AMTG-2 Accurate Magnetic Thickness Gauge

Using Manual

AT2E
6 rue des Cours Neuves
ZA de la peupleraie
77135 PONTCARRE, FRANCE
Tel: +33 01 64 66 03 02
Fax: +33 01 64 66 02 98
E-mail: info@at2e.com

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

Content
1. Introduction ..................................................................................................... 1

2. Technical data .................................................................................................. 1

3. Structure........................................................................................................... 1

4. Interfaces and connection ............................................................................... 2

5. Accessories ....................................................................................................... 3

6. Functional buttons .......................................................................................... 4

7. Magnetic Probe ................................................................................................ 4

8. Operation ......................................................................................................... 5

9. Single-Point calibration .................................................................................. 7

10. Multi-Point calibration ................................................................................... 9

11. Screens and functions ................................................................................... 17

12. Precaution for measures ............................................................................... 20

13. Communication setting ................................................................................. 22

14. Maintenance................................................................................................... 22

15. Packing list ..................................................................................................... 23

16. Optional parts ................................................................................................ 23

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

1. Introduction
AMTG-2 Accurate Magnetic Thickness Gauge adopts the magnetic probe sensing technology. It’s used for

performing the non-destructive thickness measuring of non-magnetic materials, such as plastic, glass,

aluminum, titanium, copper etc.

Basing on the principle of Hall Effect, with different sizes of steel balls and analyzing the different signal

strength feedback from the balls, the accurate measurement can be achieved.

2. Technical data
Dimension (mm): 210 x 150 x 65 (L x W x H)

Net Weight (kg) 2


Main supply (V): DC 8.4
Measuring range (mm): 0.01-6.00

3. Structure

1. Mainframe 2. Magnetic sensing probe 3. Probe support

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 1
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

4. Interfaces and connection

1 2 3 4 5

7 6 8
1) Probe cable socket (mainframe)

2) RS232 communication interface

3) Foot switch socket

4) Power socket (DC 8.4V power supply)

5) Power switch

6) Probe cable plug, for connecting to mainframe

7) Probe cable plug, for connecting to probe

8) Probe cable socket (probe)

Connect the mainframe and probe with the probe cable, please pay attention to the sockets for mainframe and

probe, it is different.

Notes: Before leaving the factory, the instrument has been well calibrated basing on a complete set of

mainframe, probe cable and magnetic probe. And they were labeled with a same serial number. So please

match the numbers when using. In case some parts are changed, it may cause the data inaccurate.

Therefore, the instrument needs to be performed a multi-point calibration for correcting the data.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 2
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

5. Accessories

Steel Balls Includes: 0.187" (4.76 mm) ball, 0.125" (3.18 mm) ball, 0.063" (1.59 mm) balls, three diameters

totally. Corresponding measure range is as below table:

Tolerance of Tolerance of Recommended


Steel ball Dia. Min. surface radius
Single-point calibration multi-point calibration Measuring range

0.187" (4.76 mm) 3% ±0.003 mm 1% ±0.003 mm 0.094 in (2.38 mm) 0.00 – 6.00 mm

0.125" (3.18 mm) 4% ±0.003 mm 2% ±0.003 mm 0.063 in (1.59 mm) 0.00 – 4.00 mm

0.063" (1.59 mm) 5% ±0.003 mm 3% ±0.003 mm 0.031 in (0.79 mm) 0.00 – 2.00 mm

With cover Without cover

Standard block gauge includes: 0.230 mm / 0.500 mm / 1.000 mm / 2.000 mm / 3.000 mm / 4.000 mm / 5.000

mm (7 pcs totally) (The following block gauges could be ordered if necessary: 1.500 mm / 2.500 mm / 3.500

mm / 4.500 mm / 5.500 mm)

Block gauge accuracy: 1 µm

Without steel ball With steel ball

Steel ball support include: 1.59 mm ball support, 3.18 mm ball support, 4.76 mm ball support. Please match

the corresponding steel ball when measuring.

Other accessories includes foot switch, power supply.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 3
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

6. Functional buttons

1) SETUP button: To enter the setting menu

2) Q-CAL button (Demagnetizing button): To clear the residual magnetism of the probe, same to the Q-CAL

button on the probe.

