SECTION 1 CHAPTER 1 Vital Signs Measurement

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 15

SECTION 1: Vital Signs and Patient Monitoring Techniques 1

CHAPTER 1: which time it begins to rise to


Vital Signs Measurement nonpregnant levels.
▪ lateral decubitus position, both systolic
INTRODUCTION and diastolic pressure decline until the
Recommended in every ED: 28th week and then begin to rise to
- Temperature nonpregnant levels.
- Pulse
- RR
- Blood pressure
- Pulse oximetry
o Vital signs may indicate the severity of illness and also
dictate the urgency of intervention.
o Emergency medical service (EMS) personnel begin
assessment of the patient’s status and vital signs in the
prehospital setting.
- Surges of epinephrine and norepinephrine commonly
occur during transport by the EMS, and these
catecholamines are known to alter vital signs and lead RESPIRATION
to increases greater than 10% in the heart rate. Breathing is initiated and primarily controlled in the medullary
o Blood pressure and pulse are frequently evaluated respiratory center of the brainstem.
together as a measure of blood volume.
The respiratory center is modulated by the pneumotaxic
o Capillary refill is discussed as an assessment of overall
center, which limits the length of the inspiratory signal and
perfusion, circulatory volume, and blood pressure.
greatly influences the RR and apneustic center in the pons.
o Body temperature is usually the last vital sign measured
during resuscitation, it has special importance for patients Increased RRs may be seen in patients with a variety of
suffering from thermal regulatory failure. pulmonary or cardiac diseases, and acidosis, anemia,
NORMAL VALUES temperature, stress, and drugs (such as stimulants and
- Normotensive or normal systolic blood pressure salicylates) can significantly alter the RR in the absence of
▪ 90 to 140mmHg cardiopulmonary dysfunction.
- Normotensive or normal diastolic blood pressure
▪ 60 to 90 mmHg INDICATIONS AND CONTRAINDICATIONS
- Heart rate The only contraindications to careful measurement of RR are
▪ 60 to 100 bpm the scenarios of respiratory distress, apnea, and upper airway
- Respiratory rate obstruction.
▪ 12 to 24 breaths/min o A prehospital RR of less than 10 breaths/min or greater
▪ 8 to 24 breaths/min than 29 breaths/min is associated with major injury in 73%
o Pregnancy results in alterations in the normal adult values of children.
of pulse and blood pressure. o Using tachypnea alone as a predictor of pulmonary
- Pregnancy is characterized by significant increases in pathology, infants with an RR higher than 60 breaths/min
minute ventilation and is thought to be due to the are found to be hypoxic 80% of the time.
combined facilitatory effects of progesterone and o An RR higher than 25 breaths/min in prehospital trauma
estrogen on central and peripheral chemoreflex patients was associated with increased mortality.
drives to breathe. PROCEDURE
- Resting pulse rate increases throughout pregnancy o Count the respirations when the patient is unaware that
from 10% to 15% over baseline values. his or her breathing is being observed.
- Norms for systolic and diastolic blood pressure are o Count for a full minute.
dependent on patient positioning. o Frequency of breathing is less regular than the pulse.
▪ sitting or standing, systolic pressure is o An infant’s RR can easily be determined by observing or
essentially unchanged. palpating the excursion of the chest or the abdominal wall.
▪ Diastolic pressure declines until o Infants should be observed for grunting respirations, which
approximately 28 weeks’ gestation, at are produced by expiration against a partly closed glottis
(an attempt to maintain positive airway pressure).

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 2

▪ Periodic breathing is considered a


benign disorder.
- Apnea is defined as a respiratory pause longer than
INTERPRETATION 20 seconds.
▪ It may be associated with bradycardia
Respiratory Rate
and hypoxia.
o Reproducibility of RR measurements may be limited by
▪ Infants with symptomatic apneic
significant interobserver variability:
episodes that result in apparent life-
- Interobserver variability may account for a difference
threatening events are thought to be at
of up to 6 breaths/min.
increased risk for sudden infant death
- Variability in the same observer may account for up to
syndrome.
5 breaths/min.
o RR is an independent risk marker for in-hospital mortality PULSE
in community-acquired pneumonia.
o The RR will generally increase in the presence of fever. Examine the pulse to establish the cardiac rate and regularity
Respiratory Pattern and Amplitude of the rhythm.
Doppler ultrasound has utility in locating a pulse, assessing
fetal heart tones beyond the first trimester of pregnancy,
evaluating peripheral lower extremity vascular insufficiency
with an ankle-brachial index, and assessing blood pressure in
infants or in patients with low-flow states.
o Blood flowing into the aorta with each cardiac cycle
initiates a pressure wave.
o Blood flows through the vasculature at approximately 0.5
m/sec, but pressure waves in the aorta move at 3 to 5
m/sec.
o Therefore palpated peripheral pulses represent pressure
waves, not blood flow.

INDICATIONS AND CONTRAINDICATIONS


No contraindications exist to assessment of the pulse rate.
Assessment of blood flow by palpation of the pulse can be used
to gauge the presence of cardiac contractility and not just the
electrical rhythm.
o Continuous monitoring is not routine but may be helpful
when the clinical situation predicts significant variability in
heart rate, as in the setting of sepsis.
o Absence of a radial pulse or absence of both radial and
femoral pulses and hypotension has been demonstrated in
hypovolemic trauma patients.
Abnormal Respiratory Patterns. o Avoid concurrent bilateral carotid artery palpation because
o Decreased RR is commonly seen with opiate toxicity. this maneuver could theoretically endanger cerebral blood
o Hyperpnea, or a normal RR but clinically significant flow.
hyperventilation secondary to increased tidal volume, may o Massage of the carotid sinus, found at the bifurcation of
be seen with salicylate poisoning. the external and internal carotid arteries at the level of the
o In infants, it is essential to distinguish normal periodic mandibular angle, may result in reflex slowing of the heart
breathing from apnea. rate.
- Periodic breathing consists of three or more - To avoid inadvertent carotid sinus massage, palpate
respiratory pauses longer than 3 seconds in duration the carotid pulse at or below the level of the thyroid
with less than 20 seconds between pauses. cartilage.
▪ There is no associated bradycardia or o In adults with atherosclerotic disease, there is a rare risk of
cyanosis. precipitating a cerebrovascular event by vigorous palpation
▪ Particular problem in preterm infants. of the carotid artery.

