CADD-Legacy PCA ™
S I M S D E LT E C , I N C .
i
The issue date of this Operator’s Manual is included on the back cover for
the clinician’s information. In the event one year has elapsed between the
issue date and product use, the clinician should contact SIMS Deltec, Inc.
to see if a later revision of this manual is available.
Technical Assistance
If you have comments or questions concerning the operation of the
CADD-Legacy™ pump, please call the appropriate number given below.
When calling, please specify your pump’s software module. This informa-
tion is located on the start-up screen.
CADD, Medication Cassette Reservoir design, CADD-Legacy, and Medication Cassette are
SIMS trademarks. (The symbol ® indicates it is registered in the U.S. Patent and Trademark
Office and certain other countries.)
DURACELL® is a registered trademark of Duracell Inc. EVEREADY® ENERGIZER® is a
registered trademark of Union Carbide Corp. Super Sani-Cloth® is a registered trademark of
Professional Disposables, Inc.
The products described are covered by one or more of the following: U. S. Patent Nos.
4,559,038; 4,565,542; 4,650,469; 5,364,242; 5,531,697; 5,538,399; 5,540,561; 5,564,915;
5,567,119; 5,567,136; 5,647,854; 5,695,473; Japanese Patent No. 2034590; European
Patent No. 0182502; other patent(s) pending, foreign patent(s) pending.
ii
Warnings
• This Operator’s Manual should be used by clinicians only. Do
not permit patients to have access to this manual, as the informa-
tion contained would allow the patient complete access to all
programming and operating functions. Improper programming
could result in death or serious injury to the patient.
iii
• When the Air Detector is turned off, the pump will not detect air
in the fluid path. Periodically inspect the fluid path and remove
any air to prevent air embolism.
iv
• When you enter a new Dose Lockout time or Doses per Hour
value, any lockout time in effect will be cleared. A Demand Dose
could be requested and delivered immediately upon starting the
pump, resulting in over-delivery.
• If the pump is dropped or hit, the battery door tabs may break.
Do not use the pump if the battery door or tabs are damaged
because the batteries will not be properly secured; this may result
in loss of power and non-delivery of drug.
• Close the fluid path tubing with the clamp before removing the
cassette from the pump to prevent unregulated gravity infusion.
• Ensure that the entire fluid path is free of all air bubbles before
connecting to the patient to prevent air embolism.
Cautions
• Do not operate the pump at temperatures below +2°C (36°F) or
above 40°C (104°F).
vi
vii
Warnings iii
Cautions vi
ix
Section 5: Reference 49
Messages and Alarms, Alphabetical List ................................... 49
Cleaning the Pump and Accessories .......................................... 53
Exposure to Radiation, Ultrasound, Magnetic Resonance
Imaging (MRI), or Use near ECG Equipment ........................ 55
Continuous Rate Scroll Ranges ................................................. 56
Demand Dose, Clinician Bolus Scroll Ranges: Milliliters .......... 56
Demand Dose, Clinician Bolus Scroll Ranges: Milligrams ......... 57
Demand Dose, Clinician Bolus Scroll Ranges: Micrograms ....... 58
Technical Description ............................................................... 59
Specifications (Nominal) ..................................................... 60
Accuracy Test Results ......................................................... 65
Safety Features and Fault Detection .................................... 67
Software Safety Features ..................................................... 69
Data Handling Software Safety Features ............................. 70
Annual Functional Inspection and Testing Procedures ........ 71
Inspection Procedures ......................................................... 71
Testing Procedures .............................................................. 72
Occlusion Pressure Range Tests .......................................... 79
Accuracy Tests .................................................................... 82
Index ........................................................................................ 87
Limited Warranty ..................................................................... 90
Introduction
The CADD-Legacy™ PCA (Patient Controlled Analgesia) ambula-
Description
General
tory infusion pump provides measured drug therapy to patients in
hospital or outpatient settings. Therapy should always be overseen
by a physician or a certified, licensed healthcare professional. As
appropriate to the situation, the patient should be instructed in
using and troubleshooting the pump.
