Some possible explanations why the error has occurred:
- No connection between transducer board and FM I/O board
- No connection between FM I/O board and protective driver board - Transducer board mounted in incorrect position on FM I/O board - Incorrect transducer ID. Please update transducer ID (I2C scan). - Transducer board is broken - Serial bus error on protective CPU board - Uncalibrated transducer. Please calibrate transducer - Broken FM I/O board
P PTRA 000 104 ERROR READING PARAMETERS FROM VENOUS PRESSURE
TRANSDUCER Unconditional error, blood pump not available
General description of conditions for occurring:
- An error occurred while reading transducer parameters during init (start-up).
Some possible explanations why the error has occurred:
- No connection between transducer board and BM I/O board - No connection between BM I/O board and protective driver board - Transducer board mounted in incorrect position on BM I/O board - Incorrect transducer ID. Please update transducer ID (I2C scan). - Transducer board is broken - Serial bus error on protective CPU board - Uncalibrated transducer. Please calibrate transducer - Broken BM I/O board
P PTRA 000 105 ERROR READING PARAMETERS FROM PH TRANSDUCER
Unconditional error, blood pump not available
General description of conditions for occurring:
- An error occurred while reading transducer parameters during init (start-up).
Some possible explanations why the error has occurred:
- No connection between MM I/O board and protective CPU board - Incorrect transducer ID. Please update transducer ID (I2C scan). - MM I/O board is broken - Serial bus error on protective CPU board - Uncalibrated transducer. Please calibrate transducer
P PTRA 000 106 ERROR READING PARAMETERS FROM ARTERIAL PRESSURE
TRANSDUCER Unconditional error, blood pump not available
General description of conditions for occurring:
- An error occurred while reading transducer parameters during init (start-up).
Some possible explanations why the error has occurred:
- No connection between transducer board and BM I/O board - No connection between BM I/O board and protective driver board - Transducer board mounted in incorrect position on BM I/O board - Incorrect transducer ID. Please update transducer ID (I2C scan). - Transducer board is broken - Serial bus error on protective CPU board - Uncalibrated transducer. Please calibrate transducer - Broken BM I/O board
HCEN12371 Rev. 10.2011 AK 96® - Error Codes 107 (115)
Program version 3.50 P PTRA 000 107 ERROR READING PARAMETERS FROM DIASCAN TRANSDUCER Unconditional error, blood pump not available
General description of conditions for occurring:
- An error occurred while reading transducer parameters during init (start-up).
Some possible explanations why the error has occurred:
- No connection between transducer board and FM I/O board - No connection between FM I/O board and protective driver board - Transducer board mounted in incorrect position on FM I/O board - Incorrect transducer ID. Please update transducer ID (I2C scan). - Transducer board is broken - Serial bus error on protective CPU board - Uncalibrated transducer. Please calibrate transducer - Broken FM I/O board
P SDT, CPU P system
P SDT 003 001 CHECK SUM ERROR Unconditional error, blood pump not available
General description of conditions for occurring:
- The check sum of the protective system presets is not correct. Enter service/ preset and preset the machine (make a default preset).
Some possible explanations why the error has occurred:
- Broken/ new CPU P safe-RAM
P SDT 004 001 CPU C DOES NOT SEND SHUT DOWN ACKNOWLEDGE
Some possible explanations why the above error has occurred:
- Internal software error
P SDT 004 002 CPU W DOES NOT SEND SHUT DOWN ACKNOWLEDGE
Some possible explanations why the above error has occurred:
- Internal software error
P SDT 004 003 SHUT DOWN FAILURE
General description of conditions for occurring:
- The AK 96 was not switched off within 5 seconds after shut down request.
Some possible explanations why the above error has occurred:
- Protective driver board broken (IC16 and components)
P SDT 004 004 CPU C DOES NOT SEND RESET ACKNOWLEDGE
Some possible explanations why the above error has occurred:
- Internal software error
P SDT 004 005 CPU W DOES NOT SEND RESET ACKNOWLEDGE
Some possible explanations why the above error has occurred:
- Internal software error
108 (115) AK 96® - Error Codes HCEN12371 Rev. 10.2011