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 3

- Minimize this risk by prior auscultation of the carotid


artery.
- If a bruit is present, gently palpate the carotid pulse
while avoiding vigorous palpation, or use a Doppler
ultrasound probe to assess carotid flow instead.
PROCEDURE
o For convenience the radial pulse at the wrist is routinely
used.
o Use the tips of the first and second fingers to palpate the
pulse. Two advantages of this technique:
- (1) the fingertips are quite sensitive, thereby enabling
the pulse to be located easily and counted, and;
- (2) the examiner’s own pulse may be erroneously
counted if the thumb is used.
o Pulses are easily palpated at the carotid, brachial, femoral,
posterior tibial, and dorsalis pedis arteries.
o Palpate the pulse at the brachial artery to appreciate its
contour and amplitude. o Pulse varies with respiration: it increases with inspiration
- Locate the pulse at the medial aspect of the elbow and slows with expiration. This is known as sinus
and note that it is more easily palpated when the dysrhythmia and is physiologic.
elbow is held slightly flexed. o Bradycardia is most commonly defined as a heart rate
- Determine the pulse rate by counting for 1 minute, lower than 60 beats/min in adults.
particularly if any abnormality is present. o Consider whether a patient’s abnormal pulse rate is a
o Common convention in the acute care setting is to count a primary or secondary condition. Examine the entire set of
regular pulse for 15 seconds and multiply the resulting vital signs when attempting to discern the cause of the
number by 4 to determine the beats per minute. abnormal rate.
o In neonates, use direct heart auscultation and umbilical o Consider the medications that the patient may be taking or
palpation as the methods of choice to determine the heart the presence of a mechanical pacemaker.
rate. - Digitalis compounds, β-blockers, and
o In unstable children, palpate the central arteries, antidysrhythmics may alter the normal heart rate and
particularly the femoral and brachial pulses, instead of the the ability of this vital sign to respond to a new
more peripheral arteries. physiologic stress.
- Sympathomimetic drugs such as cocaine and
INTERPRETATION methamphetamine increase heart rate, as do
Pulse Rate anticholinergic drugs.
o In infants and children, interpret the pulse rate with
Heart Rhythm
reference to age.
Obtain information about the regularity of the pulse by
palpation.
o An irregular pulse suggests atrial fibrillation or flutter with
variable block, and accurate assessment should be carried
out by auscultation of the apical cardiac sounds.
- The apical pulse is frequently greater than the
peripheral pulse because of inadequate filling time
and stroke volume, with resultant non-transmitted
beats.
Pulse Amplitude and Contour
o Superimposition of one pathophysiologic state on another
may modify the pulse.
- For example, sepsis may result in variable pulse
amplitudes, depending on the stage in the
development of the disease.

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 4

- Early in sepsis, cardiac output increases and vascular - Repeated measurements will provide an evaluation of
resistance decreases, causing bounding pulses. the adequacy of resuscitation in patients whose blood
- In advanced sepsis or septic shock, falling cardiac pressure cannot be auscultated by standard
output and increased vascular resistance are seen, techniques and in those in whom intraarterial blood
and pulses are diminished. pressure measurements are either contraindicated or
o Age-related changes in pulse amplitude and contour. technically unobtainable.
- Such changes are due to an increase in arterial o Placing a catheter for direct intraarterial measurement of
stiffness, resulting in increased pulse wave velocity blood pressure may be performed safely in the ED, but is
and progressively earlier wave reflection. not standard of care and has a higher risk for complications.
▪ Pulse wave analysis may be useful in - In particular, direct measurement of arterial pressure
determining arterial stiffness and the during pulseless electrical rhythms may help
likelihood of atherosclerotic disease in discriminate between a severe shock state and
a vascular laboratory setting. otherwise non-resuscitatable status.
o Weak pulses can be a significant finding in hypotensive o Alternative noninvasive devices for continuous blood
patients, or an indication of limb ischemia if isolated to one pressure measurement (CBPM) have been introduced
extremity. clinically:
o Bounding pulses can be seen with a widened pulse - One common method of CBPM uses finger cuffs
pressure. equipped with infrared (IR) photoplethysmography
o Routine measurement of pulse amplitude requires and sophisticated technology for quantification of
instrumentation. finger blood pressure levels.
Pulses During Cardiopulmonary Resuscitation o Relative contraindications to specific extremity blood
A carotid pulse is preferred when assessing the adequacy of pressure measurement include an arteriovenous fistula,
chest compressions during cardiopulmonary resuscitation. ipsilateral mastectomy, axillary lymphadenopathy,
lymphedema, and circumferential burns over the intended
ARTERIAL BLOOD PRESSURE site of cuff application.
An abrupt reduction in a patient’s arterial blood pressure EQUIPMENT
usually indicates the need for immediate intervention or Two types of blood pressure monitoring equipment are
reconsideration of therapy. currently available and used in EDs: cuff type and noninvasive
o Arterial blood pressure indicates the overall state of waveform analysis.
hemodynamic interaction between cardiac output and Cuff Type
peripheral vascular resistance. Equipment required for indirect blood pressure
o Arterial blood pressure is the lateral pressure or force measurement includes:
exerted by blood on the vessel wall. It indirectly measures - Sphygmomanometer (cuff with an inflatable bladder,
perfusion, and blood flow equals the change in pressure inflating bulb, controlled exhaust for deflation, and
divided by resistance. manometer)
o Mean Arterial Pressure: - Stethoscope
- Doppler device (for auscultation), or oscillometric
device
American Heart Association Guidelines:
INDICATIONS AND CONTRAINDICATIONS - The width of the bladder should be at least 40% of
o Patients with minor ambulatory complaints unrelated to the distance of the limb’s midpoint (i.e., from the
the cardiovascular system may not necessarily need their acromion process to the lateral epicondyle).
blood pressure measured in the ED, and those with - The length of the bladder should be 80% of the
hemodynamic instability need frequent monitoring of midarm circumference or twice the recommended
blood pressure. width.
o In general, the younger the patient, the less likely blood The manometers in common use are either aneroid, digital,
pressure will be measured. or mercury gravity column, though the mercury type is much
o In newborns, infants, and even toddlers, capillary refill is less common in modern use.
sometimes substituted for standard blood pressure PROCEDURE
measurement. o Obtain indirect blood pressure measurements at the
o In low-flow states, Doppler measurement of blood patient’s bedside by palpation, auscultation, Doppler, or
pressure may be obtained rapidly. oscillometric methods.

