Chapter 8. Disassembly and Reassembly
Chapter 8. Disassembly and Reassembly
8.1 Overview
Chapter 8 describes how to disassemble ACCUVIX V10.
Refer to this chapter when you upgrade or repair the hardware.
WARNING The system contains dangerous high voltage. Never disassemble the system.
There is a risk of electric shock and injury.
The repair of the system and the replacement of parts must be carried out by
an authorized engineer or the customer service department of MEDISON Co.,
Ltd.
The company is not responsible for any injury and damage caused by not
following this warning.
WARNING When working with the system on, do not wear a static electricity protective
wristband. There is a risk of electric shock and injury.
NOTE When disassembling or reassembling the system, wear static electricity protective
gloves and a wristband.
These will prevent any accidents due to carelessness, and damage to the system
due to static electricity.
8.2.1 Preparations
Prepare a (+) screwdriver and static electricity protective gloves.
Shut down the system referring to “3.7 System Shutdown (Power Shut Down)”.
Cover Foot
Screw
Screw
Screw
Screw
②
Screw
① ②
Screw
3) Disassemble the Cover Top Open referring to “8.2.7 Cover Top Open”.
4) Remove the 6 screws of the Handle Bar using the (+) screwdriver.
5) Hold and pull the Handle Bar to separate it.
6) Remove the 4 screws of the Cover Top using the (+) screwdriver.
7) Hold and pull the Cover Top to separate it.
Screw
8.3.1 Preparations
Prepare a (+) screwdriver and static electricity protective gloves.
Prepare an LCD Monitor Panel protective cover.
Shut down the system referring to “3.7 System Shutdown (Power Shut Down)”
WARNING Although there is a catch to secure the LCD Panel, it may come off and fall,
which will damage it.
When disassembling or assembling the LCD Monitor, it is recommended to do it
with another person.
6) Hold the LCD Panel with both hands and disassemble it.
7) Separate the cable from the AD Board.
8) Remove the 4 screws from the Cover EMI LCD using the (+) screwdriver and
separate it.
9) Remove the 7 screws from the Cover EMI LCD and AD Board using the (+)
screwdriver and separate it.
Screw
Screw
③
Screw
Screw
Cable
④ ⑤
Screw
Screw
Cable Screw
8.4.1 Preparations
Prepare a (+) screwdriver and static electricity protective gloves.
Shut down the system referring to “3.7 System Shutdown (Power Shut Down)
2) Remove the 10 screws from the PSA ASSY using the (+) screwdriver.
3) Hold the PSA ASSY with both hands and pull it to disassemble it.
WARNING
Since the edge of the PSA ASSY is sharp, disassemble it with care.
PSA ASSY
Screw
1) Disassemble the Cover Front Lower referring to “8.2.2 Cover Front Lower”.
2) Disassemble the PSA ASSY referring to “8.4.2 PSA ASSY”.
3) Hold the handle of the CW Board with both hands and pull it to separate it.
4) Hold the handle of the BF Board with both hands and pull it to separate it.
5) Hold the handle of the BE Board with both hands and pull it to separate it.
6) Hold the handle of the Motor Control Board with both hands and pull it to
separate it.
CW Board
BF Board
BE Board
Motor Control
Board
LCD IF Board
Screw
8.5.1 Preparations
Prepare a (+) screwdriver and static electricity protective gloves.
Shut down the system referring to “3.7 System Shutdown (Power Shut Down)”
8.5.2 HDD
1) Disassemble the Cover Back Upper referring to “8.2.3 Cover Back Upper”.
2) Disassemble the Cover Back Lower referring to “8.2.4 Cover Back Lower”.
3) Separate the HDD S-ATA Cable and Power Cable.
4) Remove the 4 screws of the EMI Cover HDD using the (+) screwdriver and
separate it.
5) Separate the HDD from the EMI Cover.
HDD
Screw
8.6.1 Preparations
Prepare a (+) screwdriver and static electricity protective gloves.
Shut down the system referring to “3.7 System Shutdown (Power Shut Down)”.
8.6.2 PC Part
1) Disassemble the Cover Back Upper referring to “8.2.3 Cover Back Upper”.
2) Disassemble the Cover Back Lower referring to “8.2.4 Cover Back Lower”.
3) Separate all cables connected to the PC Part.
4) Remove the 2 screws of the PC Part using the (+) screwdriver.
WARNING
Since the edge of PC Part is sharp, disassemble it with care.
Screw
Screw
Screw
Screw
Screw
VGA Card
PCI Board
Cable
Rear Panel R Rear Panel L
①
②
Screw
Screw
Screw Screw
8.7.1 Preparations
Prepare a (+) screwdriver and static electricity protective gloves.
Shut down the system referring to “3.7 System Shutdown (Power Shut Down)”.
WARNING Since the ADM is heavy, take care when disassembling it. If the ADM falls
suddenly, it may cause an injury.
AC to AC Power
Cable ①
Screw
DC to DC Power Module
Screw
PC Power Supply
Screw
8.8.1 Preparations
Prepare a (+) screwdriver and static electricity protective gloves.
Shut down the system referring to “3.7 System Shutdown (Power Shut Down)”
Screw Screw
1) Disassemble the upper board of the Control Panel referring to “8.8.2 Control
Panel”.
2) Separate all cables connected to the Key Matrix Board.
3) Take the Switch Cap off the Control Panel.
4) Remove all the screws of the Key Matrix Board using the (+) screwdriver and
separate it.
Cable
Switch Cap
Track Ball
Screw
Cable
Screw
8.9.1 Preparations
Prepare a (+) screwdriver and static electricity protective gloves.
Shut down the system referring to “3.7 System Shutdown (Power Shut Down)”.
8.9.2 DVD
1) Disassemble the Cover Front Lower referring to “8.2.2 Cover Front Lower”.
2) Disassemble the Cover Side Right & Left referring to “8.2.6 Cover Side Right &
Left.
3) Disassemble the Cover Top Open referring to “8.2.7 Cover Top Open”.
4) Disassemble the Cover Top and Handle Bar referring to “8.2.8 Cover Top &
Handle Bar”.
5) Separate the cable connected to the DVD.
6) Remove the 2 screws fixed the DVD using the (+) screwdriver.
7) Hold the DVD and lift it to separate it.
Cable Screw
8.10.1 Preparations
Prepare a (+) screwdriver and static electricity protective gloves.
Shut down the system referring to “3.7 System Shutdown (Power Shut Down)”.
2) Disassemble the Cover Side Right & Left referring to “8.2.6 Cover Side Right &
Left.
3) Disassemble the Cover Top Open referring to “8.2.7 Cover Top Open”.
4) Disassemble the Cover Top and Handle Bar referring to “8.2.8 Cover Top &
Handle Bar”.
5) Separate the cable connected to the ECG Module.
6) Remove the 3 screws fixed the ECG Module using the (+) screwdriver.
7) Hold the ECG Module and lift it to separate it.
Cable Screw
8.11.1 Preparations
Prepare a (+) screwdriver and static electricity protective gloves.
Shut down the system referring to “3.7 System Shutdown (Power Shut Down)”.
Speaker
Cable Screw