Method for evaluating performance
characteristics of ultrasonic
pulse-echo testing systems
ICS 19.100
Reference number: JIS Z 2352 : 2010 (E)
PROTECTED BY COPYRIGHT 17 S
Z 2352 : 2010
© JSA 2012
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or
utilized in any form or by any means, electronic or mechanical, including photocopying and
microfilm, without permission in writing from the publisher.
Printed in Japan
KKiAT
PROTECTED BY COPYRIGHT
Z 2352: 2010
Contents
Page
7 Periodic checks········································································································· 21
8 Record················································ ...................................................... ··················21
Annex JA (informative) Comparison table between JIS and corresponding
International Standard ....................................................... 24
(i)
PROTECTED BY COPYRIGHT
Z 2352 : 2010
Foreword
This translation has been made based on the original Japanese Industrial Standard
revised by the Minister of Economy, Trade and Industry, through deliberations at
the Japanese Industrial Standards Committee as the result of proposal of revision
of Japanese Industrial Standard submitted by the Japanese Society for Non-Destructive
Inspection (JSNDI)/Japanese Standards Association (JSA) with the draft being attached,
based on the provision of Article 12 Clause 1 of the Industrial Standardization Law
applicable to the case of revision by the provision of Article 14.
Consequently JIS Z 2352: 1992 is replaced with this Standard.
This JIS document is protected by the Copyright Law.
Attention is drawn to the possibility that some parts of this Standard may conflict
with a patent right, application for a patent after opening to the public, utility model
right or application for registration of utility model after opening to the public which
have technical properties. The relevant Minister and the Japanese Industrial Standards
Committee are not responsible for identifying the patent right, application for a patent
after opening to the public, utility model right or application for registration of utility
model after opening to the public which have the said technical properties.
(ii)
PROTECTED BY COPYRIGHT
JAPANESE INDUSTRIAL STANDARD JIS Z 2352 : 2010
In troduction
This Japanese Industrial Standard has been prepared based on the first edition of
ISO 18175 published in 2004, while modifying some of the technical contents, based
on the actual management of the standard in Japan, in such a way as to facilitate
smoother management of the standard in Japan.
The portions with dotted underlines are the matters in which the contents of the
corresponding International Standard have been modified. A list of modifications with
the explanations is given in Annex JA.
1 Scope
This Standard specifies the method for evaluating performance characteristics and
the method for periodic checks of the ultrasonic pulse-echo testing systems that are
composed of the combination of the manually controlled ultrasonic test instrument hav-
ing a basic display based on an ultrasonic pulse-echo technique and a probe.
NOTE 1 For the method for evaluating the electrical characteristics of an ultra-
sonic testing instrument, see JIS Z 2351.
NOTE 2 The International Standard corresponding to this Standard and the
symbol of degree of correspondence are as follows.
ISO 18175: 2004 Non-destructive testing-Evaluating performance char-
acteristics of ultrasonic pulse-echo testing systems without the use of
electronic measurement instruments (MOD)
The symbols which denote the degree of correspondence in the con-
tents in the corresponding International Standard and JIS are IDT
(identical), MOD (modified) and NEQ (not equivalent) according to ISO/
lEe Guide 21-1.
2 Norma ti ve references
The following standards contain provisions which, through reference in this text,
constitute provisions of this Standard. The most recent editions of the standards
(including amendments) indicated below shall be applied.
PROTECTED BY COPYRIGHT
2
Z 2352 : 2010
3.5 %fs
relative value when the full scale of the time base or normal axis of a display is taken
as 100 %
PROTECTED BY COPYRIGHT
3
Z 2352: 2010
PROTECTED BY COPYRIGHT
4
Z 2352 : 2010
Usually, measurement shall be performed at the rise position of an echo, the posi-
tion in which an echo signal exceeds one half the echo height, the peak position of an
echo, the zero cross position or the position that exceeds the threshold set beforehand.
