3-1 HbA1c Clase 1
3-1 HbA1c Clase 1
HbA1c
2013
1
Diabetes
Definition
• Diabetes is usually a lifelong (chronic) disease in which there
are high levels of sugar in the blood, either because insulin
production is insufficient, or because the body's cells do not
respond properly to insulin, or both.
POCT
• Glucose monitoring
• HbA1c
• Microalbuminurea
2
HbA1c
What is HbA1c?
Glycated hemoglobin, or glycohemoglobin.
3
HbA1c is highly glycated during diabetes
4
Hb - Hemoglobin
HbA0: non-glycated hemoglobin.
HbA1: glycated hemoglobin
HbA1a1: Glycation with Fructose 1, 6 diphosphate
HbA1a2: Glycation with Glucose 6 phosphate
HbA1b: Glycation with an unknown reaction partner
5
Why is HbA1c test important?
The amount of HbA1c reflects the blood sugar
levels for the last several weeks, typically
encompassing periods of 3-4 months.
6
What is the meaning of % for HbA1c
i-Chamber Cartrige
8
i-Chroma HbA1c
For measuring glycated hemoglobin in human blood specimen
HbA1c(in sample)
Sample
fluorescence
fluorescence
Competition HbA1c
Peptide
9
i-Chroma HbA1c
For measuring Total hemoglobin in human blood specimen
Hemoglobin assay
Sample Met-hemoglobin
potassium ferricyanide
525nm light
Measurement
10
i-Chroma HbA1c
Specifications Data
Product name iChroma™ HbA1c
Intended use Quantitative test of HbA1c % in human blood
CV % <10%
Correlation R2 >= 0. 9
Reaction time 12min
Detection Range,Ref. value 4 – 15% , 6.5%
Sample type Whole blood
Sample volume 5 uL
DB volume -
Stability 20 months, 2 ~ 8℃
Test device : 4 ~ 30℃
Storage condition
Detection buffer : 2 ~ 8℃
11
Test procedure
Preparation - 1
12
Test procedure
Preparation - 2
[Cartridge]
1. Insert a cartridge into a i-Chamber slot for
stabilizing the reaction temperature.
13
Test procedure
Preparation - 3
[Reaction Buffer]
1. Take 100uL of detection buffer from plastic
amber vial.
14
Test procedure
Blood collection
1. Collect 5 uL of whole blood from
fingertip or tube blood with capillary.
15
Test procedure
Blood collection
2. Put the capillary contains blood sample
into the tube prepared in “reaction
buffer preparation step”.
16
Test procedure
Sample processing
1. Shake the tube 15 times strongly in order
that the blood come out from capillary.
17
Test procedure
Sample processing
3. Apply 150uL(75 ul x 2 times) of the mixture
into the Hb well using a pipette
18
Test procedure
Sample processing
Apply 75ul of the mixture into the A1c well and
re-insert the cartridge in i-chamber
19
Test procedure
Reading
1. Keep the cartridge in i-Chamber for 12
minutes before scanning by i-CHROMA
Reader
20
Test procedure
Reading
2. The cover of window 2 is automatically
opened when the cartridge is inserted into
a holder of i-CHROMA reader.
21
Accuracy of i-Chroma HbA1c
• The accuracy of HbA1c(New) with i-Chamber is remarkably improved according to keeping
temperature through incubation the cartridge in the chamber.
• Test Value
• Recovery
7% Patient Sample
Cartridge
Products Data Run1 Run2 Run3 Run4 Run5 Run6 AVG CV(%)
Temp
22
Accuracy of i-Chroma HbA1c
Reproducibility(Between run, Interassay CV%)
of i-CHROMA HbA1c(new) with i-Chamber
G7 Tester
No. AVG CV(%)
(Reference) Tester 1 Tester 2 Tester 3
1 5 5.19 5.5 4.99 5.23 4.96
2 5.1 5.35 5.12 5.29 6.18 2.71
3 5.5 5.14 5.4 5.77 7.03 3.76
4 5.6 5.56 5.5 5.95 5.44 5.85
5 5.9 6.27 6.28 5.99 7.29 0.94
6 6 6.22 6.52 6.41 6.69 1.19
7 6.3 6.4 6.18 5.97 7.13 2.44
8 6.3 6.59 6.36 6.26 5.25 2.28
9 6.4 6.8 6.39 6.66 5.67 4.31
10 6.7 6.36 6.69 6.35 8.96 4.84
11 6.8 6.76 6.71 6.6 6.62 3.15
12 6.9 7.17 7 6.72 7.65 3.29
13 7 7.09 6.98 7.32 6.47 3.01
14 7.1 6.73 7.23 7.13 6.18 3.52
15 7.3 7.33 7.32 7.21 7.78 2.10
16 7.3 7.43 7.02 7.2 6.38 2.38
17 7.8 7.77 7.36 7.82 6.96 3.24
18 8 7.64 7.96 7.74 6.40 2.62
19 8.5 8.9 8.71 8.94 8.85 1.40
20 9 8.56 8.91 9.42 7.22 2.85
23
Accuracy of i-Chroma HbA1c
• Difference of Correlation(R²) among testers
• Tester 1 ; R² = 0.9333
• Tester 2 ; R² = 0.9539
• Tester 3 ; R² = 0.9462
10 y = 0.9203x + 0.5741
R² = 0.9333
9
(Tester 1)
8 y = 0.9197x + 0.5721
7 R² = 0.9539
(Tester 2)
6 Tester 1
4
(Tester 3) 선형 (Tester 1)
3 선형 (Tester 2)
선형 (Tester 3)
2
0
0 2 4 6 8 10
Tosoh G7 HBA1c (%) HPLC
24
Comparison study of i-Chroma HbA1c
• High consistency is shown ; Comparative result , R² = 0.98 in comparison
with Tosho G7(HPLC)
16.0
10.0 N= 150
8.0
6.0
4.0
2.0
0.0
0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0
25
Comparison study of i-Chroma HbA1c
• Correlation(R2) between i-CHROMA HbA1c and other POCT products is 0.933(Nycocard)
and 0.932(Labonacheck), respectively
Tosoh
No HbA1C “N” “L” "N" with i-CHAMBER HbA1c
G7
1 5.5 5.6 5.5 4.6 14
2 5.5 5.6 5.6 4.9 12 y = 0.9728x + 0.235
26