MT - 152

12.3 Checking MZ2 and MZ3 [2] Checking Motor Operation in M-Utility

● Checking MZ2 operation


[1] Analysis Flow
(1) Start the M-Utility.
(2) Type in [6][ENT] → [4][ENT] → [4][ENT] → [1][ENT] to execute “HOME POSITION” for “SUB
START READING GRIP DRIVE (MZ2)”.
(3) Make sure that “RESULT: OK” appears and that there is nothing abnormal with driving sound
of the motor.
Actuate the motor in N DRV12A board LED N Reseat the DRV12A/ N
M-Utility. The result lighting status normal? CPU12A board and check
1 (4) Navigate the menu to return to the initial screen of the M-Utility.
normal? if the error occurs again.
Y
Y [3]
[2] Y ● Checking MZ3 operation
1 (1) Start the M-Utility.
(2) Type in [6][ENT] → [4][ENT] → [7][ENT] → [1][ENT] to execute “HOME POSITION” for
N “SCANNER CLEANING (MZ3)”.
END The resistance value of Repair or replace.
the motor normal? (3) Make sure that “RESULT: OK” appears and that there is nothing abnormal with driving sound
Y of the motor.
[4]
(4) Navigate the menu to return to the initial screen of the M-Utility.

N
Cable normal? Repair or replace.
Y
[5]

Repeat DRV12A board


LED lighting status checks
five times to check if the
error recurs.

If the error recurs, replace


the boards in the order named.
1. DRV12A board replacement
2. CPU12A board replacement
3. MTH12A board replacement
FR6H0418.EPS

010-051-06 CR-IR346RU Service Manual [12.3] MT - 152


04.20.2002 FM3386