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 5

o The patient may be lying or sitting, as long as the site of o Apply the bell of the stethoscope directly over the brachial
measurement is at the level of the right atrium and the arm artery with as little pressure as possible.
is supported. - Systolic blood pressure is defined as the first
o Unless the arm is kept perpendicular to the body with the appearance of faint, clear, tapping sounds that
elbow resting on a desk, measurements will be 9 to 14 gradually increase in intensity (Korotkoff phase I).
mmHg higher, regardless of body position. - Diastolic blood pressure is defined as the point at
o Allowing the arm to be parallel to the body when supine which the sounds disappear (Korotkoff phase V).
and supporting the arm perpendicular to the body when - In children, phase IV defines diastolic blood pressure.
measuring blood pressure may create a pseudo-drop in - Phase IV is marked by a distinct, abrupt muffling of
blood pressure. sound when a soft, blowing quality is heard.
o To palpate arterial blood pressure, inflate the cuff to 30 o Alternative sites include the radial, popliteal, posterior
mmHg above the level at which the palpable pulse tibial, or dorsalis pedis arteries, although any fully
disappears. compressible extremity artery may be used.
o Palpate directly over the artery and deflate the cuff at a o Novel noninvasive continuous finger cuff technology
rate of 2 to 3 mmHg/sec. offers the benefit of uninterrupted monitoring and has the
o Report the initial appearance of arterial pulsations as the advantage over invasive techniques of being safer and
palpable blood pressure. This practice, known as the Riva- immediately available.
Rocci palpatory technique. o Wrist blood pressure has been shown to have good
o The Doppler method is preferred when determining blood average accuracy in the surgical environment when
pressure in infants. compared with oscillometric devices.
o Inadvertent prolonged application of an inflated blood
pressure cuff may result in falsely elevated diastolic
pressure and ischemia distal to the site of application.

INTERPRETATION
o Blood pressure tends to increase with age and is generally
higher in males.
o Individual factors that influence blood pressure include
body posture, emotional or painful stimuli,
environmental influences, vasoactive foods or
medications, and the state of muscular and cerebral
activity.
- Exercise and sustained isometric muscular
contraction increase blood pressure in proportion to
the strength of the contraction.
- Normal diurnal pattern of blood pressure consists of
an increase throughout the day with a significant,
rapid decline during early, deep sleep.
o Systolic blood pressure in infants and children:
- Normal lower limits of systolic blood pressure in
infants and children can be estimated by adding 2
times the age (in years) to 70 mmHg.
o When auscultating blood pressure at the brachial artery, - The 50th percentile for a child’s systolic arterial blood
apply the blood pressure cuff approximately 2.5 cm above pressure from 1 to 10 years of age can be estimated
the antecubital fossa with the center of the bladder over by adding 2 times the age (in years) to 90 mmHg.
the artery. - Children older than 2 years are considered
hypotensive when systolic blood pressure is less than
80 mmHg.

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 6

o Most adults are considered hypotensive if systolic blood - In those who have just finished exercising
pressure is lower than 90 mm H g, but some individuals - In children younger than 5 years
normally exhibit a systolic pressure in that range. o Irregular heart rates may also interfere with accurate
o In the elderly, the presence of normotension within determination of blood pressure.
defined or published limits may not be reassuring. WHY? - Take a second or third reading, with 2 minutes of
o When considering systolic blood pressure cutoffs for deflation between recordings, and obtain an average
trauma patients: when premature contractions or atrial fibrillation are
- 85 mm H g for patients aged 18 to 35 years present.
- 96 mm H g for patients aged 36 to 64 years o Hemiplegic patients may exhibit different blood pressures
- 117 mm H g for those older than 65 years in the affected and unaffected arms.
o In patients with shock, blood flow cannot be reliably - A flaccid extremity tends to yield lower systolic and
inferred from heart rate and blood pressure values. diastolic pressure.
Hypertension - A spastic extremity tends to yield higher values than
Adults are hypertensive if either systolic or diastolic pressure the extremity with normal motor tone.
consistently exceeds 140 or 90 mm H g, respectively. - It is preferable to monitor blood pressure in the
o A metaanalysis showed strong correlation of blood unaffected limb.
pressure to vascular (and overall) mortality down to at least o There is an increasing number of patients with heart failure
115/75 mmHg. treated with left ventricular assist devices (LVADs). 2 types
o Patients with hypertension require repeated of VADs:
measurements to assess whether therapy is required in the - Pulsatile
ED. Pulse and blood pressure reading in patients with
o White coat hypertension (WCH) is defined as a persistent pulsatile VADs are comparable to the general non-
elevation in blood pressure in the clinical setting only. VAD population.
- Nonpulsatile
Nonpulsatile VADs function by either centrifugal or
axial blood flow, and this has a significant impact on
the ability to detect pulses.

