Uncertainty Assessment For DC Power Supply: SKC Compliance Lab

You might also like

Download as xlsx, pdf, or txt
Download as xlsx, pdf, or txt
You are on page 1of 18

Uncertainty Assessment for DC Power supply

Name of the Laboratory SKC COMPLIANCE LAB

Parameter
Demonstration Capability
Value to be Demonstrate

Environmental Conditions

Nomenclature Make & Eq Id

Master
Equipment
Details Digital Multi meter Fluke & SCL/144

TYPE A EVALUATION OF STANDARD UNCERTAINTY :

Measured Reading in V

Number of Readings (n)


Average of the Readings taken, X1…..X5 =
Standard Deviation, s n-1 =
Degrees of freedom , ( n-1 ) =
Hence, Ua, (s n-1) / sqrt(n) =

TYPE B EVALUATION OF STANDARD UNCERTAINTY :

Ub1:Uncertainty Due to Accuracy of DC Power supply

Demo Value in Accuracy of Reading in Additional Uncert in


V % V
20.00 1.00 2
b1 =
Distribution =
Degrees of freedom =
Hence Ub1, b1/sqrt (3) =

Ub2: Measurement Uncertainty mentioned in Calibration Report of DC Power supply

Demo Value in Uncertainty in


V V
20.00 0.00014
b2 =
Coverage factor ( k ) =
Di+A42:AD50stribution =
Degrees of freedom =
Hence Ub2, b2/k : =
Ub3: Uncertainty due to Resolution error of DC Power supply

Resolution in Uncertainty in
Limit
V V
0.1 1 0.1
b3 =
Distribution =
Degrees of freedom =
Hence Ub3, b3/sqrt(3) =

Uncertainty Assessment for DC Power supply

Name of the Laboratory SKC COMPLIANCE LAB

COMBINED UNCERTAINTY : Uc = √ (Ua2 + Ub12 + Ub22 + Ub32)

Ua Ub1 Ub2

0.06224147 1.2701706 0.00007000

EFFECTIVE DEGREES OF FREEDOM :

Uc4
Veff = SQRT
(Ua)4/n-1 + (Ub1)4/∞ + (Ub2)4/∞ + (Ub3)4/∞
EXPANDED UNCERTAINTY :

Uc =
Coverage factor @ 95.45 % CL k=
Expanded Uncertainty, Ue =

REPORTED RESULT :

The Measured Voltage corresponding to


The Reported Expanded uncertainty of measurement is stated as the combined standard uncertainty in measurement m
coverage factor k=2, which for a normal distribution corresponds to a coverage probability or confidence level of 95.4

UNCERTAINTY BUDGET

Source of Uncertainty Limits

Ub1 : Uncertainty due to accuracy of Power


2.200
Analyzer

Ub2 :Uncertainty from Cal-Report of


0.000
Power Analyzer

Ub3 : Uncertainty due to Resolution of


0.100
Power Analyzer
at
at
Ua : Repeatability Error ….

Uc : Combined Uncertainty ….

Ue : Expanded Uncertainty ….

Prepared By :
Uncertainty Assessment for DC Power supply

Date
SKC COMPLIANCE LAB
Page No

Voltage
Measurement
20.00 V

Temperature : 25°C ± 2°C


Humidity : 40 % to 75 % RH

Traceable to Cal validity

IDEMI(M), CC/ECL/1375/18-19 9/23/2020

NTY :

20.18 20.19 20.00 20.30

5
20.13 V
0.1391761 V
4
Ua 0.0622
0.0622 V

NTY :

Total Uncert in
V
2.20000
2.20000 V
Rectangular
¥
Ub1 1.270
1.270 V

n Report of DC Power supply

0.00014 V
2.0
Normal
¥
Ub2 0.00007
0.00007 V
upply

0.1 V
Rectangular
¥
Ub3 0.0577
0.05773503 V

Uncertainty Assessment for DC Power supply

Date
SKC COMPLIANCE LAB
Page No

Ub12 + Ub22 + Ub32)

Ub3 Combined Uncertainty

0.05773503 Uc 1.273

F FREEDOM :

Uc 4
(Ua)4/n-1 + (Ub1)4/∞ + (Ub2)4/∞ + (Ub3)4/∞
Expanded Uncertainty
1.2730046 V
2.0
2.5460092
2.5460092 V
Ue
12.730

0.4 is 0.4 ± 2.5460


d as the combined standard uncertainty in measurement multiplied by the
onds to a coverage probability or confidence level of 95.45 %

UNCERTAINTY BUDGET

Standard Sensitivity Uncertainty


Probability Distribution
Uncertainty Coefficient Contribution
Type - B
Rectangular 1.270 1 1.270
√3
Type - B
Normal 0.000 1 0.000
k=2
Type - B
Rectangular 0.0577 1 0.0577
√3
Type - A
Normal 0.062 1 0.062
√5

