Professional Documents
Culture Documents
University of Ghana
University of Ghana
University of Ghana
gh
UNIVERSITY OF GHANA
UNIVERSITY OF GHANA
BY
ID: 10361536
OF
MASTER OF PHILOSOPHY
MEDICAL PHYSICS
July, 2018
i
University of Ghana http://ugspace.ug.edu.gh
DECLARATION
STUDENT’S DECLARATION:
I hereby declare that, with the exception of references to other people’s work which
have been duly acknowledged, this work is the result of my own original research
undertaken under supervision, and either in whole or in part has not been presented for
……………………………… ..…………………………….
(STUDENT)
SUPERVISORS’ DECLARATION:
We hereby declare that the preparation and presentation of the thesis were supervised
Ghana.
……………………………… ..…………………………….
(PRINCIPAL SUPERVISOR)
……………………………… ..…………………………….
(CO-SUPERVISOR)
……………………………… .…………………………….
(CO-SUPERVISOR)
ii
University of Ghana http://ugspace.ug.edu.gh
DEDICATION
This thesis is dedicated to my country, my family, and in particular to all those working
tirelessly for the establishment of Nuclear Regulatory Body and Atomic Energy
Commission in Benin.
iii
University of Ghana http://ugspace.ug.edu.gh
ACKNOWLEDGEMENTS
Thanks to Almighty God for assisting me through the IAEA/BEN 6006 Medical
physics programme.
All my gratitude to IAEA, for offering me two years fellowship in Medical Physics
and Two years fellowship in Nuclear science and Technology. I also thank IAEA for
I would like to thank my supervisors, Dr. Stephen Inkoom, Dr. Francis Hasford and
Dr. Edem Sosu who scarified their time to help collect data in Benin and guide me
Thanks to Prof. Amoussou K. Marcellin, the coordinator of the National project BEN
6006 for his effort to make establish in Benin nuclear medicine and radiotherapy
Thanks to Prof. Pascal Houngnandan and Prof. Felix Hontinfinde, for their advice and
support.
I also want to extend my gratitude to Benin Ministry of Health who facilitated access
to the hospitals. Thanks to staff of all the seven Hospital involved in the present study.
Thanks to Ghana Atomic Energy Commission and the Dean of School of Nuclear and
Allied
Science for their contribution to knowledge and their effort in making equipment
iv
University of Ghana http://ugspace.ug.edu.gh
TABLE OF CONTENTS
DECLARATION ..................................................................................................................... ii
ACKNOWLEDGEMENTS .................................................................................................... iv
ABSTRACT............................................................................................................................. 1
INTRODUCTION ................................................................................................................... 2
2.3 Importance of some essential quality control tests and standard limits ..................... 14
v
University of Ghana http://ugspace.ug.edu.gh
METHODOLOGY ................................................................................................................ 24
3.1.3 Collimator and beam alignment Quality Control test tool ..................................... 26
3.2.1.1 Set up for simultaneous measurement of kilovoltage peak, exposure time, half
vi
University of Ghana http://ugspace.ug.edu.gh
3.2.2.3 Reproducibility of kilovoltage peak, exposure time and dose output .................... 33
vii
University of Ghana http://ugspace.ug.edu.gh
4.1.3 Results of reproducibility of kVp, exposure time end dose output ....................... 44
4.2.2 Exposure factors (kVp, mAs), FDD and patient thickness .................................... 72
viii
University of Ghana http://ugspace.ug.edu.gh
4.3.2.2 Comparison of estimated entrance surface dose with other studies ....................... 89
REFERENCES ...................................................................................................................... 95
ix
University of Ghana http://ugspace.ug.edu.gh
LIST OF TABLES
4.1 Results of kVp accuracy for X-ray unit at Clinic Ste Anne 37
unit at CNHU 40
x
University of Ghana http://ugspace.ug.edu.gh
unit at CNHU 42
4.20 Results of kVp reproducibility for X-ray unit at Clinic Ste Anne 44
Ste Anne 48
Roseraie 49
xi
University of Ghana http://ugspace.ug.edu.gh
Senande 49
Mènontin 50
Porto Novo 50
Suru Léré 51
4.34 Results of dose output reproducibility for X-ray unit at Clinic Ste
Anne 52
Roseraie 53
Senande 53
Mènontin 54
Novo 54
Léré 55
xii
University of Ghana http://ugspace.ug.edu.gh
4.48 Results of specific dose-mAs linearity for X-ray unit at Clinic Ste
Anne 63
Roseraie 63
Senande 64
Mènontin 64
Novo 64
Lere 65
4.55 Results of leakage of X-ray tube housing for X-ray unit at CNHU,
Lere 66
projection 73
xiii
University of Ghana http://ugspace.ug.edu.gh
A.1 Results of specific dose - kVp2 linearity for X-ray unit at Clinic Ste
Anne 101
A.2 Results of specific dose - kVp2 linearity for X-ray unit at CNHU 101
A.3 Results of specific dose - kVp2 linearity for X-ray unit at Clinic
Roseraie 101
A.4 Results of specific dose - kVp2 linearity for X-ray unit at Clinic
Senande 102
A.5 Results of specific dose - kVp2 linearity for X-ray unit at CS-
Mènontin 102
A.6 Results of specific dose - kVp2 linearity for X-ray unit at HZ-Porto
Novo 102
A.7 Results of specific dose - kVp2 linearity for X-ray unit at CHU-
B.1 Results of dose output for X-ray unit at Clinic Ste Anne 104
B.3 Results of dose output for X-ray unit at Clinic Roseraie 104
B.4 Results of dose output for X-ray unit at Clinic Senande 105
B.6 Results of dose output for X-ray unit at HZ-Porto Novo 105
xiv
University of Ghana http://ugspace.ug.edu.gh
B.7 Results of dose output for X-ray unit at CHU-Suru Lere 105
xv
University of Ghana http://ugspace.ug.edu.gh
LIST OF FIGURES
2.1 Interpretation of the image of the steel ball in the beam alignment
test tool 19
4.1 Specific dose versus kVp2 for X-ray unit at Clinic Ste Anne 59
4.3 Specific dose versus kVp2 for X-ray unit at Clinic Roseraie 60
4.4 Specific dose versus kVp2 for X-ray unit at Clinic Senande 60
4.6 Specific dose versus kVp2 for X-ray unit at HZ-Porto Novo 61
4.7 Specific dose versus kVp2 for X-ray unit at CHU-Suru Lere 62
4.8 Results of X-ray beam alignment test for X-ray unit at Clinic Ste 67
Anne
4.9 Results of X-ray beam alignment test for X-ray unit at CNHU 67
xvi
University of Ghana http://ugspace.ug.edu.gh
4.10 Results of X-ray beam alignment test for X-ray unit at Clinic 67
Roseraie
4.11 Results of X-ray beam alignment test for X-ray unit at Clinic 67
Senande
4.12 Results of X-ray beam alignment test for X-ray unit at CS- 68
Mènontin
4.13 Results of X-ray beam alignment test for X-ray unit at HZ-Porto 68
Novo
4.14 Results of X-ray beam alignment test for X-ray unit at CHU- 68
Suru Lere
4.15 Specific dose output versus kVp for X-ray unit at Clinic Ste Anne 69
4.16 Specific dose output versus kVp for X-ray unit at CNHU 69
4.17 Specific dose output versus kVp for X-ray unit at Clinic Roseraie 70
4.18 Specific dose output versus kVp for X-ray unit at Clinic Senande 70
4.19 Specific dose output versus kVp for X-ray unit at CS-Mènontin 71
4.20 Specific dose output versus kVp for X-ray unit at CS-Mènontin 71
4.21 Specific dose output versus kVp for X-ray unit at CHU-Suru Lere 72
recommended limit 79
recommended limit. 80
limit.
xvii
University of Ghana http://ugspace.ug.edu.gh
recommended limit. 82
recommended limit. 83
HVL.