PROTECTED BY COPYRIGHT
5
Z 2352: 2010
echo
Adjustment Adjustment
method 1 method 2 10~~-+--~~-+--~~~~1~~ M
%fs %fs %fs %fs /W"
9 ~-+---+------1f----+---+----+--+V"""7"'-{4,
/·~v·
Set point (80, 9)
1 0 0 8 -Error line for /V/ I I
o d .. h l//);,/
~ 7 etermmmg t _et---;"'</"[l'ih<'--7"-"I-<:::---+---+----I
2 20 10
~ -lin~arity range/, ~V . . . . Theoretical line
3 40 20 ~ 6 '\~ / L'I' Error
4 60 30 ~ 5 '""'" /j/;/ I
<5 ~~/
5 80 40
4 /~/
6 100 50 3 / /. / Set point (20, 3)
2 /'././ 1 1 1 'I
7 60
~7
1()-<---'------'---"-----'------'----'----+---+---'-----L...~
8 70 o 10 20 30 40 50 60 70 80 90 100 (%fs)
9 80
Figure 1 Example of graphical
10 90
presentation of linearity
11 100
of the time base (in the
Note a) The maximum continuous area stayed in
case of adjustment
the error range determined beforehand
shall be taken as the linearity range. method 2)
PROTECTED BY COPYRIGHT
6
Z 2352 : 2010
6.2.2.2 Measurement
The measurement shall be performed as follows.
a) Connect the probe contacted to the test block via couplant or a signal generator
to the flaw detector (see figure 2), and display the echo or dummy echo from the
signal generator on the display.
b) Set the other controls of the flaw detector to the conditions simulating the actual
use.
c) Adjust the gain controls of the ultrasonic test instrument to make the height of
the echo or dummy echo 100 %fs.
d) Reduce the gain in 2 dB steps by gain controls and read the echo height in %fs.
Continue this procedure until 26 dB is read.
e) The difference between the theoretical value and the measured value shall be des-
ignated as the error. The "positive" maximum error (+hMAX) and the "negative"
maximum error (-hMAX) shall be obtained.
f) Continue to reduce the gain to 30 dB and observe to judge whether echo can be
clearly seen.
g) When setting of the echo height of 100 %fs in the measurement above-mentioned
is difficult, set to the value which is nearest to 100 %fs but not more than 100 %fs,
and read the echo height.
b) If the echo is clearly observed at the reduced gain of 30 dB, record as "echo does
not disappear". If the echo disappears at 30 dB or lower, record the amount of
gain reduced at the disappearance.
PROTECTED BY COPYRIGHT
7
Z 2352: 2010
Ultrasonic Ultrasonic
test test
instrument instrument ~---~
Dummy
Calibration
signal
block
generator
6.2.3.2 Measurement
Measurement shall be performed as follows.
a) Put the probe where two echoes have the echo height of about 2 to 1, and allow
the larger echo height (HA ) to be able to change from not more than 10 %fs to
100 %fs with a gain control.
b) Adjust the gain control and set HA and HB to 60 %fs and 30 %fs, respectively. When
it is difficult to perform accurate adjustment, set to any of the initial echo height
values (HAS, H BS ) given in table 3.
c) Maintain a state in which an echo is stable, change HA by 10 % by operating the
gain control from 10 %fs to 100 %fs and read the value of HA and H B. When HA
cannot be set to 10 %fs, measurement may be performed from 10 %fs or lower.
PROTECTED BY COPYRIGHT
8
Z 2352 : 2010
~
D
r
Test
surface
1----+-+----'--...,
Location Dimension
Unit: mm
Tolerance
A 32 ±1
~ B 25 ±1
C 19 ±1
-r- ~ d1
D 25 ±1
~ E 19 ±1
-'-- r---< d2 H 75 ±1
T 25 ±0.2
W 50 ±1
w d" d2 ¢1.2 ±0.1
Smoothness ±0.02
Parallelism ±0.02
Surface roughness (fAm) Rz 6.3
HAs %fs 60 65 64 63 62 61 60 59 58 57 56 55
H Bs %fs 30 30 to 29 to 29 to 28 to 27 to 27 to 27 to 26 to 26 to 25 to 25 to
36 36 35 34 34 33 33 32 32 31 31
E3(44,100) F3(S6,100)
Example of setting of limit line
when the initial values of echo height
(HAS, H Bs ) are not 60 %fs and 30 %fs
Unit: %fs
Coordinate HB axis HA axis
symbol
Echo
height E1 -6 0
HA (%fs) F1 6 0
E2 H Bs-1 HAS
F2 HBs+1 HAS
E3 100HBS IHAS -6 100
F3 100HBS IHAS +6 100
60
Echo height HB (%fs)
PROTECTED BY COPYRIGHT
9
Z 2352: 2010
6.2.4.2 Measurement
Measurement shall be performed as follows.