Measurement Errors
o Falsely low blood pressure may be caused by using an Pulse Pressure
overly wide cuff, by placing excessive pressure on the head o The difference between systolic and diastolic pressure.
of the stethoscope, or by rapid cuff deflation. o Increased pulse pressure (i.e., ≥60 mm H g) is commonly
o Falsely high blood pressure may be caused by the use of observed with:
an overly narrow cuff, anxiety, pain, tobacco use, exertion, - Anemia
an unsupported arm, or slow inflation of the cuff. - Exercise
o There appears to be a statistically significant difference in - Hyperthyroidism
the error rate associated with patients weighing more than - Arteriovenous fistula
95 kg. - Aortic regurgitation
o The use of small cuffs was associated with a mean error of - Increased intracranial pressure
8.5 and 4.6 mmHg in systolic and diastolic pressure. - Patent ductus arteriosus
o Hypotensive patients have unreliable Korotkoff sounds, o Narrowed pulse pressure (≤20 mm Hg) may be a
but Doppler measurements are well correlated with direct manifestation of:
arterial systolic pressure measurements in these patients. - Hypovolemia
o An auscultatory gap can be appreciated in hypertensive - Increased peripheral vascular resistance (as seen in
patients. early septic shock)
- It is heard during the latter part of phase I and should - Decreased stroke volume
not be confused with diastolic readings. - A narrowed pulse pressure is classically noted in
o Measurement of diastolic blood pressure should occur at aortic stenosis and pericardial tamponade.
Korotkoff phase IV: o There is evidence that an early indicator of reduced central
- In patients with aortic insufficiency blood volume in the presence of stable vital signs is the
- Hyperthyroidism reduction in pulse pressure.

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 7

Differential Brachial Artery Pressure - Initially, the arterial pulse will be heard only during
o The presence of a systolic blood pressure difference of 10 expiration and will disappear during inspiration.
to 20 mmHg between the arms suggests a normal Deflate the cuff further until arterial sounds are heard
condition. throughout the respiratory cycle.
- If greater, it may indicate advanced focal - Palpation at the radial or femoral arteries may yield
atherosclerosis, coarctation of the aorta proximal to complete disappearance during inspiration. When
the left subclavian artery, type A aortic dissection, present, this technique is a quick bedside
aortic arch syndromes, or other vascular processes confirmation of the possibility of severe tamponade.
preferentially affecting one extremity.
Pulse-Pressure Variation
o Fluid resuscitation is an integral piece of the management
of patients with circulatory failure.
o Fluid responsiveness: ability of the left ventricle to increase
stroke volume in response to fluid administration.
- This concept is based on the Frank-Starling curve and
the knowledge that pulse pressure (systolic pressure
minus diastolic pressure) is directly proportional to
stroke volume.
o The variation in pulse pressure seen with the respiratory
cycle, or pulse-pressure variation, reflects the magnitude of
respiratory change on stroke volume.
- This is best demonstrated by the influence of
mechanical ventilation on right ventricular preload.
- Pulse-pressure variation of greater than 13% is highly
predictive of fluid responsiveness in mechanically
ventilated patients.
o Standard method to assess fluid responsiveness is done
with passive leg raising (PLG).
- This “self-volume challenge” increases preload
through translocation of venous blood from the lower
extremities to the thorax.
- Patient who exhibits a rise of more than 10% in their o An alternative method for determination of pulsus
aortic blood flow (measured with esophageal paradoxus is by visually observing the loss of the pulse
Doppler) or cardiac index (measured with oximetry waveform and then its reappearance.
thermodilution) is considered a “fluid responder,” o If the difference between inspiratory and expiratory
which is indicative of the need for further fluid pressure is greater than 12 mm H g, the paradoxical pulse
administration. is abnormally wide.
Pulsus Paradoxus - Most patients with proven tamponade have a
o Normal respiration briefly decreases systolic blood difference of 20 to 30 mm H g or greater during the
pressure by approximately 10 mm H g during inspiration. respiratory cycle.
o Pulsus paradoxus occurs when there is a greater than 12- - This may not be true of patients with very narrow
mmHg decrease in systolic blood pressure during pulse pressures (typical of advanced tamponade),
inspiration. who have a “deceptively small” paradoxical pulse of
o Occur in patients with chronic obstructive pulmonary 5 to 15 mmHg.
disease, pneumothorax, severe asthma, or pericardial o Pulsus paradoxus has been correlated with the level of
tamponade. impairment of cardiac output by tamponade.
- To measure a paradoxical pulse, have the patient lie - In an uninjured patient with pericardial effusion, a
comfortably in the supine position at a 30- to 45- pulsus paradoxus greater than 25 mmHg (in the
degree angle and breathe normally in an unlabored absence of relative hypotension) is both sensitive and
fashion. specific for moderate or severe versus mild
- Inflate the blood pressure cuff well above systolic tamponade.
pressure and slowly deflate it until the systolic sounds
that are synchronous with expiration are first heard.

