Professional Documents
Culture Documents
VSM Moviplan Technical Manual ENG PDF
VSM Moviplan Technical Manual ENG PDF
Technical Manual
Release 3 October 2007 (Rev. 7)
TECHNICAL MANUAL
Revision history
Revision history
2 21.01.02 i, ii, 3-1, 6-13, 11-1, 11-13, New Potter Bucky SD version
11-23, 11-25, 11-27, 11-29, (120/230V version).
11-31, 11-33, 11-35, 11-37, Schematics update.
11-39, 11-41, Addition of "Potter Bucky schematic
from 12-1 to 12-22 diagram" and "Spare parts" chapter.
(Ref. RDM 5131, 5227, 5239, 5279)
Contents
1. INTRODUCTION 1-1
1.1 Icons appearing in the manual ............................................................ 1-1
2. SAFETY ASPECTS 2-1
2.1 Warnings ............................................................................................. 2-2
2.2 Radio-protection instructions .............................................................. 2-4
2.3 Environmental risks and disposal........................................................ 2-5
2.4 Symbols used ...................................................................................... 2-6
3. DESCRIPTION 3-1
3.1 Identification labels ............................................................................. 3-1
3.2 Configuration of the MOVIPLAN 800S.................................................. 3-2
4. TECHNICAL DATA 4-1
4.1 Dimensions.......................................................................................... 4-6
4.2 Applied safety regulations.................................................................... 4-7
5. PRE-INSTALLATION 5-1
5.1 Electrical wiring set-up ........................................................................ 5-2
5.2 Packaging ............................................................................................ 5-3
5.3 Space requirements ............................................................................. 5-3
6. INSTALLATION 6-1
6.1 Unpacking and components inspection ............................................... 6-1
6.2 Rails assembly..................................................................................... 6-2
6.3 Table positioning and fastening ........................................................... 6-3
6.4 Column assembly ................................................................................ 6-5
6.5 Tubehead, Control Panel and Collimator assembly .............................. 6-8
6.6 Cables connection.............................................................................. 6-10
6.6.1 Connection from column ................................................................ 6-10
6.6.2 Connection MOVIPLAN 800S table - Cabinet................................... 6-10
6.6.3 Other connections .......................................................................... 6-10
6.7 Table top assembly ............................................................................ 6-12
6.8 Pedals connections ............................................................................ 6-12
6.9 Final balancing and adjustment ........................................................ 6-13
6.10 Electrical connections........................................................................ 6-14
7. INSPECTION, CALIBRATION AND ADJUSTMENT 7-1
7.1 Inspection of the A3 logic card power supply ....................................... 7-1
7.2 Inspection of angle reference potentiometer ......................................... 7-1
7.3 Column stop angles control in tomography.......................................... 7-2
7.4 Inspection and calibration of tomographic angles ................................ 7-3
1. INTRODUCTION
NOTE:
* The present manual is updated for the product it is sold with in order to
grant an adequate reference in performing diagnostics and repair
operation normally carried out by the service engineer.
The manual may not reflect changes to the product not impacting service
operations or use mode.
The scope of this Manual is to supply user instructions for a safe and
efficient operation of the device described.
This icon indicates a WARNING: the items marked by this icon refer to
safety aspects of the patient and/or of the operator.
2. SAFETY ASPECTS
WARNING:
All the information provided in this chapter must be read carefully, and,
where so required, must be implemented to prevent damage to the
equipment and injury to people.
VILLA SISTEMI MEDICALI designs and builds its devices complying with
the related safety requirements; furthermore it supplies all information
necessary for a correct use.
• use of the MOVIPLAN 800S device for other than its intended use,
• damage to the device or injury to the operator or patient due to
incorrect installation and maintenance performed other than in
compliance with the procedures defined in the Technical Manual
provided with the device or due to incorrect operating techniques,
• mechanical and/or electrical changes made during and after
installation.
Only the authorised personnel can remove the covers and/or have
access to the components under voltage.
2.1 Warnings
The MOVIPLAN 800S is suitable for use in a hospital environment and
cannot be connected to a "domestic" electrical system.
The device has not been designed to be used in the presence of steam,
anaesthetic mixtures which are flammable in air, oxygen or nitrous
oxide.
The operator MUST ALWAYS PAY THE UTMOST CARE when the
motorised mechanisms are activated such as the vertical movement of
the table top or of the column. The red EMERGENCY button on the front
of the device can be pressed in case of danger. This button immediately
stops all movements and all functions and provides the operator with an
important safety tool in addition to the other intrinsic safety devices of
the device.
