Point of ● Run CPU ● DI120~150

TCD 2,3 Unidentified Alarm Occurrence / Run CPU Address Error Application
Detection ○ Speed CPU ● DI180~420

[Meaning] ■TCD 2: When un identified codes are read [Symptoms]


by ALARM Interrupt(IR1 of 82C59A) 3 or more consecutive times. 1.Check ALARM LED(LED3) of DOC-200 PCB flickering status.
[When an alarm interrupt occurs without an address error.] ● Normal →OFF
● Abnormal(Address ERROR) →ON
■TCD 3: When unidentified address by address map of
DOC-200 PCB access is detected 3 or more consecutive times. [Actions to take]
1. Check +5[V] voltage.
Must be at least +4.8[V].
[Diagram] 1) +5[V] : Connector MG/1, MG/6
2) GND : TP16(Located just above Connector MG)

2. As to TCD 2, GAL(U30), 82C59A(U26), or EPLD(U11) of


DOC-200 PCB may be failed.
: Replace GAL first. If TCD 2 error still exists, replace EPLD.
* Be careful not to damage socket during EPLD
replacement.(Be careful for electrostatics.)

3. As to TCD 3, DOC-200 PCB FLASH MEMORY(U12, U13) or


GAL inside program of the FLASH MEMORY may be failed.
Replace FLASH MEMORY.
Alarm Interrupt * Be careful not to damage or bend pins while replacing
read_alarm_code() FLASH MEMORY.(Be careful for electrostatics.)
4. If TCD 2 or 3 error exists even after above actions
SIS_ADDR_ERROR ERR_ADDR(TCD 3)
have been taken, replace DOC-200 PCB.
system_down()

Count >= 3
INC
Count

usr_reboot()

TCD 2

Undefined_Alarm_Count >3
TCD 3

INC
Undefined_Alarm_Count