4.30 Correlation of the specific dose versus kVp square with linear 85
trendline.
recommended limit
ray unit. 87
xviii
University of Ghana http://ugspace.ug.edu.gh
ABSTRACT
growth of population and Benin is not an exception. Achieving good quality image
acceptable level has become a prerequisite for radiology department in their effort to
comply with radiation protection principle of optimization. The present research work
south of Benin. In addition, patient entrance surface dose was also estimated in the
seven hospitals. RDS-120 universal survey meter, multifunction detector (Piranha) and
beam alignment test tool were used to perform quality control tests on the seven X-ray
units. The method used as well as the interpretation of the results was based on
AAPM, FDA, HARP, IPEM, IAEA and S.C 35 recommendations. The quality control
results showed that all X-ray equipment investigated were within standard limits for
accuracy of exposure time below 10 ms; reproducibility of kVp, exposure time and
dose output; specific dose-kVp2 linearity; specific dose-mAs linearity and leakage of
X-ray tube housing. 5/7 of diagnostic X-ray machines passed quality control tests such
as X-ray beam alignment, exposure time above 10 ms and kVp accuracy. For the seven
X-ray units, the maximum average estimated entrance surface dose for chest PA, Skull
AP, abdomen AP, pelvis AP, and lumber spine AP were respectively, 1.3922 ± 0.0217
mGy; 10.5709 ± 0.4549 mGy ; 11.2909 ± 0.5324 mGy; 10.1398 ± 0.2322 mGy;
16.1073 ±0.6931 mGy and 1.4317 ± 0.0161 mGy. The choice of radiographic
1
University of Ghana http://ugspace.ug.edu.gh
CHAPTER ONE
INTRODUCTION
problem, relevance and justification of the study, scope and limitation and the
1.1 Background
Germany [1]. X-rays are ionizing radiations. All forms of such radiation have
sufficient energy to ionize atoms that may destabilize molecules within cells and lead
to tissue damage. The harmful effects of ionizing radiation became apparent after the
severe burns as a result of holding the energized X-ray tube in his hands. In May 1896,
a man who had a head radiograph suffered skin burns and loss of hair [2]. From the
beginning of the use of X-rays for diagnosis, harmful effects were observed and
(IAEA), recommend that proper management systems that include quality assurance
(QA) programme encompasses activities involving the use of ionizing radiation [3].
adequate confidence that an item, process or service will satisfy given requirements
for quality [4]. Also the World Health Organization (WHO) defines a QA programme
that the diagnostic images produced are of sufficiently high quality so that they
2
University of Ghana http://ugspace.ug.edu.gh
consistently provide adequate diagnostic information at the lowest possible cost and
with the least possible exposure of the patient to radiation [5]. As part of QA, quality
control (QC) intends to verify that structures, systems and components corresponding
used in the monitoring (or testing) and maintenance of the technical elements or
components of an X-ray system. The QC techniques are concerned directly with the
equipment performance that can affect patient dose and the quality of the radiographic
image [6]. X-rays machines are used in diagnostic radiology departments to produce
radiology equipment are made of an X-ray tube mounted in such a way that allows the
(kilovoltage peak, milliamperes, exposure time) and the ability to choose between
small and large focal spot. The QC tests related to those radiographic parameters as
well as the measurement of the focal spot size is of importance to ensure safe use of
the machine and protection of patients, workers and the public. The quality and
quantity of X-ray beam that is produced should allow good image to be obtained while
Benin Republic, is a West African country that shares boarder with Nigeria in
the east, Togo in the west, Niger and Burkina Faso in the north. The Health system of
the country is composed in decreasing order of the Ministry of health, twelve (12)
departmental divisions of health and hospitals. Public as well as private hospitals use
centres are concentrated in Cotonou and Porto Novo. Because of the lack of
3
University of Ghana http://ugspace.ug.edu.gh
overlooked. The equipment are repaired by biomedical engineers only when there is
the state of the equipment. In order to solve this problem, a law was enacted in October
2017 by the Parliament for regulation of activities involving the use of ionizing
radiation. A national project designed with the support of IAEA is aimed at building
There are several research works which have been done on quality control of
diagnostic X-ray equipment and patient entrance surface dose assessments. The QC
tests are performed and evaluated based on kVp and exposure time accuracy
(Akpochafor et al., 2016); output linearity with mAs (Rasuli et al., 2015); beam
alignment and congruence (Ismail et al., 2015); focal spot size; linearity of output with
kVp2 (Azzoz et al., 2014); screen-film contact uniformity (Begum et al., 2011);
reproducibility of kVp and output (Korir et al., 2011). Entrance surface dose was
measured either with a direct measurement method using a TLD (Jibiri et al., 2016),
or the indirect estimation from X-ray equipment output variation (Ackom et al., 2017;
selected hospitals in South of Benin, assess their performances and establish baseline
data for future work. The study also assesses patient radiation dose in terms of entrance
surface dose.
1.2 Objectives
achieve the above objective the following specific objectives will be addressed.
4
University of Ghana http://ugspace.ug.edu.gh
i) Perform QC checks with emphasis on the following tests and tube QC tests.
In Benin Republic, more than one hundred X-ray machines are in operation
system in the country, regarding the use of these machines, acquisition of such
equipment is not under proper control. Most of the equipment are not subjected to
acceptance test and the X-ray machines are not inspected during operation. The
reproducibility and accuracy of selected imaging parameters are known when regular
QCs are performed. Due to the lack of QCs performed on the equipment a big doubt
exists on the level of doses received by patients, staff and the public as satisfying the
quality, using minimum necessary dose to the patient and surrounding individual,
The use of X-ray machines for radiography is good for patient’s diagnosis, but
when not held in check, it can induce severe health effects to patients, radiographer
5
University of Ghana http://ugspace.ug.edu.gh
and the general public. The adverse health effects associated with ionizing radiation
do not allow it to be used without balancing between risk and benefit. To achieve
proper use of those machines, protocols and guidelines have been developed at both
international and national levels for diagnostic equipment QC. In Benin no QC tests
have been performed for decades, and the machines are only repaired when there are
mechanical and electrical defaults. This study is relevant and justified in the sense that
it will provide for the first time, information regarding the state of some X-ray
i) X-ray systems’ quality control tests except focal spot size determination
ii) The entrance surface dose will be estimated from radiation output
measurement.
The present research thesis is composed of five chapters. Chapter one covers
Introduction. Chapter two provides an overview of X-ray machine quality control and
review of work done on patient radiation dose assessment. Chapter three outlines the
materials and the experimental methods used to perform quality control of X-ray
equipment in South of Benin. Chapter four covers results obtained from data
collection, their interpretation and discussion drawn from them. In chapter five,
conclusion is made and recommendations are proposed. References used as well as the
6
University of Ghana http://ugspace.ug.edu.gh
CHAPTER TWO
LITERATURE REVIEW
This chapter provides an overview of X-ray machine quality control and review
of work done on patient radiation dose optimization through quality control. The
chapter also focuses on studies done regarding entrance surface dose in radiographic
procedures.
collimator system and image receptor. Other parts of the machine include the tube
support mechanism that allows movement of the tube to be correctly positioned at the
The generator is the source of electrical energy to the X-ray tube and ensures
how long an exposure will take. The tube of the diagnostic X-ray machine is made of
the filament (cathode); target material (anode), focusing cup, and the glass envelop
[9]. By thermionic emission, electrons are expelled from the cathode, concentrated in
the focusing cup, these electrons are accelerated by an electrical field applied between
cathode and anode. Electrons interact with the anode material and X-rays are
produce breaking X-ray radiation [10]. The production of X-rays depends on the
by the filament circuit and its product with the exposure time (mAs) determines the
quantity of the diagnostic X-ray machine output. The kilovoltage peak (kVp) supplied
by the cathode–anode circuit, creates the electrical field that accelerates electron
7
University of Ghana http://ugspace.ug.edu.gh
between anode and cathode [11]. Because of the need of uniform electrical field to
transformer and rectifying circuit to correct the alternative current supplied by the
generator. The focal spot, point of interaction of electron beams with anode material
inside the X-ray tube, releases the X-rays produced and its size influence the image
resolution. During the production of X-ray heat is generated more than the X-ray itself.
For example at 100 kVp, only about 1% X-rays are produced in form of combination
of Bremsstrahlung and characteristic X-ray [12]. The pulse duration timer controls the
exposure time to optimize patient dose, film exposure and heat generated in the tube.
Low energy X-rays are attenuated from the beam spectrum by aluminum filters to
minimize patient entrance surface dose. The collimator system confines the radiation
only to the region of interest so that unnecessary exposure to other parts of patient
body is eliminated. Anti-scatter grid removes scatter radiation that will reach the film
and affect image quality. The intensifying screen and the film combination allow the
detection of the beam transmitted through the patients and the formation of the image.
Diagnostic X-ray machines are man-made sources of radiation that are commonly
used in hospitals and radiological research institutions. The use of such machines that
deliver ionizing radiation may induce health effects not only to patients but also to the
radiographer as well as the surrounding individuals when not held in control. Rules
World health Organization (WHO) to ensure that X-ray system perform satisfactorily
in service [3]. To achieve the goal of good image at low patient dose, regular quality
8
University of Ghana http://ugspace.ug.edu.gh
machine defects and to initiate corrective actions in order to ensure a safe working
and status tests. Acceptance tests are performed at equipment installation or after
test is applied to already installed equipment and is similar to acceptance test. Routine
equipment. Those parameters are tested to verify change that may occur during
equipment operation [13]. To comply with the principles of optimization which imply
obtaining a good image quality at low patient dose, a lot of studies have been done
In 2016, Akpochafor et al, performed kVp accuracy test in ten X-ray centres. This
study conducted in Nigeria, analysed 40 kVp accuracy results. Some machine were
above acceptable accuracy limit of ± 5%. One fifth of those equipment that failed the
test were at least ten years old [14]. In the same country in 2015, Godfrey et al., worked
The importance of kVp is brought out in the sense that if kVp is out of range or limits,
it can bring about either over exposure or under exposure which in turn increases the
level of rejects, retake or unwanted exposure to radiation. The study was conducted on
five X-ray machines, and it was found out that none of the machines demonstrated
[15].
In Iran, Gholami et al., in 2015, assessed tube voltage and exposure time of
hospitals. A total of seven X-ray units in Lorestan province were involved in the study.