Before activating the movement of the column, check that the patient is
placed correctly on the table top with the legs and arms inside the
support area.
Although the MOVIPLAN 800S table was designed with a high degree of
protection against electromagnetic interference, it is necessary to install
it a certain distance away from electric energy transformers, continuity
static groups, pack units (walkie-talkies) and cellular phones. Use of the
latter is only permitted at a distance from the device as provided in the
table below.
Electromagnetic
Conformity EMC Environment
Emissions
All parts that come into contact with the patient must be cleaned,
and where necessary, disinfected.
2.2 Radio-
Radio-protection
protection instructions
Villa Sistemi Medicali designs and manufactures device in compliance
with the safety requirements. Furthermore, the Company supplies all the
necessary information and warnings relative to the dangers associated
with devices connected to X-ray generators.
Before activating the X-rays from the generator table, the operator shall
always check the alignment of the tubehead with respect to the image
receptor and should set the maximum aperture of the collimator in
accordance both with the format of the radiography cassette used and
with the distance of the cassette from the focal point of the tubehead.
Use the farthest focal point-skin distance compatible with the X-ray
examination to be performed.
WARNING:
Before dismantling the column, lock the column's tube-carrier arm. Then
simultaneously remove the radiogenic-collimator tube and the counter-
weights of the sliding mechanism inserted in the column.
The crossed out wheeled bin symbol placed on the equipment or its
package means that the equipment must be collected separately from the
other waste.
The separate collection of this equipment at the end of its life cycle is
managed by the manufacturer/distributor. The user that want to discard
this equipment should contact the local distributor and follow the system
he adopted for the separate collection.
The correct separate collection for the subsequent recycling, treatment
and environmentally compatible disposal contributes to avoid negative
effects on the environment and on the health and promotes the reuse
and/or recycling of the materials that the equipment is made of.
The illegal disposal of the equipment made by the user will cause the
application of sanctions according to the local regulations.
Symbol Description
∼ Alternating current
Protection grounding
Operation grounding
Dangerous voltage
3. DESCRIPTION
2
4
1 3
MOVIPLAN 800S Potter Bucky SD
characteristics label characteristics label
2 4
Manual Collimator (Villa model) Collimator screening
characteristics label (*) label
Grids
• Standard (Ratio 10, focal length 120 cm, 34 pl/cm) (code
4695102600).
• High performance (Ratio 12, focal length 120 cm, 36 pl/cm) and low
absorption (code 4695101600).
Accessories
The following accessories are available for all models on request:
• Single leg rest (code 7173142000).
• Band stretcher with compression band (code 7173129000).
• External side cassette carrier (code 7173115000).
4. TECHNICAL DATA
General characteristics
Type of device MOVIPLAN 800S models SF-SA
Manufacturer Villa Sistemi Medicali S.p.A.
20090 Buccinasco (MI) Italy
Type of device and classification in accordance Class I with type B applied
with Standard IEC 60601-1 parts
Type of device and classification Class I - CE 93/42
Degree of protection in accordance with IP00
standard IEC 60529
Operating mode Continuous
Mechanical characteristics SF SA
Table weight 175 kg 205 kg
Table height 790 mm
Table length with support centred 2180 mm
Table width with support centred 1030 mm
Column stand weight 290 kg 315 kg
Column height 2400 mm
Column width (base) 1070 mm
Column depth (with arm extended) 1400 mm
Electrical characteristics SF SA
Voltage 230 Vac
Frequency 50/60 Hz
Maximum current 2.8 A 4.2 A
Device protection fuse 3AT
Potter protection fuse 0.2 A T
Collimator lamp protection fuse 6.25 A T
Power 645 VA 965 VA
Line impedance < 1.0 Ω
Voltage drop in load <1% <2%
Movements SF SA
Manual longitudinal movement of the potter yes
with electromagnetic braking
Manual longitudinal movement of the table top yes
with electromagnetic braking
Manual transversal movement of the table top yes
with electromagnetic braking
Motorised vertical movement of the table top no yes
(elevator) with pedal command
Motorised adjustable Tomo Layer height yes
Motorised longitudinal movement of the yes
column with electromagnetic braking