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 8

o The absence of a paradoxical pulse does not rule out - One crystal transmits the signal and the other
tamponade. receives it.
o In the pediatric population, pulsus paradoxus has been o Probes with a frequency of 2 to 5 MHz are best for
studied to determine the severity of obstructive and detecting fetal heart sounds.
restrictive pulmonary disease, most commonly asthma. o Frequencies of 5 to 10 MHz are appropriate for limb
- A value of 15 mmHg or greater correlates well with arteries and veins.
the clinical score, peak expiratory value, flow rate, PROCEDURE
oxygen saturation, and subsequent need for o Place the Doppler probe against the skin with an acoustic
admission. gel used as an interface. The gel ensures optimal
Shock Index transmission and reception of the ultrasound signal and
o The ratio of pulse rate to systolic blood pressure has been protects the crystals.
suggested as a measure of clinical shock. - In an emergency, water-soluble lubricant (e.g.,
o The shock index (SI) has a normal range of 0.5 to 0.7. Surgilube or K-Y Jelly) may be substituted for
- An SI above 0.85 to 0.90 suggests acute illness in commercial acoustic gel.
medical patients and a marked increase in the o Angle the probe at 45 degrees along the length of the
potential for gross hemodynamic instability in trauma vessel to optimize frequency shifts and signal amplitude.
patients. o To evaluate peripheral perfusion:
o Although the SI appears to correlate with the left - Place a sphygmomanometer cuff proximal to where
ventricular stroke work index, it has little correlation with the arterial pulse is being evaluated and inflate it.
systemic oxygen transport in patients with hemorrhagic - Place the probe over the arterial pulse and slowly
and septic shock. deflate the cuff.
DOPPLER ULTRASOUND FOR EVALUATION OF PULSE - The pressure at which flow is first heard is the systolic
AND BLOOD PRESSURE pressure under the cuff, and not the pressure at the
PRINCIPLES OF DOPPLER ULTRASOUND level of the Doppler probe.
o In the evaluation of peripheral vascular disease, one may
o The frequency of sound waves varies depending on the
determine the ankle-brachial index.
speed of the sound transmitter in relation to the sound
- Examine both brachial arteries at the medial aspect of
receiver.
the antecubital fossa.
o Doppler devices transmit a sound wave that is reflected by
- Angle the probe until the most satisfactory signal is
flowing erythrocytes, and the shift in frequency is detected.
obtained.
o Frequency shift can be detected only for blood flow
- Inflate the cuff and slowly deflate it until the systolic
greater than 6 cm/sec.
pulse is heard.
- Repeat the procedure for the posterior tibial and
dorsalis pedis arteries of both lower extremities.
INDICATIONS AND CONTRAINDICATIONS
Doppler ultrasound is commonly used in the ED for the
measurement:
- blood pressure in low-flow states
- evaluation of lower extremity peripheral perfusion
- assessment of fetal heart sounds after the first
trimester of pregnancy
o Doppler’s sensitivity allows detection of systolic blood
pressure down to 30 mm H g in the evaluation of a patient
in shock.
o In a patient with peripheral vascular disease in whom there
is concern about the adequacy of peripheral perfusion, the
ankle-brachial index provides a rapid, reproducible, and
standardized assessment.
o The use of Doppler ultrasound for the evaluation of deep
venous thrombosis is a valuable tool.
EQUIPMENT
o A nondirectional Doppler device has a probe that houses
two piezoelectric crystals:

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 9

o In evaluating fetal heart tones, an examination of several


locations and angles over the uterus must be performed in
search for the optimal signal due to the variable positioning
of the fetus.
- It is best to begin in the mid-suprapubic area and then TYPICAL PRESSURES IN NORMAL INDIVIDUAL:
explore the uterus via angulation of the probe.
- Once tones are located, move the probe along the
abdomen to reach a position closer to the origin of
the sound.
- Distinguish fetal heart tones from placental flow by
differentiating the quality of the fetal heart tones,
which will not match the maternal pulse. The
placental flow and maternal pulse should be identical.
INTERPRETATION
o In low-flow states Doppler ultrasound can detect blood
pressure as low as 30 mmHg.
o Calculate the ankle-brachial index of each limb by dividing
the higher systolic pressure of the posterior tibial or the
dorsalis pedis artery of the limb by the higher of the systolic
pressures in the brachial arteries.
- In normal individuals, the index should be greater
than 1.0.
- Patients with mild to moderate claudication have
values between 0.4 and 0.9. Normal findings:
- Values lower than 0.4 indicate severe impairment 1. ankle-to-brachial pressure index of 1.0 or higher
and are consistent with critical limb ischemia 2. segmental pressure gradients lower than 30 mm H g
o Segmental lower extremity pressure measurements may 3. upper thigh pressure at least 40 mm H g above
help identify the level of the obstruction. brachial pressure

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 10

- hemodynamic monitoring (e.g., monitoring of central


TYPICAL PRESSURES IN A PATIENT WITH OBSTRUCTION venous pressure)
OF THE RIGHT POPLITEAL OR TIBIAL ARTERIES o The ideal test for determining volume status would rapidly
and accurately detect 5% or greater depletion of volume
with a noninvasive technique. At present, no such test
exists.
ORTHOSTATIC VITAL SIGNS MEASUREMENT
o Orthostatic vital signs have historically been used to
evaluate patients with fluid loss, hemorrhage, syncope, or
autonomic dysfunction.
o Although orthostatic testing is commonly cited as a method
to detect hypovolemia, it is frequently misleading and has
less clinical value.
o In patients with an acute loss of less than 20% of total blood
volume, orthostatic vital signs have been shown to lack
both sensitivity and specificity.
PHYSIOLOGIC RESPONSE TO HYPOVOLEMIA
o Acute blood loss or severe hypovolemia related to
dehydration decreases venous return.
- This can be seen with acute blood loss (usually greater
Significant findings: than 20% of blood volume), severe burns or
1. ankle-to-brachial pressure index less than 0.9 in the prolonged vomiting or diarrhea that depletes body
right leg fluids.
2. abnormally high gradient from the ankle to below the - As a result, cardiac output falls and clinical
knee and again from below to above the knee in the manifestations of shock ensue.
right leg o Several compensatory mechanisms are initiated by acute
3. upper thigh pressure 50 mm H g higher than brachial hypovolemia:
pressure, consistent with normal flow at the aorta-
iliac level
*Findings are suggestive of right popliteal occlusion or right
anterior and posterior tibial occlusion, or both
o When the lower extremity has been amputated or injured,
brachial-brachial indices can be used (i.e., comparison of
systolic blood pressure in the injured or diseased upper
extremity with the other extremity).
o Patients with ankle-brachial index values of 0.9 or lower
have increased cardiovascular morbidity and mortality.
o Obese patients, diabetic patients, or those with calcified
vessels that are not compressible may have abnormally
high systolic pressure (e.g., 250 to 300 mm H g) and indices
that do not accurately reflect flow.
VITAL SIGN DETEMRINATION OF VOLUME STATUS
- The dominant compensatory mechanism in shock is a
- evaluation of skin color reduction in carotid sinus baroreceptor inhibition of
- skin turgor sympathetic outflow to the cardiovascular system.
- skin temperature This increased sympathetic outflow results in several
- supine, serial, and orthostatic vital signs effects:
▪ Serial vital sign measurements have been 1. arteriolar vasoconstriction, which greatly
used for assessing blood loss, but they do increases peripheral vascular resistance
not reliably detect small degrees of blood 2. constriction of venous capacitance vessels, which
loss. increases venous return to the heart
- neck vein status
- transcutaneous oximetry