Indications
The CADD-Legacy™ PCA pump is indicated for intravenous, intra-
arterial, subcutaneous, intraperitoneal, epidural space, or subarach-
noid space infusion. The pump is intended for therapies that require
a continuous rate of infusion, patient-controlled demand doses, or
both (such as patient-controlled analgesia).
Epidural/Subarachnoid Administration
The selected drug must be used in accordance with the indications
included in the package insert accompanying the drug. Administra-
tion of any drug by this pump is limited by any warnings, precau-
tions, or contraindications in the drug labeling.
Analgesics
Administration of analgesics to the epidural space is limited to use
with indwelling catheters specifically indicated for either short-or
long-term drug delivery.
Anesthetics
Administration of anesthetics to the epidural space is limited to use
with indwelling catheters specifically indicated for short-term drug
delivery.
Description
General
WARNING:
• Do not administer drugs to the epidural space or subarachnoid
space unless the drug is indicated for administration to those
spaces. Drugs not intended for epidural or subarachnoid space
infusion could result in death or serious injury to the patient.
• To prevent the infusion of drugs that are not indicated for
epidural space or subarachnoid space infusion, do not use ad-
ministration sets that incorporate injection sites. The inadvertent
use of injection sites for infusion of such drugs could result in
death or serious injury to the patient.
• If a Medication Cassette™ Reservoir, CADD® Extension Set or
CADD® Administration Set is used for drug delivery into the
epidural or subarachnoid space, clearly differentiate them from
those used for other routes of infusion, for example, by color
coding, or other means of identification. Drugs not intended for
epidural or subarachnoid space infusion could result in death or
serious injury to the patient.
Symbols
Description
Accessory Jack
General
! Attention, consult accompanying documents
(read Instructions for Use)
K Class II Equipment
J Type CF Equipment
D Date of Manufacture
SN Serial Number
Pump Diagram
Display
Description
Power Jack
General
Accessory Jack
Front
AC Indicator
View
Light
Air Detector
®
Keypad
Cassette
Power Jack
symbol
Threaded
Mounting Accessory
Hole Jack symbol
Battery Rear
Compartment
View
Cassette Lock
Description
to power pump.
General
Display
The Liquid Crystal Display (LCD) shows programming information
and messages. In this manual, the term “display” is synonymous
with display panel or LCD.
Keypad
The keys on the keypad are described below. A key beeps when
pressed if it is operable in the current lock level.
⁄ starts and stops pump delivery; silences alarms.
¤ is used to enter (save) a new value in the pump’s memory when
programming doses or pump settings or to clear values from
record-keeping screens. It is also used to return from the
Biomed Functions to the main screen (Section 4).
‹ is used to fill the tubing and to remove air bubbles from the
fluid path.
Œ is used to view or change the pump’s current lock level. Lock
levels are used to limit patient access to certain programming
and operating functions. (See Lock Levels, this section.) This
key is also used to access the Clinician Bolus while the pump is
running.
„ is used to move from one programming screen to the next
without changing the setting or value displayed; silences
alarms.
´ allows you to “scroll up” or increase a value, or scroll through
Biomed Function settings.
Î allows you to “scroll down” or decrease a value, or scroll
through Biomed Function settings.
Å is used to put the pump into a low power state when not in
use or back into full power.
Í is used in the PCA delivery mode. It allows the patient to
deliver a programmed amount of medication upon request.
Description
General
Power Jack
You may plug an AC Adapter into the Power Jack as an alternate
source of power. The indicator light on the front of the pump will
illuminate when the AC Adapter is in use.
Accessory Jack
The accessory jack is used for attaching a Remote Dose Cord for
remote operation of the dose key and for accessory cables. See the
Instructions for Use supplied with those accessories.
Air Detector
The Air Detector is on the pump in the area shown in the diagram.
If air is detected in the part of the tubing that passes through the Air
Detector, an alarm sounds and delivery stops. (See Section 5 for Air
Detector specifications.) If an Air Detector is not required, it may be
turned off. (See Section 4, Biomed Functions.)
WARNING: When the Air Detector is turned off, the pump will not
detect air in the fluid path. Periodically inspect the fluid path and
remove any air to prevent air embolism. Air embolism could result
in death or serious injury to the patient.