Manual vertical movement of the column tube yes
carrier arm with electromagnetic braking
Manual transversal movement of the column yes
tube carrier arm with electromagnetic braking
and central mechanical block
Manual rotation movement of the column with yes
mechanical stops
Manual rotation movement of the column tube yes
carrier arm with electromagnetic braking
Movements distance SF SA
Potter longitudinal movement 540 mm
Table top longitudinal movement 415 mm head end
415 mm feet end
Table top transversal movement 126 mm internal
126 mm external
Table top vertical movement (elevator) min. NA 460 mm
distance Table top-Floor
Table top vertical movement (elevator) max. NA 790 mm
distance Table top-Floor
Column longitudinal movement 1710 mm
Tube carrier arm vertical movement 1470 mm
Minimum focal point-floor distance (in function 430 mm
of tube used)
Maximum focal point-floor distance (in 1900 mm
function of tube used)
Maximum focal point-film distance (in function 1200 mm 1530 mm
of tube used)
Tube carrier arm transversal movement 190 mm
Column rotation movement 180° (± 90°)
Tube carrier arm rotation movement 360° (± 180°)
Tomography SF SA
Linear plane-plane tomography yes
Tube-potter connection With tomography bar
Tomographic movement From head end to feet end
Focal length - fixed film in tomography 1210 mm
Tomography not set signal Buzzer
Tomographic layer movement Motorised
Layer movement range From 0 to 250 mm
Indication of layer height on patient Luminous spot
Display of layer height Display
Tomography angles 5° - 20° - 30° - 45°
Tomography speed 3
Tomography examination enable Only with all settings activated
Tomography times See table
Tomography times table
Accessories SF SA
Handles 1
Leg rests for gynaecology examinations Optional
Compression band with band stretcher Optional
Cassette support for side projections Optional
Environmental conditions SF SA
Operating environmental conditions Temperature: from +10° to +40° Celsius
Humidity: from 30% to 75 %
Pressure: from 700 hPa to 1060 hPa
Environmental conditions for Temperature: from -20° to +70° Celsius
transportation and storage Humidity: ≤ 95% non condensing
Pressure: > 630 hPa
4.1 Dimensions
Figure 4-1
• General safety:
IEC 601-1
IEC 601-2-32
• Electromagnetic compliance:
IEC 601-1-2
5. PRE-
PRE-INSTALLATION
The following pages provide instructions for the correct installation of the
MOVIPLAN 800S.
The supplier can provide the necessary assistance and technical
servicing as from the pre-installation phase.
The room must be prepared for the installation of the MOVIPLAN 800S
as indicated in the Figure 5-1.
The standard installation of the MOVIPLAN 800S requires the cables to
be enclosed in channels. Please note the maximum distance allowed
between distribution box "A" and distribution box "B" (centre of the
electric cabinet which is available only in the version with elevator).
When installation is required with distances greater than those shown in
the figure, electric connecting cables of suitable lengths must be
requested when issuing out the order.
Requirements for a correct installation are:
• Room minimum height of 3 m.
• For transportation of the machine inside the room, the door sizes
must be as follows:
– 700 mm for the column
– 1200 mm for the packaged materials.
• To obtain a proper cable setting of cables coming from the column,
the table must be positioned close to a wall.
A plastic duct (section 100x100) has to be fixed on the wall aligned
with the center of the rail: its lower end must be corresponding to the
"B" collector on the floor; its upper end must be high enough so that
the cables do not interfere along the entire range of motion of the
column along the rail.
6x6" (150x150)
(0.5m)
Cabinet
40.5x21.6x20.5"
(1030x550x520)
Cable entry
6x6" (150x150)
14" (355)
39.5" (1002)
min 6.5ft
118.5" (3010)
(2m)
Figure 5-1
WARNING:
The general ground system must comply with the laws in force.
Insufficient grounding of the device can be extremely dangerous for
the operator and the patient and can cause malfunctions in the
electrical equipment connected to the MOVIPLAN 800S.
The device's power cable must either be connected to the X-rays
generator in series connection or to a differential thermal magnetic
switch with the following characteristics:
250V - 6A - 0.03A.
A connecting line for the emergency switch of the MOVIPLAN 800S
table with the radiology room's general mains switch must also be
installed.
See chapter 4 “Technical data” for current and absorbed power data.
5.2 Packaging
The MOVIPLAN 800S is packed and shipped in one single case.
