Excerpt from the Proceedings of the 2011 COMSOL Conference in Stuttgart
Abstract: A drift tube Ion Mobility ionization leading to ultra-low detection limits
Spectrometer (IMS) is a device to analyze [3] for most trace gases. Nevertheless, the
different components of a gas mixture. One analytical power of an IMS is limited by a
particular feature of this technology is its ultra- comparably low selectivity making it difficult
low detection limit. However, due to a limited to separate ions with similar K or to identify
resolution required to differ between different individual components in a complex matrix of
components an improved IMS field geometry multiple substances. This limited separation
would help to increase resolution and to power is mainly caused by inhomogeneities of
analyze even complex mixtures of multiple the electrical field within the IMS, the
gases. Furthermore - especially for novel, non- diffusion of ions, coulomb repulsion and the
radioactive ionization sources [1] - the extent of the volume in which the molecules
influence of Coulomb repulsion needs to be are initially ionized.
taken into account. In this work a transient and In this work we present a transient model
parametric model of a drift tube IMS based on of a drift tube IMS which is considering these
COMSOL 4.1 is presented in good agreement effects to design an optimized geometry for an
with the experimental results. The exact shape IMS with improved resolution.
of the electrical field, the loss of ions at
metallic surfaces and the timing of ion 2. Experimental Setup
generation and injection is considered.
Furthermore, coulomb repulsion caused by the In fig. 1 the experimental setup is shown.
transient charge distribution is taken into The model is based on this setup. The sample
account. gas is ionized by an electron emitter. The
ionization depth can be calculated by the
Keywords: Drift Tube Ion Mobility kinetic energy of emitted electrons. The ions
Spectrometer, Transient Simulation of Ions are injected into the drift tube by an injection
under Atmospheric Pressure Conditions, Ion voltage (Usource = Uinjection) at t = t0. The
Drift Tube, Coulomb Repulsion. duration of the ion injection is tinjection.
1. Introduction
5. Results
6. Conclusions