18 inch Combo CALIBRATION REPORT
TABLE OF CONTENTS
Page 2 of 19
18 inch Combo CALIBRATION REPORT
Page 3 of 19
18 inch Combo CALIBRATION REPORT
1.1. Purpose
To ensure that specified and declared tool accuracy and detection values are
met during the actual in-line pipeline inspection.
1.2. Scope
This procedure is applicable to working processes of QA/QC, Production, Data
Evaluation and Project Departments.
1.3. Responsibility
QA/QC, Production, Data Evaluation Department Managers are responsible
for implementation of this procedure.
1.4. Definitions
Page 4 of 19
18 inch Combo CALIBRATION REPORT
∙ Connection cables;
∙ Pulling rope;
∙ Hoisting facilities;
∙ Accessories;
∙ Metal processing tools;
∙ Measurement instruments including handheld UT scanner.
1.6. Procedure
Pull-Through Test is carried out by pulling the magnetic flux leakage or
geometry inspection vehicle through the test pipe by means of a hydraulic or
manual winch. The tool is moved at a constant velocity of 0.3-3.0 m/sec. On
the test pipe, artificial models of defect types of different sizes are applied.
Dimensions of the defects are accurately measured. The tool is switched to
operating mode during the test and after the test, the received data is
analyzed.
Geometry tool is calibrated on a stand-by means of a special calibration ruler
in a static condition for building calibration curves.
1.7. Preparation
Joint measurements of all the test pipe anomalies provided may be checked
by client representative with our QA/ QC personnel, with inspection tools and
instruments, prior to Pull through Test and to be recorded.
Operator downloads the data to PC, records on a portable memory unit and
handed over to Evaluation department for analysis.
Page 5 of 19
18 inch Combo CALIBRATION REPORT
1.11. QA/QC
QA/QC department representative controls the whole process and
documentation is checked and approved by the QA/QC Manager. Pull through
tests are obligatorily attended by QA/QC Production representative. Pull
Through Test & Calibration reports are prepared and maintained by the
Evaluation and Projects Department.
Page 6 of 19
18 inch Combo CALIBRATION REPORT
Page 7 of 19
18 inch Combo CALIBRATION REPORT
Page 8 of 19
18 inch Combo CALIBRATION REPORT
Pull Trough Test was conducted in a spool (test pipe) with artificial defects
according to the drawings attached in Chapter 2.
Magnetization level in test pipe with 9.5 mm wall thickness reached 20 kA/m.
According to the POF specification, this level of pipe wall magnetization guaranteed
good sensitivity and accuracy of the MFL ILI tool.
For the PTT run the defects will be analyzed against actual parameters.
MFL log presents PTT run data. The vertical axis on the log represents degree
position of the defects (0-360 degree corresponding to 12:00 o’clock clockwise
positions). The horizontal axis on the logs represents recorded distance aligned
with the starting of the PTT pipe - first spool.
Page 9 of 19
18 inch Combo CALIBRATION REPORT
Page 10 of 19
18 inch Combo CALIBRATION REPORT
Page 11 of 19
18 inch Combo CALIBRATION REPORT
Page 13 of 19
18 inch Combo CALIBRATION REPORT
1 5 9 13 17 25 29
33 35 38
2 6 10 14 18 26 30
37
21
22 27 31
3 7 11 15 19
23
34 36 39
24
4 12 20 28 32
8 16
Page 14 of 19
18 inch Combo CALIBRATION REPORT
1 5 9 13 17 25 29
33 35 38
2 6 10 14 18 26 30
37
21
27 31
3 7 11 15 19 22
23
34 36 39
24
4 12 20 28 32
8 16
Page 15 of 19
18 inch Combo CALIBRATION REPORT
The Calibration Test was conducted on a stand. The table represents the dependence of the signal value on the arms’
deviation.
L (mm) Sen1 Sen2 Sen3 Sen4 Sen5 Sen6 Sen7 Sen8 Sen9 Sen10 Sen11 Sen12 Sen13 Sen14 Sen15 Sen16
245 519 406 293 248 264 270 187 214 230 230 281 337 303 272 484 266
240 679 567 438 384 421 418 344 389 392 384 437 465 483 429 633 406
235 808 712 589 539 552 557 471 527 518 514 584 584 621 585 770 538
230 949 845 714 663 683 682 594 662 646 648 716 729 776 696 892 660
225 1072 971 838 780 821 802 724 787 768 766 828 856 919 855 1006 780
220 1185 1085 943 898 941 929 847 914 873 874 940 960 1030 965 1136 891
215 1301 1203 1052 987 1055 1031 963 1012 972 985 1051 1094 1153 1085 1261 1002
210 1414 1315 1159 1107 1175 1136 1065 1126 1082 1085 1139 1119 1271 1195 1326 1102
205 1520 1419 1257 1204 1267 1231 1165 1226 1191 1186 1246 1298 1382 1279 1430 1196
200 1622 1508 1359 1299 1370 1329 1257 1349 1283 1295 1337 1361 1488 1384 1541 1299
195 1727 1621 1456 1394 1490 1424 1362 1438 1373 1383 1428 1446 1594 1481 1621 1394
190 1824 1718 1547 1489 1587 1521 1454 1537 1480 1486 1520 1581 1696 1592 1718 1490
185 1920 1813 1640 1568 1676 1610 1560 1629 1562 1572 1610 1597 1798 1701 1804 1580
180 2011 1914 1727 1669 1777 1701 1658 1723 1650 1662 1698 1746 1894 1757 1895 1673
175 2107 2003 1821 1756 1864 1782 1734 1817 1744 1744 1784 1838 1990 1851 1981 1755
170 2196 2093 1912 1841 1945 1872 1831 1901 1824 1839 1870 1873 2084 1960 2064 1840
165 2287 2186 2005 1929 2034 1952 1923 1999 1911 1930 1953 1986 2177 2030 2169 1922
160 2372 2265 2082 2010 2122 2042 1998 2076 1986 2012 2038 2078 2268 2120 2252 2008
Page 16 of 19
18 inch Combo CALIBRATION REPORT
Figure 1 shows dependence of measured signal of each sensor (vertical axis) on arm’s deviation (horizontal axis).
Page 17 of 19
18 inch Combo CALIBRATION REPORT
Number of runs :3
Tool specifications:
The 18-inch Combo Tool has been tested and accordingly calibrated.
The MFL tool sensitivity should be Offset=2 & Gain=2, and optimum sampling rate should be
2 ms.
Page 18 of 19
18 inch Combo CALIBRATION REPORT
Chapter 4. Contact Us
E-mail : info@linscaninspection.com
Website : www.linscaninspection.com
Page 19 of 19