Professional Documents
Culture Documents
Activity 1A Familiarization With Ac Instruments: The Oscilloscope
Activity 1A Familiarization With Ac Instruments: The Oscilloscope
Activity 1A Familiarization With Ac Instruments: The Oscilloscope
ACTIVITY 1A
FAMILIARIZATION WITH AC INSTRUMENTS: THE OSCILLOSCOPE
b. ability to design and conduct experiments, as well as to analyze and interpret data
d. ability to function on multidisciplinary teams
e. ability to identify, formulate, and solve engineering problems
g. ability to communicate effectively
A typical dual-channel oscilloscope has the following functions. From the oscilloscope,
identify the different parts or component’s mechanism with the corresponding functions.
1. Cathode Ray Tube (CRT) – it is used to display the image of input waveforms. The
control knobs and switches are used to adjust and stabilize the image in the screen. The
screen is divided by a grid consisting of horizontal and vertical scale lines called the
graticule, which is used to measure the displayed waveforms. The side of each square,
or division, is one centimeter.
2. Power Switch – used to turn the oscilloscope power on and off.
3. Light – indicate whether the power is on or off.
4. Intensity Control – usually labeled INTEN or INTENSITY, controls the brightness of
the trace.
5. Focus Control – usually labeled FOCUS, adjusts the sharpness of display.
6. Trace Rotation Control – usually labeled ROTATION, aligns the trace with the
horizontal graticule lines.
The level control, usually labeled LEVEL, sets the amplitude point on the trigger signal at
which the oscilloscope is triggered. There are three level settings: -, +, and the PRESET.
The – and + settings are variables, for examples;
a. When the level control is set in the – range, the oscilloscope is triggered at a point
on the negative portion of the trigger signal.
b. When the control is set in the + range, the oscilloscope is triggered at some point
of the positive portion of the trigger signal.
c. When the control is in the PRESET position, the Oscilloscope is triggered at the
point where the trigger signal amplitude is close to zero volt. Initially, the level
control should be set to PRESET.
29. Hold off Control -- usually labeled HOLDOFF, is used to synchronize the oscilloscope
to input waveforms that do not repeat an identical pattern. Leave this control in the NORM
position.
30. Calibrator – usually labeled CAL or CALIBRATOR, is a square wave output signal
produced by the oscilloscope. It is used to properly calibrate the x10 input probes. You
should calibrate the x10 probes before each major use.
The four procedures that follows allow you to determine and demonstrate the proper setup
and use of the oscilloscope.
1A.5 Materials/Equipment
1- F. A. C. E. T. Base Unit
1- AC 1 FUNDAMENTALS Circuit Board
1- Analog Oscilloscope, dual trace
1A.6 Procedure/s
A. INITIAL SETUP. The following procedure demonstrates how the oscilloscope controls
should be set before each use.
NOTE: If you have trouble locating the following controls on your oscilloscope, refer to
the user’s manual.
1. With the power off, connect x10 probes to the channel 1 and channel 2 input jacks, and
adjust controls to the following settings. Check each box corresponding to each control as
you adjust it.
2. Turn on the oscilloscope power. After the CRT has a few seconds to warm up, a straight
line should appear near the center of the screen.
3. Adjust the channel 1 position control to until the display line rests on the center
horizontal graticule line. What voltage does this line represent?
4. Set the vertical mode control to CH2, and adjust the channel 2 position control so that
the line rests on the center horizontal graticulate. Does adjusting the Channel 2 position
control affect the Channel 1 adjustment?
5. Use a screwdriver to turn the trace rotation control in a clockwise and then in a counter
clockwise direction. How does the display react when the rotation control is turned
clockwise?
6. Adjust the trace rotation control so that the trace is properly aligned with the center
horizontal graticule line.
7. Return the vertical mode control to the CH1 position, and turn down the intensity until
no display is present.
Remember that the x10 probe makes the input’s amplitude look ten times smaller than it
really is. For example, suppose the attenuator control is set to 0.2V/div, a x10 probe is in
use, and the display rises 2 divisions above the ground reference when a dc voltage is
applied to the input. The voltage value is 6 Vdc, determined as follows: V= (0.2V/div) (3
PROCEDURE B:
1. Turn off the power sources. Insert the AC1 FUNDAMENTALS circuit board into the
base unit. Turn on the power source.
2. Use a two-post connector to apply power from the positive variable supply to resistor R1
on the DC/AC WAVEFORMS circuit block.
3. Use a multitester to adjust the positive supply for 2Vdc across R1.
4. Use terminal posts to connect the Channel 2 x10 probe across R1. Turn up the intensity
control to about midpoint.
5. Switch the Channel 1 vertical coupling control to DC. How many divisions did the line
shift above the reference? ____________ divisions.
VR1 = ______V
8. Switch the vertical coupling to AC. Did the display return to the center graticule line?
_____________________________________________________________________
9. Return the vertical coupling to GND, and turn down the intensity until no display is
present. Turn off the power sources to the F. A. C. E. T. base unit, and remove any
connections from the AC 1 FUNDAMENTALS circuit board.
1. Switch the vertical coupling to AC and connect the Channel 1 x10 probe to the calibrator.
Turn up the intensity until a clearly defined trace appears.
2. Adjust the Channel 1 attenuator and time base controls. As you turned the time base
control counter clockwise, did the number of cycles displayed increase or decrease?
________________________________________________________________________
3. Check the box that corresponds to the way your x10 probe is adjusted.
Undercompensated
4. Locate the adjusting screw on the x10 probe; it is located on the probe connector or on
the probe itself. If your probes do not have a method of adjustment, perform step 6 and go
on to PROCEDURE D.
NOTE: You should check your x10 probes for proper adjustment before each major use.
6. Return the Channel 1 vertical coupling to GND and the Channel 1 attenuator to 0.1
V/div.
7. Connect the channel 2 x10 probe to the calibrator, switch the vertical mode to CH 2, and
set the Channel 2 vertical coupling to AC.
8. Adjust the Channel 2 x10 probe as you did to Channel 1 x10 probe in Step 5. Check the
box below that corresponds to the condition of your channel 2 x10 probe prior to
adjustment.
Undercompensated
9. Return the Channel 2 vertical coupling to GND and the channel 2 attenuator to 0.1 V/div.
Return the vertical mode to CH 1. Turn down the intensity until no display is present.
PROCEDURE D:
1. Turn up the intensity to about midpoint. Use the Channel 1 position control to move the
channel 1 ground reference up to the second graticule line form the top of the oscilloscope
screen.
2. Switch the vertical mode to CH 2. Use the Channel 2 position control to move the
channel 2 ground reference down to the second graticule line from the bottom of the
oscilloscope screen.
3. Set the vertical mode to ALT. Which channel does the upper half of the oscilloscope
screen represent? __________
4. Connect the x10 probes to the calibrator, and switch both vertical coupling controls to
Ac. Adjust the channel attenuator and the time base control. Do not allow either channel
display to cross the center horizontal graticule line, both attenuators should have identical
settings. Are the channel displays identical? _____________.
5. Switch the vertical mode to ADD; a single waveform should appear. Since both channel
inputs were of equal size and added together, is the resulting waveform twice the size of
either channel input? _______________________
6. Return the vertical mode switch to ALT. Activate the channel 2 invert switch. Did the
channel 2 waveform invert? _______________________________
7. Switch the vertical mode to ADD. Were the channel inputs added or subtracted?
8. Turn off power to all equipment, and remove all connections from the AC1
FUNDAMENTALS circuit board.
1A.7.1 Calculations