9
University of Ghana http://ugspace.ug.edu.gh
The detector used was a Barracuda dosimeter. From their work evaluation of kVp
accuracy showed values out of therequired limit of accuracy Exposure time accuracy
evaluation showed that some hospitals complied with the requirements. The highest
X-ray tube output was 147×10-3 mGy/mAs. From their conclusion, age of X-ray
machines and the rate of their maintenance are likely the causes that affect patient
radiation dose and radiographic image quality [16]. Mehrdad Gholami et al work done
Iranian Journal of medical physics in September 2015. They performed ten standard
Iran. Reproducibility test for kVp, dose exposure output and time as well as linearity
test were successful, for all devices. There was poor beam alignment in 3/5 of the units.
The results revealed, X-ray machines met the standard criteria despite the fact that the
machines were relatively old with high workload. Such result were due to the
the province [17]. In the same country Iran but in a third province, Golestan et al 2013,
performed quality control on forty-four X-ray units. kVp accuracy and reproducibility;
beam quality, and beam alignment were assessed. 100% of equipment kVp were
Units at Khartoum State Hospitals. The QC tests performed were, kilovoltage peak
and exposure time reproducibility and accuracy; mAs linearity; of light beam
coincidence with radiation beam and fog level in darkroom. CONNY II QC dosimeter
made by PTW was used to achieve measurements. It was found out that two of
10
University of Ghana http://ugspace.ug.edu.gh
hospitals had a problem in mAs linearity, also two out of eighteen unit had a problem
in kVp accuracy and one had a problem in kVp reproducibility. Light field and
radiation beam alignment were successful for 15 machines. Only few darkrooms are
free from fog level problems; time accuracy and time reproducibility were in the
acceptable limit. From their results it was necessary to spread the quality control
In the same year 2015, Ngoye et al. published in the Journal of Medical Imaging
and Radiation Sciences, a research work done to verify how quality control measures
are implemented in Tanzania by radiographers. From the study it is found out lack in
In Egypt, Azzoz et al. 2014, evaluated ten stationary X-ray units in eight hospitals.
The quality control tests performed included, beam Alignment and collimator accuracy
exposure per mAs for small and large focus; beam quality; focal spot size;
scattered radiation. From their results the total beam filtration was 2.5 mm aluminum.
Their study revealed that optimization of radiographic parameters is important for dose
radiology machines. The study was based on experience gained in Ghana. From their
reduction, image quality, medical costs reduction and improving diagnostic radiology
11
University of Ghana http://ugspace.ug.edu.gh
QC procedures all the quality control tests performed in Ghana were documented in
the regulations that allow those tests to be performed anywhere ionizing radiation is
used [6].
In Bangladesh, Begum et al, 2011, assessed forty X-ray units. The tests that were
carried out are light field and radiation beam congruence; focal spot size; half value
layer and film-screen contact test. Each parameter was assessed with a specific quality
control tests tool. The method used in their work was based on existing standard
quality control protocol. Investigation of half value layers (HVL) revealed all the forty
diagnostic X-ray machines were out of range. Only 7.5% machines failed the focal
spot size test. Congruence between light field and radiation beam was satisfactory in
more than 75% of cases while 65% of machines achieved the expected screen film
contact uniformity. It is clear that from Begum et al studies, quality control actions
need to be performed regularly for safe and proper operation of the X-ray units [22].
Exposure time and kVp accuracy, mAs linearity with exposure, and exposure
obtained revealed many faults on equipment. It was recommended not to use the
devices in the regions where those errors were found and it was suggested that some
over a seven year period in the state of São Paulo were analysed. Results on half value
89% in 2000 to 94% in 2006. The satisfactory results obtained throughout the period
of study was attributed to the compliance with the state regulation. Worker integrated
12
University of Ghana http://ugspace.ug.edu.gh
quality assurance in their working culture, this induced natural observance of the
In Kenya, Korir et al., 2011 conducted a research work on the protection of patients
by establishing QA baseline for diagnostic radiology. The study was done on four X-
ray machine. Each machine was used over 1 month with 400-speed film/screen system.
Reproducibility, accuracy, linearity and beam quality are some of the QC tests
performed. The results revealed that the machines were in good status with faults
observed for the light/radiation beam alignment where the bottom and top as well as
cathode side showed deviation from the standard limits. From their conclusion patient
In Nigeria, Akaagerger et al, 2015, assessed HVL and beam alignment using
collimator tests tool. The study was made on two X-ray machines, using DIAVOLT
universal model 43014 dosimeter. It was found out that light-radiation field
Comparing the two diagnostic X-ray machines, one of them had higher probability of
Sungita et al, 2006, worked on QC of X-ray machines in Tanzania. This study was
conducted because of the increasing number of X-ray machines in the country and the
little availability of technical support to operate and maintain them adequately. Four
QC tests were performed on 196 diagnostic X-ray units. The tests included beam
alignment and collimation test made on 80 X-rays units; timer accuracy test performed
on 120 X-ray units; radiation leakage test assessed on 47 X-ray units and preventive
maintenance tests evaluated on each of the 196 diagnostic X-ray units. The outcome
revealed that 41% of equipment passed kVp tests, 43% passed exposure time accuracy,
13
University of Ghana http://ugspace.ug.edu.gh
60% satisfied conformity with beam alignment tests standards, and 20% failed the
radiation leakage of tube housing test. From their research report, governmental
ray units. Five groups of five X-ray equipment were involved in the study. It was found
out that the absorbed dose to different organs was a function of the X-ray tube loading
factors (kVp and mAs). Leakage radiation tests were acceptable and the measured at
1 m was far below the acceptable limit of 1 mGy/h at 1m. Tests of X-ray field
alignment with light field as well as focal spot position were satisfactory. Accuracy
tests for kVp showed the deviation was within standards limit [28].
diagnostic X-ray units to optimize radiation dose. The QC tests carried out included,
test as well as image quality was performed in five hospitals on ten diagnostic
radiology X-ray units. The objective was to determine the tube loading factors that will
be suitable for quality control tests [8]. The parameters setting values of 70 kVp-(20,
40, 50mAs) were recommended for the acquisition of high quality image
2.3 Importance of some essential quality control tests and standard limits
tube current and exposure time. The kVp is very important parameter, since it account
for X-ray beam quantity and quality. Variation of kilovoltage peak influences the mean
energy of photon emitted and thus affects the subject contrast of radiographic image
produced. It is therefore necessary that kVp be accurate as much as possible. The dose
14
University of Ghana http://ugspace.ug.edu.gh
output is proportional to the square of kVp. The product of tube current and exposure
measured exposure time is lower than the selected, then mAs is small and this induces
quantum mottle which affects radiograph by noise. When the exposure time is greater
between indicated and measured kVp and exposure time for diagnostic X-ray
parameters (kilovoltage peak/ time) and the indicated or selected values. According to
AAPM [31], FDA [32] and IPEM [33], the measured kVp accuracy limit is set to be
±5% of the indicated or selected kVp . Based on the Healing Arts Radiation Protection
Act, the kVp should also be assessed at most commonly used milliamperage stations.
According to Canadian’s Safety Code S.C. 35, deviation of the measured exposure
time should not exceed ±10% +1 ms of the time selected [29]. Similarly H.A.R.P Act
set the limit to be ±10% [34]. More precision was provided by AAPM report 74 which
specify limit for exposure below 10 ms and exposure above 10 ms. Above 10 ms, the
limit for time accuracy is set to be ±5% while below 10 ms it is ±10%. The tube
current, limit was set to be 20% while that for mAs is 10% +0.2 mAs [31].
dose output) are taken in the same conditions, the degree of agreement between
radiographic parameters can be repeated without major deviation. The same or very
close exposure can be obtained each time from an X-ray machine. Reproducibility is
important because it guarantees that the X-ray machine is capable of delivering dose
15
University of Ghana http://ugspace.ug.edu.gh
to the specifications for which it is rated and help to detect operator error when the X-
coefficient of variations which is the standard deviation divided by the mean of ten
detector distance.
Limit for reproducibility was recommended by the United State Food and Drug
5% and each measurement should not exceed 15% of the average value. H.A.R.P is in
agreement with S.C.35 and AAPM coefficient of variation but allows 20% deviation
milliamperage and exposure time, a constant amount of radiation is produced [30]. The
electrons that interact with the target material (anode). Thus mAs is related directly to
electrons with both target material electrons and nucleus are probabilistic, it is
expected that the radiation output to be the same for a constant film density. In the
variation of X-ray output should be obtained. Linearity tests are essential in the
detection of change in image quality and monitoring of patient dose in the sense that
poor quantity of X-ray beam reaching the film will induce- noise in the image and this
will produce image reject and repeat of patient examination. According to Canada
safety code 35 and FDA, for a fixed value of kVp, the difference between two specific
16
University of Ghana http://ugspace.ug.edu.gh
Filtration exists to cut off low energy X-rays by the mean of a layer of material.
The purpose is to reduce dose to the skin and shallow tissues induced by low energy
X-rays which cannot reach image receptor. They increase patient dose unnecessarily
and contribute nothing useful to the image. Radiographic images are obtained by
transmission of X-ray beam through patient. It is important that the X-ray beam has
just enough ability to penetrate deeper into tissue so that patient dose can be reduced
and image contrast to be maintained [1]. Total filtration consists of inherent filtration
and added filtration. Attenuation by housing oil; the thickness of glass envelop of the
tube and collimator assembly field light mirror are the inherent filtration components.