3) CAL button: To enter the calibration function.

4) CHECK DATA button: To check the saved data.

5) SAVE button: Save data, save the current data by pressing this button.

6) SEND button: Send out data, send out the current data by pressing this button.

7) MIN button: To display the minimum value.

8) MAX button: To display the minimum value.

9) GRAPH Button: To display the graph, minimum value, maximum value and average value.

10) button: To activate the “Delayed start & Auto-stop” function.

11) button: To stop the “Delayed start & Auto-stop” cycle.

7. Magnetic Probe

Magnetic sensing tip Demagnetizing button

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 4
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

8. Operation
1) Well connect the mainframe and probe, switch on the instrument.

2) After the screen display the measurement information, choose the proper steel ball and put it into the

corresponding steel ball support for preparing the single point calibration. (Caution: please don’t put the

steel ball support on the probe at this moment)

3) Press the “CAL” button on the panel.

4) And the “BALL OFF” marking will show on the screen. It means to press the “CAL” button again to

confirm when the probe is in no-load status, so that the mainframe can detect the data when the probe is

no-load.

5) Then the “BALL ON” marking will show on the screen, it means to put the steel ball (the type that is

going to be used for measuring) onto the probe. Now please put the steel ball support together with the

steel ball onto the probe gently (in avoiding the sensing tip damaged by fierce impact). The steel ball will

be jack-up a little by the probe. That means the probe and steel ball are in contact. Then press the “CAL”

button to confirm, and the system will return to the measuring screen automatically. Please take out the

steel ball support from the probe at this moment.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 5
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

6) After entered the measuring screen, it is ready to perform the measurement. To carry out a measurement,

simply place the probe tip on one side of the material, then place the steel ball (the one that calibrated in

the previous operation) on the other side near the probe tip. The ball will be attracted to the probe tip by

magnetic field. Once the ball is within range, the system will begin to display thickness values providing

the tip and ball are both in contact with the material. The value display on screen is the actual value. Or to

recheck the accuracy of the instrument, please use the steel ball to measure the equipped standard block

gauge to compare the value.

*Measuring the sample

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 6
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

*Measuring the 3.000 mm standard block gauge to recheck the instrument

Notes: During the measuring, the steel ball and magnetic probe will bring some magnetism to the probe

tip, which will affect the measures. Therefore, please do not perform the measure in a rapid and

continuous way. Please wait for 5 seconds before measuring a new sample or changing a new standard

block gauge. Or press the “QCAL” button located on the side of the probe to demagnetize.

9. Single-Point calibration
The Single-Point calibration, it is to let the instrument to detect the zero point when the probe is in no-load

status and the steel ball type that is being used (whether it is 4.76 mm, 3.18 mm or 1.59 mm steel ball). So the

Single-Point calibration is not a linear one, and it does not lead to any linear correction.

This is a simple and convenient way of operation. Granted that a wrong calibration has been done, it just needs

to simply do it again to correct it with a few steps in couple seconds. So generally, it is suggested that users to

operate in this mode.

Notes: During the measuring, the steel ball and magnetic probe will bring some magnetism to the probe

tip, which will affect the measures. Therefore, please do not perform the measure in a rapid and

continuous way. Please wait for 5 seconds before measuring a new sample or changing a new standard

block gauge. Or press the “QCAL” button located on the side of the probe to demagnetize.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 7
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual
Press the “CAL” button on front panel to enter the Single-Point calibration screen.

1) On the screen, it shows the “Calibration Step 01”. Please press the “CAL” button to confirm when the

probe is in no-load status.

2) Then the screen will enter step 02, please put the steel ball onto the ball support and put them together onto

the probe gently and press the “CAL” button to confirm.

