ARCHITECT c8000 System

Service Training
Handouts

2005 Abbott Laboratories, Abbott Park IL


200402-102 Sept 2005 Page 1 of 9
ARCHITECT® c8000® System Service Training Handouts
The Abbott ARCHITECT® System is distributed by Abbott Laboratories, USA, Abbott Park, IL 60064

Document Control Number Revision Date Content or Pages


Revised, Added, or Deleted
200402-101 May, 2005 New Release
200402-102 Sept, 2005 Page 9 of 9 revised and page #s added to all pages

©Copyright 2005 Abbott Laboratories


ARCHITECT is a registered trademark of Abbott Laboratories.

Any product information in this document should be used in conjunction with the latest version of the
Operation’s Manual and Service and Support Manual. If any discrepancies in information exist within
this document or any other, the latest version of the Operation’s Manual and/or Service and Support
Manual takes precedence.
All samples (printouts, graphics, displays or screens, etc.) are for information and illustration
purposes only and shall not be used for clinical or maintenance evaluations.
Each person assumes full responsibility and all risks arising from use of the information. The
information is presented “AS IS” and may include technical inaccuracies or typographical errors.
Abbott Laboratories reserves the right to make additions, deletions, or modifications to the
information at any time without any prior notification

2005 Abbott Laboratories, Abbott Park IL


200402-102 Sept 2005 Page 2 of 9
ARCHITECT® c8000® System Service Training Handouts
Main Transformer T1

T1 T.F. 200V
Main Switch (rear of P/S) Line 240 VAC
Power 1 1
Filter
Inlet Plug 230 VAC
Power Cable LN L L
0V
I O 220 VAC 2 2
15A
LN N A
E D 208 VAC T.F. 100V
Plug 3 3
200 VAC
4 4
5 5

1 to Emergency 6 6
Refrigerator OFF Button
4 Emergency 7 7
in front 0V
8 8

N
100 VAC

TB1 DISTRIBUTION
1 +5
Power Supply to CNN PCB
1 1 4 +5 to
L 1 +5 7 +5 CPU Master Card 1 +5
2 2 5 VDC Cage
1 1 0
3 3 0 2 GND 2 GND
PS1
2 2 4 4 Master 5 GND 3
V Card
100 VAC N 8 GND 4
3 3 5 5 A Cage
C GND 3
4 4 6 6
7 7 6
Preliminary 9 TBC
Power Switch 8 8
CP2 10A +5 to Heater Assy
–front of 9 9 ISH GND
P/S +5
10 10 +5 1
N GND GND
Connectors on side P/S 11 11 4
Unit Cooler
12 12 3
DISTRIBUTION
CPU 13 13 6
TBC
14 14
+24 DC 2
LAMP 15 15
5
POWER SAVE 16 16
100V~
CPU
Terminal
Block
Connectors on front of P/S 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5
AC/DC Driver

+11.5 Lamp
+12 Pelt

GND
+12

+24

–15
GND

GND

GND

GND

GND
+15
+5
STR Driver
Refrigerator/Emergency

Power Indicator LEDs CNN1 Power Indicator LED PCB CNN2

5 VDC
+11.5 Lamp
-15
+15
+12 Peltier
+24
+12
+5

2005 Abbott Laboratories, Abbott Park IL


200402-102 Sept 2005 Page 3 of 9
ARCHITECT® c8000® System Service Training Handouts
c8000 (>C800223)
Power Supply

Connectors on side P/S


Unit Cooler
DISTRIBUTION
CPU
TBC
+24 DC
LAMP
POWER SAVE Power Indicator LEDs
100V~ +11.5 Lamp

Total +12 VDC


CPU -15
+15
+12 Peltier
Connectors on front of P/S +24
AC/DC Driver +12
STR Driver +5
Refrigerator/Emergency

2005 Abbott Laboratories, Abbott Park IL


200402-102 Sept 2005 Page 4 of 9
ARCHITECT® c8000® System Service Training Handouts
Main Transformer T1

T1 T.F. 200V
Main Switch (rear of P/S) Line 240 VAC
Power 1 1
Filter
Inlet Plug 230 VAC
Power Cable LN L L
0V
I O 220 VAC 2 2
15A
LN N A
E D 208 VAC T.F. 100V
Plug 3 3
200 VAC
4 4
5 5
6 6
7 7
0V
8 8

100 VAC

to
TB1 CPU Master Card
Cage

1 1 1
2 2 4
1 1
3 3 7
STRDRIVER
2 2 4 4 2

3 3 5 5 5 1
100 VAC 8
4 4 6 6 2
7 7 3 +15 +15
3 +15
Preliminary 6 GND GND
Power Switch 8 8 4 GND
–front of 9 9 9 –15 –15
5 –15
P/S
Connectors on side P/S 10 10 Transformer T2
Unit Cooler 11 11 +15 G –15
DISTRIBUTION N
12 12 3.15 A 24 VAC D1 3 2 1
CPU
TBC 13 13 ~ + CNN3 +15
12 12 3 1
+24 DC 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5
14 14 6.3 A 18.3 V 12 ~ - 12 GND
LAMP 6 CNN2 2