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 11

3. an increase in the heart rate and force of - Orthostatic hypotension caused by autonomic insufficiency
contraction, helping maintain cardiac output is not usually accompanied by tachycardia, and the
despite significant loss of volume. orthostatic hypotension produced by acute volume
o 30% to 40% of blood volume can be lost before death depletion is commonly accompanied by a pronounced
occurs. reflex tachycardia.
- When sympathetic reflexes are absent, a loss of only - Even in normal individuals, passive tilting generates a high
15% to 20% of blood volume may cause death. incidence of orthostatic syncope.
o Increased sympathetic nerve activity results in the - Because of decreased vasomotor tone, limited
commonly recognized physical signs of shock, including chronotropic response, and other factors, the elderly have
pallor, cool clammy skin, rapid heart rate, muscle a higher incidence of orthostatic hypotension leading to
weakness, and venous constriction. syncope and fall-related injuries.
o An inadequate immediate compensatory response will - Patients with hypertension may also have abnormal
result in dizziness, altered mental status, or loss of vasomotor responses to tilt testing and demonstrate more
consciousness. instability.
o The central nervous system response to ischemia further - Chronic anemia patients, who exhibit compensated blood
stimulates the sympathetic nervous system after arterial volume, seem to have the same postural response as
pressure falls below 50 mmHg. normal individuals.
o Subsequent compensatory mechanisms that work to - Ethanol ingestion exaggerates postural pulse changes and
restore blood volume to a normal level include the release mimics the hemodynamic changes seen with acute blood
of angiotensin and antidiuretic hormone (vasopressin). loss.
- This causes arteriolar vasoconstriction, conservation - Another complicating factor in interpreting orthostatic vital
of salt and water by the kidneys, and a shift in fluid signs is the development of paradoxical bradycardia in the
from the interstitium to the intravascular space. presence of blood loss.
PHYSIOLOGIC RESPONSE TO CHANGES IN POSTURE - This paradoxical bradycardia may be more frequently
When an individual assumes the upright posture, complex associated with rapid and massive bleeding.
homeostatic mechanisms compensate for the effects of gravity - Patients with more gradual blood loss tend to have a
on the circulation to maintain cerebral perfusion with minimal more typical tachycardic response.
change in vital signs. These responses include: INDICATIONS AND CONTRAINDICATIONS
1. Baroreceptor-mediated arteriolar vasoconstriction Many factors influence orthostatic blood pressure, including
2. venous constriction and increased muscle tone in the age, preexisting medical conditions, medications, and
legs and the abdomen to augment venous return autonomic dysfunction.
3. sympathetic-mediated inotropic and chronotropic o Contraindications to orthostatic measures:
effects on the heart - supine hypotension
4. activation of the renin-angiotensin-aldosterone - the clinical syndrome of shock
system - severely altered mental status
o When a normal individual stands, the pulse increases by an - the setting of possible spinal injuries
average of 13 beats/min, systolic blood pressure falls - lower extremity or pelvic fractures
slightly or does not change, and diastolic pressure rises - use of medications that block the normal vasomotor
slightly or does not change. and chronotropic response to orthostatic tests is also
o In patients with vasodepressor syncope, the normal a contraindication to using this test for the
compensatory reflexes that preserve cerebral perfusion assessment of volume status
with changes in posture are altered. o Orthostatic vital signs are often used to assess a patient’s
- The normally increased sympathetic tone on standing response to therapy. In patients receiving intravenous
is paradoxically inhibited, and an exaggerated rehydration therapy, serial orthostatic vital signs are widely
enhancement of parasympathetic activity used to judge the end point of therapy before release.
(bradycardia) occurs and can lead to syncope. TECHNIQUE

VARIABLES AFFECTING ORTHOSTATIC VITAL SIGNS


Most of the conditions that affect postural blood pressure
regulation involve a pathologic condition that affects the
sympathetic nervous system or the use of medications that
alter normal cardiovascular compensatory functions.