…. …. …. 1.273

k=2 …. …. 2.546

Verified By:
15-Nov-2019
1 of 2

19.97

V
V

15-Nov-2019
2 of 2

ncertainty

836.62399151608
certainty

Degrees of
freedom


Uncertainty Assessment for DC Power supply

Name of the Laboratory SKC COMPLIANCE LAB

Parameter
Demonstration Capability
Value to be Demonstrate

Environmental Conditions

Nomenclature Make & Eq Id

Master
Equipment
Details Digital Multi meter Fluke & SCL/144

TYPE A EVALUATION OF STANDARD UNCERTAINTY :


Measured Reading in A

Number of Readings (n)


Average of the Readings taken, X1…..X5 =
Standard Deviation, s n-1 =
Degrees of freedom , ( n-1 ) =
Hence, Ua, (s n-1) / sqrt(n) =

TYPE B EVALUATION OF STANDARD UNCERTAINTY :


Ub1:Uncertainty Due to Accuracy of DC Power supply

Demo Value in Accuracy of Reading in Additional Uncert in


A % A
2.0 2.00 2
b1 =
Distribution =
Degrees of freedom =
Hence Ub1, b1/sqrt (3) =

Ub2: Measurement Uncertainty mentioned in Calibration Report of DC Power supply

Demo Value in Uncertainty in


A A
2.0 0.0302
b2 =
Coverage factor ( k ) =
Di+A42:AD50stribution =
Degrees of freedom =
Hence Ub2, b2/k : =
Ub3: Uncertainty due to Resolution error of DC Power supply
Resolution in Uncertainty in
Limit
A A
0.1 1 0.1
b3 =
Distribution =
Degrees of freedom =
Hence Ub3, b3/sqrt(3) =

Uncertainty Assessment for DC Power supply

Name of the Laboratory SKC COMPLIANCE LAB

COMBINED UNCERTAINTY : Uc = √ (Ua2 + Ub12 + Ub22 + Ub32)

Ua Ub1 Ub2

0.07071068 1.1777945 0.01510000

EFFECTIVE DEGREES OF FREEDOM :

Uc4
Veff = SQRT
(Ua)4/n-1 + (Ub1)4/∞ + (Ub2)4/∞ + (Ub3)4/∞

EXPANDED UNCERTAINTY :

Uc =
Coverage factor @ 95.45 % CL k=
Expanded Uncertainty, Ue =

REPORTED RESULT :

The Measured Current corresponding to


The Reported Expanded uncertainty of measurement is stated as the combined standard uncertainty in measurement m
coverage factor k=2, which for a normal distribution corresponds to a coverage probability or confidence level of 95.4

UNCERTAINTY BUDGET

Source of Uncertainty Limits

Ub1 : Uncertainty due to accuracy of Power


2.040
Analyzer

Ub2 :Uncertainty from Cal-Report of


0.030
Power Analyzer

Ub3 : Uncertainty due to Resolution of


0.100
Power Analyzer
at
at
Ua : Repeatability Error ….

Uc : Combined Uncertainty ….

Ue : Expanded Uncertainty ….

Prepared By :
Uncertainty Assessment for DC Power supply

Date
SKC COMPLIANCE LAB
Page No

Current
Measurement
2.0 A

Temperature : 25°C ± 2°C


Humidity : 40 % to 75 % RH

Traceable to Cal validity

IDEMI(M), CC/ECL/1375/18-19 9/23/2020

NTY :
2.1 2.2 1.9 1.8

5
2.0 A
0.1581139 A
4
Ua 0.0707
0.0707 A

NTY :

Total Uncert in
A
2.04000
2.04000 A
Rectangular
¥
Ub1 1.178
Ub1 1.178
1.178 A

n Report of DC Power supply

0.03020 A
2.0
Normal
¥
Ub2 0.01510
0.01510 A
upply

0.1 A
Rectangular
¥
Ub3 0.0577
0.05773503 A

Uncertainty Assessment for DC Power supply


Date
SKC COMPLIANCE LAB
Page No

Ub12 + Ub22 + Ub32)

Ub3 Combined Uncertainty

0.05773503 Uc 1.181

F FREEDOM :

Uc 4
(Ua)4/n-1 + (Ub1)4/∞ + (Ub2)4/∞ + (Ub3)4/∞

Expanded Uncertainty
1.1814234 A
2.0
2.3628469
Ue
2.3628469
2.3628469 A
Ue
118.142

0.4 is 0.4 ± 2.3628


d as the combined standard uncertainty in measurement multiplied by the
onds to a coverage probability or confidence level of 95.45 %

UNCERTAINTY BUDGET

Standard Sensitivity Uncertainty


Probability Distribution
Uncertainty Coefficient Contribution
Type - B
Rectangular 1.178 1 1.178
√3
Type - B
Normal 0.015 1 0.015
k=2
Type - B
Rectangular 0.0577 1 0.0577
√3
Type - A
Normal 0.071 1 0.071
√5

…. …. …. 1.181

k=2 …. …. 2.363

Verified By:
15-Nov-2019
1 of 2

2.0

A
A

15-Nov-2019
2 of 2

ncertainty

558.30453733333

certainty

A
A

Degrees of
freedom

You might also like