Professional Documents
Culture Documents
Datasheet Logiq F8 Expert
Datasheet Logiq F8 Expert
LOGIQ F8 Expert
Ultrasound System
Specification Sheet
The LOGIQTM F8 Expert is the multipurpose
ultrasound imaging system designed for
User Interface
Abdominal, Obstetrical, Gynecological, Small Parts, Operator Keyboard
Musculoskeletal, Vascular/Peripheral Vascular, Ergonomic full size keyboard
Urological, Pediatric, Transcranial and Cardiac 8 TGC pods
applications 8.4” (213.4 mm) LED touch screen
Colorize
CINE Memory/Image Memory
M-Mode
Gain: –20 -20 dB, 1 dB step
Dynamic Range: 36 to 96dB, 36-48/78-96 6dB step; 48-78
3dB step.
Gray Scale Map: 6 or 8 types (Probe dependent)
Colorization: 9 types
Continuous Wave Doppler (Option) LOGIQ View (Option)
Gray Scale Map: 8 types Extended Field of View Imaging
Baseline: 11 steps Available on 4C-RS, L6-12-RS, 8C-RS, 3Sc-RS, E8C-RS, E8Cs-
Angle Correct: ±90˚, 1˚ step RS, RAB2-6-RS, L8-18i-RS, 6S-RS, BE9CS-RS probes
Spectral Color: 6 types For use in B-Mode
Invert: On/Off CrossXBeam is available on linear probes
Spectral Averaging: 5 steps Auto detection of scan direction
Gain: 0 – 85 dB, 1 dB steps Post-process zoom
Wall Filter: 5.5 – 5000Hz, 27 steps, dependent on probe and Rotation
application Auto fit on monitor
CW-Mode includes: Measurements in B-Mode
• Transmit Frequency Up to 160cm scan length
• CW Colorization
• Velocity Scale Range: 0.2-6105cm/s Easy 3D (Option)
• Spectrum Inversion Allows unlimited rotation and planar translations
• Trace Method 3D reconstruction from Cine sweep
• Doppler Auto Trace Utilities : Average Off/Average Light/Average Medium/
• Trace Direction Average strong
• Trace Sensitivity GreySurface 0-100%
Threshold1 0-255
Auto Optimization Threshold2 0-255
Optimize B-Mode image to improve contrast resolution ScanDistance 1.0-15.0
Selectable amount of contrast resolution improvement (low, Colorize 0-360
medium, high)
Auto-Spectral Optimize adjusts Advanced 3D (Option)
• Baseline DefineAxis: select two points as start and end point of
• Invert long Axis.
• PRF (on live image) Group Planes: OFF/Main/Parallel/Angular
Reslice:Cube/Virtual Rescan/Cubic Plane
Coded Harmonic Imaging Tile:1/2/4/6
Coded Phase Inversion Harmonic Imaging
Available on all Probes Static 3D/ Realtime 4D (option)
Line Density: 5 steps Available on RAB2-6-RS probe
Line Density Zoom: 5 steps Acquisition Modes: Realtime 4D mode,
Suppression: 6 steps Static 3D mode
Edge Enhance: 7 steps Visualization Modes:
Gray Scale Map: 6 types • 3D Rendering (Diverse surface and intensity
projection modes)
Tint Map: 9 types
• Sectional Planes (3 Section planes perpendicular to
Gain: 0 – 90 dB, 1 dB step each other)
Dynamic Range: 36 to 96dB 3dB step; 36-48/78-96 Render Mode: Surface texture, Surface Smooth, max-,
6dB step; min-, X-ray, Mix Mode of two render Modes
Rejection: 6 step Display Format:
Frequency: Up to 4 steps, probe dependent • Quad: A-/B-/C-Plane/3D
• Dual: A-Plane/3D
• Single: 3D
Curved 3point Render start
Static 3D/ Realtime 4D (Option) Continued... Elastography (Option) Continued...