Added filtration use metal filters, intentionally insert in the X-ray field to modify it
effective energy. Generally the filters used are plates made of aluminum to remove
measuring the half value layer (HVL). HVL of an X-ray beam is the thickness of filter
that will halve the initial intensity of the beam. X-ray beam quality test is done to check
the minimum half value layer necessary to remove low energy radiation.
According to FDA the following minimum HVL are defined for various kVp setting
(table 2.1).
17
University of Ghana http://ugspace.ug.edu.gh
Collimators localize and modify the dimensions and form of the X-ray field rising from
the tube port. In the apparatus design, collimator assembly is hooked up to the tube
housing with a swivel joint. A rectangular X-ray field is set by means of lead shutters.
Collimator housing includes a light bulb and mirror. Le light emitted by the bulb is
reflected by a mirror to simulate the X-ray emission from the tube [35]. Collimator’s
shadows help identify the collimation of the X-ray field. Congruence between light
field and radiation field help practitioner to position well the anatomical part to be
18
University of Ghana http://ugspace.ug.edu.gh
According to safety code S.C 35 and AAPM report n° 74, the limit that should not be
exceeded is 2% deviation of the radiation field perimeter from the light field perimeter.
This limit is in an agreement with FDA and H.A.R.P policies. For X-ray beam
alignment the following criterion is applied for a source-table distance of 100 cm [35].
Figure 2.1 Interpretation of the image of the steel ball in the beam alignment test tool
The X-ray tube housing shield, supports and protects the X-ray tube insert. All X-rays
in the direction other than the exiting window are attenuated by a lead shielding. A
small fraction of these X-rays, known as leakage radiation may escape from the
housing [36]. Leakage radiation quality control test is necessary to minimize the risk
of 1.0 mGy/h at a distance of 100 cm from the position of the focal spot and at
maximum specified energy input. In standby the limit is 20 μGy/h at a distance of 5cm
19
University of Ghana http://ugspace.ug.edu.gh
not physically able to support themselves, comforters may also wear protective
greater than 670 mm2 the protective devices should not be used. If the cracks or holes
are close to the thyroid or gonads, it should not exceed 5 mm in diameter. When not
Entrance surface dose or entrance surface air kerma is the absorbed dose to air
at point of intersection of X-ray beam central axis with patient or phantom. It takes
account of incident and backscatter radiation. There are two ways of establishing
entrance surface dose including, the direct method and the indirect assessment method.
recorded tube loading factors and specific dose variation curve with kVp to estimate
including 39.3% males and 376 (60.7 %) females with age ranging from 18 to 82 years
20
University of Ghana http://ugspace.ug.edu.gh
old were investigated. Indirect assessment method was used to estimate the entrance
Commission in 1996. From their study it was found out that, calculated mean ESD
were within the recommended references values except chest PA and range from 0.29±
0.0041 mGy to 6.08 ±0.55 mGy for all the diagnostic examinations including chest
PA, lumber spine AP and LAT, skull LAT and AP, cervical spine LAT and AP, and
abdomen AP. Their results compared with publications from Inkoom et al, IAEA and
Public Health of UK respectively, revealed that values they obtained were lower than
In 2016, Jibiri et al, conducted research work on investigating patient dose for
diagnostic X-ray equipment were involved in the study. Seven different types of
NRPBHPA 2010 review for UK, lower values of entrance surface dose for pelvis AP
and lumber spine AP. For the other five examinations, ESD were higher. This
difference in results was related to the tube loading (mAs) used [39].
In 2014, Ofori et al., estimated entrance surface dose for patients aged over 18
years old. A total of 320 patients were considered in the research work. The mean
entrance surface dose (ESD) was estimated using caldose_x 5.0 software. Input data
were patients’ data (age, sex) and exposure parameters. It was found out that the
highest mean ESD was 3.25 mGy for lumber spine examination. For all examinations,
ESD ranged from 0.27 mGy to 3.25 mGy. The entrance skin doses for thorax was
slightly higher than published works. From the study it was confirmed that
21
University of Ghana http://ugspace.ug.edu.gh
optimization of technical and clinical factors will reduce substantially patient doses
[40].
In the same year 2014, Taha et al, investigated entrance skin dose for 500
Six different examinations were considered in the study. The methods applied was
indirect assessment using mathematical model. The input for the model included X-
ray dose output, backscatter factor, tube current-time product, focus to skin distance
patient thickness and kilovoltage peak. Taha et al study was within the range of
chest, skull, abdomen and pelvis radiography. Five radiology centres were considered
in the research, with 450 patients with ages ranging from 0 to 15 years old. In their
work, indirect assessment method was used to estimate entrance skin dose (ESD)
based on standard exposure data because of the lack of TLD system. The results
revealed that there was no standard procedure used in the hospitals due to the wide
variation in technical parameters (kVp and mAs) for the same examination and
indirect method. Satisfactory results were obtained during the study [43].
In Serbia and Montenegro, Ciraj et al, 2004, estimated patient dose for eleven
different X-ray examinations. A total of 419 patients were considered in the study.
Results revealed that only entrance surface dose from chest PA was above stated
reference level for plane film examination. The study underlined the importance of the
survey data in the establishment of national radiation protection and quality control
22
University of Ghana http://ugspace.ug.edu.gh
examination, d is the distance at which dosimeter reading was made, d FTD is focus to
23
University of Ghana http://ugspace.ug.edu.gh
CHAPTER THREE
METHODOLOGY
The main objective of this study was to perform quality control of seven X-ray
machines in south of Benin. This chapter outlines the materials and the experimental
methods used to achieve this objective. The methods used for each quality control test
(AAPM), Healing Arts Radiation Protection Act (H.A.R.P), Safety code 35,
International Atomic Energy Agency (IAEA), United States, Food and Drug
3.1 Equipment
Seven x-ray units were used. Their specifications are indicated in table 3.1
Table 3.1: X-ray equipment specifications
Tube Range
Tube Type of
Manufacturer Manufacture
Hospitals serial X-ray
& Country date
number system* kVp mAs
SIEMENS
Ste ANNE 579502 Not visible SF 40-125 0.5-800
Germany
SHIMADZU
CNHU November
Japan 75178 CR 40-150 0.5-800
2007
COMET
ROSERAIE ROHRE 348482 Not readable SF 40-125 0.1-400
Germany
PERLONG
MENONTIN 820017131 June 2017 DR 40-130 0.4-360
China
BMI
SURU-LERE 12799 January 2007 CR 40-150 0.5-630
Italy
(*) SF: Screen Film; CR: Computed Radiography; DR: Digital Radiography
24
University of Ghana http://ugspace.ug.edu.gh
Piranha is a solid-state detector designed to detect mostly X-ray radiation (figure 3.1).
It is manufactured by RTI group based in Sweden. The black Piranha used in the
was used for seven diagnostic radiography equipment to measure kilovoltage peak,
exposure time, half value layer and dose output of the equipment simultaneously. The
OCEAN Software associated with the detector allows the selection of the
3.2 below
25
University of Ghana http://ugspace.ug.edu.gh
The serial number of RDS-120 universal survey meter used is 20563054.It was
dose rate in the range 0.05μSv/h − 10Sv/h. RDS-120 universal survey meter, is
designed for X-ray and gamma radiation detection. When used with beta probe it can
detect beta radiation. The RDS-120 universal survey meter can detect photon energy
ranging from 50 keV to 3 MeV. The maximum X-ray energy generated by the
diagnostic X-ray equipment involved in the present research work is 150 keV. The
survey meter was suitable for measuring dose rate for leakage radiation of X-ray tube
housing.
The test tool allows the performance both QC for light field and X-ray field congruence
and central beam perpendicularity with image receptor. The test tool consists of flat
plate (20 × 25 cm) made of brass and contains rectangular outline and marking
etched. The main component of beam alignment tool is the sixteen centimeter high
26
University of Ghana http://ugspace.ug.edu.gh
acrylic cylinder. Each end of the cylinder holds one steel ball. The superimposition of
the steel ball determines the alignment of central X-ray beam. The test tool also
includes a leveller for checking whether patient table top is horizontal (figure 3.3) [47].
Tape measure with minimum count of 0.1 cm was used in the present research work
(figure 3.4). It was used in focus to table top distance (FTD) checking, measurement
27
University of Ghana http://ugspace.ug.edu.gh
At each diagnostic radiology department, at least one lead apron was available (figure
3.5). The radiation leakage test was performed in the X-ray room outside the
radiographers bunker. Lead apron was used to protect against X-ray in case of leakage.
The types of X-ray systems used in the study include Screen-Film (SF), Computed
Radiography (CR) and Digital Radiography (DR) system. The CR plate and the
cassette used for light field and X-ray field alignment tests was the plate or cassette
available that can contain the alignment test tool (figure 3.6). For screen film system
the film used were processed manually in two centres and automatically in one centre.
For CR and DR systems, the obtained image of the tests tool setting was printed on
radiographic film.