Cassette
The cassette is the part of the Medication Cassette™ Reservoir or
CADD® Administration Set that attaches to the bottom of the
pump. The following single-use products are compatible with the
CADD-Legacy™ pump:
Description
General
either an integral or an Add On Anti-Siphon Valve to
protect against unregulated gravity infusion that can result
from an improperly attached cassette. Unregulated gravity
infusion could result in death or serious injury to the pa-
tient.
Battery Compartment
Two AA batteries fit into the battery compartment. The AA batter-
ies serve as the primary source of power, or as a backup when an
AC Adapter is in use.
Cassette Lock
This attaches the cassette (the part of the Medication Cassette™
Reservoir or CADD Administration Set that attaches to the pump)
to the pump. This allows you to secure the cassette to the pump
using the key provided. If the cassette becomes unlocked while the
pump is running, delivery will stop and an alarm will occur. If the
cassette becomes unlocked while the pump is stopped, an alarm will
occur.
“Upstream Occlusion.”
ing volume, inspect the fluid path for kinks, a closed clamp, or
General
When stopped:
STOPPED
Status of pump
Lock Levels
Description
General
Lock levels are used to limit patient access to certain programming
and operating functions. The table on the next page lists the func-
tions that are accessible in Lock Level 0 (LL0), Lock Level 1 (LL1),
and Lock Level 2 (LL2). When a function is accessible, the key
associated with the function beeps when pressed. If a function is not
accessible, the pump ignores the key press and a beep does not
sound. Section 2, Pump Setup and Programming, describes how to
change the lock level.
Security Codes
The following security codes are preset by the manufacturer for the
clinician’s use:
Stopped Running
Pump Operations
and Programming LL0 LL1 LL2 Any
Lock Level
Stop/Start the pump Yes Yes Yes Yes
Reset Reservoir Volume Yes Yes Yes No
Prime Yes Yes No No
Change the lock level Yes, w/code Yes, w/code Yes, w/code No
Start a Demand Dose No No No Yes
Start a Clinician Bolus No No No Yes
Change Units Yes No No No
Change Concentration Yes No No No
Change Continuous Rate Yes Up to LL0 value No No
Change Demand Dose Yes Up to LL0 value No No
Clear Doses Given Yes Yes No No
Clear Doses Attempted Yes Yes No No
Clear Given amount Yes Yes No No
Biomed Functions
Access to Functions Yes, w/code No No No
Air Detector On/Off Yes w/code View only View Only View Only
Upstream Occlusion
Sensor On/Off Yes w/code View only View Only View Only
10
11
1. Push down and hold the arrow button while sliding the
door off.
Pump Setup &
Programming
NOTE:
12
13
14
NOTE:
• The life of the batteries is dependent on the amount of
medication delivered, delivery rate, battery age, and the
temperature.
Programming
• The power of the batteries will be quickly depleted at tem-
peratures below +10°C (50°F).
15
Watching Power Up
When you install the batteries, the pump will start its power up
sequence during which it performs self-tests and displays pro-
grammed values. Watch for the following:
16
17
18
Delivery Methods
The CADD-Legacy™ PCA pump offers three methods of delivery
alone or in combination:
• Continuous Rate (up to 50 ml per hour)
• Demand Dose
• Clinician Bolus
Programming
Operating the Pump. Ranges and programming increments are
listed in the Specifications in Section 5.
Clinician Bolus
(used here as a
loading dose) Continuous Rate
Demand Doses
Dosage
Time
19
Milligrams
Concentration Concentration
(mg/ml or mcg/ml) 1.0 mg/ml
20
Doses Attempted
Doses Attempted
2 doses
Given Given
(ml, mg, or mcg)
2.50 ml
Reservoir Volume
Enter the volume of fluid contained in a filled fluid container. The
Reservoir Volume value decreases as the pump delivers fluid or as
you prime the tubing. When you change the fluid container, reset
the reservoir volume on this screen. If you do not wish to use the
Reservoir Volume feature, scroll down to “Not In Use” (located
before 1 and after 9999 in the range of values).