TUI: Transverse ultrasound imaging Map 8
Display Format: 1*1/1*2/2*2/3*3 FrameAverage 0-10
Slices: 3-19 LineDensity 0-4
Slices Distance: 0.5-40mm Soft Compress 0-10
3D Movie Hard Compress 0-10
LoopSpeed:6-400
Run/Stop TVI (Option)
Scalpel: 3D Cut tool Myocardial Doppler Imaging with color overlay on tissue
ut Mode: Inside Contour/OutSide Contour/Inside Box/
C image
Outside Box/Small Eraser/Big Eraser Available on the sector probes
Cut Depth: Full/UserDefine Tissue color overlay can be removed to show just the 2D
Depth: 0-1000 image, still retaining the tissue velocity information
3D Rotation Cine Q-Analysis: Multiple Time Motion trace display from selected
points in the myocardium
Rotational Angle:30-360
Step Angle1-15
Contrast Agency Detection (Option)
Rotation Axis: X/Y
Available on 4C probes
3D Volume Review
Line Density: 0, 1, 2, 3, 4
Start Frame: define start Frame
SRI HD: 0, 1, 2
End Frame: define end Frame.
Suppression: 0, 1, 2, 3, 4, 5
Run/Stop
Focus Width: 0, 1
Frame Average: 0, 1, 2, 3, 4, 5, 6, 7
4D Volume Cine
Dynamic Range: 36-96DB
LoopSpeed:6-400
Rotation: Up/Down
Run/Stop
Modes: General, Resolution and Penetration; Frequencies
isplay Tissue Image and Contrast-enhanced Image
D
Scan Coach simultaneously in split screen
Scan Coach is a contextual reference tool that provides 2 Contrast Timers
clinical guidance for scan plane acquisition and references
for anatomical structures. Clinical reference images and Timed Updates: 0.05 – 10 seconds
animations depict information related to each step and can Accumulation mode, 8 steps
be displayed on demand by the user. It includes abdominal, Maximum Enhance Mode: On/off
obstetric, gynecologic, carotid and cardiac applications.
Gray Scale Map: 10 types
Colorization: On/off, 6 steps
My Trainer Flash
Abstracted from basic user manual, it lists FAQs from Time Intensity Curve (TIC) Analysis
customers and instructs users how to solve problems on
their own. Auto MI control
Time Trigger Scan: 0.3 & 0.5 – 10 sec., 0.5 sec. step
Elastography (Option) Flash/Burst Mode
Available on L6-12-RS Contrast Tech: AM, HRes, CHA
Semi-Quantification Time Intensity Curve (TIC) Analysis
Frame Reject 0-8 Single view/Dual view
Axial Smoothing 0-4 Visualization: Contrast, Tissue, Hybrid Contrast
Noise Reject 0-8 Hybrid Map: 6 types
Sample Vol 0-2
Lateral Smoothing 0-4
Window 0-8
Stress Echo Controls Available While “Live” Continued...
Advanced and flexible stress-echo examination capabilities Transmission Focus Position
Provides exercise and pharmacological protocol templates Transmission Focus Number
8 default templates Line Density Control
T emplate editor for user configuration of existing templates Sweep Speed for M-Mode
or creation of new templates Number of Angles for CrossXBeam
Reference scan display during acquisition for stress level PW-Mode
Comparison (dual screen) Gain
Baseline level/Previous level selectable Acoustic Output
Raw data continuous capture over 87.5 seconds available Transmission Frequency
Wall motion scoring (bulls-eye and segmental) Scale
Smart stress: Automatically set up various scanning Wall Filter
parameters (for instance geometry, frequency, gain, etc.)
Sample Volume Gate
according to same projection on previous level
Length
Depth
Virtual Convex
Velocity Scale
Provides a convex field of view
Color Flow Mode
Compatible with CrossXBeam
CFM Gain
Available on linear and Sector transducers
CFM Velocity Range
Acoustic Output
SRI-HD
Wall Echo Filter
High Definition Speckle Reduction Imaging Provides multiple
levels of speckle reduction Packet Size
Compatible with Side by Side DualView Display Frame Rate Control
Compatible with all linear, convex and sector transducers CFM Spatial Filter
Compatible with B-Mode, 3D/4D imaging CFM Frame Averaging
Frequency/Velocity Base Line Shift
CrossXBeam
Provides 3, 5, 7, 9 of spatial compounding Controls Available on “Freeze” or Recall
Live Side by Side DualView Display Automatic Optimization
Compatible with: SRI-HD
• Color Mode CrossXBeam – Display non-compounded and compounded
image simultaneously in split screen
• PW
3D reconstruction from a stored Cine loop
• SRI-HD
B/M/CrossXBeam Mode
• Coded Harmonic Imaging
Gray Map Optimization
• Virtual Convex
TGC
Available on 4C-RS, L6-12-RS, E8C-RS, 8C-RS, RAB2-6-RS, L8-
18i-RS, BE9CS-RS, E8CS-RS Colorized B and M
Frame Average (loops only)
Controls Available While “Live” Dynamic Range
Write Zoom Anatomical M-Mode
B/M/CrossXBeam-Mode Sweep Speed
Gain Gray Map
TGC Post Gain
Dynamic Range Baseline shift
Acoustic Output Sweep Speed
Controls Available on “Freeze” or Recall Continued... General Doppler Measurements/Calculations Continued...
