Magmeter Accuracy Graph

+/-% of Rate Accuracy

4.0

3.5

3.0

2.5

2.0

1.5

1.0

0.5

0.0
0.000 100.000 200.000 300.000 400.000 500.000 600.000 700.000 800.000 900.000 1000.000

Column C

GENERAL DATA
Project: APDR
Tag No: FT-200-A
Model Numbers
Transmitter: 8732EMT1A1
Flowtube 8705TSA060C1W0B3Q4DT
% Flow USGPM ft/sec Accuracy(%) % Flow USGPM ft/sec Accuracy (%)
100% 900.000 9.99 0.30 30% 270.000 3.00 0.36
95% 855.000 9.49 0.30 25% 225.000 2.50 0.38
90% 810.000 9.00 0.30 20% 180.000 2.00 0.41
85% 765.000 8.50 0.31 15% 135.000 1.50 0.47
80% 720.000 8.00 0.31 10% 90.000 1.00 0.58
75% 675.000 7.50 0.32 9% 81.000 0.90 0.61
70% 630.000 7.00 0.32 8% 72.000 0.80 0.66
65% 585.000 6.50 0.30 7% 63.000 0.70 0.72
60% 540.000 6.00 0.30 6% 54.000 0.60 0.80
55% 495.000 5.50 0.31 5% 45.000 0.50 0.91
50% 450.000 5.00 0.32 4% 36.000 0.40 1.07
45% 405.000 4.50 0.32 3% 27.000 0.30 1.34
40% 360.000 4.00 0.33 2% 18.000 0.20 1.89
35% 315.000 3.50 0.34 1% 9.000 0.10 3.53
This report is provided according to the terms and conditions of the Instrument Toolkit(TM) End-Use Customer License Agreement.