The dimensions, net weight, gross weight and contents of this case are
listed below:
Weight
Case type Contents
Net Gross
- Table base + Potter Bucky
- Table top
- Front floor rail
Moviplan 800 - Rear floor rail
SA and SF - Tube column + tube arm 680kg 900kg
330x120x110 - Collimator
- Push button control panel
- Cables
- Cabinet
6. INSTALLATION
6.1 Unpacking
Unpacking and components inspection
Upon receipt of the machine, check that the packaging has not been
damaged.
Wait for the Villa Sistemi Medicali technician or area representative
before opening the packaging and check the following:
• the goods correspond to the items ordered
• no damage was suffered during transportation.
If there is any damage or missing parts inform the shipper, Villa Sistemi
Medicali or Villa Sistemi Medicali's area representatives immediately,
giving full details of the problem.
The unit is fully tested in the factory to reduce installation time and to
optimise its functions.
NOTE:
* The device is shipped in pre-assembled groups. The mechanical assembly
consists exclusively in the assembly of these groups. Regulation of the
bearings, brakes and clamp torques is therefore made before delivery.
Any unnecessary intervention on these parts can cause malfunction of
the equipment.
NOTE:
* The operations described in this chapter must be performed for both
versions of the MOVIPLAN 800S.
Position the front rail in the room as shown in the figure, taking into
account the size of the equipment and the cable incoming distribution
box as indicated in the pre-installation (chapter 5).
Fasten the rail to the floor with the expanding plugs provided.
Check that the rail is level and if not, level using the spacers provided.
Fasten the back rail to the front rail, level and fasten to the floor using
the ø8 plugs provided.
Rear rail
Fixing point
between rails
Front rail
Bolts Ø8
580 (22.8") 300 (11.8") 300 (11.8") 580 (22.8") (Ø.31")
= =
Figure 6-1
Rail covers
Table
Cable exit
Cover plate
Figure 6-2
NOTE:
* The table is positioned asymmetrically with respect to the front rail to
allow the column to travel outside the space of the table on one side. It is
possible to choose the side towards which the column can exit the table
dimension, by positioning the table towards one or the other side.
The length of the table covered by tomography function is different in the
two position of about 20cm.
Mounting the table towards the left the max tomography range is
achieved.
Position the table in correspondence to the holes on the front rail. Fasten
the table to the floor with the plugs provided and to the rail with the two
TSEI M8 x 25 screws.
Table
Front rail
Fixing points
Table-Rail
Figure 6-3
Check that the table is level. If necessary, use the spacers provided.
Verify the alignment of all parts, paying especially attention to:
• the alignment of the front rail with the rear rail
• the alignment of the front rail with the table
• the levelling of both rails and the table on the floor.
WARNING:
Remove the two supports, that are necessary during the shipment,
positioned on the lateral parts of MOVIPLAN 800S base. Do not re-install
the relevant screws.
NOTE:
* On the back of the tomography tower seat, there are two threaded holes
where the two clips used to store the tomography shaft lock, when not in
use, have to be screwed in. The clips are shipped in a plastic bag
positioned close to the tomography tower seat.
Wrapped to the tomography shaft, you find another clip that must be
screwed in the back of the column, just below the counterweights cover,
used to store the tomography shaft when not in use.
End stop
Figure 6-4
• remove the external cover plate from the rear of the column
• set two counter-weights in position in the column and check the
balancing of the tube cross-arm movement up and down the
column
• lift and secure the cross-arm in its highest position with a clamp
so that the counter-weights are at their lowest position inside the
tube column. In this way, there is less danger of the counter-
weights falling when you position the column.
Cover
Counterweight
Figure 6-5
2. Slide the column onto the rear rail making sure that the bearings
(two for each side of the column) locate on the guide and the column
wheel slides smoothly on the front guide. There is no need to make
any adjustment to the bearings as they are factory tested before
shipping to ensure correct operation.
NOTE:
* If further regulation is needed, a special hexagonal wrench, 17 mm, with
thickness lower than 4.5 mm must be used.
Column
bearing
Side
bearings
Figure 6-6
End stop
Security pin
Figure 6-7
End run up
Figure 6-8
WARNING:
On the column it is possible to use X-ray tubes with a maximum weight
of 25kg.
1. To fit the tube and housing on the column cross-arm, first take out
the locking pin from the tube assembly. Next insert the tube
assembly on the cross-arm pin and re-fit the locking pin temporarily.
Cross-arm
pin
Tube
assembly
Locking
pin
Figure 6-9
Base +
Push button panel
Fixing cone
Cable
passage
Figure 6-10
NOTE:
* Please take care that international regulatory standards require that total
filtration of X-ray beam must be greater than 2.5mm Al.eq.