+11.5 Lamp
+12 Pelt

–15

GND
+12

+24
GND

GND

GND

GND

GND
to Master

+15
+5
POWER SAVE 15 15 RECTIFIER –15 CPU
1 3 Card Cage
100V~
16 16 3.15A 18 V
CPU 4 1
Terminal 2 1 1 Power Indicator LED PCB 4
CNN1 CNN2
Connectors on front of P/S Block 3.15A 18 V LM
5 2 2 IC1 2
AC/DC Driver 350T
STR Driver CNN1 3 3
OV 5
Refrigerator/Emergency 4 1 3 11.45
LM
DC Power 5 2 IC2 6 GND
Power Indicator LEDs 350T
Supply 6 3
+11.5 Lamp 1 1
-15 7 LAMP +11.45 VDC
+15 2 2
+12 Peltier 8
+24
+12
+5

DC Lamp

2005 Abbott Laboratories, Abbott Park IL


200402-102 Sept 2005 Page 5 of 9
ARCHITECT® c8000® System Service Training Handouts
c8000 (>C800223)
Power Supply

Connectors on side P/S


Unit Cooler
DISTRIBUTION
CPU
TBC
+24 DC
LAMP
POWER SAVE Power Indicator LEDs
100V~ +11.5 Lamp
CPU -15

100 VAC
+15
+12 Peltier
Connectors on front of P/S +24
AC/DC Driver +12
STR Driver +5
Refrigerator/Emergency

2005 Abbott Laboratories, Abbott Park IL


200402-102 Sept 2005 Page 6 of 9
ARCHITECT® c8000® System Service Training Handouts
c8000 (>C800223)
Power Supply

Connectors on side P/S


Unit Cooler
DISTRIBUTION
CPU
TBC
+24 DC
LAMP
POWER SAVE Power Indicator LEDs
100V~ +11.5 Lamp
CPU -15
+15

200 VAC
+12 Peltier
Connectors on front of P/S +24
AC/DC Driver +12
STR Driver +5
Refrigerator/Emergency

2005 Abbott Laboratories, Abbott Park IL


200402-102 Sept 2005 Page 7 of 9
ARCHITECT® c8000® System Service Training Handouts
c8000 (>C800223)
Power Supply

Connectors on side P/S


Unit Cooler
DISTRIBUTION
CPU
TBC
+24 DC
LAMP
POWER SAVE Power Indicator LEDs

+24 VAC
100V~ +11.5 Lamp
CPU -15
+15
+12 Peltier
Connectors on front of P/S +24
AC/DC Driver +12
STR Driver +5
Refrigerator/Emergency

2005 Abbott Laboratories, Abbott Park IL


200402-102 Sept 2005 Page 8 of 9
ARCHITECT® c8000® System Service Training Handouts
11 0507
0507 77 0524 0524 13 0616 19
13 0616 19 11191119 25 25 0603
0603 31 31 0702
0702 CUVETTE
CUVETTE WASHER
WASHER
22 0128
0128 88 1026 1026 14 0202 20
14 0202 20 09240924 26 26 0808
0808 NOZZEL PUMP VACUUM PUMP
33 0908
0908 99 0130 0130 15 1016 21
15 1016 21 07290729 27 27 0418
0418
44 0816
0816 10 10 1006
1006 16 0728 22
16 0728 22 03210321 28 28 0922
0922 1 Cuvette Wash N/A (uses HC Waste Pump)
55 1006
1006 11 11 1107
1107 17 0614 23
17 0614 23 11281128 29 29 1102
1102 2, 3 Wash Solution 2
66 0222
0222 12 12 0513
0513 18 0416 24
18 0416 24 07010701 30 30 0221
0221
4, 5, 6 Cuvette Wash 2
Note: date codes must be used in accordance with information contained in the
ARCHITECT c8000 System Service and Support Manual P-201 7, 8 N/A 1
REAGENT PREP TEMPERATURE

SOLUTION PREPARATION STABILITY AREA DETECTOR LOCAL RANGE Procedure AC/DC CONT. POT**

0.5% 995 mL Reagent Grade Type II Water Same as on Water Bath Bottom Water Bath at 36.8-37.20C P-106 VR5
Acid Wash 5 mL Acid Wash Acid Bottle R2 Bath additive area

R1 Front of Carousel 2-100C P-102 VR1


10% 90 mL Reagent Grade Type II Water 14 Days
Detergent B 10mL Detergent B R2 Front of Carousel 2-100C P-103 VR2
LLS VOLTAGE CARD CAGE BOARDS Samp. Crsl Left side of Carousel 5-200C P-104 VR3
Test Voltage Potentiometer Ambient On Heater Assembly 15-300C P-105 VR4
Point
AC/DC CONTROLLER

**NOTE: Adjustment of AC/DC Cont. Pot only affects the DISPLAYED temperature and
SMC
DAQ

CPU
PM

B6 0.3VDC VR4 does not change the AREA temperature output


B10 0.4VDC VR3 CHECK Heating Devices Valves CONCENTRATED
SYRINGE ASSAYS
T6 0.3VDC VR6 VALVE
Temp Detector Fluidics Valves
T10 0.4VDC VR5 ICT Orange Urea, CO2, Phos,
Heating Element ICT Aspiration Valve Alkaline Phos
A6 0.5VDC VR1 Wash Solution Blue
Thermal Fuse ICT DATE CODE (JUNE 7 2001)
A10 2.0VDC VR2
A A 1 06 07 30
A11 -0.5VDC VR7 ICT Ver yr Mo day SN

ARCHITECT c8000 S Y S T E M NOTE: Information applies to SN/>c800223 and above

MAIN POWER SUPPLY MAIN POWER SUPPLY

2005 Abbott Laboratories, Abbott Park IL


ARCHITECT® c8000® System Service Training Handouts 200402-102 Sept. 2005 Page 9 of 9