa
u
etld
Hadr
CS Pressure: Min: 30 Max: 90 psig X
Ppei
Peak Consumption: 96 kBTU/hr
*Max Consumption per Load: 21 kBTU
Utility: Chilled Water Return Pressure difference P CS/CR must be nIsa
u
etld
Hadr
CR Connection Size/Type: 1/2 inch / MNPT greater than 7 PSIG. Ppei X
Utility: Cold Water Cold water is used to prime the vacuum
Connection Size/Type: 1/2 inch / MNPT pump. Optimal temperature is 60°F. Waetr
CW Pressure: Min: 30 Max: 70 psig shod u
ble
so
tnfeo
0d
6
t-0ppm Bo.aw
lcfk F
exl X
Peak Consumption: 2.2 gal/min perveonim
tu bsetuepd
ib
pcyluo
stm ase,r
*Max Consumption per Load: 2.6 gal op
clo
e
adrle.
ESTIMATED EMISSIONS

CONTINUED ON PAGE 6

Beli ed BELIMED MST-V 606-VS2


DOUBLE DOOR STEAM STERILIZER
TECHNICAL SPECIFICATIONS
Dwg. No. :
CUT-606-VS2
Rev. :
F
Infection Control Technical Specifications for the installation Drawn By :
A. JONES
Date :
05/11/18
1535 Hobby Street, Charleston, SC 29405 of a Belimed MST-V 606-VS2 Sterilizer. Scale : Page :
Tel: (800) 457-4117 Web: www.belimed.com NONE 5 of 6
*Max Consumption per Load: 2.6 gal
ESTIMATED EMISSIONS
Utility: Open Drain
Connection Size: 1-1/4 inch Provide flush mounted floor sink and trap 1C
4
-/"opper
DR Discharge Flow Rate: 4.0 gal/min per local code. pe
i
X
*Consumption per Load: 45 gal
Medium: Noise Radiation In accordance with ISO3746: 1995 Testing
Radiation (Net): 65 dBA Standards.
Medium: Heat Radiation Heat dissipated into service room with
Radiation (Net): 1.5 kWh ambient temperature of 86°F.
PHYSICAL SPECIFICATIONS
Parameter: Carry-In Dimensions
Height: 81 in Machine Alone
Width: 39 in (i.e. removed from packaging)
Length: 44.1 in
Parameter: Chamber Usable Dimensions
Height: 26 in
Width: 26 in
Length: 27.5 in
Parameter: Weights
Transport Weight: 1800 lbs
Operating Weight: 2160 lbs
OSHPD NUMBER: OPM-0131-13
*Consumptions calculated under pre-vac cycle at 270°F and water temperature of 63°F. Average load/cycle time one hour.

Additional Information:
- MNPT = Male pipe thread -FNPT = Female pipe thread
- Others must provide service valves and strainers for steam and water supply.
- All utility supplies and connections are customer's responsibility. TS-606-VS2_RevH

Beli ed BELIMED MST-V 606-VS2


DOUBLE DOOR STEAM STERILIZER
TECHNICAL SPECIFICATIONS
Dwg. No. :
CUT-606-VS2
Rev. :
F
Infection Control Technical Specifications for the installation Drawn By :
A. JONES
Date :
05/11/18
1535 Hobby Street, Charleston, SC 29405 of a Belimed MST-V 606-VS2 Sterilizer. Scale : Page :
Tel: (800) 457-4117 Web: www.belimed.com NONE 6 of 6