Villa collimator has a filtration of 0.5mm Al.eq.; Ralco collimator has a
filtration of 1.0mm Al.eq.
Taking into account inherent filtration of the X-ray tube, as reported on
its label, add necessary filtration using the aluminium disks provided in
the kit P/N 6674893000 with the system.
These disks can be inserted removing the ring "A" in Figure 6-10.
Collimator
Locking
points
Cross-arm Tube
assembly
Figure 6-11
NOTE:
* In order to grant adequate correspondence between light field and X-ray
beam the distance between focal spot of the X-ray tube and contact plane
of the collimator with the X-ray tube cone must be 80mm.
If needed use 1mm spacers provided in the kit P/N 6674893000 with the
system to bring this distance to the correct value.
NOTE:
* The current position of the collimator may not grant perfect alignment of
the anode focus and the collimator itself.
A regulation may be needed as described in paragraphs 7.10 or 7.11.
• Cable X15 from control panel to the cabinet (A4 board – connector
X15)
• cable X27 from Control Panel to cable X27 from MOVIPLAN 800S
table (Tomo device) using the extension cable (code 6271812700) if
necessary
• cable PRL9-1 and PRL9-2 from column to cabinet (X0 terminal board
– pins X0.12 and X0.13)
• cable X16 from column to the cabinet (A4 board – connector X16)
• cable X31 and ground cable from the tube arm to PRL9 terminal
board on the top of the column
• ground cables to the ground connections in the cabinet.
• Cable X14 from MOVIPLAN 800S table (A5 board – connector X14)
to cabinet (A4 board – connector X14)
• cable X17 from MOVIPLAN 800S table (A5 board – connector X17) to
cabinet (A4 board – connector X17)
• cables X0 from MOVIPLAN 800S table (A5 board) to cabinet (X0
terminal board)
• ground cables to the ground connections in the cabinet.
• HT cables, rotating anode cable and ground cable from X-ray tube to
generator
• for Villa collimator, collimator cable between 9 pins connector and
X30 on control panel (see instruction note 39119001). For Ralco
collimator, collimator cable between X30 on control panel and
terminal board on collimator (see specific Manual).
NOTE:
* Make reference to the System Integration Manual for electrical
connection to the generators and the other components.
Column
connections
HT cables
Control Panel
X15
Control Panel
cable
X31
X14= Table
X15= Control Panel
X16= Column
X17= Table
X27= Layer height potentiometer
PRL9 = Column motor
Cabinet
A4 board
X0
A5 board
HT cables to Generator
Figure 6-12
Table top
Locking end
End stop stop screws
Locking ratchet
Figure 6-13
Table base
Fixing points
Pedal control
Figure 6-14
Cover
Counterweight
Figure 6-15
Rail
.078-.12" (2-3mm)
Adjusting screws
Figure 6-16
WARNING:
Before electrically connecting the MOVIPLAN 800S, check that all the
instructions and warnings given in paragraph 5.1 have been complied
with.
Disconnect from the mains before performing any operations
connected with the electrical wiring system.
Table 1
WARNING:
The Potter Bucky call signal must be set at 24Vdc. If the generator
does not provide the Bucky call signal at 24Vdc it is necessary to
generate the signal at 24Vdc making use of an external relays.
After having connected the power cables, they should be tied together to
prevent accidental contact in case of wire breakage.
7. INSPECTION, CALIBRAT
CALIBRATION
ION AND
ADJUSTMENT
The device is precalibrated and all its functions are checked during
inspection and testing. A number of adjustments however, require a final
calibration that can only be made once installation has been completed.
Example:
Perform a calibration of the column stop angle of -27° connecting the digital
voltmeter to test points TP2 - TP6 of the logic card.
If the reading obtained is other than 2V as indicated in Table 2, adjust
trimmer P9.
Tomography
Test point Voltage Trimmer
angle
Table 2
NOTE:
* The stop points are so determined which allow all the tomography angles
to be effected.
NOTE:
* The negative angles are with the column on the left.
It is not necessary to move the column during the calibration phases as
the trimmers are always power supplied.
Terminal board X1
Function Voltage Trimmer
card A2
Manual one way 7-1 1.3V P1
Manual return 7-5 -1.3V P5
Tomo return 7-6 -1.5V P6
1st. Tomo speed 7-2 2.0V P2
2nd. Tomo speed 7-3 2.5V P3
3rd. Tomo speed 7-4 3.6V P4
Table 4
CORRECT INCORRECT
WARNING:
Calibration of the R2 must be made exclusively with the grid in the still
position.