The reservoir volume could be set higher than the capacity of the
fluid container. Be sure to program the reservoir volume to reflect
the actual volume of the medication being used.
Units
Enter the programming units. Possible settings are milliliters,
milligrams, and micrograms. When you change the Units, the pump
requires you to enter or verify the Continuous Rate and Demand
Dose. If the units are mg or mcg, you must also enter the Concen-
tration. Changing the Units clears the amount Given.
Concentration
If Units are mg or mcg, enter the concentration of drug in mg/ml or
mcg/ml. When you enter a new Concentration, the pump requires
you to enter or verify Continuous Rate and Demand Dose.
21
Continuous Rate
Enter the continuous rate of medication delivery (in mg/hr, ml/hr, or
mcg/hr, depending on the units). The maximum rate is 50 ml/hr or
the mg or mcg equivalent. If the prescription does not call for a
Continuous Rate, enter zero.
NOTE: If you intend to run the pump in Lock Level 1 so the
Continuous Rate can be varied, you should enter the maximum
allowable rate while programming in Lock Level 0. After program-
ming, you may then change to Lock Level 1 and decrease the rate
to its starting value. See Programming with Upper Limits, Adjusting
Pump Setup &
Programming
Demand Dose
Enter the amount of drug to be delivered when the patient presses
Í (or the Remote Dose Cord button if attached). If the prescrip-
tion does not call for a Demand Dose, enter zero.
NOTE: If you intend to run the pump in Lock Level 1 so the
Demand Dose can be varied, you should enter the maximum
allowable dose while programming in Lock Level 0. After program-
ming, you may then change to Lock Level 1 and decrease the dose
to its starting value. See Programming with Upper Limits, Adjusting
Doses in LL1 at the end of Section 2.
Dose Lockout
If you programmed a Demand Dose, enter the minimum amount of
time that must elapse between the time one Demand Dose starts
and the time the next Demand Dose starts. This lockout period is
unaffected if the batteries are removed or if the pump is stopped.
22
Given
This screen shows the total amount of drug delivered since the last
time this value was cleared. The amount shown is rounded to the
nearest 0.01 mg, ml, or mcg. If this value reaches 99999.95 or
99999.99, depending on units and concentration, it automatically
returns to 0 and continues counting. The Given amount does not
include drug used when priming the tubing.
23
24
• Press „ to begin.
• Press ¤.
• Press „.
• Press ¤.
25
• Press „.
• Press ¤.
• Press „.
Pump Setup &
Programming
• Press ¤.
• Press „.
• Press ¤.
• Press „.
• Press ¤.
26
• Press „.
• Press ¤.
WARNING: When you enter a new Doses per Hour value, any
lockout time in effect will be cleared. A Demand Dose could be
requested and delivered immediately upon starting the pump,
resulting in over-delivery, which could result in death or serious
injury to the patient.
• Press „.
27
• Press „.
• Press „.
• Press „.
• Press „.
28
• Press „.
29
Removing a Cassette
WARNING: Close the fluid path tubing with the clamp before
removing the cassette from the pump to prevent unregulated gravity
infusion, which could result in death or serious injury to the patient.
3. Insert the key into the lock and turn it clockwise. The lock
will pop out when you unlock the cassette.
30
Attaching a Cassette
Obtain a new, filled Medication Cassette™ Reservoir, or CADD®
Administration Set attached to a non-vented, flexible IV bag.
2. Insert the cassette hooks into the hinge pins on the pump.
1999-08-17 D. Zurn
«Lgc Attach Cass Key»
31
4. Insert the key into the lock, push in, and turn counterclock-
wise until the line on the lock lines up with the arrow on the
side of the pump and you feel the lock click into place.
1998-12-17 D. Zurn
«Lgc Lock Key»
Pump Setup &
Programming
1
«
1999-01-23 D. Zurn
«Lgc Twist 50 ml»
32
NOTE: If you are not changing the fluid container but wish to
prime the fluid path, you may follow the same procedure.