a) Connect the flaw detector, probe, connecting cable and test block as shown in fig-
ure 5, and display the echo. In the case of the double probe technique, use the
initial transmitted pulse, and in the case of the single probe technique, use the
first back wall echo.
b) With the attenuator set to about 30 dB, perform adjustment of sweep and gain,
display the echo on the screen centre and set the echo height to 50 %fs.
c) Change the echo height to the full scale by decreasing the attenuation of the ex-
ternal attenuator by 1 dB. At this time, record the echo height at each step in
%fs.
d) Return the attenuator to 30 dB, and from the state where the echo height has
become 50 %fs, make attenuation increase in five steps of 2 dB and then in steps
of 4 dB, and continue until the echo practically disappears. Record the echo height
at each step.
e) When setting of the echo height of 50 %fs in the measurement above-mentioned
is difficult, set to the value which is nearest to 50 %fs but not more than 50 %fs,
and read the echo height (Hs).
Flaw detector
Transmitting terminal
o Receiving terminal
Termina-
'D~tion
Coaxial cable iI resistor
~
Attenuator
PROTECTED BY COPYRIGHT
10
Z 2352 : 2010
PROTECTED BY COPYRIGHT
11
Z 2352: 2010
PROTECTED BY COPYRIGHT
12
Z 2352 : 2010
110
VU3 l 6, 106)
100 J. Vs3(6, 100)
90 /l0VI>3( 6,94)
Example of setting of limit lines
~V when the initial value of
~f(N
80 Theoretical line
echo height (Hs) is arbitrary %fs.
70 LI. v
60
A~ Example of limit Ii nes Coordinate dB axis Echo height
symbol (%fs)
Echo 50 ~0,5I) U1 LO (1I10)Hs+ 1
~f
height S2(0, 50)
D2(0, 49) 81 LO 0
(%fs) 40
~ ~j>' D1 LO -(1I10)Hs-1
30
I'.C~
U2 0
,IIV/'
11"11" ./ V
Hs+1
20 82 0 Hs
v~ 1
II"~ 1/
ItVii D2 0 Hs-1
Ul (LO, 6)
SI (LO, 0)
Dl (LO, -6)
"I
~
iI~
I"1'987654 3 2 1 0 1 2 3 4 5 6
U3
83
6
6
(21/10)Hs+ 1
(20/10)H s
14 10 Increase in Decrease in
30 12 D3 6 (19/10)Hs -1
20 16 attenuation attenuation
L0 (dB) (dB)
18
PROTECTED BY COPYRIGHT
13
Z 2352: 2010
Unit: mm
RB-RA No.OOO
PROTECTED BY COPYRIGHT
14
Z 2352 : 2010
a) Adjust the probe position and the gain controls of the ultrasonic test instrument
so that the height of the echo from the flat-bottomed hole of the test block (see
figure 8) is maximum and the echo height is 100 %fs.
b) In the case of RB-RB type reference block, examine if the flat-bottomed echo can
be separated from the back wall echo at 3 %fs.
c) In the case of RB-RC type reference block, examine if the flat bottomed-echo can
be separated from the echo from the test surface at 3 %fs.
d) The minimum value of the distance between the flat-bottomed hole and the back
wall, obtained using the RB-RB type reference block, shall be the far surface reso-
lution.
e) The minimum value of the distance between the flat-bottomed hole and the test
surface, obtained using the RB-RC type reference block, shall be the near surface
resolution.
f) Dimensional tolerance on the test block shall be in accordance with the middle grade
as given in JIS B 0405.
Unit: mm
RB-RB No.OOO RB-RB No.OOO
-1 -2
o Test
- r-- r-
00 0
00
M l(")
surface
Sur- Rz 6.3
face
~!
rough- Bottom
II
I I I
ness face
(fAm)
Rz 6.3
Others
Rz 25
Unit: mm
RB-RC No.OOO
-1
Test surface
Sur- Rz 6.3
face
rough- Bottom face
ness Rz 6.3
(~tm) Others
Rz 25
Figure 8 RB-RB type and RB-RC type reference blocks for resolution
measurement
PROTECTED BY COPYRIGHT
15
Z 2352: 2010
a) Locate the probe at the position in which the echo height from the flat-bottomed
hole of ISO-RA type reference block (see figure 9) becomes maximum, and adjust
the flaw detector to set the echo height to 80 %fs.