3) After the calibration is done, system will return to the measuring screen automatically. And the screen

should display “0.000 mm”

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 8
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

4) Use the steel ball which has been calibrated just now for the measures. Measuring range refers to the

below table:

Recommended measuring range after Accuracy of single-point


Steel ball dia.
single-point calibration calibration

0.187" (4.76 mm) 0.00 – 6.00 mm 3% ±0.003 mm

0.125" (3.18 mm) 0.00 – 4.00 mm 4% ±0.003 mm

0.063" (1.59 mm) 0.00 – 2.00 mm 5% ±0.003 mm

10. Multi-Point calibration


Multi-Point calibration is an advanced calibration mode that base on the Single-Point calibration, to build up a

detail linear section, divided the measure range up to 13 segments and calibrates them respectively, which to

make the whole data section highly accurate.

Standard block gauge thickness (units: mm)


Ball Spec.
0.000 0.230 0.500 1.000 2.000 3.000 4.000 5.000

1.59 mm A1 B1 C1 D1 E1

3.18 mm A2 B2 C2 D2 E2 F2 G2

4.76 mm A3 B3 C3 D3 E3 F3 G3 H3

Please perform the multi-point calibration according to the above table and select the corresponding steel ball
to measure the sample within the corresponding measuring range as below table. Please note that the accuracy
will not be ensured if the measurement is out of measuring range as below table. (Caution: the accuracy is
only basing on the correct method of the multi-point calibration. Otherwise, the accuracy may not be
ensured. Please contact us for advice if there is any doubt of the multi-point calibration process.)

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 9
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

Recommended measuring range after Accuracy of multi-point


Steel ball dia.
multi-point calibration calibration

0.187" (4.76 mm) 0.00 – 6.00 mm 1% ±0.003 mm

0.125" (3.18 mm) 0.00 – 4.00 mm 2% ±0.003 mm

0.063" (1.59 mm) 0.00 – 2.00 mm 3% ±0.003 mm

The calibration data is saved in the system, only when the Multi-Point calibration is performed will the data of

A-H be re-written to the table. Referring to the table data, system will detect the strength of magnetism and

convert to the corresponding thickness value.

Therefore, when performing the Multi-Point calibration, it MUST be very careful. If there is any mistake

during the process, it will cause the data error to the table and finally lead to the inaccurate measurement.

Generally, the above data table will be used when performing the single-point calibration. Performing the

multi-point calibration is attended to modify the data of the above table. If the measurement is accurate in the

Single-Point calibration mode, there is no need to perform the Multi-Point calibration. Unless there is a large

deviation occurred in the Single-Point calibration mode, it is necessary to perform the Multi-Point calibration

for rebuilding the table data.

In addition, the Multi-Point calibration to every type of steel ball is independent from each other. That means

when calibrating the 4.76 mm steel ball, it will only rewrite the A3, B3, C3, D3, E3, F3, G3, H3 data, while the

data of 3.18 mm and 1.59 mm steel balls will not be changed.

When finishing a Multi-Point calibration, user can use the standard block gauges to perform a comparison test

to check whether the calibration data has been well set.

More standard block gauges could be ordered to build up a more accurate linear section, such as: 1.500 mm,

2.500 mm, 3.500 mm, 4.500 mm, 5.500 mm standard block gauges. With these additional standard block

gauges, the multi-point calibration shall be carried out from the thin standard block gauge to the thick ones.

For example, when performing the multi-point calibration with 4.76 mm magnetic steel ball, please use the

standard block gauges as the following sequence: 0.230 mm, 0.500 mm, 1.000 mm, 1.500 mm, 2.000 mm,

2.500 mm, 3.000 mm, 3.500 mm, 4.000 mm, 4.500 mm, 5.000 mm, 5.500 mm. Including the 0.000 mm data,

the maximum data which will be written into the data table is 13.

Notes: During the measuring, the steel ball and magnetic probe will bring some magnetism to the probe

tip, which will affect the measures. Therefore, please do not perform the measure in a rapid and

continuous way. Please wait for 5 seconds before measuring a new sample or changing a new standard

block gauge. Or press the “QCAL” button located on the side of the probe to demagnetize.
AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 10
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual
Procedure of Multi-Point calibration (For example: 4.76 mm steel ball. According to the above table, the

following standard block gauges are needed: 0.230 mm / 0.500 mm / 1.000 mm / 2.000 mm / 3.000 mm /

4.000 mm / 5.000 mm.)