WARNING: Do not prime the fluid path with the tubing connected
to a patient as this could result in overdelivery of medication or air
embolism, which could result in death or serious injury to the
2. Press and hold ‹. You will hear a single beep, and the
word “Prime” will appear on the display.
3. After “Prime” and three sets of dashes appear, and you hear
three beeps, release ‹.
4. Press and hold ‹ again to fill the fluid path and to elimi-
nate air bubbles. The screen displays “Priming . . .” and you
will hear a short beep each time the pump goes through a
delivery cycle.
NOTE:
• The air detector alarm is automatically disabled when
priming.
• Fluid delivered during priming is subtracted from the
Reservoir Volume, but is not added to the Given screen
since this fluid is not delivered to the patient.
33
WARNING: Ensure that the entire fluid path is free of all air
Pump Setup &
Programming
34
2. Place the tubing over the groove in the air detector and tuck
it under the catch.
Catch
35
3. To seat the tubing into the groove, gently pull the tubing,
until it is under the retention nubs and flat in the groove.
Pump Setup &
Programming
Retention
nubs
WARNING: Ensure that the entire fluid path is free of all air
bubbles before connecting to the patient to prevent air embo-
lism. Air embolism could result in death or serious injury to the
patient.
36
NOTE: You may change the lock level at any time by stopping the
pump and following the procedure below.
37
38
Operating
the Pump
To start the pump
1. Press and hold ⁄.
39
40
Operating
the Pump
To start a Clinician Bolus
1. Make sure the pump is running (in any lock level). Start the
pump if necessary.
2. Press Œ.
4. Press Œ again or ¤.
6. Press ¤ or Í.
41
42
Operating
2. Press „ to display the Reservoir Volume screen. the Pump
3. Press ¤ to reset the volume to the programmed volume.
43
44
45
WARNING: When the Air Detector is turned off, the pump will not
detect air in the fluid path. Periodically inspect the fluid path and
remove any air to prevent air embolism. Air embolism could result
in serious injury or death to the patient.
46
47
48
5.0 Reference
49
50
Power lost while The pump was running when power was
pump was on removed. Stop the pump before changing
TWO-TONE ALARM the batteries or removing the power
source. Press ⁄ or „ to silence the
alarm.
Reference
51
52
Routinely clean the pump to keep it free of dirt, liquids, and foreign
objects.
Use any of the following solutions to clean the pump and accesso-
ries:
• Soap solution
• Benzalkonium Chloride concentrate (0.13%)
• Glutaral Concentrate, USP (2%)
• 10 percent solution of household bleach (one part house-
hold bleach to nine parts water)
• Alcohol, USP (93%)
• Isopropyl Alcohol, USP (99%)
• Chlorohexidine (70%)
• PDI — Super Sani-Cloth®
• Mada Medical — MadaCide
Reference
2. Wipe the entire surface dry with another soft, lint-free cloth.
Allow the pump to dry completely before use.
53
54
55
Starting
Units Value Increment Maximum
Milliliters
Demand Dose Clinician Bolus
increment max. increment max.
0.05 9.9 0.05 20
Reference
56
Concentration Milligrams
mg/ml Demand Dose Clinician Bolus
increment max. increment max.
0.1 0.01 0.99 0.01 2
0.2 0.02 1.98 0.02 4
0.3 0.03 2.97 0.03 6
0.4 0.04 3.96 0.04 8
0.5 0.05 4.95 0.05 10
1 0.05 9.9 0.05 20
2 0.10 19.8 0.10 40
3 0.15 29.7 0.15 60
4 0.20 39.6 0.20 80
5 0.25 49.5 0.25 100
10 0.50 99.0 0.50 200
15 0.75 148.5 0.75 300
20 1.00 198.0 1.00 400
25 1.25 247.5 1.25 500
30 1.50 297.0 1.50 600
35 1.75 346.5 1.75 700
40 2.00 396.0 2.00 800
45 2.25 445.5 2.25 900
50 2.50 495.0 2.50 1000
55 2.75 544.5 2.75 1100
60 3.00 594.0 3.00 1200
65 3.25 643.5 3.25 1300
70 3.50 693.0 3.50 1400
Reference
57
Micrograms
Concentration
mcg/ml Demand Dose Clinician Bolus
increment max. increment max.