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 12

CAPILLARY REFILL
The capillary refill test is a measurement of the interval of time
from the release of pressure on the nail bed or soft tissue
sufficient to blanch the nail bed or superficial soft tissue, until
the return to normal coloration.
o Delayed capillary refill is an indication of reduced skin
turgor, often as a result of volume depletion or limited
perfusion.
o Skin elasticity is the characteristic that allows skin to spring
back to its original shape after it has been deformed, and
the speed of refilling the capillary bed after compression is
responsible for the return of color to the skin.
o To obtain orthostatic vital signs, record the blood pressure
and pulse after the patient has been in the supine position
for 2 to 3 minutes.
INDICATIONS AND CONTRAINDICATIONS
o Next, ask the patient to stand and be prepared to assist if o CRT should not be determined from a dependent
severe symptoms or syncope develop. extremity, from a recently burned or injured extremity, or
- A supine-to-standing test is more accurate than a at the site of an infection or acute injury.
supine-to-sitting one. If severe symptoms develop on o Useful bedside assessment of perfusion and dehydration
standing, defined as syncope or extreme dizziness when used in conjunction with other objective signs of the
requiring the patient to lie down, the test is adequacy of perfusion.
considered positive and should be terminated. o It should not be considered accurate as a stand-alone tool.
- If not symptomatic, allow the patient to stand for 1 PROCEDURE
minute and then record the blood pressure and pulse. o The preferred sites for determining CRT are the nail bed,
INTERPRETATION the thenar surface of the palm, and the heel.
The most sensitive criteria for orthostasis are tachycardia or o Regardless of the site chosen, position the extremity at
symptoms of cerebral hypoperfusion (e.g., near-syncope). approximately the level of the right atrium.
- BLOOD PRESSURE is too variable to be an indicator of o The minimum pressure necessary to produce blanching
loss of blood volume yields the most reproducible values. Release the nail bed
o In the setting of suspected blood loss, if the patient has a and begin timing with a stopwatch or simply by counting
rise in pulse of 30 beats/min or manifests severe symptoms out “one-thousand-one, one-thousand-two” for an
and if other complicating factors have been excluded, approximation of the interval.
blood loss is highly likely (2% false-positive rate). o Stop the clock when the nail bed becomes pink again.
o Orthostatic changes in the SI were no more sensitive than INTERPRETATION
established tilt test criteria in discriminating normal o The normal CRT increases with age and is slightly longer in
individuals from those with moderate acute blood loss (450 female patients.
mL). o It is further increased by degrees of dehydration or
o Criteria for significant orthostatic changes in blood hypoperfusion.
pressure cannot be definitively set for the following o INCREASE CRT:
reasons: - Hypothermia
1. Large variability in postural blood pressure has been - Hyponatremia
found in the adult ED population - Congestive heart failure
2. the results of studies using passive tilt tables cannot be - Malnutrition
extrapolated to the bedside use of orthostatic vital - edema
signs o Environmental conditions such as ambient air temperature
3. studies using healthy patients with acute blood loss can falsely alter capillary refill.
may not reflect the orthostatic changes seen in the o Fever alone did not appear to prolong or shorten CRT in
elderly, those with chronic bleeding, dehydration, children, and a study of healthy adults found a 5% decrease
other medical problems, or in association with certain in CRT for each degree Celsius rise in patient temperature.
medications o Measurements obtained from the nail bed were more
4. many studies of orthostatic changes did not use a reproducible than those from the heel.
criterion standard in their determinations o The presence of a 2-second or longer delay in CRT when
combined with any two or more of the findings of absent

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 13

tears, dry mucous membranes, or ill general appearance o Measurement of actual core body temperature requires
predicted clinical dehydration (>5% deficit in body weight) the placement of invasive monitors, such as an esophageal
in children (1 month to 5 years of age) with 87% sensitivity or pulmonary artery (PA) probe.
and 82% specificity. o Ingestion of hot or cold beverages has shown they can alter
o Frequent monitoring of capillary refill may be useful in oral temperature readings for 5 to 30 minutes, and can
assessing responses to rapid fluid resuscitation in children. falsely elevate a normal temperature or mask a fever.
- A normal CRT of 2 seconds or less has been shown to MEASUREMENT SITES
correlate with superior vena cava oxygen saturation Core Body Temperature
of 70% or higher in critically ill children The following sites accurately reflect core body temperature
o When an abnormal CRT was defined as greater than 4.5 and changes in it:
seconds, coupled with extremity coolness, these - the distal third of the esophagus
parameters identified patients who had been - the tympanic membrane (TM) (with a direct
hemodynamically stabilized and continued to have more thermistor in contact with the anterior inferior
severe organ dysfunction and higher lactate levels. quadrant of the TM)
o In adults, CRT was found to be less sensitive and less - the PA (pulmonary artery)
specific than orthostatic vital signs in detecting 450-mL
blood loss during blood donation. Other sites may represent core body temperature under
TEMPERATURE certain conditions; for example:
An inability to maintain normal body temperature is indicative - the rectum when the temperature is obtained at least
of a vast number of potentially serious disorders, including 8 cm from the anus with an indwelling thermistor and
infections, neoplasms, shock, toxic reactions, and the body temperature is relatively constant.
environmental exposures. - the bladder when measured with an indwelling
thermistor.
o Under normal conditions, the temperature of deep
o Rectal temperature is often considered the criterion
central body tissues (i.e., core temperature) remains at
standard for body temperature in ambulatory patients and
37°C ± 0.6°C (98.6°F ± 1.08°F).
it is often used routinely in children younger than 3 years
o Environmental temperature varies from as much as 13°C to
- Disadvantages include longer intervals for
60°C (55°F to 140°F).
measurement, safety concerns, and inconvenience.
o Mean oral temperature is 36.8°C ± 0.4°C (98.2°F ± 0.7°F).
o Neutropenia and recent rectal surgery represent relative
o Maintenance of normal body temperature requires a
contraindications to measurement of rectal temperature.
balance of heat production and heat loss.
o Thermistor probes are available for measurement of
- Heat loss occurs by radiation, conduction, and
esophageal, bladder, and rectal temperature.
evaporation by approximately 60%, 18%, and 22%,
Peripheral Body Sites Approximating Core Temperature
respectively.
o A body temperature measurement by IR radiation can be
- Heat loss is increased by wind, water, and lack of
detected from the ear, including the auditory canal and
insulation (e.g., clothing).
TM; it is easy to use, hygienic, convenient, and quick.
- Sweating, vasodilation, and decreased heat
- It is possible that alterations of regional blood flow
production serve to decrease temperature.
accompanying critical illness (TMs may behave as an
- Piloerection, vasoconstriction, and increased heat
extension of the skin or the mucous membrane in the
production serve to increase body temperature.
critically ill) and the peripheral vasoconstriction that
- Heat production is increased by shivering, fat
occurs with inotropes and some forms of shock may
catabolism, and increased thyroid hormone
occur in the TM, making such measurements less
production.
accurate in this population.
o Temperature is controlled by feedback mechanisms
- A theoretical disadvantage of TM temperatures
operating through the hypothalamus.
might be a falsely elevated estimate of the core
- Heat-sensitive neurons in this area increase their rate
temperature in the presence of otitis media.
of firing during experimental heating.
o Rectal temperature measurement as the “gold standard”
- Receptors in the skin, spinal cord, abdominal viscera,
for detecting fever in the pediatric population.
and central veins primarily detect cold and provide
o Prehospital providers who might wish to measure IR TM
feedback to the hypothalamus that signals an
temperature at low ambient temperatures should be
increase in heat production.
aware that below 24.6°C, the TM readings will greatly
INDICATIONS AND CONTRAINDICATIONS
underestimate core temperatures.

