Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Dymind software release note—DM1105 A10.

DH76 Software Release Note—A10.4 Version

Date of release: 26th Mar 2020

1. Software version information

2. Software design change point

1. Fix the problem which the mode change from CBC to CBC+DIFF when the sample
data transfer to the third party LIS software;
2. Fix the issue that the IPU software crash occasionally;
3. Remove a fault alarm that there is an Insufficient Aspiration Volume prompted when
the machine detect insufficient cleanser;
4. Address the problem that the machine get zero in WBC differentiation parameter
occasionally;
5. Decrease the alarm of HGB voltage abnormal;

6. Resolve the issue that the machine classify Mon and Lym incorrectly;
7. Improve the performance of PLT of the sample with small MCV value, which
effectively decrease the possibility of PLT bad repeatability of the sample with
microcyte;
8. Decrease the possibility of low Eos;
9. Add two more languages: Russian and French;
10. Remove the function button of Flush, which we can find it in Service-Maintenance
previously;
11. Increase the QC print items to 31;
12. Fix the problem that the machine failed to exit sleep mode.
Dymind software release note—DM1105 A10.4

3. Software upgrading process and compatibility

Current version Upgrading process Remark

Host A9.6 A9.6 → A10.4 Support

Host A10.0 A10.0 → A10.4 Support

Host A10.3 A10.3 → A10.4 Support

Host A9.6 A9.6 → A10.4 Support

Host A10.0 A10.0 → A10.4 Support

Host A10.3 A10.3 → A10.4 Support

4. Upgrading file

Function File name

IPU software
DM1105G001-003[A10.4]setup.exe
upgrading

Host upgrading file upgrade_1105_v965_202003021650.tar.bz2

Upgrading tool for


upgradestudio_v13132_201912251528.zip
Host upgrading

Note: Please use the correct upgrading tool when doing the HOST upgrade

You might also like