1 0.05 9.9 0.05 20
2 0.10 19.8 0.10 40
3 0.15 29.7 0.15 60
4 0.20 39.6 0.20 80
5 0.25 49.5 0.25 100
10 0.50 99.0 0.50 200
15 0.75 148.5 0.75 300
20 1.00 198.0 1.00 400
25 1.25 247.5 1.25 500
30 1.50 297.0 1.50 600
35 1.75 346.5 1.75 700
40 2.00 396.0 2.00 800
45 2.25 445.5 2.25 900
50 2.50 495.0 2.50 1000
55 2.75 544.5 2.75 1100
60 3.00 594.0 3.00 1200
65 3.25 643.5 3.25 1300
70 3.50 693.0 3.50 1400
75 3.75 742.5 3.75 1500
80 4.00 792.0 4.00 1600
85 4.25 841.5 4.25 1700
90 4.50 891.0 4.50 1800
95 4.75 940.5 4.75 1900
Reference
58
Technical Description
Standards used in Development of the Pump
The following standards were used in whole or part in the development of
the pump.
Medical Electrical Equipment
IEC 60601-1, Medical Electrical Equipment, Part 1: General
Requirements for Safety.
IEC 60601-1-1, Medical Electrical Equipment, Part 1: General
Requirements for Safety. Collateral Standard: Safety requirements
for medical electrical systems.
IEC 60601-1-2 (1993), General Requirements for Safety, Part 2:
Electromagnetic Compatibility – Requirements and Tests.
IEC 60601-1-2, (2nd Edition, 01-June-1996) Medical Electrical
Equipment, Part 1: General Requirements for Safety. 2. Collateral
Standard: Electromagnetic Compatibility – Requirements and Tests.
IEC 60601-1-4, Medical Electrical Equipment, Part 1: General
Requirements for Safety – 4. Collateral standard: Programmable
electrical medical systems.
IEC 60601-2-24, Medical Electrical Equipment, Part 2: Particular
Requirements for the Safety of Infusion Pumps and Controllers.
Electromagnetic Compatibility
RTCA/DO -160C, Radiated Emissions Only, Category A & Z
Limit.
CISPR11, Radiated and Conducted Emissions, Group 1, Class B.
Reference
59
Specifications (Nominal)
General Pump Specifications
Resolution .......................... Medication Cassette™ Reservoir or
CADD® Administration Set, 0.050 ml per
pump stroke nominal
Size ...................................... 4.1 cm × 9.5 cm × 11.2 cm [1.6 in. × 3.8 in.
× 4.4 in.] excluding cassette or other
accessories
Reference
60
61
System Operating
Temperature ...................... +2°C to 40°C (35°F to 104°F)
System Storage
Temperature ...................... -20°C to 60°C (-4°F to 140°F)
System Delivery
Accuracy ............................ ± 6% (nominal). At low infusion rates,
this accuracy may not be achieved for
short periods. During the total infusion
time, the accuracy averages out (see
accuracy curves, pages 65 and 66)
System Definition .............. System is defined as a CADD-Legacy™
pump with an attached Medication
Cassette™ Reservoir and CADD®
Extension Set with integral anti-siphon
valve, or an attached CADD®
Administration Set with integral or add-
on anti-siphon valve.
High Pressure Alarm ........ 26 ± 14 psi [1.79 ± 0.97 bar].
Air Detector Alarm ........... Single bubble
Low sensitivity = greater than 0.250 ml
High Sensitivity = greater than 0.100 ml
Multi-bubble = 1.0 ml nominal
Bolus Accuracy at Set
Value of 0.1 ml .................. ± 6%
Bolus Accuracy at Set
Value of 6.0 ml .................. ± 6%
Maximum Volume
Reference
62
minutes
5 minutes between 20 minutes and 24
hours
Default: 24 hours
Doses Per Hour ................. 1 – 12 doses in 1 dose increments (will
also be limited by the Dose Lockout
value)
Default: 1
Doses Given .................... 0 to 999
Doses Attempted ............. 0 to 999
63
Biomed Functions
Air Detector ....................... Off
On-Low
On-High
Default: On-High
Upstream Sensor ............... Off
On
Default: On
Reference
64
48
36
Flow 24
(ml/hr)
12
0
0 10 20 30 40 50 60 70 80 90 100 110 120
T (min)
siphon valve
10
5
Epmax
% Error 0
of Flow Epmin
-5
4 10 22 38 62 p
-10
2 5 11 19 31 min
T (min)
65
Flow
(ml/hr)
T (min)
% Error
of Flow
T (min)
66
expecting it, the microprocessor traps the event, sounds the audible
alarm and displays an error message on the LCD.