Invert Spectral wave form TAMAX (Time Averaged Maximum Velocity)
Compression Volume Flow (TAMEAN and Vessel Area)
Rejection Heart Rate
Colorized Spectrum PI (Pulsatility Index)
Display Format RI (Resistivity Index)
Doppler Audio
Angle Correct Real-Time Doppler Auto Measurements/Calculations
Quick Angle Correct PS (Peak Systole)
Auto Angle Correct ED (End Diastole)
Overall Gain (loops and stills) MD (Minimum Diastole)
Color Map PI (Pulsatility Index)
Transparency Map RI (Resistivity Index)
Frame Averaging (loops only) AT (Acceleration Time)
Flash Suppression ACC (Acceleration)
CFM Display Threshold PS/ED (PS/ED Ratio)
Spectral Invert for Color/Doppler ED/PS (ED/PS Ratio)
Anatomical M-Mode on Cine loop HR (Heart Rate)
TAMAX (Time Averaged Maximum Velocity)
Aortic Valve
Aortic Valve Cusp Separation (AV Cusp)
Aortic Valve Area Planimetry (AVA Planimetry)
(Trans AVA)
Left Atrium Pulmonic Valve
Left Atrium Diameter (LA Diam) Pulmonic Valve Area (PV Planimetry)
LA Length (LA Major) Pulmonic Valve Annulus Diameter (PV Annulus Diam)
LA Width (LA Minor) Pulmonic Diameter (Pulmonic Diam)
L eft Atrium Diameter to AoRoot Diameter Ratio Right Atrium
(LA/Ao Ratio) Right Atrium Diameter, Length (RAD Ma)
Left Atrium Area (LAA(d), LAA(s)) Right Atrium Diameter, Width (RAD Mi)
L eft Atrium Volume, Single Plane, Method of Disk (LAEDV Right Atrium Area (RAA)
A2C, LAESV A2C) (LAEDV A4C, LAESV A4C)
Right Atrium Volume, Single Plane, Method of Disk (RAAd)
ight Atrium Volume, Systolic, Single Plane, Method of
R
Left Ventricle Disk (RAAs)
Left Ventricle Mass (LVPWd, LVPWs)
Left Ventricle Volume, Teichholz/Cubic (LVIDd, LVI Ds) Right Ventricle
Left Ventricle Internal Diameter (LVIDd, LVI Ds) Right Ventricle Outflow Tract Area (RVOT Planimetry)
Left Ventricle Length (LVLd, LVLs) Left Pulmonary Artery Area (LPA Area)
Left Ventricle Outflow Tract Diameter (LVOT Diam) Right Pulmonary Artery Area (RPA Area)
Left Ventricle Posterior Wall Thickness (LVPWd, LVPWs) Right Ventricle Internal Diameter (RVIDd, RVIDs)
Left Ventricle Length (LV Major) Right Ventricle Diameter, Length (RVD Ma)
Left Ventricle Width (LV Minor) Right Ventricle Diameter, Width (RVD Mi)
Left Ventricle Outflow Tract Area (LVOT) Right Ventricle Wall Thickness (RVAWd, RVAWs)
L eft Ventricle Area, Two Chamber/Four Chamber/Short Axis Right Ventricle Outflow Tract Diameter (RVOT Diam)
(LVA (d), LVA (s))
Left Pulmonary Artery (LPA)
Left Ventricle Endocardial Area, Width (LVA (d), LVA(s))
Main Pulmonary Artery (MPA)
Left Ventricle Epicardial Area, Length (LVAepi (d), LVAepi (s))
Right Pulmonary Artery (RPA)
Left Ventricle Mass Index (LVPWd, LVPWs)
Ejection Fraction, Teichholz/Cube (LVIDd, LVIDs)
System Inferior Vena Cava
L eft Ventricle Posterior Wall Fractional Shortening
Systemic Vein Diameter (Systemic Diam)
(LVPWd, LVPWs)
Patent Ductus Arterosis Diameter (PDA Diam)
L eft Ventricle Stroke Index, Teichholz/Cube (LVIDd, LVIDs and
Body Surface Area) Pericard Effusion (PEs)