For the versions with the elevator, position the table top to the highest
limit, then at a distance of 100 cm between the focal point of the X-ray
source and the film in the potter bucky under the table, apply the
centimetre scale on the front surface of the column so that the 100 cm
notch corresponds to the lower edge of the slide.
For the versions with the elevator, position the table top to the maximum
highest limit, driving the motor via the controls located at the foot of the
table. Insert the tomographic shaft stop arm in position in the potter
under the table. Insert the shaft and lock the transversal tube carrier
arm of the column.
This inspection must be made every time the collimator lamp is changed.
Switch on the collimator light at a distance of 100 cm from the table top.
Adjust the knobs to 35 x 43 cm at 100 cm.
Measure the size of the light image starting from about halfway into the
penumbra. The sizes must be equal to the set values, ± 8 mm.
Check the size of the light field in the various formats indicated by the
knobs for a distance of 100 cm.
WARNING:
Only in the case of light bulb replacement: adjust the position of the
collimator lamp until these measurements are obtained.
<5mm
7.11.1 Alignment of X-
X-ray beam with the collimator shutters
NOTE:
* Before aligning the X-ray beam with the current procedure it is necessary
to make sure that the column is perfectly vertical and the X-ray tube is
horizontal; these checks can be easily done using a bubble level.
Tube focus
Collimator center
Correct centering
Figure 7-1
NOTE:
* This procedure applies to system carrying Villa collimator. In case of
Ralco collimator make reference to the dedicated Manual.
NOTE:
* The light beam of the collimator is calibrated in the factory so that once
the alignment between the tube and the collimator is achieved, the
alignment of the X-ray beam and the light field should be granted.
1. In order to check the alignment of the X-ray beam with the light field
of the collimator turn on the collimator light and position 4 radio
opaque objects on the borders of the light field.
2. Make an exposure on a large film positioned on the table top and
check the position of the objects; the max deviation must be less
than 3% of the in each direction and the sum of the deviation along
the two directions must be less than 4%.
3. If the alignment is not achieved within the specified limits (i.e. the
X-ray focus distance from the image plan is different from the
distance of the slot of the lamp) it is possible to act on the lamp of the
collimator, remembering that moving the lamp closer to the mirror
the light field gets smaller, while moving the lamp further from the
mirror the light field gets larger.
Figure 7-2
4. Once the optimal alignment between light field and X-ray beam has
been achieved it is possible to complete the alignment between the
X-ray beam and the image receptor (Potter Bucky – see paragraph
7.12).
NOTE:
* This procedure applies to system carrying Villa collimator. In case of
Ralco collimator make reference to the dedicated Manual.
Both versions with our without elevator (SA and SF) have a vertical
column with a motorised longitudinal movement.
This characteristic makes it possible to perform tomographic
examinations via a coordination device, the "tomographic shaft" with a
variable motorised fulcrum incorporated in the table.
The device has been designed for a life duration of 10 years, when
properly maintained.
WARNING:
Read the instructions given in chapter 2 before using the table.
The size of the irradiated field, the reference axes and the luminous line
for the centering of the column with respect to the potter bucky can be
identified on the table top when the collimator light is switched on.
Tube assembly
Cross-arm
Pushbutton panel
Collimator
Column
Table top
Table
Figure 8-1
To load a cassette:
1. Turn the lever anti-clockwise.
2. Open the clamps.
3. Position the cassette by coinciding the center notch of the cassette.
4. Position the magnetic reference in the notch relative to the format of
the inserted cassette.
5. Turn the lever clockwise.
6. Push the clamps so that they fit tightly around the edge of the
cassette.
Table Emergency
push button
Bucky
tray
Potter Release
push button
Cassette
Centre
notch Magnetic lock
Locked
Jaws
Lever
Figure 8-2
020 mm
Figure 8-3
8.5 Column
The column can make the following movements:
• Longitudinal
By pressing the button, the longitudinal movement brakes are
released and the column can be moved along the rail via the knob on
the control panel.
Figure 8-4
Figure 8-5
Figure 8-6
ANGLE METER
Figure 8-7
Unlocking
pedal
Figure 8-8
Motor connected
Motor not
connected
Figure 8-9
NOTE:
* See dedicated Manual if the system comes with "Ralco" collimator.