67
68
69
70
Inspection Procedures
Visual Inspection
• Visually inspect the pump for any damage to the LCD,
occlusion sensor seals, valves and expulsor, pump hinge
Reference
71
Mechanical Inspection
• Press each key on the keypad. Each key should have a
distinctive dome feeling. The keys should not feel flat.
• Attach the battery door. The battery door should fit snugly
in place when it is closed on the pump.
Testing Procedures
Functional Testing
Reference
Power-up Check
• Insert batteries or press Å and observe the LCD during
power up. The first screen will display the serial number,
model number, and software number with revision level.
The second screen will display 32 character blocks. (If
“LEC” (Last Error Code) and four digits appear prior to the
pump displaying the 32 character blocks, the pump has
experienced an electrical or mechanical fault and should be
returned for service.) If no error message is immediately
72
Lock Check
• Attach a 50 or 100 ml Medication Cassette™ Reservoir or
CADD® Administration Set to the pump. The line on the
lock should be aligned with the arrow on the side of the
pump.
The following three checks (LCD, Motor and Gear Train and
Reservoir Volume Empty Alarm Check) should be performed in the
sequence shown.
LCD Check
Reference
73
74
75
GIVEN Check
Reference
76
77
78
Description
Pressure is generated by activating the pumping mechanism with an
attached filled, clamped Medication Cassette™ Reservoir. The pump
is started and a Demand Dose is given until the high pressure alarm
sounds.
Equipment needed
50 or 100 ml Medication Cassette™ Reservoir containing distilled
water
Procedure
1. Insert two AA batteries or turn the pump on and wait for
the pump to power up.
4. Close the slide clamp on the distal end of the tubing near the
female luer of the Medication Cassette™ Reservoir.
79
Description
An adjustable metered pressure source is connected to the Medica-
tion Cassette™ Reservoir tubing. The pressure is slowly increased
until the high pressure alarm sounds.
Equipment needed
Pressure gauge, 40 psi ± 1 psi [2.76 bar ± 0.07 bar]
Procedure
1. Insert two AA batteries and wait for the pump to power up.
NOTE: The pressure from the source must be zero when the
cassette is attached.
80
1998-11-22 D. Zurn
«Lgc Occlusion Setup»
40
PSI
Regulator Pressure Gauge
81
Accuracy Tests
Gravimetric Accuracy Testing
Description
A Medication Cassette™ Reservoir is partially filled with water and
weighed, then attached to a pump that is set to deliver a certain
amount of water. The Medication Cassette™ Reservoir is then
removed and weighed again. The amount of water delivered is
compared to the amount that the pump should have delivered.
Equipment needed
50 or 100 ml Medication Cassette™ Reservoir
50 or 60 ml syringe
Procedure
1. Fill the 50 or 60 ml syringe with 40 ml of water. Transfer
the water into a Medication Cassette™ Reservoir.
82
is filled with fluid to the end of the extension set luer lock
connector.
3. Secure the slide clamp as close to the extension set luer lock
connector as possible. This should assure a minimum water
loss from the tubing when the syringe is removed.
83
percentage.
84
Description
A predetermined amount of water is delivered into a collection
device such as a burette or graduated cylinder. The amount of water
delivered is compared to the amount that the pump is programmed
to deliver.
Equipment needed
50 or 100 ml Medication Cassette™ Reservoir
50 or 60 ml syringe
Procedure
Reference
85
device.
