ANALEXrs

On-line Oil Condition Monitoring


The Ultimate Gearbox Oil Condition Monitoring Solution

• ANALEXrs (remote )sensor)


• Total ferrous debris sensor
• Oil condition sensor
• Moisture sensor
• Particle content sensor

Real-time measurements From multiple locations


With flexible output options
ANALEXrs Particle Content Sensor

Measurement Principle
• Magnitude and direction of magnetic flux measured as metallic particles move
through a magnetic field
• Metallic particles passing through a magnetic field cause a spike in the field
• The number of spikes, their magnitude and direction are translated into particle type,
size and count
• Unique, patented technology
• Rugged & IP65 rated
• Flow 1.3 to 21 l/min

Performance
• Detects:
• >40µm (Fe)
• >120µm (Metallic Non-Fe)
• -20°C to 80°C (oil temp)
ANALEXrs Particle Content Sensor

Particle Size Validation


• “Worst Case” Sperical Particles Tested

Particle “Thin Edge” through field = Hardest Detection

Steel Gear and Bearing Wear Particles


(Mag. X200)

Particle “Thick Edge” through field = Easiest Detection

Spherical Particles give Worst Case.


72µm Sphere Vs 25µm platelet
Sensor Classified to Spheres
ANALEXrs Particle Content Sensor

Particle Size Validation Particle Material and Size Classification

• Precision made, spherical particles of Peak


Material Distribution /µm
different materials P8-01172-1 SS420 44
P8-01172-2 SS420 105
• Particles optically characterised to accurately P8-01172-3 W 149
P8-01172-4 Misc Metallic 44
determine size P8-01172-5 Cu 106
P8-01172-6 Cu 150
P8-01172-7 Cu 250
P8-01172-8 Bronze 500

P8-01172-2 2 P8-01172-2 1
ANALEXrs Particle Content Sensor

Particle Size Validation


• Particle manipulated into test rod
• Test rod material selected for magnetic
properties close to lubricating oils
• Particle sealed into rod using epoxy
ANALEXrs Particle Content Sensor

Particle Size Validation


• Test rods moved through sensor on test rig
• Output to Oscilliscope and PC

1, 0, 0, 1, OK, 4956, 0, 0, 0, 0, 0, 0, 0, 0, 0, 2, 1.35


1, 0, 1, 1, OK, 4960, 1, 0, 0, 0, 0, 0, 0, 0, 0, 2, 1.69
ANALEXrs Particle Content Sensor
ANALEXrs Particle Content Sensor

Spark Erosion

Fatigue Fatigue debris showing pitted surface and radial cracks


ANALEXrs Particle Content Sensor

Diamond Indent

Particle showing Fatigue Striations


ANALEXrs Particle Content Sensor

3D Surface Measurements