b) In the case of the measurement section of far surface resolution, examine whether
the echo from the flat-bottomed hole can be separated from the back wall echo at
not more than 20 %fs.
c) In the case of the measurement section of near surface resolution, examine whether
the echo from the flat-bottomed hole can be separated from the echo from the test
surface at not more than 20 %fs.
d) Take the minimum interval between the separable flat-bottomed hole and the
bottom face as far surface resolution.
e) Take the minimum interval between the separable flat-bottomed hole and the test
surface as near surface resolution.
f) When the measurement exceeding the measurement range of the test block men-
tioned-above is required, prepare the ISO-RA type reference block of which the beam
path difference is changed.
Unit: mm
Target Dimen- Toler-
location sion ance
L 200 0.5
L T 82.5 0.5
<6
A A A A A A W 50 0.5
C 25 0.5
~r j/ ~ j/ j/ ~ ~ S 100 0.5
o.y o.y 0.)1-' o.y 0.;/ o.y A 25 0.5
E 25 0.5
w
d 1 to d6 1.2 0.01
k:~
II II II ul r,
"
"
----'-'----
"
Bl
B2
2.5
5
0.1
0.1
For near surface
resolution B3 7.5 0.1
For far surface F, 7.5 0.1
~ r:e: ~
resolution F2 12.5 0.1
II II
F3 17.5 0.1
S
Surface roughness (~tm)
PROTECTED BY COPYRIGHT
16
Z 2352 : 2010
b) For the report, record all of the reference blocks, test holes, probes and test pa-
rameters used.
6.4.2 Measuremen t
Measurement shall be performed as follows.
a) Connect the instrument as shown in figure II.
b) Couple the probe via the couplant to the RB-RD type reference block, and adjust
the position of the probe so that the heights of echoes from adjacent two steps of
the reference block are maximum and equal.
c) Adjust the gain controls to set these echo heights to 100 %fs.
d) Obtain the minimum step where the level of the valley of the two echoes will not
exceed 3 %fs (-30 dB) of the scale. This shall be the resolution of angle beam test-
ing.
e) When the measurement of the resolution that exceeds the maximum step of the
RB-RD type reference block is needed, the step section of 9 mm of the RB-HS type
reference block shall be used.
f) Dimensional tolerance for the test block shall be in accordance with the middle
grade as given in JIS B 0405.
PROTECTED BY COPYRIGHT
17
Z 2352: 2010
Unit: mm
135
40
Unit: mm
Used part
,-'--.,
30 30
Ultrasonic
test
instrument
Angle probe
6.5.2 Measurement
The measurement shall be performed as follows.
PROTECTED BY COPYRIGHT
18
Z 2352 : 2010
a) With the probe not coupled with the test block, adjust the gain controls to get the
highest sensitivity which can maintain the noise level (peak value) on the display
within 10 %fs. Take the control position of the gain controls at this time as So [see
figure 12 a)].
b) Contact the probe via the couplant to the test block, and adjust the probe posi-
tion to get the maximum height of the echo from the reference hole or an artifi-
cial flaw. Adjust the gain control so that the echo height is 50 %fs and obtain the
control position Sl of the gain controls [see figure 12 b)].
b) If other test block than calibration block is used, record the result calculated by
formula (2).
S = So - Sl±A (dB) .............................................................. (2)
where, A: the ratio (dB value) of the height of echo from the
reference hole of STB-G VI5-5.6 to the height of echo
from an artificial flaw in the test block used. However,
(+) sign indicates that the height of echo from an arti-
ficial flaw is lower than that of echo from STB-G V15-
5.6 and (-) sign indicates that the echo height of an
artificial flaw is higher than that of STB-G VI5-5.6.
Ultrasonic Ultrasonic
test test
instrument instrument
PROTECTED BY COPYRIGHT
19
Z 2352: 2010
6.6.2 Measurement
Measurement shall be performed as follows.
a) At sensitivity
1) Connect the probe to the flaw detector, and adjust the gain control to get the
sensitivity as high as possible within 10 %fs noise level on the display with the
probe not coupled with the test block. Take the control position of the gain control
at this time as Ao.