28
University of Ghana http://ugspace.ug.edu.gh
3.2 METHODS
Below are the equipment used for simultaneous measurement of kilovoltage peak,
3.2.1.1.2 Procedures
mA, 100 ms); (50kVp, 100 mA, 200 ms); (70kVp, 100 mA, 200 ms)] and with
4 Collimator of X-ray machine was set at 0° to face the patient table in such a
way that X-ray source assembly was centred over the table.
5 Focus to table top distance (FTD) was set to be 100 cm and it was verified
29
University of Ghana http://ugspace.ug.edu.gh
6 X-ray field was collimated where possible to just cover the sensitive area of
Figure 3.7: Set up for kVp, time, HVL and dose measurement
5 Tape measure
3.2..2.2 Procedures
1 Patient table was levelled to verify its horizontality and X-ray tube was centred
to the table so that the central beam was perpendicular to the table.
2 Loaded cassette or CR plate was placed on the table in the center of the light
30
University of Ghana http://ugspace.ug.edu.gh
3 The congruency tool was placed on the cassette in the center of the light field
and the perpendicularity tool in the center of the congruency tool (figure 3.8).
4 Collimator shutters were adjusted so that edges of the rectangular field coincide
4 Tape measure
3.2.1.3.2 Procedures
1 The collimator was shut off such that no primary X-ray was emitted.
3 The tape measure was used to define the position of 1 meter distance in front,
behind, left, right and on top away from the focal spot (figure 3.9).
31
University of Ghana http://ugspace.ug.edu.gh
For each quality control, radiation detector was exposed with exposures parameters
Accuracy of kilovoltage peak was performed on all the seven diagnostic X-ray
machines involved in the present study. The radiographic parameter mAs, was kept at
a constant value (20 mAs), and clinically used kilovoltage peak accuracy was
investigated. The value of kVp tested ranged from 50 kVp to 120 kVp. For each kVp
value, three measurement were recorded. The focus size was the frequently used in the
radiology department. Accuracy of the measured values were calculated by the means
̅̅̅̅
Xm − Xs
E=| | × 100% ( 3.1) [17]
Xs
32
University of Ghana http://ugspace.ug.edu.gh
Exposure time accuracy was conducted on six diagnostic X-ray machines. The QC test
takes account of two intervals of exposure time, exposure time below 10 ms and
exposure time above 10 ms. For each interval, three values of exposure times were
selected and each of them was measured three times. The mean exposure time was
̅̅̅̅
tm − ts
E=| | × 100% (3.2)
ts
Measurements were done at a constant and most commonly used kVp. The focal size
was which is frequently used in the radiology department. Results of the tests are
At the frequently used focal size, ten consecutive exposures of the Black Piranha were
performed at constant mAs, kVp and exposure time. The most commonly used kVp
was the one used for the reproducibility test. All seven X-ray units kilovoltage peak as
well as dose output and exposure time reproducibility were tested. For the Unit at
radiology department of Mènontin it was not possible to select exposure time. But at
the constant parameter setting exposure time was measured and its reproducibility was
checked. At Senande clinic exposure was display by the X-ray units after each
exposure. 122 ms was the exposure time used at this radiology department to perform
33
University of Ghana http://ugspace.ug.edu.gh
n
SDx 1 xi 2
CVx = =√ [∑ ( − 1) ] (3.3) [17]
x̅ n−1 x̅
i=1
Where CVx is coefficient of variation, x̅ is the mean value of measurement, SDx is the
standard deviation, n is the total number of measurement, xi is the ith measurement and
Deviation of each measurement from the mean measurement was also calculated based
on the equation:
xi
E = | − 1| × 100% (3.4)
x̅
Half value layer was measured by direct reading for all seven X-ray units. The tube
current and exposure time product were kept constant. Kilovoltage peak was varied in
increments of 10 from 40 to 90 kVp. For each kVp value the half value layer was
recorded three times. The mean of the three measurements was the HVL at the specific
kVp. Results of the tests are indicated from tables 4.41 to 4.47
mAs was kept constant. kVp was varied from 40 to 90 with increments of 10. The
dose output and kVp were measured. The specific dose expressed in mGy/mAs was
plotted against the square of measured kVp. Results of the tests are indicated from
34
University of Ghana http://ugspace.ug.edu.gh
Kilovoltage peak was kept constant (the most commonly used kVp) and mAs was
varied. Dose output was recorded for each mAs value and specific dose output (𝑫′𝒊 )
was calculated by dividing the dose output value by the corresponding mAs value. To
check for linearity, the linearity coefficient (LC) was calculated using the equation 3.5
|D′ i − D′ i+1 |
𝐿𝐶 = ′ (3.5)
|D i + D′ i+1 |
Where D′i is the ith specific dose, D′i+1 is the i + 1th specific dose
Results of the tests are indicated from tables 4.48 to 4.54
Leakage test was done at 80 kVp and mAs ranging from 60 to 63 for six X-ray units.
The X-ray unit at Ste Anne collimator was not able to select field size. The collimator
cannot be shut off. It was clear that the parameter selected for leakage tests did not
reflect the standard recommendation, but because on the X-ray unit console indication
of error appear at the highest kVp and mAs when exposure was undertaken, the
leakage tests was done with these factors to have at least an idea on leakage value for
At each radiology department the exposure factors for extremity radiography were
used to expose the collimator and beam alignment test tool. Beam alignment was
performed in all centre while for light field and radiation field congruence the image
obtained cannot allow to distinguish the radiation field edges (we should use coins to
35
University of Ghana http://ugspace.ug.edu.gh
delimit the edges for a better results). Results of the beam alignment tests are indicated
Measurement was done on all seven X-ray machines. After setting up the Piranha
detector, different exposure factors were selected to expose the detector. The X-ray
tube output graph was generated by plotting the specific dose against the kVp. Results
The chief radiographer in each diagnostic radiology department was asked to provide
the range of kVp and mAs used for six radiography examinations. The examination
included chest PA, abdomen AP, pelvis AP, skull AP, lumber spine AP and foot AP.
The Focus to detector distance was also provided for each examination. Twenty
patients thicknesses were measured for each radiography projection. To calculate the
entrance surface dose, average kVp value and average mAs value were introduced in
2
−1 )
d
ESD = B × Output(mGy ∗ mAs × mAs × ( ) (3.6) [41]
dFDD − t p
examination, d is the distance at which dosimeter reading was made, dFTD is focus to
Results of the estimated entrance surface dose are indicated from tables 4.58 to 4.64
36
University of Ghana http://ugspace.ug.edu.gh
In the present study, dose is measured in mGy and specific dose which is dose
The choice of focal spot size was based on the most commonly used in the radiology
department. For Clinic Ste Anne, CNHU, Clinique Senande and HZ-Porto Novo, broad
focus was used while for Clinic Roseraie and CHU-Suru Léré, fine focus was used.
The test was undertaken using 20 mAs. Tables 4.1 to 4.7 indicate kVp accuracy results.
