amx 4 cal service password

MAS up, MAS down, KV down and KV up. Press these 4 buttons in this order and that is your
password.

You need to flip a toggle switch located inside the unit,behind a film bin on the left side, and turn a
key switch on. You will the see an"ENS SERVC MODE" message on the display. By pressing a kVp
upper arrow kew scroll up until you see the prompt "CALIBRATE SYSTEM" and then press mAs down
arrow key, to confirm selection.

If I remember correctly, test 4 error means that your battery charger need calibration.

Hope it helps,

Power up error at test 4 is calibration data base checksum error. Calculates check sum for calibration
data base and compares it with the stored value. Test fails if they are not equal. Entering calibration
will give a more specific error code for the check sum failure. You can also do a checksum by-pass.

Checksum by-pass: Power off. Turn service switch as described by Val Vaysberg to "RUN" mode [up].
Power on. Turn service switch to "SERVICE" mode [down], before "test 04-failed". Turn service
switch back to "RUN" mode [up], before "testing complete". You are in application mode with all
functions enabled.

In application mode, the application error code will let you know how to troubleshoot the problem.