FUNCTIONAL TEST RECORD SHEET: 1010X (STERRAD 100S)

PAGE 1 OF 2

MACHINE SERIAL # DATE TESTED

AUTOTEST SOFTWARE:

INJECTOR PUMP SERIAL #

CONTROL ENCLOSURE SERIAL #

LFPS2 ENCLOSURE SERIAL #

SECTION FEATURE REQUIREMENTS ACTUALS SIGNATURE

START 60 psi ± 3 psi psi


12.1.10 COMPRESSOR
SHUT-OFF 80 psi ± 3 psi psi

12.1.11.3 DOOR LOCK STAYS CLOSED YES

12.1.16.2 CONN. DIST. PWA-Point #3 (PS1) +5.20 VDC ± 0.02 VDC

CONN. DIST. PWA-Point #4 (PS2) +12.85 VDC ± 0.05 VDC

CONN. DIST. PWA-Point #2 (PS1) -12.0 VDC (REF) VDC

12.1.16.3 +12V AT BARATRON FROM PS1 +12.0 VDC (REF) VDC

12.1.16.4 ADJUST A/D SETPOINT BETWN PINS 2.048 VDC ± 0.001 VDC
TP1 (+) & TP2 (-).

12.3.5.12 VAPORIZER HEATER VOLTAGE 20-24 VAC (Loaded)


VAC

12.3.5.18 CASSETTE DOOR VOLTAGE +12.0 VDC ± 1 VDC VDC

12.3.6.4 CHAMBER HTR VOLTAGE (Nom) 208 VAC  10%


VAC

12.3.6.4 DOOR HEATER VOLTAGE (Nom.) 208 VAC  10%


VAC

12.3.7.2 DOOR REVERSAL FORCE 14 Kg30 lbs)


Ibs

12.3.8.3 CHAN O PRESSURE INPUT 4.000 (nom)


VDC

12.3.8.15 #.### V, HIGH PRESSURE ½ OF CHAN 0

CO-07507-7 FORM 15-01578-001 REV L,


(3/07)

Proprietary Information: This document and the contents hereof are considered proprietary and confidential information of ADVANCED
STERILIZATION PRODUCTS a division of Johnson & Johnson Medical, Inc. (ASP) and disclosure to unauthorized individuals or dissemination, publication,
or copying is prohibited without prior written consent by ASP. *Trademark © Johnson & Johnson Medical, Inc. 1995
FUNCTIONAL TEST RECORD SHEET: 1010X (STERRAD 100S)

VDC

12.3.11.3 FANS OPERATING PROPERLY? YES / NO YES

12.3.9.1 LFPS2 DELIVERED (LCD) LFPS2 DELIVERED SIGNATURE


RF TEST (3040 mVDC Min.) (PWR. VER. MTR) (2.89
AIR VDC to 3.52 VDC)

AFTER 5
MIN. mVDC VDC

12.3.9.1 LEAK BACK RATE #1 < 25 mTorr/min mTorr

12.3.9.2 LFPS2 DELIVERED (LCD) LFPS2 DELIVERED


RF TEST (3040 mVDC Min.) (PWR. VER. MTR) (2.89
PEROXIDE VDC to 3.52 VDC)

AFTER 5
MIN. mVDC VDC

12.3.9.2 LEAK BACK RATE #2 < 25 mTorr/min mTorr

TP-35104- PNEUMATIC INTERGRITY TEST, < 2 PSI DECAY/5 min PSI


001 SECTION 13

THIS TEST DATA IS GENERATED AT FINAL TEST, NO CHANGES, ADJUSTMENTS OR MODIFICATION HAVE
BEEN MADE BETWEEN THE DATE THIS DATA WAS GENERATED AND THE DATE OF THIS SUBMISSION TO
Q.A., EXCEPT AS FULLY NOTED BELOW:

BY: DATE:

BY: DATE:

BY: DATE:

CO-07507-7 FORM 15-01578-001 REV L,