*Corresponding author: Leila Ibrahimi Ghavamabadi. Department of Environmental Management-HSE, Ahvaz Branch, Islamic Azad University, Ahvaz,
Iran, Tel/Fax: +98-61337385318, Email: ebrahimi.ghavam@gmail.com
+982188221219
Ibrahimi Ghavamabadi et al X-Ray Radiation Levels in Radiology Departments
equivalent range to be 0.107±0.003 and 0.108±0.003 patient observation window, staff rest room, office,
μSv/h for indoor and outdoor, respectively [15]. patient waiting room, behind the door of the X-ray
In a study carried out by Basirjafari et al. (2014), the room (entrance door for patients), and outdoor (Figure
mean outdoor gamma dose rate was 0.094±0.024 1).
μSv/h [16]. In another study conducted by Okoye et al. The monitor was held one meter above the
(2013) in a medical center, the results showed that the terrestrial level in all the six locations except for behind
average range for indoor measurement within X-ray the patient observation window, which was 1.5 meter
department, locations within the hospital, and the high. The equivalent dose (in µSv/h) was measured in
departments within this center were 0.14±0.02 μSv/h, all locations investigated in this study. The United
0.14±0.02 μSv/h, and 0.13±0.02 μSv/h, respectively [2]. Nations Scientific Committee on the Effects of Atomic
With this background in mind, this study aimed to Radiation (UNSCEAR ) 2000 Report recommended the
measure the indoor and outdoor X-ray radiation level in occupancy factors of 0.2 and 0.8 for outdoor and indoor,
six radiology departments of two educational hospitals respectively [3]. The occupancy factor is the proportion
in Ahvaz, Iran. of the total time during which an individual is exposed
to a radiation field. In this study, due to major work
Materials and Methods shifts and overtime, 4600 h/y was used instead of 8760
This cross-sectional study was conducted in six h/y, which is generally used by the UNSCEAR.
radiology departments of two hospitals during May- To convert micro Sievert per hour (μSv/h) into mili
July 2015. The radiation levels were inspected in six Sievert per year (mSv/y), the following equations were
locations of the respective radiology departments in applied:
both on and off status of X-ray generating device, using 1) Annual indoor equivalent dose rate (mSv/y) =
the calibrated digital Geiger-Muller counter (S.E. indoor equivalent dose rate (μSv/h) × 4600 (h/y) × 0.8
International Inc., USA). The instrument was calibrated (indoor occupancy factor) × 0.001
electronically according to the instruction of the 2) Annual outdoor equivalent dose rate (mSv/y) =
manufacturer. The detector is able to measure dose outdoor equivalent dose rate (μSv/h) × 4600 (h/y) ×
rates from 0.01 μSv/h to 1000 μSv/h with accuracy of 0.2 (outdoor occupancy factor) × 0.001
15%. The measurement was repeated at each location The irradiation parameters of X-ray generating
for at least five times with 3 min intervals. lamp included the voltage of 100 or 150 kV, time of 2 s
The device was placed with the end window facing (dead time of survey meter is longer than 0.1 s), and
the area where count rates were taken. We recorded tube current of 50 mA. The data analysis was
the mean displayed by the detector during the X-ray performed using the descriptive statistics and one-
imaging with high exposure (lumbar spine and sample t-test through SPSS version 22.0 (IBM SPSS Inc.,
abdominal imaging with 0.7 and 1.2 mSv, respectively 1989-2013). P-value less than 0.05 was considered
[17]). The measured locations included behind the statistically significant.
88
Iran J Med Phys, Vol. 14, No. 2, June 2017
X-Ray Radiation Levels in Radiology Departments Ibrahimi Ghavamabadi et al
The results showed no significant difference observation window and behind the door of the X-
between the background radiation and the X-ray ray room) (P<0.001; Table 1).
radiation levels in the radiology departments while Furthermore, there was a significant difference
the X-ray generating equipment was off (P>0.365). between the indoor and outdoor measured levels in
Nevertheless, there was a significant difference both hospitals (P<0.001; Table 2). The statistical
between the background radiation and the X-ray comparison between the two hospitals for indoor
radiation levels during the radiography procedure in and outdoor levels showed no significant
two locations of the radiology departments in relationship. Figure 1 presents the comparison
hospital 2, radiology department 4 (i.e., patient between the recorded dose rates and the standard
levels for indoor measured levels.
Table 1. Average levels of radiation levels in six locations of two hospitals (µSv/h)
Behind the
Equipment Patient observation Staff rest Patient
Office door of X-ray Outdoor
status window room waiting room
room
Radiology on 0.128±0.014 0.128±0.018 0.132±0.001 0.132±0.001 0.132±0.001
0.085±0.002
department 1 off 0.126±0.01 0.126±0.02 0.130±0.002 0.130±0.002 0.130±0.002
Radiology on 0.131±0.002 0.119±0.024 0.098±0.036 0.131±0.002 0.131±0.002
Hospital 1 0.08±0.02
department 2 off 0.130±0.01 0.118±0.01 0.097±0.03 0.129±0.002 0.130±0.002
Radiology on 0.136±0.022 0.131±0.002 0.098±0.036 0.131±0.002 0.131±0.002
0.092±0.001
department 3 off 0.134±0.02 0.129±0.01 0.097±0.03 0.128±0.01 0.130±0.001
Radiology on 0.365±0.11* 0.128±0.018 0.131±0.002 0.132±0.001 0.367±0.12*
0.065±0.03
department 4 off 0.131±0.015 0.127±0.019 0.130±0.002 0.130±0.002 0.124±0.03
Radiology on 0.122±0.023 0.132±0.18 0.131±0.002 0.117±0.022 0.117±0.022
Hospital 2 0.085±0.002
department 5 off 0.122±0.01 0.131±0.02 0.130±0.001 0.115±0.02 0.115±0.02
Radiology on 0.128±0.024 0.098±0.036 0.098±0.036 0.132±0.001 0.132±0.001
0.085±0.002
department 6 off 0.125±0.03 0.097±0.03 0.097±0.03 0.130±0.002 0.130±0.002
*Above the standard level
Table 2. Mean values of indoor and outdoor radiation dose rates while the X-ray generating equipment was working
Measurement unit Indoor Outdoor P-value
µSv/h 0.130±0.002 0.085±0.002
Radiology department 1 0.001
mSv/y 0.48±0.007 0.08±0.0018
µSv/h 0.122±0.014 0.08±0.02
Hospital 1 Radiology department 2 0.001
mSv/y 0.45±0.05 0.074±0.0184
µSv/h 0.125±0.02 0.09±0.001
Radiology department 3 0.003
mSv/y 0.46±0.074 0.083±0.001
µSv/h 0.214±0.09 0.065±0.03
Radiology department 4 0.000
mSv/y 0.76±0.017 0.06±0.03
µSv/h 0.122±0.008 0.085±0.002
Hospital 2 Radiology department 5 0.001
mSv/y 0.45±0.03 0.08±0.0018
µSv/h 0.124±0.015 0.085±0.002
Radiology department 6 0.003
mSv/y 0.46±0.055 0.08±0.0018
Figure 2. Comparison of the standard level with the measured levels at six radiology departments
90
Iran J Med Phys, Vol. 14, No. 2, June 2017
X-Ray Radiation Levels in Radiology Departments Ibrahimi Ghavamabadi et al