Professional Documents
Culture Documents
Drawings: Model ZR202G - L Integrated Type Zirconia Oxygen / High Temperature Humidity Analyzer
Drawings: Model ZR202G - L Integrated Type Zirconia Oxygen / High Temperature Humidity Analyzer
Ø123
170
122 125
Flange
L 338 to 351
t
Rc1/4 or 1/4NPT
Reference gas inlet
C
ØA
Ø50.8
25
ØB
Rc1/4 or 1/4NPT
Calibration gas inlet
Flange
153 to 164
2-G1/2,2-1/2NPT etc.
252 to 265 49 Cable connection port
Unless otherwise specified,differences in the dimensions are specified as: General tolerance = 6 (Criteria of tolerance class it18 in JIS BO401-1986) - 2.
P-
2
48.5
Ø123
170
Flange 122 125
L 342±4
t
Rc1/4 or 1/4NPT
Reference gas inlet
C
*1 Ensure that the flange gasket
ØA does not block the reference gas outlet
on the detector flange.
25
Flange Weight
Model, Code L PIPING (kg)
Specification A B C t
ZR202G-040--L 400 Approx. 9
ZR202G-070--L 700 Approx. 10
ZR202G-100--L 1000 Approx. 11
1500 JIS 10K-65-FF 175 140 4-Ø19 18 A
ZR202G-150--L Approx. 13
ZR202G-200--L 2000 Approx. 15
ZR202G-250--L 2500 Approx. 16
ZR202G-300--L 3000 Approx. 18
HOOD (Option / H)
274 ±4 150 ±3
150 ±3
Unless otherwise specified,differences in the dimensions are specified as: General tolerance = 6 (Criteria of tolerance class it18 in JIS BO401-1986) - 2.
SD 11M12B02-08E 2/3
1st Edition : Oct. 2000 (YK)
P- 4th Edition : Feb. 2015 (YK)
3
• Wiring Diagram
Integrated type Zirconia Oxygen/High Temperature Humidity Analyzer
Contact input 2
1 2 3 4 5 6 7
DI-1 DI-2 DI-C DO-1 DO-1 DO-2 DO-2
8 9 10 11 12 13 14
FG AO AO L N G FG
(+) (-)
The protective grounding for the analyzer shall be connected either the protective
ground terminal in the equipment or the ground terminal on the case.
Standard regarding grounding: Ground to earth, ground resistance: 100Ω or less.
Unit : mm
• Specified by Option Code
Approx. 44
Approx. 100
40 Ø48
19
Full open height
Calibration gas inlet 58
Rc1/4 or 1/4NPT Calibration gas inlet Rc1/4 or 1/4NPT
With stop valve (option : /SV) weight : Approx. 200g Check valve (option : /CV) weight : Approx. 90g
Unless otherwise specified,differences in the dimensions are specified as: General tolerance = 6 (Criteria of tolerance class it18 in JIS BO401-1986) - 2.
SD 11M12B02-08E 3/3
1st Edition : Oct. 2000 (YK)
4th Edition : Feb. 2015 (YK)
P-