1) Press the “SETUP” button on the front panel to enter the menu and select “CALIBRATION” to go ahead.

2) In the “CALIBRATION” screen, press option “MULTIPOINT” to change its status from “OFF” to “ON”,

which means to activate the Multi-Point calibration mode. Since the Multi-Point calibration mode is not a

frequent-used mode and in case the users forget to turn it off after the calibration, the default option of

“MULTIPOINT” is “OFF”. When the instrument is power off and restarted, the status of option

“MULTIPOINT” will be “OFF”. If users need to perform the Multi-Point calibration, please activate it in

the menu.

3) Press button “ESC” to return to the measuring screen.

4) After returning to the measuring screen, press “CAL” button to enter the Multi-Point calibration screen.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 11
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

5) In the “BALL OFF” screen, please press the “CAL” button to confirm when the probe is in no-load status.

6) When screen shows the “BALL ON” marking, please put the steel ball onto the ball support and put them

together onto the probe gently and press the “CAL” button to confirm. It equals to set the probe zero and

input the A3 data on the table. After pressing the “CAL” button, please take out the ball support from the

probe.

7) After entering the step 3, it will need to use the standard block gauges for calibrating. Regarding to the

4.76mm steel ball, we offer the 0.230 mm standard block gauge as the first standard block gauge. If users

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 12
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual
have their own qualified block gauges, they may replace the supplied one. The calibration principle is to

measure from the thinnest one to the thickest one, the data order must be from small to large.

8) Put the standard block gauge onto the probe gently, then put the steel ball into the standard block gauge.

The “Measure Value” will display the value before calibration. As per below, corresponding to the 0.230

mm standard block gauge, the “Measure Value” is 0.251 mm which is far larger than 0.230 mm. So, it has

to be corrected. Please press the right side of “Setting Value” (yellow frame area of below photo) and use

the numeric keyboard on the right side of screen to input 0.230 mm then press button “ENTER” to

confirm. Then press “CAL” button to save the setting and move to next step automatically. It equals to

input the B3 data on the table.

9) Follow the above procedure and continue to set the step 4 for 0.500 mm standard block gauge. Press

“CAL” button to move to next step after the setting is done. It equals to input the C3 data on the table.

10) Follow the above procedure and continue to set the step 5 for 1.000 mm standard block gauge. Press

“CAL” button to move to next step after the setting is done. It equals to input the D3 data on the table.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 13
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

11) Follow the above procedure and continue to set the step 6 for 2.000 mm standard block gauge. Press

“CAL” button to move to next step after the setting is done. It equals to input the E3 data on the table.

12) Follow the above procedure and continue to set the step 7 for 3.000 mm standard block gauge. Press

“CAL” button to move to next step after the setting is done. It equals to input the F3 data on the table.

13) Follow the above procedure and continue to set the step 8 for 4.000 mm standard block gauge. Press

“CAL” button to move to next step after the setting is done. It equals to input the G3 data on the table.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 14
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

14) Follow the above procedure and continue to set the step 9 for 5.000 mm standard block gauge. Press

“CAL” button to move to next step after the setting is done. It equals to input the H3 data on the table.

Since the 5.000 standard block gauge is the final one for 4.76 mm steel ball, please press button “ESC” to

quit the multi-point calibration process and be back the main interface.

15) Return to the measuring screen.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 15
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual
Notice:

*Regarding to other dia. Steel ball, the multi-point calibration process is same as the above. Please just

pay attention to use the corresponding standard block gauges as recommendation.

*For each dia. of steel ball, up to 12 standard block gauges could be used to perform the multi-point

calibration. It will return to the main interface automatically after finishing calibrating the 12th

standard block gauge.

*Please remember to set the “MULTIPOINT” to be “OFF” in menu after multi-point calibration. This

is to let the instrument to be back to single-point calibration status. Otherwise, the instrument may go

to multi-point calibration again when pressing the “CAL” button.