Left Ventricle Fractional Shortening (LVIDd, LVIDs) Patent Foramen Ovale Diameter (PFO Diam)
Left Ventricle Stroke Volume, Teichholz/Cubic (LVIDd, LVIDs) Ventricular Septal Defect Diameter (VSD Diam)
L eft Ventricle Stroke Index, Single Plane, Two Chamber, Interventricular Septum (IVS) Fractional Shortening
Method of Disk (LVI Dd, LVIDs, LVSd, LVSs) (IVSd, IVSs)
L eft Ventricle Stroke Index, Single Plane, Four Chamber,
Method of Disk (LVI Dd, LVIDs, LVSd, LVSs) Tricuspid Valve
L eft Ventricle Stroke Index, Bi-Plane, Bullet, Method of Disk Tricuspid Valve Area (TV Panimetry)
(LVAd, LVAs)
Tricuspid Valve Annulus Diameter (TV Annulus Diam)
Interventricular Septum (IVS)
Left Ventricle Internal Diameter (LVI D)
Left Ventricle Posterior Wall Thickness (LVPW)
M-Mode Measurements
Mitral Valve Aorta
Mitral Valve Annulus Diameter (MV Ann Diam) Aortic Root Diameter (Ao Root Diam)
E-Point-to-Septum Separation (EPSS) Aortic Valve
Mitral Valve Area Planimetry (MVA Planimetry) Aortic Valve Diameter (AV Diam)
Aortic Valve Cusp Separation (AV Cusp)
Aortic Valve Ejection Time (LVET)
Left Atrium Aortic Valve Continued...
L eft Atrium Diameter to AoRoot Diameter Ratio Aortic Valve Mean Pressure Gradient (AV Trace)
(LA/Ao Ratio) Aortic Valve Peak Pressure Gradient (AR Vmax)
Left Atrium Diameter (LA Diam) Aortic Insufficiency Peak Velocity (AR Vmax)
Left Ventricle Aortic Insufficiency End-Diastolic Velocity (AR end Vmax)
Left Ventricle Volume, Teichholz/Cubic (LVIDd, LVI Ds) Aortic Valve Peak Velocity (AV Vmax)
Left Ventricle Internal Diameter (LVIDd, LVI Ds) Aortic Valve Peak Velocity at Point E (AV Vmax)
Left Ventricle Posterior Wall Thickness (LVPWd, LVPWs) Aorta Proximal Coarctation (Coarc Pre-Duct)
Left Ventricle Ejection Time (LVET) Aorta Distal Coarctation (Coarc Post-Duct)
Left Ventricle Pre-Ejection Period (LVPEP) Aortic Valve Insufficiency Pressure Half Time (AR PHT)
Interventricular Septum (IVS) Aortic Valve Flow Acceleration (AV Trace)
Left Ventricle Internal Diameter (LVI D) Aortic Valve Pressure Half Time (AV Trace)
Left Ventricle Posterior Wall Thickness (LVPW) Aortic Valve Acceleration Time (AV Acc Time)
Aortic Valve Deceleration TIme (AV Dec Trace)
Mitral Valve Aortic Valve Ejection Time (AVET)
E-Point-to-Septum Separation (EPSS) ortic Valve Acceleration to Ejection Time Ratio (AV Acc Time,
A
Mitral Valve Leaflet Separation (D-E Excursion) AVET)
Mitral Valve Anterior Leaflet Excursion (D-E Excursion) Aortic Valve Area according to PHT
Mitral Valve D-E Slope (D-E Slope)
Mitral Valve E-F Slope (E-F Slope) Left Ventricle
L eft Ventricle Outflow Tract Peak Pressure Gradient
Pulmonic Valve (LVOT Vmax)
QRS complex to end of envelope (Q-to-PV close) Left Ventricle Outflow Tract Peak Velocity (LVOT Vmax)
Right Ventricle L eft Ventricle Outflow Tract Mean Pressure Gradient
(LVOT Trace)
Right Ventricle Internal Diameter (RVIDd, RVIDs)
Left Ventricle Outflow Tract Mean Velocity (LVOT Trace)
Right Ventricle Wall Thickness (RVAWd, RVAWs)
L eft Ventricle Outflow Tract Velocity Time Integral