2) In connection of the instrument to be used which is shown in figure 13 a), con-
tact the angle probe via. couplant to the STB-Al test block. Turn the probe to
the RI00 surface, and make the probe index coincide with the RI00 mark of the
test block keeping the ultrasonic beam direction in parallel with the side face
of this block. Adjust the gain control to set the echo height from the RI00 sur-
face to 50 %fs at this time and take the control position of the gain control at
this time as A 1.
b) A2 sensitivity
1) Obtain gain Ao in the same way as Al sensitivity.
2) In connection of the instrument to be used which is shown in figure 13 b), con-
tact the angle probe via. couplant to the STB-A2 test block and aim the through-
hole of 1.5 mm diameter with one skip. At this time, adjust the gain control to
set the echo height from the through-hole to 50 %fs after the echo height is
maximized by adjusting the position of the probe. Take the control position of
the gain control at this time as A 2 •
c) Record the type name of test block and probe, serial number and the control po-
sition of the flaw detector.
PROTECTED BY COPYRIGHT
20
Z 2352 : 2010
Ultrasonic Ultrasonic
test test
instrument instrument
Calibration block
STB-Al type
6.7 Dead zone after transmitter pulse with straight beam testing technique
6.7.2 Measuremen t
Measurement shall be performed as follows.
a) In setting in figure 14, contact the probe via couplant to the location of the thick-
ness of 48 mm of the test block and adjust the gain control to set the back wall
echo height (B l ) of this part to 100 %fs.
b) Move the probe from the thick portion of the block to the thin portion via couplant,
and obtain the thickness of the plate (t) just before the echo height of Bl becomes
80 %fs or less.
PROTECTED BY COPYRIGHT
21
Z 2352: 2010
Unit: mm
Ultrasonic
test
instrument
Normal probe
7 Periodic checks
The periodic checks shall be performed periodically at least once a year in order to
------------------------------------------------------------------------------------------------------------------------------------------------------
~~~fi:~!!l:_~_1?:~~__~~_~_J~~_~_~~_~~~~~_I?:~_~_J?~~f<?~~~_I?:~_~__?K_~_1?:~_~~_~~~~_?_l!t~_~_~_~~_t~_~~~_~~~!!_~_~_~ __l!!:~t~_~
tained. The frequency of checks and items to be checked shall be decided by the user
------------------------------------------------------------------------------------------------------------------------------------------------------------
of this Standard. The periodic check items shall include the followings.
-------------------------------------------------------------------------------------------------------------------------------------
a) Vertical linearity.
-----------------------------------------
8 Record
The record of performance measurements shall include all the relevant information
in sufficient detail, such as the procedure, instrument, control position of the flaw de-
tector, etc. so that the measurements could be duplicated later if desired. For peri-
odic checking, a format of recording should be fully defined in order to facilitate
comparison between multiple checking results.
Required items shall be recorded according to the following recommended format.
a) Date of measurement
b) Person in charge of measurement
c) Instruments
1) Test block or dummy signal generator which has been used for the measurement
2) Measured objects
2.1) Type and serial number of ultrasonic test instrument
2.2) Type name and serial number of probe
2.3) Type and length of probe cable
PROTECTED BY COPYRIGHT
22
Z 2352 : 2010
d) Measurement conditions
1) Test frequency in measurement of JIS Z 2344 (except for the case of presenta-
tion of linearity of time base)
2) Controls of the ultrasonic test instrument
3) Method of reading of echo position and echo height
4) Other informative matters
e) Measurement results
1) Linearity of the time base For the linearity of the time base, the positive
and negative maximums of error shall be recorded in %fs.
When the table or graph of measurement results is created, the table or graph
and the linearity range shall be recorded.
2) Vertical linearity of normal axis For the vertical linearity, the positive and
negative maximums of error shall be recorded in %fs.
When the table or graph of measurement results is created, the table or graph
and the linearity range shall be recorded.
3) Resolution of straight beam testing technique For the resolution of the
straight beam testing technique, the kind of measurement method applied shall
be clearly described, and it shall be as follows.
3.1) In the case of resolution measurement method A The far surface reso-
lution obtained by the RB-RA type reference blocks shall be recorded in
millimetres.
When the minimum step section is separated, the record shall state "less
than 4 mm", and when the maximum step section cannot be separated, it shall
state "more than 9 mm".
3.2) In the case of resolution measurement method B The far surface reso-
lution and near surface resolution obtained by the RB-RB type reference blocks
and RB-RC type reference blocks shall be recorded in millimetres.