*Please perform the multi-point calibration in the below situation:

A. There is a big deviation between the measuring value and standard value.

B. In case of the probe has been changed, dropped or touched the strong magnetic object.

C. In case of the probe has been abrasive wear due to long time of using.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 16
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

11. Screens and functions

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 17
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual
1) SETUP MENU: Main menu

- Entering menu, press the “SETUP” button in the measuring screen to enter the “SETUP MENU”

2) MEASUREMENT: Measurement setting

- UNIT: select units (optional for mm or inch)

- ID NUMBER: the serial number of saved sample. For setting the serial number of saved samples

which display on the measuring screen. For example, if users want the serial number started with 0001,

please set this data as 0001, then the system will save the samples with serial number from 0001.

- RESOLUTION: Accuracy selection. Optional for “HIGH” with resolution 0.001 mm and “LOW” with

resolution 0.01 mm

3) GENERAL: Common setting

- BRIGHTNESS: Set the brightness of the screen from 0-100. The screen is brighter when the setting is

higher. Default setting is 60.

- FOOTSWITCH: foot switch, optional for “SAVE” (save data), “SEND” (send out data via

communication port) and “CYCLE” (to start and stop the “Delayed start & Auto-stop” mode)

- SAVE/SEND: Primary (data of primary window);Secondary (data of secondary window). Generally,

the data of primary window is real time value, and the data of secondary window is minimum value. If

selected “Primary” in this option, when press the “SAVE” or “SEND” button, system will save or send

out the data in primary window, likewise, if selected “Secondary” in this option, system will save or

send out the data in secondary window.

- CHECK DATA: Review the saved data.

4) CALIBRATION: Calibration setting

- AUTO QCAL: automatic demagnetize. “ON” means is activate, “OFF” means deactivate. During the

measuring, the magnetism on the probe will keep accumulating which may cause the measure value

keep shifting. By activating this function, the system will auto demagnetize in a fixed period when

there is no ball on the probe tip. This function is the same as pressing on the “QCAL” button located

on the side of the probe

- QCAL interval: auto demagnetize interval which could be set from 1-999 seconds. Default setting is 5

seconds. It means if there is no measurement over 5 seconds, the system will perform the

demagnetizing automatically for one time, to ensure the probe will work in the best status.

- MULTIPOINT: Multi-Point calibration. “ON” means is activate, “OFF” means deactivate. It was

defaulted as “OFF” in the system, every time when the instrument is power off and restarted, the

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 18
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual
option will be changed to “OFF” automatically.

5) DATABASE RESET: Erase the data.

- Before carry out this operation, please consider it carefully. After reset, all the settings will return to

default factory setting. The default settings may not be the best setting for measurements in all

conditions. So after reset, please perform the Multi-Point calibration for all three types of steel balls to

let the instrument achieve the best result. Normally, it is not recommended to use this function.

(Password to confirm reset: 8888)

6) DIFF/ALARM: Correction and alarm setting

- DIF_ENABLE: value difference function. Defaulted as “OFF”. The following “REF_VALUE” will be

available only when the “DIF_ENABLE” function set to “ON”. The measure value will compare to

the setting value in “REF_VALUE” for a value difference comparison.

- REF_VALUE: reference value for making the value difference comparison. Defaulted as “0.000”. For

example, if set it as 1.000, and the measure value is 2.500 mm, then the screen will display 1.500 mm

which is the difference between reference value and measure value.

- ALARM_ENABLE: activate or deactivate the alarm. “ON” means is activate, “OFF” means

deactivate.

- LOW_VALUE: lower alarming value which could be set from 0.000-9.999 mm. When the measure

value is lower than this value, system will sound the alarm and the color of value display will be

changed from green to yellow.

- HIGH_VALUE: upper alarming value which could be set from 0.000-9.999 mm. When the measure

value is higher than this value, system will sound the alarm and the color of value display will be

changed from green to red.

7) DIAGNOSTIC: Data diagnosis, default setting is “OFF”.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 19
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual
- Use for assisting to judge the magnetism strength of the probe when failure occurs. This function is

only applicable to the manufacturer or authorized reseller for assisting to services. When the setting is

“ON”, there are 3 lines of data on the left side of main screen. They are used to indicate the magnetism.