Table 4.1: Results of kVp accuracy for X-ray unit at Clinic Ste Anne
37
University of Ghana http://ugspace.ug.edu.gh
Table 4.3: Results of kVp accuracy for X-ray unit at Clinic Roseraie
Table 4.4: Results of kVp accuracy for X-ray unit at Clinic Senande
38
University of Ghana http://ugspace.ug.edu.gh
Table 4.6: Results of kVp accuracy for X-ray unit at HZ-Porto Novo
Table 4.7: Results of kVp accuracy for X-ray unit at CHU-Suru Léré
39
University of Ghana http://ugspace.ug.edu.gh
Quality control tests were performed using 60 kVp for Clinic Ste Anne and 80 for the
others X-ray units. Tables 4.8 to 4.13 indicate exposure time less than 10 ms accuracy
results
Table 4.8: Results of accuracy of exposure time less than 10 ms for X-ray unit at
Table 4.9: Results of accuracy of exposure time less than 10 ms for X-ray unit at
CNHU
Table 4.10: Results of accuracy of exposure time less than 10 ms for X-ray unit at
HZ-Porto Novo
40
University of Ghana http://ugspace.ug.edu.gh
Table 4.11: Results of accuracy of exposure time less than 10 ms for X-ray unit at
Clinic Roseraie
Set Time Measured Time Average time Deviation
(ms) (ms) (ms) (%)
Table 4.12: Results of accuracy of exposure time less than 10 ms for X-ray unit at
Clinic Senande
Table 4.13: Results of accuracy of exposure time less than 10 ms for X-ray unit at
CHU-Suru Léré
41
University of Ghana http://ugspace.ug.edu.gh
The test was performed at constant kVp, 60 kVp for one hospital and 80 for the other
facilities. Tables 4.14 to 4.19 indicate exposure time greater than 10 ms accuracy
results
Table 4.14: Results of accuracy of exposure time greater than 10 ms for X-ray unit at
Table 4.15: Results of accuracy of exposure time greater than 10 ms for X-ray unit at
CNHU
Table 4.16: Results of accuracy of exposure time greater than 10 ms for X-ray unit at
Clinic Roseraie
42
University of Ghana http://ugspace.ug.edu.gh
Table 4.17: Accuracy of exposure time greater than 10 ms for X-ray unit at Clinic
Senande
Table 4.18: Accuracy of exposure time greater than 10 ms for X-ray unit at HZ-Porto
Novo
Table 4.19: Accuracy of exposure time greater than 10 ms for X-ray unit at CHU-Suru
Léré
43
University of Ghana http://ugspace.ug.edu.gh
Constant kVp and 20 mAs was used to perform reproducibility tests. 60 kVp was
used at clinic Ste Anne and 80 for the others X-ray units. 100 ms was used for the
Tables 4.20 to 4.26 indicate results of kVp reproducibility of the seven x-ray units
Deviation Deviation
Measured kVp (M) from Measured kVp (M) from
mean (%) mean (%)
44
University of Ghana http://ugspace.ug.edu.gh
Deviation Deviation
Measured kVp (M) from Measured kVp (M) from
mean (%) mean (%)
45
University of Ghana http://ugspace.ug.edu.gh
reproducibility for X-ray unit at CS- reproducibility for X-ray unit at HZ-
Deviation Deviation
Measured kVp (M) from Measured kVp (M) from
mean (%) mean (%)
46
University of Ghana http://ugspace.ug.edu.gh
CHU-Suru Lere
Deviation
Measured kVp (M) from
mean (%)
M1 80.91 0.26
M2 80.93 0.29
M3 80.46 0.30
M4 80.45 0.31
M5 80.41 0.36
M6 81.30 0.74
M7 81.19 0.61
M8 80.77 0.09
M9 80.39 0.39
M10 80.18 0.64
Mean 80.70
𝑺𝑫𝒙 𝟎. 𝟑𝟖
𝑪𝑽𝒙 (%) 𝟎. 𝟒𝟕
47
University of Ghana http://ugspace.ug.edu.gh
Tables 4.27 to 4.33 indicate results of exposure time reproducibility of the seven x-ray
units
Table 4.27: Results of exposure time Table 4.28: Results of exposure time
48
University of Ghana http://ugspace.ug.edu.gh
Table 4.29: Results of exposure time Table 4.30: Results of exposure time
49
University of Ghana http://ugspace.ug.edu.gh
Table 4.31: Results of exposure time Table 4.32: Results of exposure time
reproducibility for X-ray unit at CS- reproducibility for X-ray unit at HZ-
50
University of Ghana http://ugspace.ug.edu.gh
CHU-Suru Lere
M1 100.39 0.16
M2 99.89 0.34
M3 99.89 0.34
M4 100.39 0.16
M5 100.36 0.13
M6 100.34 0.11
M7 100.39 0.16
M8 100.38 0.15
M9 100.39 0.16
M10 99.89 0.34
Mean 100.23
𝑺𝑫𝒙 𝟎. 𝟐𝟒
𝑪𝑽𝒙 (%) 𝟎. 𝟐𝟒
51
University of Ghana http://ugspace.ug.edu.gh
Tables 4.34 to 4.40 indicate results of dose output reproducibility of the seven x-ray
units
Table 4.34: Results of dose output Table 4.35: Results of dose output
52
University of Ghana http://ugspace.ug.edu.gh
Table 4.36: Results dose output Table 4.37: Results of dose output
53
University of Ghana http://ugspace.ug.edu.gh
Table 4.38: Results of dose output Table 4.39: Results of dose output
reproducibility for X-ray unit at CS- reproducibility for X-ray unit at HZ-
54
University of Ghana http://ugspace.ug.edu.gh
CHU-Suru Lere
M1 0.8882 0.0281
M2 0.8888 0.0472
M3 0.8887 0.0353
M4 0.8878 0.0723
M5 0.8885 0.0142
M6 0.8876 0.0878
M7 0.8883 0.0065
M8 0.8885 0.0142
M9 0.8885 0.0142
M10 0.8885 0.0142
Mean 0.8884
𝑺𝑫𝒙 𝟎. 𝟎𝟎𝟎𝟒
𝑪𝑽𝒙 (%) 𝟎. 𝟎𝟒𝟒𝟒
55
University of Ghana http://ugspace.ug.edu.gh
HVL was measured at 20 mAs. Tables 4.41 to 4.47 indicate X-ray beam filtration
56
University of Ghana http://ugspace.ug.edu.gh
57
University of Ghana http://ugspace.ug.edu.gh
58
University of Ghana http://ugspace.ug.edu.gh
Figures 4.1 to 4.7 indicate the results of specific dose-kVp2 linearity tests for the
Figure 4.1: Specific dose versus kVp2 for X-ray unit at Clinic Ste Anne
Figure 4.2: Specific dose versus kVp2 for X-ray unit at CNHU
59
University of Ghana http://ugspace.ug.edu.gh
Figure 4.3: Specific dose versus kVp2 for X-ray unit at Clinic Roseraie
Figure 4.4: Specific dose versus kVp2 for X-ray unit at Clinic Senande
60
University of Ghana http://ugspace.ug.edu.gh
Figure 4.5: Specific dose versus kVp2 for X-ray unit at CS-Mènontin
Figure 4.6: Specific dose versus kVp2 for X-ray unit at HZ-Porto Novo
61
University of Ghana http://ugspace.ug.edu.gh
Figure 4.7: Specific dose versus kVp2 for X-ray unit at CHU-Suru Lere
The tests were performed at constant kVp. 60 kVp was used for clinic Ste Anne and 80
kVp for the other X-ray units. Tables 4.48 to 4.54 indicate results of specific dose-
Table 4.48: Results of specific dose-mAs linearity for X-ray unit at Clinic Ste Anne
62
University of Ghana http://ugspace.ug.edu.gh
Table 4.49: Results of specific dose-mAs linearity for X-ray unit at CNHU
Table 4.50: Results of specific dose-mAs linearity for X-ray unit at Clinic Roseraie
Table 4.51: Results of specific dose-mAs linearity for X-ray unit at Clinic Senande
63
University of Ghana http://ugspace.ug.edu.gh
Table 4.52: Results of specific dose-mAs linearity for X-ray unit at at CS-Mènontin
Table 4.53: Results of specific dose-mAs linearity for X-ray unit at HZ-Porto Novo
Table 4.54: Results of specific dose-mAs linearity for X-ray unit at CHU-Suru Lere
64
University of Ghana http://ugspace.ug.edu.gh
For all the five X-ray units involved in the leakage test, 80 kVp was used to perform
the tests. The mAs was set to be 63 at CNHU and 60 for the others X-ray units. Leakage
radiation was measured in μSv/h. Background dose rate was 0.05 μSv/h except at
Table 4.55: Results of leakage of X-ray tube housing for X-ray unit at CNHU, Clinic
65
University of Ghana http://ugspace.ug.edu.gh
Figure 4.8: Results of X-ray beam Figure 4.10: Results of X-ray beam
alignment test for X-ray unit at alignment test for X-ray unit at Clinic
Clinic Ste Anne Roseraie
Figure 4.9: Results of X-ray beam Figure 4.11: Results of X-ray beam
alignment test for X-ray unit at alignment test for X-ray unit at Clinic
CNHU Senande
66
University of Ghana http://ugspace.ug.edu.gh
Figure 4.12: Results of X-ray beam Figure 4.14: Results of X-ray beam
alignment test for X-ray unit at CS- alignment test for X-ray unit at CHU-
Mènontin Suru Lere
67
University of Ghana http://ugspace.ug.edu.gh
Figures 4.15 to 4.21 indicates the results of the seven X-ray units dose output curve
Figure 4.15: Specific dose output versus kVp for X-ray unit at Clinic Ste Anne
Figure 4.16: Specific dose output versus kVp for X-ray unit at CNHU
68
University of Ghana http://ugspace.ug.edu.gh
Figure 4.17: Specific dose output versus kVp for X-ray unit at Clinic Roseraie
Figure 4.18: Specific dose output versus kVp for X-ray unit at Clinic Senande
69
University of Ghana http://ugspace.ug.edu.gh
Figure 4.19: Specific dose output versus kVp for X-ray unit at CS-Mènontin
Figure 4.20: Specific dose output versus kVp for X-ray unit at HZ-Porto Novo
70
University of Ghana http://ugspace.ug.edu.gh
Figure 4.21: Specific dose output versus kVp for X-ray unit at CHU-Suru Lere
71
University of Ghana http://ugspace.ug.edu.gh
Table 4.56 below shows information obtained from radiographers at each Radiology
Department.
Table 4.56: Frequently used exposure factors and FDD for radiography projection
Departmen
Factors Projections
Radiology
&
t
72
University of Ghana http://ugspace.ug.edu.gh
Patient thickness
Projection Range Mean SD
Chest PA 19.50 - 26.00 22.75 1.20
Abdomen AP 21.00 - 26.00 23.50 1.60
Pelvis AP 20.00 - 26.00 23.00 1.80
Skull AP 17.00 – 23.00 20.00 0.90
Lumber Spine
21.00 - 26.00 23.50 1.60
AP
Foot AP 7.00 - 9.00 8.00 0.50
Tables 4.48 to 4.64 show the results of estimated entrance surface dose.