86
C
A cassette, 4, 6
AC Adapter, 6, 7, 61 attaching, 31
AC indicator light, 4, 5 cautions, vi
accessory jack, 4, 6 removing, 30
accuracy, 65 warnings, v
accuracy tests, 82 cassette lock, 4, 7, 32
gravimetric, 82 cleaning pump, accessories, 53–54
volumetric, 85 cautions, vii
Administration Set, 6, 7 Clinician Bolus, 10, 19, 64
cautions, vi starting, 41
warnings, iv stopping, 43
Air Detector, 4, 6, 16, 35, 46, 64 Clinician Bolus Code, 9, 41
inserting tubing into, 35 concentration, 10, 21, 63
status screen, 23, 28 programming, 26
testing, 77 Continuous Rate, 10, 19, 21, 22,
warning, iv 38, 56, 63, 64
alarms, 49, 61 programming, 26, 38
analgesics, 1
anesthetics, 2 D
caution, vii
Anti-Siphon Valve, 7 delivery mode, 19
warning, iv Demand Dose, 10, 19, 22, 23, 26,
arrow keys, 5 56, 63
Reference
87
88
symbols, 3
syringe
warning, iii
system definition, 62
T
testing procedures, 72–78
turning pump on/off, 40
U
ultrasound, 55
caution, vii
units, 10, 20, 21, 25
Upstream occlusion sensor, 7, 24,
29, 47
testing, 77–78
W
warnings and cautions, iii–vii
Reference
89
Limited Warranty
SIMS Deltec, Inc. (the “Manufacturer”) warrants to the Original Purchaser that the infusion
pump (the “Pump”), not including accessories, shall be free from defects in materials and work-
manship under normal use, if used in accordance with this Operator’s Manual, for a period of
one year from the actual date of sale to the Original Purchaser. THERE ARE NO OTHER
WARRANTIES.
This warranty does not cover normal wear and tear and maintenance items, and specifically
excludes batteries, administration sets, extension sets or any other accessory items or equip-
ment used with the Pump.
Subject to the conditions of and upon compliance with this Limited Warranty, the Manufac-
turer will repair or replace at its option without charge (except for a minimal charge for postage
and handling) any Pump (not including accessories) which is defective if a claim is made during
such one-year period.
The following conditions, procedures, and limitations apply to the Manufacturer’s obligation
under this warranty:
A. Parties Covered by this Warranty: This warranty extends only to the Original Purchaser of
the Pump. This warranty does not extend to subsequent purchasers. The Original Purchaser
may be a patient, medical personnel, a hospital, or institution which purchases the Pump for
treatment of patients. The Original Purchaser should retain the invoice or sales receipt as proof
as to the actual date of purchase.
B. Warranty Performance Procedure: Notice of the claimed defect must be made in writing or
by telephone to the Manufacturer as follows: SIMS Deltec 1265 Grey Fox Road, St. Paul MN
55112 U.S.A., 1-800-426-2448 or SIMS Graseby Ltd. WD2 4LG UK, +44 (0)1923 246434.
Notice to the Manufacturer must include date of purchase, model and serial number, and a
description of the claimed defect in sufficient detail to allow the Manufacturer to determine and
facilitate any repairs which may be necessary. AUTHORIZATION MUST BE OBTAINED
PRIOR TO RETURNING THE PUMP. If authorized, the Pump must be properly and care-
fully packaged and returned to the Manufacturer, postage prepaid. Any loss or damage during
shipment is at the risk of the sender.
C. Conditions of Warranty: The warranty is void if the Pump has been 1) repaired by someone
other than the Manufacturer or its authorized agent; 2) altered so that its stability or reliability
is affected; 3) misused; or, 4) damaged by negligence or accident. Misuse includes, but is not
limited to, use not in compliance with the Operator’s Manual or use with nonapproved accesso-
ries. The Pump is a sealed unit, and the fact that the seal has been broken will be considered
conclusive evidence that the Pump has been altered or misused. Removal or damage to the
Reference
90
Reference
91
European Representative:
SIMS Graseby Ltd.
WD2 4LG UK
+44 (0)1923 246434
2000-03
40-3933-51D