Step 1: Selection of measurement point & pipe preparation Use the cursor keys to select
Setup Wizard Single for single
● Avoid installation of sensors in the vicinity of deformations and
channel measurement (one set
pipe defects, near welding seams or where deposits could have
of clamp-on sensors).
accumulated. Select Setup Wizard Dual, for
● Select a measuring point with sufficient straight pipe to obtain dual channel measurement (two
accurate measurements, generally a length of ten times the sets of clamp-on sensors).
pipe diameter on the inlet of the measuring point, and a five Confirm selection by pressing
times pipe diameter on the outlet. <ENTER>.
● For a horizontal pipe, mount the sensors on the side of the If the sensors are recognised,
pipe. For a vertical pipe, mount the sensors at a location where the serial number will be shown.
the liquid flows upwards. (Pic 1) If they are not recognised or not
connected, they may be selec-
● Mount the sensors in the direction of the flow. (Pic 2)
ted from a list.
● Clean the pipe at the measuring point. Remove loose paint and
rust with a wire brush or file. Select the units of measurement
● Apply coupling paste to the bottom of the clamp-on sensors using cursor keys and confirm
before installation. Leave straps slightly loose so sensors can by pressing <ENTER>.
be moved along the pipe during the positioning process.
1/3
Subject to change without prior notice. I Copyright © Katronic Technologies Ltd. 2010. I Issue: KF150QuickStartV22E0710
KATflow 150 Quick Start Manual
Katronic Technologies Ltd I 23 Cross St I Leamington Spa I UK I Ph. +44(0)1926 882954 I info@katronic.co.uk I www.katronic.co.uk
Step 3: Quick Start Function and Setup Wizard (Continued) Sensor Positioning Screen
Select pipe lining material using Mount the transducers with the
cursor keys. suggested spacing. This has
Confirm by pressing <ENTER>. been determined by the flow
meter on basis of to the entered
parameters.
The suggested sensor spacing
is the distance from the inside
Select transducer configuration edge of each sensor.
(number of passes) using cursor Use the middle line between the
keys. signal indication bars for the fine
Auto: Automatic selection by adjustment of the sensor moun-
meter according to the entered ting position. A central position
parameters (number of passes of this line is desired.
shown at later stage on Sensor If the line between the bars is to
Positioning Screen). the left-hand side of the lower
1: 1 pass, diagonal mode middle line, the sensors are too
2: 2 passes, reflection mode close to each other. If the line
3: 3 passes, diagonal mode between the bars is to the right,
4: 4 passes, reflection mode the sensors are too far apart.
5: 5 passes, diagonal mode Observe the upper bar (signal-
6: 6 passes, reflection mode to-noise ratio) and lower bar
..etc. (signal quality). These should be
Confirm by pressing <ENTER>. of identical length and as far to
the right as possible.
Confirm by pressing <ENTER>
1. Reflection Mode to obtain measurements.
The flow meter uses an even number of passes. This is the most
convenient mounting arrangement, as the transducer separation
Process Value Display, Totaliser and Datalogger
distance can be measured very easily and the sensors can be
accurately aligned. Use whenever possible. (Pic 3) The main process value can be
changed in the Quick Start or
2. Diagonal Mode
Output section (Output-Display-
The signal travels on an odd number of passes through the pipe. Units). Press <ESC> to return to
Used for larger pipes where greater signal attenuation might the main menu. Change to the
occur. The sensor distance on this mounting configuration can be diagnostic display by pressing
negative (sensors overlapping). (Pic 4) <DISP>. Cycle through available
flow channels using <MUX>.