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 14

PROCEDURE o Based on measurement of temperatures in normal, healthy


o Begin the temperature measurement by selecting the body infants, it is recommended that fever be defined as a rectal
site. temperature of:
o Insert the temperature probe and allow the probe to - 38°C or higher in infants younger than 30 days,
equilibrate with the temperature of the local body tissues. - 38.1°C or higher in infants 30 to 60 days (1 to 2
- Obtain sublingual oral temperatures in either the months)
right or the left posterior sublingual pocket with the - 38.2°C or higher in infants 60 to 90 days old (2 to 3
mouth closed. months)
- Obtain a rectal temperature with the patient in the o Hypothermia has been defined as a core body temperature
left or right lateral decubitus position and advance lower than 35°C (<95°F) with accompanying symptoms and
the probe gently to a depth of 3 to 5 cm to ensure signs.
accurate, atraumatic results. o Hyperthermia has been defined as a core body
o Complications associated with rectal temperature temperature higher than 41°C (>105.8°F) with
measurements are extremely rare and include rectal accompanying symptoms and signs.
perforation, pneumoperitoneum, bacteremia,
dysrhythmias, and syncope.
- Falsely low supranormal rectal temperature
measurements may be seen during shock.
- Rectal temperature may also lag behind changes in
core temperature.
o Digital electronic probes are commonly used for the
measurement of oral temperature in ambulatory patients.
o Disposable single-use oral thermometers are now
available and are as reliable as mercury or TM
thermometers.
o The average time needed to record a reading with a
pacifier thermometer is 3 minutes and 23 seconds, thus
making its application in emergency medicine limited.
o TM perforation and pain have been reported as
complications of placement of the thermistor probe in the
auditory canal.
INTERPRETATION o Interpretation of temperature measurements during
Normal values for body temperature are affected by the clinical assessment must consider the use of antipyretics,
following variables: level of activity, pregnancy, environmental exposure, and
1. site and methods used for measurement patient age.
2. perfusion o Body temperature is increased during sustained exercise,
3. environmental exposure pregnancy, and the luteal phase of the menstrual cycle.
4. pregnancy o Temperature also increases in the late afternoon because
5. activity level of diurnal variation.
6. time of day o Body temperature is generally reduced with advanced age,
Although core body temperature remains nearly constant and age may have an impact on the magnitude of fever.
(37.0°C ± 0.6°C or 98.6°F ± 0.18°F), surface temperature rises o Axillary temperatures have a low sensitivity but a high
and falls with changes in ambient temperature, exercise, and specificity for fever. Axillary temperatures should not be
time of day. used to screen for fever.
o The definition of fever varies by the site of measurement o Oral temperature measurements are affected by the
and is defined by a temperature greater than 2 standard ingestion of hot or cold liquids,227 tachypnea,257 and cold
deviations above the mean. ambient air.
- Fever has been defined as an oral temperature of o When an elevated temperature is suspected or crucial in
37.8°C or higher (100.0°F) decision making and not evident with an oral thermocouple
- rectal temperature of 38.0°C or higher (100.4°F) probe or IR TM thermometer, measurement with a
- An IR ear temperature of 37.6°C or higher (99.6°F) mercury in- glass thermometer is indicated.

Chapter 1: Vital Signs and Measurement


SECTION 1: Vital Signs and Patient Monitoring Techniques 15

o When rapid changes in body temperature occur, oral and


TM temperature measurements appear to be more reliable
than rectal temperature.
o Infrequently, ED patients require constant monitoring of
temperature (e.g., in cases of profound hypothermia or
hyperthermia). This can usually be performed by using a
bladder or esophageal probe attached to a potentiometer.
o Interpretation of ear IR temperatures requires knowledge
of the mode of thermometer operation and ambient
temperature.
- Occlusion of the ear canal by cerumen may produce a
falsely low reading.
PAIN AS A VITAL SIGN
o Pain is viewed as a mechanism to provide a mechanical
warning of actual or potential damage to cells and tissues
in a specific area.
o Pain can cause sympathetically mediated changes in vital
signs, but standard vital signs (pulse, respiration, blood
pressure) do not meaningfully correlate with the level of
perceived pain.
o Reliably quantifying pain should be the goal of ED clinicians
and is an appropriate step in the triage process.
PROCEDURE/INTERPRETATION
A patient’s self-report of the pain is considered the gold
standard for the initial assessment of pain and tracking of a
response to interventions, although such reporting does not
mandate specific interventions.
o Common instruments include the verbal rating scale,
numerical rating scale, visual analog scale (VAS), and
graphic rating scales.
o The most used pain scale in the ED is the 1 to 10 VAS.
- The VAS uses a 10-cm line bounded on each end by
perpendicular stops and descriptors.
- Zero equates to no pain, and 10 equates to the worst
pain ever experienced.
o There is wide variability in this technique, and at least a 13-
mm to 30-mm (1.3-cm to 3-cm) change on the scale is
required to validate clinically relevant worsening or relief.
o Graphic rating scales are useful for patients with limited
cognitive and expressive ability, especially children.
OVERALL GOAL OF PAIN RELIEF
The goal of pain management in the ED is to adequately relieve
or control pain without compromising diagnosis, treatment
plans, or the safety of the patient and population.
o Theoretically ideal objective is to totally relieve pain, this
goal is difficult to consistently achieve in the complex ED
milieu, and should not be the standard of care.

Chapter 1: Vital Signs and Measurement

You might also like