Factors Affecting Measurement of Equilibration Time


of Dry Goods Loads in Autoclaves
Soham Shah, Sara Alayli and Yiming Peng
PDA Journal of Pharmaceutical Science and Technology February 2019, pdajpst.2018.009225; DOI:
https://doi.org/10.5731/pdajpst.2018.009225

Article References Info & Metrics  PDF

Abstract

Equilibration time of an autoclave is defined as the period of time that elapses between the attainment of
sterilization temperature at the reference measurement point of the autoclave chamber and the attainment of
sterilization temperature at all measurement probes within the dry goods load placed within the autoclave during a
sterilization cycle. Equilibration time is an indicator of the ability of the autoclave to remove air from and
subsequently sterilize the dry goods load placed within the autoclave chamber. Therefore equilibration time is an
indicator of the efficacy of an autoclave cycle. Periodic measurements of equilibration time and evaluation of the
equilibration time data can assure continued efficacy of the autoclave cycle. For autoclaves that support
biopharmaceutical manufacturing plants, equilibration time is generally measured by placing heat penetration probes
within dry goods load items placed within the autoclave during a sterilization cycle. However, equilibration time
measured in this fashion yields results that are highly variable and difficult to replicate. This paper presents a study
that attempts to understand the impact of the following autoclave variables on consistent measurement of
equilibration time and reduction of equilibration time: 1. Type of dry good load item placed inside the autoclave 2.
Location of the heat penetration probe placed within the dry good load item 3. Location of the dry good load item
inside the autoclave chamber 4. Orientation of the dry good load item inside the autoclave chamber 5. Number of
coils of the tubing elements of a dry good load item 6. Total thermal mass of the dry good load items placed inside
the autoclave 7. Temperature of the dry good load items at the start of the sterilization cycle 8. Presence of residual
moisture in the dry good load items at the start of sterilization cycle 9. Number of steam and vacuum pulses used
for pre-conditioning in the autoclave recipe 10. Depth of steam and vacuum pulses used for pre-conditioning in the
autoclave recipe 11. Hold time of steam and vacuum pulses used for pre-conditioning in the autoclave recipe The
focus of this paper is on tests performed to evaluate these variables followed by analysis and interpretation of test
data to identify key variables that have the highest impact on measurement of equilibration time. The conclusion of
this paper presents a process to achieve shorter equilibration times in autoclaves through control of key variables
identified.

Autoclave Autoclave Equilibration Time Equilibration Time PDA TR1 PDA TR48 Sterilization of Dry Goods

Received July 2, 2018.

Accepted January 15, 2019.

Copyright © 2019, Parenteral Drug Association

PDA members receive access to all articles published in the current year and previous volume year. Institutional
subscribers received access to all content. Log in below to receive access to this article if you are either of these.  

If you are neither or you are a PDA member trying to access an article outside of your membership license, then you
must purchase access to this article (below). If you do not have a username or password for JPST, you will be
required to create an account prior to purchasing. 

Full issue PDFs are for PDA members only.

NOTE TO PDA.ORG USERS

The PDA and PDA bookstore websites (www.pda.org and www.pda.org/bookstore) are separate websites from the
PDA JPST website. When you first join PDA, your initial UserID and Password are sent to HighWirePress to create
your PDA JPST account. Subsequent UserrID and Password changes required at the PDA websites will not pass on
to PDA JPST and vice versa. If you forget your PDA JPST UserID and/or Password, you can request help to retrieve
UserID and reset Password below.. 

Log in using your username and password

Username Password

Log in

You might also like