WARNING:
Before making any type of examination ensure that:
• a loaded cassette is in the potter or in one of the supplementary
accessories (teleradiographer, side cassette carrier, etc.)
• the collimator is adjusted in accordance with the set focal point-film
distance and the format of the cassette used. When doing this, it is
possible to use the wound meter integrated in the collimator and the
projection of the light field corresponding to the irradiated field
• the radiogenic beam is correctly centered and aligned with the
cassette
• there are no foreign objects between the X-ray source and the
cassette.
WARNING:
The duty cycle of the collimator lamp must not be longer than 2
consecutive activation cycle separated by 180 seconds of de-activation
(lamp off).
This duty cycle assure that the parts which may come in touch with the
patient/user do not reach a too high temperature due to the heating
effect of the halogen lamp.
NOTE:
* When the tomography function is not required, the tomography shaft and
the shaft lock pin have to be removed.
On the back of the tomography tower seat there are two dedicated clips
to store the shaft lock pin; the tomo shaft can be stored on the back of
the column using the dedicated clip.
When the table is to be used as a normal table, the X-ray beam must be
aligned with the potter and focal point-film distance most suitable to the
type of examination to be performed must be set.
NOTE:
* Always select the greatest focal point-film distance compatible with the
examination to be performed.
In the elevator version, the height of the table with respect to the floor
can be changed using the pedals in order to set the most comfortable
working conditions both for the patient and for the operator.
WARNING:
The elevator pedals commands directly the elevator motor. Take care to
avoid unwanted movement of the table top caused by accidental pressure
of the elevator pedals.
WARNING:
During the operation of the elevator make sure that no obstacle is
present in the range of the elevator movement.
Figure 8-11
3. Center the X-ray beam with the film format moving the potter
carriage lengthways and releasing the brake via the switch located on
the table.
Bucky tray
Figure 8-12
Figure 8-13
When the tube up and down movement brake is released via the relative
button on the column control panel, the desired focal-point-film distance
is obtained.
Figure 8-14
The focal point-film distance can be measured using the meter in the
collimator. The measurement must be made with respect to the table top.
NOTE:
* The distance between the table top and the cassette inserted in the potter
is about 80 mm. The focal point-film distance can be obtained by
summing this distance, 80 mm and the distance obtained on the
collimator meter or by adjusting the column slide in accordance with the
meter applied on the column. For example, to set a DFF of 100 cm the
tubehead must adjusted at a distance of 92 cm from the table top.
WARNING:
The scale is graduated for a few specific focal spot-film distances. For
distance different from the ones provided, the extrapolated formats may
not be reliable.
Once the irradiated surface has been collimated, the device is ready for
radiography.
8.7.2 Preparation
Preparation of the table for a lateral radiography
Figure 8-15
NOTE:
* Moving the column towards the outside of the table, it is possible to lower
the tube below the level of the table top and take exposures of the legs
(knee and ankle) under load with the patient standing close to the Chest
Stand, usually associated with the Bucky table.
2. Insert the layer shaft lock arm in the back of the potter.
COLUMN
7. Adjust the height of the layer by pressing
WARNING:
Disconnect the device from the mains before cleaning.
Ensure that no water or other liquids penetrate inside the device in order
to prevent corrosion or short circuits.
To clean the painted surfaces, accessories and cables, use only a damp
cloth and a neutral detergent and dry thoroughly. Do not use corrosive or
abrasive solvents (alcohol, petrol and trichloroethylene).
Disinfect the table top, the potter handle, the control panel handle and
any other accessories that come into contact with the patient or the
operator, with a 2% glutaraldehyde water solution.
10. MAINTENANCE
As with all electrical apparatus, this unit not only needs to be used
correctly, but also needs maintenance operations and inspections made
at regular intervals.
Such precautions ensure the safe and efficient functioning of the device
and serve to prevent all possible risks to the patient and to the operator.
WARNING:
Metallic friction noises issued from the column may be a symptom of
wear of the counterweight support cables. If this should occur, stop
using the device immediately and contact the Villa Sistemi Medicali
Technical Service Department.
NOTE:
* When components that have a direct or indirect influence on safety are
replaced, only original spare parts should be used.
The maintenance intervention and replacement of defective parts must be
made exclusively by Villa Sistemi Medicali authorised personnel.