Table 4.58: Estimated entrance surface dose for Clinic Ste Anne
Chest Abdomen Pelvis Skull Lumber
Projection Foot AP
PA AP AP AP Spine AP
Mean kVp 117.0000 85.5000 83.0000 73.5000 85.5000 63.0000
Mean mAs 15.0000 35.0000 32.5000 20.5000 35.0000 4.7500
Specific Dose
0.1332 0.0702 0.0661 0.0516 0.0702 0.0376
(mGy/mAs)
FSD (cm) 177.2500 76.5000 77.0000 80.0000 76.5000 92.0000
ESD (mGy) 0.8585 5.6678 4.8914 2.2313 5.6678 0.2849
Error 0.0122 0.2411 0.2321 0.0518 0.2411 0.0033
73
University of Ghana http://ugspace.ug.edu.gh
74
University of Ghana http://ugspace.ug.edu.gh
75
University of Ghana http://ugspace.ug.edu.gh
4.3 DISCUSSION
The results obtained in assessing the kilovoltage peak accuracy in the seven radiology
departments revealed that five out of the seven diagnostic radiology departments
passed the quality control tests (table 4.65). The kVp deviation range for all kVp
selected for these five departments were far below the 5% limit recommended by
international organizations. These five X-ray units, when efficiently used would
produce the desired good quality X-ray for a given examination. 28.57% of the X-ray
units that were involved in the kVp accuracy tests have all or at least one of their
accuracy deviation values above 5%. Radiographic image contrast may be affected
because selected kVp is no longer the optimum kVp for a specific examination as the
radiographer thought. In Mènontin Hospital, it was found out that as the kVp increased
the deviation also increased and the inaccuracy is mainly observed at high kVp values
above 70 kVp. The minimum deviation calculated represent 69% of required limit.
The mean deviation of all kVp measured in this hospital, represent 94.8% of the
required limit. In clinic Senande, the kVp measurement showed inaccuracy for all kVp.
76
University of Ghana http://ugspace.ug.edu.gh
Because of lack of QC equipment, cross check after repair is not possible. It was
proposed to the Radiology Department the use of calibration curve to adjust the kVp
77
University of Ghana http://ugspace.ug.edu.gh
recommended limit.
Accuracy of exposure time below 10 ms was investigated for six X-ray units
including the X-ray unit at Clinic Senande where exposure time selection was not
possible. At Senande Radiology Department the control console indicates the exposure
time after exposure. The measurement done with the Piranha multifunction meter was
compared with the exposure time display on the X-ray unit control console. For all
measurements of exposure time below 10 ms, the X-ray units were within required
limit set by Safety Code S.C 35 (figure 4.24). For 66.67% of diagnostic radiology X-
ray units, the AAPM recommended limit was above all measured exposure time
deviation, meaning that in those X-rays units exposure time was accurate according to
78
University of Ghana http://ugspace.ug.edu.gh
AAPM standard limit. Two X-ray units present both accuracy and inaccuracy for some
limit.
radiology departments showed that 66.67% of the X-ray units passed the quality
control test of exposure time accuracy. The five X-ray units’ average deviation are all
HARP (10%), and Safety Code S.C35 (10% + 1ms) as indicated in figure 4.25. Two
X-ray units failed the QC test. Clinic Senande, X-ray unit accuracy tests revealed that
exposure time display by the control console was far different from the measured one.
79
University of Ghana http://ugspace.ug.edu.gh
4.3.1.3 Reproducibility
For all X-ray units involved in the present study, kilovoltage peak coefficient
of variation ws within standard limit specified by AAPM (5%) as shown in figure 4.26.
Average deviation of each measurement from the mean measurement ranged from
0.06% to 1.36%.This is far below the deviation limit recommended by S.C 35 (15%)
and HARP (20%). All the X-ray units passed the reproducibility tests. This mean that
in all the radiology department kVp can be repeated without any major deviation.
80
University of Ghana http://ugspace.ug.edu.gh
recommended limit.
value was 0.36% for all X-ray units. It represent 2.4% and 1.8% of limit respectively
recommended by AAPM (figure 4.27). All the X-ray units successfully passed the
81
University of Ghana http://ugspace.ug.edu.gh
limit.
Dose output reproducibility QC tests revealed that all the X-ray units involved
in the present research work, reproduced dose output within the specified limit
recorded was 1.49% (figure 4.28). This value represent 29.8% of AAPM limit.
Deviation between mean dose output calculated and each measured dose output range
from 0.03% to 1.36%. The maximum deviation of measurements from their mean
82
University of Ghana http://ugspace.ug.edu.gh
According to FDA, HVL which determines beam quality should not be below
a specific minimum value of HVL for a given kVp. In the figure 4.29 above, the red
points indicate the minimum values. All average HVL measured at kVp ranging from
40 to 90, for the seven diagnostic X-ray units are above the minimum required value
of FDA. All X-ray units involved in the present study passed successfully the QC
tests of beam quality. Entrance surface dose that a patient received from dose
machine will mainly depend on operator errors for a specific examination because
soft X-rays are efficiently removed from the beam spectrum before they reach the
patient.
83
University of Ghana http://ugspace.ug.edu.gh
Figure 4.30: Correlation of the specific dose versus kVp2 with linear trendline.
The kVp squared versus specific dose was plotted. Linear trendline was selected to see
how linear the variation of specific dose with kVp square was. For all the seven
diagnostic X-ray units, very high correlation was obtained as indicated figure 4.30.
The square of correlation coefficient ( R2) when taken at two decimal places, is equal
to 1 for all diagnostic radiology X-ray units. This means that variation of specific dose
with kVp square was linear for all kVp selected between 40 and 90. The error on
linearity ranged from 0.01% to 0.18% for all seven X-ray units. Specific dose from
84
University of Ghana http://ugspace.ug.edu.gh
recommended limit
mAs, are less than 0.1. All seven X-ray units in the diagnostic radiology departments
passed the specific dose-mAs linearity test (figure 4.31). This means that for screen
film system the same density can be achieved if the mAs is kept constant. According
to HARP regulation if linearity does not meet this standard, the unit must be serviced
and retested.
85
University of Ghana http://ugspace.ug.edu.gh
Figure 4.32: Comparison of leakage of X-ray tube housing from the seven X-ray
unit.
Five hospitals were involved in leakage of X-ray tube housing verification. The
highest dose rate at 1m from focal spot was recorded at Clinic Senande at the left side
of the collimator assembly (figure 4.32). This maximum dose rate of 23.24 μSv/h
obtained at 80 kVp and 60 mAs represent 2.32% of the recommended limit if the
measurement was done at the highest kVp (125) and mAs (500) of the X-ray unit at
Clinic Senande. For all other four X-ray units considered for leakage test in the present
study all dose rates recorded from all side of the X-ray units were below 2.5 μSv/h. It
is highly possible that leakage radiation level be below the required limit of 1 mSv/h
if the measurements was done at the highest parameter setting. Operators are safe from
86
University of Ghana http://ugspace.ug.edu.gh
Figure 4.33: Comparison of X-ray beam misalignment degree with acceptable limit
of 1.5°.
From beam alignment quality control test of the seven X-ray units, 71.43% of the
misalignment. The best beam alignment among the seven units was obtained at the
main public hospital (CNHU) which has the highest workload of about 80 patients a
day. Two X-ray units ( CS-Mènontin and CHU Suru Lere) are out of acceptable range
equipment cross check after repair, maintenance cannot be performed. Staff at those
two hospitals have to work with the image distortion due to misalignment. The highest
misalignment were observed with the digital X-ray unit. This may be caused by the
fact that the leveller indicated some inclination of the image receptor.
87
University of Ghana http://ugspace.ug.edu.gh
The specific dose output curve generated for each X-ray unit, was a power curve that
estimated the dose output per mAs as proportional to a power function of the kVp.
This power curve was confirmed by the correlation coefficient ranging from 99.87%
to 99.99%. Thus from the equations, for any kVp a corresponding specific dose can be
generated to calculate the entrance surface dose with very small error.
88
University of Ghana http://ugspace.ug.edu.gh
Figure 4.34 to 4.39 indicate the comparison of ESD with others studies
89
University of Ghana http://ugspace.ug.edu.gh
90
University of Ghana http://ugspace.ug.edu.gh
departments involved in the present research work compared with UK, Australia, Taha
et al and Inkoom et al ESD values reveal that higher entrance surface doses are
91
University of Ghana http://ugspace.ug.edu.gh
the higher result was due to the fact that the radiographer uses kVp below 100 and
higher mAs for all chest examinations. In the main public hospital, CNHU, where the
workload was the highest, patient doses from chest radiography projection represent
6.75 times the UK ESD value. These higher values were caused by the higher mAs
used by radiographers during chest X-ray examination. For clinic Ste Anne the result
could be attributed to both the large field size and the higher mAs used for chest X-ray
examination. It is urgent for the radiology department to solve the collimator blades
movement to allow X-ray field selection. At CHU-Suru Lere, the selected mAs values
ranged from 3.2-6.3 mAs and the mean entrance surface dose represent 1.63 times the
ESD value from Taha et al and Inkoom et al. This observation showed that patient
reduced.
