Professional Documents
Culture Documents
Neonatal Phototherapy Radiometer Current Performance Characteristics and Future Requirements.
Neonatal Phototherapy Radiometer Current Performance Characteristics and Future Requirements.
Show more
Abstract
Hand held radiometers provide a convenient means of monitoring the
output of neonatal phototherapy treatment devices as part of planned
programs of device maintenance and output monitoring. It was
considered appropriate to determine the wavelength and angular
response of a selection of such meters and compare their indicated values
with that derived from spectral analysis of phototherapy light sources.
This was undertaken using a Bentham DMc150 double grating
spectroradiometer and a series of 10 nm band pass optical filters in the
range 400–640 nm used in conjunction with a fiber optic light source.
Specific meters investigated included a GE Biliblanket Light Meter II, a
NeoBLUE radiometer and a Bio-TEK radiometer 74345 device.
Comparisons were made of measurements made using the hand held
meters and the Bentham DMc 150 system for a range of neonatal
phototherapy treatment devices. The use of such meters is discussed in
relation to applicable equipment standards and recommendations of
intensive phototherapy from clinical groups such as the American
Academy of Pediatrics and a specification for a spectroradiometer based
measurement system is proposed.
Keywords
Neonatal phototherapy
Hyperbilirubinemia
Neonatal jaundice
1. Introduction
0.9 mW cm within the active bandwidth. Maisels and McDonagh [4] identify
−2
specific pathways of bilirubin metabolism and indicate that the most effective
wavelengths are probably between 460 nm and 490 nm through no specific
evidence is cited for this observation. McDonagh and Lightner [5] describe the
complex modes of photo disruption of bilirubin in vivo, indicating the numerous
factors which distinguish in vitro from in vivo wavelength responses. A recent
study [6] has indicated a linear response of the rate of serum bilirubin reduction
with delivered light dose, with no indication of a ‘plateau effect’ with
progressively higher levels of incident light levels.
A Gaussian response function G (λ) with peak sensitivity at 460 nm and 50%
430–490
response within the specific inclusive wavebands were also identified. Fig.
1 indicates characteristics of TH (λ), G (λ) and G (λ).
460–490 430–490 460–490
Fig. 1. Specific functions of ‘top hat’ 460–490 nm, G 430–490 (λ) and G 460–490 (λ). The function
TH (λ) is not shown.
430–490
2. Method
2.1. Determination of angular response of phototherapy
radiometers
The angular response of the indicated meters and the Bentham DMc150
system with a D6 detection head were determined using light from a fiber
optic light source with a 4 mm cross section presented to the detector
surface at angles from −90° to +90° at intervals of 10°. The 4 mm fiber
optic light source provided a convenient means of generating high levels
of optical output within the wavelength sensitivity range of the
phototherapy meters. The distance between the tip of the fiber light
source and the surface of the specific detector was 22 cm. This distance
was identified as a compromise between allowing the phototherapy
meters to register a meaningful signal from the light source (shorter
distance better) and allowing improved angular resolution of
measurements (greater distance better).
The factor f [7], [8] is used as a ‘quality factor’ for determination of the
2
where this value is normalized to the value 2 cos(α) sin(α) summed over the same
angular range.
Fig. 2. Measurement configuration of fiber optic light source, bandpass optical filter
and plane of measurement.
were derived from values of R(j) using Eq. (5) and the corrected value of
R calculated. This process was undertaken for the range of coefficient
430
values. In this process the value of central irradiance is reduced but with
inclusion of contributions from other terms in adjacent wavebands. This
correction factor did not, however, significantly alter the wavelength
sensitivity characteristics of the meter devices. For each meter device, the
normalized wavelength response Resp_norm(j) with correction for
sidebands was determined for each waveband.
A estimation of meter reading, Meter , was derived as indicated in Eq.
Est2
the meter units for this device are μW cm . This function is essentially deriving
−2
3. Results
with increasing angle to the vertical where, for example, the cumulative %
f value at 50% includes the sum of all contributions between 0° and 50°.
2
5.2 μW cm nm and the NeoBLUE meter values of 5.2 μW cm nm and
−2 −1 −2 −1
for non-contact geometry were usually higher than values of Meter for obs
the GE Biliblanket Light Meter II and the NeoBLUE hand held radiometer
devices though typically within a margin of 10%. The lower values of
Meter relative to Meter , could have been introduced through the loss of
obs Est2
Meter for non-contact lamp systems. Some observed values of Meter for the
Est2 obs
Bio-TEK unit were higher than Meter (Giraffe Photospot (40%), NeoBLUE
Est2
(low) 64%, GE Lullaby (low) 34%). The cosine response of the Bio-TEK meter
would tend to result in values of Meter less than that of Meter . The Bio-TEK
obs Est2
device was not able to record the irradiance value of several of the phototherapy
lamps since the indicated irradiance value was greater than its maximum
displayable value.
Measurements made on fiber optic pads showed more variation between
phototherapy meters and the Bentham DMc 150 measurements. A component of
this variability relates to the difficulty of selecting identical treatment areas on
the surface mats using different detection devices.
3.4. Mapping of phototherapy lamp outputs to defined bandwidth
functions
The predicted irradiance in specific waveband outputs as indicated in Table 4
was derived from the measured spectral output of the phototherapy devices and
the previous identified wavebands indicated in Fig. 1. Values in units of
μW cm nm were in turn calculated using the bandwidth values of the respective
−2 −1
devices.
Table 4. Estimated irradiance within spectral functions TH430–490(λ), G430–490(λ), TH460–490(λ)
and G (λ) based on Bentham DMc 150 measurements with values indicated in
460–490
4. Discussion
Hand held phototherapy radiometers provide a useful role in measuring outputs
in order to maintain treatment levels within product specifications. Their primary
role, however, is typically for use with specific phototherapy light sources. It is
customary in device validation procedures of phototherapy radiometers to present
a narrow bandwidth light source of known irradiance to the unit being tested at
normal incidence and at the peak of wavelength sensitivity. Such a process is
therefore neither checking the spectral response nor the cosine response function
of the device, both of which are elements which can significantly affect the
performance of the device.
The light sensitive area of each hand held meter detector was not directly
investigated. It is likely, however, that the relative sensitivity of such devices
varies within the light sensitive area of each device and which would affect the
measurement of light fields of varying uniformity.
Oláh et al. [11] and Csoma et al. [12] highlight the potential long term
risks of neonatal blue light therapy with respect to development of
melanocytic naevus and also ophthalmic changes associated with
structures such as iris freckles, iris nevi and choroidal nevi. Previous
studies such as that of Mahé et al. [13] have identified no causal link
between prevalence of atypical nevi and previous exposure to neonatal
phototherapy. The more recent study indicated an increased occurrence
of these characteristics within twins who had received blue light
phototherapy at birth. While it is acknowledged that the risks of
hyperbilirubinemia are immediate and in extreme cases irreversible, the
authors indicate the need to provide treatment only where indicated. In
this context, however, this implies an increased importance for accurate
determination of delivered levels of phototherapy which includes spectral
information.