By the changing of these 3 lines of data, the status of the probe could be estimated and judged.

12. Precaution for measures


1) Correct measure:

2) The concave position is too deep that the probe tip cannot contact to the material surface, measure data

will be inaccurate, and larger than the actual thickness.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 20
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual
3) The steel ball run into the obstacle and cannot move to the right position.

4) The steel ball encountered the bending position, cannot be attracted by the probe.

5) The probe tilts and results in the measure data inaccurate.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 21
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

13. Communication setting


AMTG-2 has the data output function. Press “SEND” button to send out the data to PC or data collection

device, its communication protocol is as below:

1) Baud rate: 9600

2) Data bit: 8

3) Check bit: none

4) Stop bit: 1

14. Maintenance
1) The instrument is using the magnetic sensing as measuring medium, so the storing and operating

environment of the instrument has a strict requirement for the magnetism. In avoiding the interference,

the instrument cannot store or operate in the environment that has strong magnetism, or in the place that

has the possibility to create electromagnetic effect, like the nearby of large motor, magnet, loudspeaker

and other instruments that with magnetism.

2) Probe operation. The probe tip is very important component, during the daily measure, please do not drag

the sample fiercely when it contacting with the probe tip, in case of the sample scratches on the tip. When

moving the samples, the better way is to lift up the object, then move to the target position, so that to

reduce the friction on the tip and extend its service life.

3) Inside the mainframe, there is a built-in rechargeable lithium battery, please ensure there is sufficient

battery power to support the instrument during the daily operation. Generally, it is suggested that to use

the battery when measuring, after the measure is done, connect the charger to recharge the battery. By this

means, it is able to avoid the interference from unstable external electric source, so that avoid the

interference to the measures. If the instrument is not using for a long time, it is recommended that to

switch on the instrument for a least half an hour each month, and recharge the battery appropriately to

activate it, in avoiding the battery failure due to the long time storing.

4) The instrument is not designed to water proof. Please keep the instrument away from water source when

operating or storing to avoid any damage caused by water.

AT2E www.at2e.com
2-6 Rue des cours neuves, ZA – 77135 PONTCARRE FRANCE info@at2e.com Tél. : +33 (0)1 64 66 03 02 22
AMTG-2 Accurate Magnetic Thickness Gauge
Using Manual

15. Packing list


Part No. Description Qty. Remark

AMTG-01 Mainframe 1 set

AMTG-02 Probe 1 pc

AMTG-03 Probe cable 1 pc

AMTG-04 Probe support 1 pc

AMTG-05 8.4V 1A Power supply charger 1 pc

AMTG-06 Foot switch 1 pc

AMTG-07-1.59 Steel ball 1.59 mm 1 box

AMTG-07-3.18 Steel ball 3.18 mm 1 box

AMTG-07-4.76 Steel ball 4.76 mm 1 box

AMTG-08-1.59 Steel ball support 1.59 mm 1 pc

AMTG-08-3.18 Steel ball support 3.18 mm 1 pc

AMTG-08-4.76 Steel ball support 4.76 mm 1 pc

AMTG-09-0.230 Standard block gauge 0.230 mm 1 pc

AMTG-09-0.500 Standard block gauge 0.500 mm 1 pc

AMTG-09-1.000 Standard block gauge 1.000 mm 1 pc

AMTG-09-2.000 Standard block gauge 2.000 mm 1 pc

AMTG-09-3.000 Standard block gauge 3.000 mm 1 pc

AMTG-09-4.000 Standard block gauge 4.000 mm 1 pc

AMTG-09-5.000 Standard block gauge 5.000 mm 1 pc

\ Using manual 1 pc

16. Optional parts


Part No. Description Qty. Remark

AMTG-09-1.500 Standard block gauge 1.500 mm 1 pc

AMTG-09-2.500 Standard block gauge 2.500 mm 1 pc

AMTG-09-3.500 Standard block gauge 3.500 mm 1 pc

AMTG-09-4.500 Standard block gauge 4.500 mm 1 pc

AMTG-09-5.500 Standard block gauge 5.500 mm 1 pc

AMTG-10 “QUALIMESURE” Software 1 pc