DELTA OPTICAL
GENETIC PRO
User Manual
1
Contents DELTA OPTICAL GENETIC PRO
contents………………………………………………………………
……………………………………………………………… 1
User Notices…………………………………………………………
………………………………………………………… 2
1 Components Name……………………………………………………
…………………………………………………… 4
2 Installation …………………………………………………………… 5
………………………………
3.1 Adjustment System Installing Illustration 10
2
Use Notices DELTA OPTICAL GENETIC PRO
、Safety note
I
1. Carefully open the box, avoid the accessories, like lens, dropping to ground and being damaged.
2. Do keep the instrument out of direct sunlight, high temperature or humidity, dusty and easy
shaking environment. Make sure the stage is smooth, horizontal and firm enough.
3. When moving the instrument, please use two hands to grip with the two sides of the microscope
body.
4. When running, the lamp house and nearby parts will be very hot. Please ensure there is enough
cooling room for them.
5. Make sure the instrument is earthed, to avoid lighting strike.
6. For safety, be sure the main switch is in “O”(off) state before replace the halogen (LED) lamp or
the fuse, then cut off the power, and do the operation after the lamp bulb and the lamp house
completely cool.( specified lamp: Halogen Lamp 6V/20W or LED lamp 3W)
7. Check the input voltage: be sure the input voltage which signed in the back of the microscope is
consistent with the power supply voltage, or it will bring a serious damage to the instrument.
8. Use the factory supplied power cord, please.
II 、Maintenance
1. All the lenses have been well checked and adjusted. It is forbidden to disassemble them yourself.
2. The nosepiece and coarse/fine focus unit have a compact and precise frame, please don’t
disassemble them as possible as you can.
3. Keep the instrument clean, wipe dust regularly, and be attention to avoid contaminating the
optical elements especially.
4. The contaminations on the prism, as finger mark and oil, could be gently wiped with a piece of soft
cloth or tissue paper, gauze which has been immersed in pure alcohol or xylene. (note that the
alcohol and the xylene are all burned easily, do not let them near the fire, and use them in a
drafty room as possible as you can.)
5. Don’t use organic solvent to wipe the non-optical elements, when you need to clean, use the soft
detergent, please.
6. When using, if the microscope is splash by liquid, cut off the power at once, and wipe up the
moisture.
7. Do not disassemble any parts of the microscope. That will affect the function or decline the
performance of the microscope.
3
8. Place the instrument in a cool, dry position. After using the microscope, remember to cover it
with dust helmet. Do wait for the lamp house cooling completely before cover.
4
1. Name of Components DELTA OPTICAL GENETIC PRO
Eyepiece Binocular
Nosepiece
Objective
condenser
Collector lens
Body
5
2.
.Installation DELTA OPTICAL GENETIC PRO
2.1 Installing Illustration
The following shows the installing order of the spare parts. The number expresses the installing process.
Before the installing, make sure there is no dust and stain on the spare parts. No marks by
outside force on the surface of the spare parts or glass.
Keeping the hexagon- spanner well, when change the spare parts, you will use it.
④
物镜
Objective
电源线
Cable Wire
显微镜体
Body
6
DELTA OPTICAL GENETIC PRO
2.2 Installing Setps
①.
Fig.1
Fig.2 Intention:
microscope:
Fig.4
7
DELTA OPTICAL GENETIC PRO
in operation.
Fig.6
②
1.Rotating the condenser holder① out as the
Fig.7 shows.
①
2.Put the filter② into the condenser holder,
8
DELTA OPTICAL GENETIC PRO
③ ②
★ The voltage of the machine can be
selected between 110V and 220V. (when
shipped from the factory, the switch is set
Fig.9
to 110V and 220V)
9
DELTA OPTICAL GENETIC PRO
Fig.14
Fig.15
11
DELTA OPTICAL GENETIC PRO
3.2 Operation
3.2.1 Adjusting the brightness (Fig.16)
75mm.
14
4.
.Technical Specifications DELTA OPTICAL GENETIC PRO
一、 Main specifications
Coaxial Coarse and Fine Focusing System, Sensitivity and Graduation of Fine
Focus System
Focus: 0.004mm, range 24mm
Stage Double layer mechanical stage, area: 132×142mm, movement range: 75×40mm
二、Eyepiece、Objectives
1. Objectives
Numerical Value
Cover Glass Working Working
Magnification Aperture Focus f(mm)
Thickness (mm) Distance(mm) Mode
Diaphragm(N.A)
4× 0.10 0.17 31.05 18 dry
10× 0.25 0.17 17.13 6.5 dry
40× 0.65 0.17 4.65 0.53 dry
100× 1.25 0.17 2.906 0.13 oil
2. Eyepiece
Kind Magnification Focus f(mm) Linear Field Of View(mm)
15
5.
.Standard Outfit table DELTA OPTICAL GENETIC PRO
Frame 1 〇
Condenser holder 1 〇
Achromatic 4× 1 〇
Achromatic 10× 1 〇
Objective
Achromatic 40× 1 〇
16
5.
.Trouble shooting DELTA OPTICAL GENETIC PRO
I、 Optical Part:
Adjust it into the located
The nosepiece is not in the located
position( turning the objective to let it
1、 The edge of the position( The objective is not in the
in the center of the light path
field of view has center of the light path )
correctly)
shadow or the
The filament shadow not in center Adjust it to center
brightness is
asymmetry The surface of the lens has
contaminant (condenser、objective、 Clean the lens
eyepiece、 Collector lens)
The surface of the lens has
contaminant (condenser、objective、 Clean the lens
2、 Find dust and eyepiece、 Collector lens)
stain in the field of There are stains on the slide Clean the lens
view Loosen the bolt of the condenser,
The position of condenser is too low adjust its position and tighten it
again
No cover glass on the specimen Add cover glass
The cover glass is too thick or too Use normal thickness cover
thin glass(0.17mm)
The specimen is on the reverse side Turn it around
Oil the dry bjective(especially easy
Clean the objective
to happen on 40X)
The surface of the lens has
contaminant (condenser、objective、 Clean the lens
eyepiece、 Collector lens)
3、 Bad image
quality No oil with oil objective Use oil
(low-resolution, bad
There are air bubbles in the oil Eliminate the bubbles
Contrast)
Use the unspecified oil Use the specified oil
The opening of Aperture diaphragm
turn it down to the proper size
is too large
There are stains on the incidence
Clean the lens
lens of the binocular head
The opening of Aperture diaphragm
Opening it to the proper size
is too small
17
The condenser is not in the center of Reset the condenser and adjust the
the field or the condenser incline Central bolt of the condenser carefully
4、 The image one The nosepiece is not in the located Rotate the nosepiece to the required
side is clear and the position position
other side is faint
The specimen is in the floating
Fix it firmly
state
II、Mechanical Part:
18
III、Electric Part:
Check the power cord, and connect
No power supply
them exactly
1、The lamp can’t
light The installation of the bulb is wrong Install the bulb correctly
19