Entrance surface dose estimated for all the six X-ray projections, showed that
the choice of radiographic parameter was the main reason for high ESD value obtained
92
University of Ghana http://ugspace.ug.edu.gh
5.1 Conclusion
The present study investigated seven diagnostic radiography X-ray machines in South
requirement and their influence on radiation dose received by patient while undergoing
some radiography examination. All the X-ray machines investigated have shown
time below 10 ms; reproducibility of kVp, exposure time and dose output; specific
dose-kVp square linearity; specific dose-mAs linearity and leakage of X-ray tube
housing. 2/7 of diagnostic X-ray machines failed quality control tests such as X-ray
beam alignment, exposure time above 10 ms and kVp accuracy. Estimated entrance
surface dose revealed that patients received compared with others publication very
high radiation dose in some Benin hospitals for chest PA, skull AP, abdomen AP,
pelvis AP, lumber spine AP and foot AP. The choice of radiographic parameters need
radiation. The good result obtained in some hospitals can be used as reference in others
93
University of Ghana http://ugspace.ug.edu.gh
5.2 Recommendations
1. Ensure that safe light and door interlock are fixed and functioning
2. Avoid several patients in the examination room, for chest X-ray radiography
facilities
6. Perform daily, weekly, monthly, quarterly, and yearly quality control tests to detect
that provide lower average entrance surface dose to patients. For chest PA, abdomen
AP, foot AP and skull AP, radiographic parameter used at CHU-Suru Lere could be
applied to the other centres, if image quality will not be affected. Radiography
regulation
3. Ministry of Health should pay more attention on radiology practice to held in patient
94
University of Ghana http://ugspace.ug.edu.gh
REFERENCES
[1] Stewart Carlyle Bushong, Radiologic Science for Technologists, physics, biology,
[2] H. Cember and T. E. Johnson, Introduction to Health physics, vol. 102, page 337,
2012.
[4] International Atomic Energy Agency. Terminology Used in Nuclear Safety and
1982
2004; http://www-pub.iaea.org/MTCD/publications/PDF/te_1423_web.pdf.
[8] Sezdi M., Dose Optimization for the quality control tests of X-Ray Equipment.
[9] P. P. Dendy, B. Heaton, Physics for Diagnostic Radiology, page 36, 2012
[12] Joseph John Bevelacqua, Health Physics in the 21st Century, page 203, 2008
95
University of Ghana http://ugspace.ug.edu.gh
Exposure Parameters Including Tube Voltage and Exposure Time in Private and
2015.
2013
[20] Wilson M. Ngoye, Jenny A. Motto, Wilbroad E. Muhogora, PhD. Quality Control
Quality control systems for X-Ray machines at different Hospitals using patient’s
96
University of Ghana http://ugspace.ug.edu.gh
[24] Maria Lucia Nana I Ebisawa, Maria de Fatima A Magon, Yvone M Mascarenhas
[25] G.K. Korir, J.S. Wambani, I. K. Korir, Establishing a quality assurance baseline
Parameters of Half Value Layer, Beam Alignment and Collimator Test Tools on
[27] Yesaya Y. Sungita, Simon S.L. Mdoe, and Peter Msaki, Diagnostic X-ray
[28] G.M. Hassana, N. Rabiea, K.A. Mustafab, S.S. Abdel-Khalikb. Study on the
quality assurance of diagnostic X-ray machines and assessment of the absorbed dose
to patients, 2012
Procedures for the Installation, Use and Control of X-Ray Equipment in Large Medical
[30] American Association of Physicist in Medicine report N°4, Basic Quality Control
97
University of Ghana http://ugspace.ug.edu.gh
[32] Food and drug administration, Performance Standard for Diagnostic X-ray
Systems and their Major Components, Title 21, Code of Federal Regulations, Sections
1020.30, 2005
the routine performance testing of diagnostic X-ray imaging systems, Report NO. 77,
1998
[34] Healing Arts Radiation Protection Act, R.R.O. 1990, Regulation 543 X-ray safety
code, 2011
www.flukebiomedical.com/rms
[36] J.T. Bushberg, J.A. Seibert, E.M Leidholdt, J.M. Boone, E.J. Goldschmidt. The
2014
[38] Daniel Ackom, Edem Sosu, Stephen Inkoom, Cyril Schandorf, Estimation of
[39] N.N. Jibiri , C.J. Olowookere, Patient dose audit of the most frequent radiographic
examinations and the proposed local diagnostic reference levels in south western
Emmanuel O. Darko, Estimation of adult patient doses for selected X-ray diagnostic
examinations, 2014
98
University of Ghana http://ugspace.ug.edu.gh
[41] M.T. Taha, F.H. Al-Ghorabie , R.A. Kutbi , W.K. Saib, Assessment of entrance
skin doses for patients undergoing diagnostic X-ray examinations in King Abdullah
[42] Ademola, A. K., Obed, R. I., Adejumobi, C. A., Abodunrin, O. P., Alabi, O. F.,
chest, skull, abdomen and pelvis of children in five selected hospitals in Nigeria, 2013
[43] Jumaa Yousif Tamboul, Mohamed Yousef, Khadija Mokhtar, Ahmed Alfaki,
Abdelmoneim Sulieman, Assessment of entrance surface dose for the patients from
[44] Olivera Ciraj, Srpko Marković, Duško Košutić, Patient dose from conventional
[45] RTI, Black Piranha – Easy & Fast X-ray Quality Control, 15 March 2018 at 7
PM www.rtigroup.com
[47] Gammex a sun nuclear company, Collimator and Beam Alignment QC Tools,
2015
[48] Inkoom, S., Togobo, J., Emi-reynolds, G., Oddoye, A., Ntiri, T. O., and Gyekye,
99
University of Ghana http://ugspace.ug.edu.gh
APPENDIX A
Table A.1: Results of specific dose - kVp2 linearity for X-ray unit at Clinic Ste Anne
Table A.2: Results of specific dose - kVp2 linearity for X-ray unit at CNHU
Table A.3: Results of specific dose - kVp2 linearity for X-ray unit at Clinic Roseraie
100
University of Ghana http://ugspace.ug.edu.gh
Table A.4: Results of specific dose - kVp2 linearity for X-ray unit at Clinic Senande
Table A.5: Results of specific dose - kVp2 linearity for X-ray unit at Clinic CS-
Mènontin
Table A.6: Results of specific dose - kVp2 linearity for X-ray unit at HZ-Porto Novo
101
University of Ghana http://ugspace.ug.edu.gh
Table A.7: Results of specific dose - kVp2 linearity for X-ray unit at CHU-Suru Lere
102
University of Ghana http://ugspace.ug.edu.gh
APPENDIX B
Table B.1: Results of dose output for X-ray unit at Clinic Ste Anne
Set kVp Set mAs Dose output (D) Specific dose (𝑫𝒔 = 𝑫/𝒎𝑨𝒔)
55 40 1.17 0.029
60 32 1.15 0.036
66 22 0.97 0.044
70 20 1.00 0.050
77 16 0.97 0.061
81 10 0.67 0.067
90 6.3 0.51 0.081
96 4 0.37 0.092
Set kVp Set mAs Dose output (D) Specific dose (𝑫𝒔 = 𝑫/𝒎𝑨𝒔)
55 40 1.15 0.029
60 32 1.09 0.034
66 22 0.96 0.044
70 20 0.96 0.048
77 16 0.92 0.058
81 10 0.64 0.064
96 4 0.35 0.089
102 2 0.20 0.098
Table B.3: Results of dose output for X-ray unit at Clinic Roseraie
Set kVp Set mAs Dose output (D) Specific dose (𝑫𝒔 = 𝑫/𝒎𝑨𝒔)
55 40 1.18 0.029
60 32 1.15 0.036
66 22 0.96 0.044
70 20 0.98 0.049
77 16 0.95 0.060
80 10 0.64 0.064
90 6.4 0.52 0.081
96 4 0.37 0.091
103
University of Ghana http://ugspace.ug.edu.gh
Table B.4: Results of dose output for X-ray unit at Clinic Senande
Set kVp Set mAs Dose output (D) Specific dose (𝑫𝒔 = 𝑫/𝒎𝑨𝒔)
56 40 0.69 0.017
60 35 0.73 0.021
66 25 0.69 0.028
70 20 0.65 0.032
76 15 0.60 0.040
80 10 0.47 0.047
95 4 0.30 0.075
Set kVp Set mAs Dose output (D) Specific dose (𝑫𝒔 = 𝑫/𝒎𝑨𝒔)
55 40 0.54 0.013
60 32 0.51 0.016
70 22 0.50 0.023
80 10 0.29 0.029
90 5 0.18 0.036
Table B.6: Results of dose output for X-ray unit at HZ-Porto Novo
Set kVp Set mAs Dose output (D) Specific dose (𝑫𝒔 = 𝑫/𝒎𝑨𝒔)
55 32 0.70 0.022
60 25 0.67 0.027
70 20 0.75 0.037
80 16 0.79 0.049
90 8 0.50 0.062
Table B.7: Results of dose output for X-ray unit at CHU-Suru Lere
Set kVp Set mAs Dose output (D) Specific dose (𝑫𝒔 = 𝑫/𝒎𝑨𝒔)
55 40 0.77 0.019
60 32 0.75 0.024
70 25 0.82 0.033
80 20 0.87 0.044
90 6.3 0.36 0.057
104
University of Ghana http://ugspace.ug.edu.gh
105