Right Ventricle Outflow Tract Diameter (RVOT Diam) (LVOT Trace)
Right Ventricle Ejection Time (RVET) Left Ventricle Ejection Time (LVET)
Right Ventricle Pre-Ejection Period (RVPEP)
Mitral Valve
System Mitral Valve Regurgitant Flow Acceleration (MV Trace)
Pericard Effusion (PE (d)) Mitral Valve Regurgitant Mean Velocity (MV Trace)
Mitral Regurgitant Mean Pressure Gradient (MR Trace)
Tricuspid Valve Mitral Regurgitant Velocity Time Integral (MR Trace)
QRS complex to end of envelope (Q-to-TV close) Mitral Valve Mean Velocity (MV Trace)
Mitral Valve Velocity Time Integral (MV Trace)
Doppler Mode Measurements Mitral Valve Mean Pressure Gradient (MV Trace)
Aortic Insufficiency Mean Pressure Gradient (AR Trace) Mitral Valve Peak Pressure Gradient (MR Vmax)
Aortic Insufficiency Peak Pressure Gradient (AR end Vmax) Mitral Regurgitant Peak Velocity (MR Vmax)
E8Cs-RS
L6-12-RS
Endo Micro Convex Probe
Linear Probe
Applications: OB/Gyn, Urology, Transvaginal,Transrectal
Applications Vascular, Small Parts, Pediatrics
Number of Element: 128
Number of Element: 128
Convex Radius: 8.73 mm
Footprint: 38.4 x 6.0 mm
FOV: 168°
-Mode Imaging Frequency:
B
6.0, 8.0, 10.0, 11.0 MHz Footprint: 21.1 X 10.55 mm
armonic Imaging Frequency:
H -Mode Imaging Frequency:
B
6.0, 8.0, 10.0, 12.0, 13.0 MHz 6.0,8.0,10.0 MHz
CFM/PDI/PWD Frequency: 4.0, 5.0, 6.0 MHz armonic Imaging Frequency:
H
7.0,8.0,10.0 MHz
Steered Angle : +/-20°°
CFM/PDI/PWD Frequency: 4.0,5.0,6.0 MHz
Biopsy Guide: Multi Angle, Reusable Bracket
Biopsy Guide: Fixed Angle,Disposable, or Reusable Bracket
L8-18i-RS
BE9CS-RS
Linear Probe
Bi-plane, micro convex probe
Applications MSK Superficial
Applications: Urology, Transrectal
Number of Element: 168
Number of Element: 96
Footprint: 34.8 X 11.1 mm
Convex Radius: 9 mm
-Mode Imaging Frequency:
B
8.0,12.0,14.0,16.0 MHz FOV: 133°
armonic Imaging Frequency:
H Footprint: 19 x 18.7 mm
9.0,15,18.0 MHz -Mode Imaging Frequency:
B
CFM/PDI/PWD Frequency: 6.7,8.3,10 MHz 6.0,8.0,10.0 MHz
Steered Angle : +/-15°° armonic Imaging Frequency:
H
8.0,9.0,10.0 MHz
Biopsy Guide: not available
CFM/PDI/PWD Frequency: 4.0,4.2,5.0,6.3 MHz
Biopsy Guide: Fixed Angle,Disposable, or Reusable Bracket
8C-RS RAB2-6-RS
Micro Convex Probe Convex Volume Probe
Applications: Pediatrics Applications: Abdomen, OB/GYN, Urology
Number of Element: 128 Number of Element: 128
Convex Radius: 10.73 mmR Convex Radius : 47.1 mmR
FOV: 131° Footprint: 62.2 x 34.0 mm
Footprint: 22.0 x 12.0mm Volume Sweep Radius: 24.11 mm
B-Mode Imaging Frequency: 6.0, 8.0, 10.0 MHz FOV: 70 (B), 84° x 70° (Volume scan)
armonic Imaging Frequency:
H B-Mode Imaging Frequency: 3.0, 4.0, 5.0 MHz
6.0, 7.0, 8.0, 10.0 MHz Harmonic Imaging Frequency: 4.0, 5.0, 6.0 MHz
CFM/PDI/PWD Frequency: 4.2, 5.0, 6.3 MHz CFM/PDI/PWD Frequency: 2.0, 3.0, 4.0 MHz
Biopsy Guide: Not Available Biopsy Guide: Multi Angle, Reusable Bracket
GE Healthcare
Pollards Wood, Nightingales Ln
Chalfont St Giles
Buckinghamshire HP8 4SP
United Kingdom
Phone:+44 1494 544000
www.gehealthcare.com