11. SCHEMATICS
SCHEMATICS AND DRAWINGS
1. Moviplan 800S – Components plate layout SA version
8. A4 Board schematic
MOVIPLAN 800 S
Components plate layout 1
SA version
Code 72711405 - Rev. 1 Page 1 of 1
MOVIPLAN 800 S
Functional drawing 2
SA version
Code 39719119 - Rev. 4 Page 1 of 4
MOVIPLAN 800 S
Functional drawing 2
SA version
Code 39719119 - Rev. 4 Page 2 of 4
MOVIPLAN 800 S
Functional drawing 2
SA version
Code 39719119 - Rev. 4 Page 3 of 4
MOVIPLAN 800 S
Functional drawing 2
SA version
Code 39719119 - Rev. 4 Page 4 of 4
MOVIPLAN 800 S
Components plate layout 3
SF version
Code 72711420 - Rev. 2 Page 1 of 1
MOVIPLAN 800 S
Functional drawing 4
SF version
Code 39719127 - Rev. 3 Page 1 of 4
MOVIPLAN 800 S
Functional drawing 4
SF version
Code 39719127 - Rev. 3 Page 2 of 4
MOVIPLAN 800 S
Functional drawing 4
SF version
Code 39719127 - Rev. 3 Page 3 of 4
MOVIPLAN 800 S
Functional drawing 4
SF version
Code 39719127 - Rev. 3 Page 4 of 4
MOVIPLAN 800 S
Potter Bucky 5
functional drawing
Code 39209043 - Rev. 0 Page 1 of 1
MOVIPLAN 800 S
A3 Control board 6
components layout
Code 58714014 - Rev. 2 Page 1 of 1
MOVIPLAN 800 S
A4 Board 7
components layout
Code 58714012 - Rev. 1 Page 1 of 1
MOVIPLAN 800 S
A4 Board 8
schematic
Code 39719121 - Rev. 1 Page 1 of 1
MOVIPLAN 800 S
A5 Board 9
components layout
Code 58711425 - Rev. 0 Page 1 of 1
MOVIPLAN 800 S
A5 Board 10
schematic
Code 39719126 - Rev. 0 Page 1 of 1
MOVIPLAN 800 S
A10 Push-button board 11
components layout
Code 58748697 - Rev. 1 Page 1 of 1
MOVIPLAN 800 S
A10 Push-button board 12
schematic
Code 39749204 - Rev. 1 Page 1 of 5
MOVIPLAN 800 S
A10 Push-button board 12
schematic
Code 39749204 - Rev. 1 Page 2 of 5
MOVIPLAN 800 S
A10 Push-button board 12
schematic
Code 39749204 - Rev. 1 Page 3 of 5
MOVIPLAN 800 S
A10 Push-button board 12
schematic
Code 39749204 - Rev. 1 Page 4 of 5
MOVIPLAN 800 S
A10 Push-button board 12
schematic
Code 39749204 - Rev. 1 Page 5 of 5
MOVIPLAN 800 S
Potter Bucky board 13
components layout
Code 58201208 - Rev. 0 Page 1 of 1
MOVIPLAN 800 S
Potter Bucky board 14
schematic
Code 39209044 - Rev. 0 Page 1 of 1
MOVIPLAN 800 S
A9 Board 15
layout and schematic
Code 58112192 - Rev. 0 Page 1 of 1
MOVIPLAN 800 S
A2 Board 16
components layout
Code 58659215 - Rev. 0 Page 1 of 1
1 – TABLE
2 – CABINET
3 – COLUMN
4 – POTTER BUCKY
1 - TABLE
D 4233811100 Microswitch X
F 4271126100 Reflector X X
E A D B
F C H
G G
L 4591605200 Bulb X X
H3 4271316600 Buzzer X X
R1 6171319300 Potentiometer R1 X X
R2 6171313700 Potentiometer R2 X X
A B
L1 4192204500 Inductance L1 X
T1 4492818300 Transformer T1 X
V3 4165121100 Rectifier V3 X
2 - CABINET
C 4890306300 Wheel X
L1 4192204500 Inductance L1 X
R1 4192108200 Resistor R1 X
T1 4492818300 Transformer T1 X
V1 ÷ V3 4165121100 Rectifier V1 ÷ V3 X
3 - COLUMN
NOTE:
* If the system comes with "Ralco" collimator make reference to the
dedicated Manual of "Ralco" collimator for Spare Parts list.
2 6174862100 Inclinometer X X
10 6374850500 Knob X X
4
3 11
2
10
5 9
6
8
4 – POTTER BUCKY
E 4291305400 Microswitch S1 X X
F 4291305500 Microswitch S2 X X
A B
D C E F H G