When the minimum path difference is separated, the record shall state "less
than 1 mm", and when the maximum path difference cannot be separated, it
shall state "more than 15 mm".
3.3) In the case of resolution measurement method C The far surface reso-
lution and near surface resolution obtained by the ISO-RA type reference blocks
shall be recorded in millimetres.
4) Resolution of angle beam testing technique The resolution of the angle
beam testing technique obtained shall be recorded in millimetres.
In the RB-RD type reference blocks, when the minimum path difference is
separated, the record shall state "less than 2 mm", and when the maximum path
difference cannot be separated, it shall state "more than 7 mm". When the step
of 9 mm in level difference in the RB-HS type reference blocks cannot be sepa-
rated, the record shall state "more than 9 mm".
PROTECTED BY COPYRIGHT
23
Z 2352: 2010
Bibliography
JIS Z 2351 Method for assessing the electrical characteristics of ultrasonic testing
instrument using pulse echo technique
PROTECTED BY COPYRIGHT
Annex JA (informative) Nt-:l
t-:l~
Comparison table between JIS and corresponding International Standard W
01
t-:l
t-:l
JIS Z 2352 :2010 Method for evaluating performance characteristics of ultrasonic ISO 18175: 2004 Non-destructive testing-Evaluating performance o
t--L
pulse-echo testing systems characteristics of ultrasonic pulse-echo testing systems without the o
use of electronic measurement instruments
(I) Requirements in JIS (II) Inter- (III) Requirements in Inter- (IV) Classification and details of (V) Justification for
national national Standard technical deviation between JIS the technical deviation
Standard and the International Standard by and future measures
number clause
No. and title of Content No. of I Content Classifi- Detail of technical
"'d clause clause cation by I deviation
~
o clause
~
M 1 Scope Method for evaluating 1 Approxima tely Alteration, Alteration of composi- There is no technical
o
~ performance characteris- identical with JIS. Deletion tion of standard deviation caused by
M
U tics of testing systems However, JIS has this alteration of com-
ttl
and method for periodic an independent position. The matter
~ checks are specified. clause for listing concerning the method
o the measurement of periodical checks
o items of perfor- will be suggested to
~ mance characteris- ISO.
~
tics. ISO specifies
Q the measurement
::r:
~ method of accuracy
of gain controls in
6.6, whereas JIS
deletes this matter
and incorporates it
into JIS Z 2351.
3 Terms and defini - I Necessary terms and 3 Reference to ISO Addition JIS added the defini- JIS permits the use of
tions definitions are specified, 5577, ASTM E tions of "digital flaw digital flaw detector,
in addition to referring to 1316, IEEE Std detector", "dummy dummy signal and
JIS Z 2300. 100. signal" and "position position of reading of
of reading of the time the time base. Sug-
base". gestion will be made to
ISO.
(I) Requirements in JIS (II) Inter- (III) Requirements in Inter- (IV) Classification and details of (V) Justification for
national national Standard technical deviation between JIS the technical deviation
Standard and the International Standard by and future measures
number clause
No. and title of Content No. of Content Classifi- Detail of technical
clause clause cation by deviation
clause
4 Measurement Measurement items of 1 ISO lists the mea- Alteration
i terns of performance performance characteris- surement items in
characteristics tics are specified. the scope.
5 Matters concern- Qualification require- 6.1 Approxima tely Alteration JIS added the require- Suggestion will be
ing measurement ments of person in identical with JIS. ment based on the made to ISO.
"'d that are to be noted charge of measurement, permission to use the
:;:d
o selection of measurement digital flaw detector.
~
t:rj items, method for read- JIS describes the
Q
~
ing data, accuracy, etc. qualification of a per-
t:rj are specified. son in charge of mea-
t:j
surement.
Cd
~ 6 Methods for As a measurement item, 6.2 Approxima tely Alteration JIS omitted the time Although equipment
Q evaluating perfor- measurement method of identical with JIS. base display limit. malfunction can be
o mance characteris- linearity of the time base checked by linearity,
~
:;:d
tics and vertical linearity suggestion will be
(5 6.1 Evaluation of range is specified. made to ISO.
~ performance charac-
~
teristics concerning
time base
6.2 Measurement of As measurement items, 6.3.1, As measurement Alteration JIS deleted the nor- Although equipment
performance charac- vertical linearity, vertical 6.3.2, items, normal axis mal axis display range malfunction can be
teristics concerning linearity range and dis- display limit, verti- in figure 1. checked by linearity,
6.3.3.4
normal axis appearance of echo are cal linearity and suggestion will be
specified. vertical linearity made to ISO.
range are specified. N
~
6.3 Resolution of As measurement meth- 6.1.5, Approximately Alteration Alteration in composi- Alteration made for W
straight beam test- ods, three kinds of meth- identical with JIS. tion of standard. No the purpose of users' 01
6.4, ~
ing technique ods are specified. technical deviation. convenience.
6.4.2 to ~
6.4.6 o
I---l~
001
Nt-:l
(I) Requirements in JIS (II) Inter- (III) Requirements in Inter- (IV) Classification and details of (V) Justification for t-:l0':l
national national Standard technical deviation between JIS the technical deviation W
01
Standard and the International Standard by and future measures t-:l
number clause t-:l
o
No. and title of Content No. of Content Classifi- Detail of technical t--L
o
clause clause cation by deviation
clause
6.4 Resolution of The measurement 6.8 Identical with JIS. Identical -
angle beam testing method using RB-RD and
technique RB-HS type reference
blocks (semicircular col-
umn step is used) is
"'d
specified.
~
o 6.5 Margin of sensi- The measurement 6.5.2, Approxima tely Alteration JIS deleted the speci- Adoption of the ASTM
~
M tivity of straight method of margin of sen- 6.5.3, identical with JIS. fication of ASTM E E 127 test blocks into
o beam testing tech- sitivity using calibration 127 type test block, JIS specification will
~ 6.5.4,
M nique block STB-G is specified. and adopted the alter- be considered after the
U 6.5.6 ative method of ISO. variation of results, or
ttl other matters associ-
~
ated with the said test
o blocks are studied.
o
~ 6.6 Al sensitivity The measurement 6.8 Identical with JIS. Identical -
~
and A2 sensitivity of method of margin of sen-
Q
::r: angle beam testing sitivity using Al and A2
~
technique blocks is specified.
6.7 Dead zone after The measurement 6.7 Identical with JIS. Identical -
transmitter pulse method of dead zone af-
with straight beam ter transmitter pulse
testing technique with straight beam test-
ing technique is speci-
fied.
(I) Requirements in JIS (II) Inter- (III) Requirements in Inter- (IV) Classification and details of (V) Justification for
national national Standard technical deviation between JIS the technical deviation
Standard and the International Standard by and future measures
number clause
No. and title of Content No. of Content Classifi- Detail of technical
clause clause cation by deviation
clause
7 Periodic checks It is specified that the 5 Approximately Alteration JIS gives clear specifi- It is necessary to de-
check frequency and the identical with JIS. cation of default. termine the period of
check items shall be de- time of the default.
termined by the user of Suggestion will be
this Standard. When made to ISO.
"'d there is no flaw detection
:;:d test to be assumed, the
o default (within one year)
~
t:rj
Q is specified.
~
t:rj 8 Record Information to be re- 7 Identical with JIS. Identical -
t:j
corded is specified.
Cd
~
Q
o Overall degree of correspondence between JIS and International Standard (ISO 18175: 2004): MOD
~
:;:d NOTE 1 Symbols in sub-columns of classification by clause in the above table indicate as follows:
(5 Identical: Identical in technical contents.
~
~ Deletion: Deletes the specification item(s) or content(s) of International Standard.
Addition: Adds the specification item(s) or content(s) which are not included in International Standard.
Alteration: Alters the specification content(s) which are included in International Standard.
NOTE 2 Symbol in column of overall degree of correspondence between JIS and International Standard in the above table indicates as follows:
- MOD: Modifies International Standard.
N
~
W
01
~
~
o
I---l~
0-,,:]
Errata for JIS (English edition) are printed in Standardization and Quality Control, published
monthly by the Japanese Standards Association, and also provided to subscribers of JIS
(English edition) in Monthly Information.
Errata will be provided upon request, please contact:
Standards Publishing Department, Japanese Standards Association
4-1-24, Akasaka, Minato-ku, Tokyo, 107-8440 JAPAN
TEL. 03-3583-8002 FAX. 03-3583-0462
PROTECTED BY COPYRIGHT