System Not
da Vinci Standard: Press SYSTEM Button Call 1-800-876-1310
da Vinci S: Press POWER Button
Powering On
(Int: +41.21.821.20.20)
da Vinci Si: Press any System POWER Button
For Assistance
System Powers On
System Yes
Successfully Completes Continue OR Set-up
Homing
No
Call 1-800-876-1310
(Int: +41.21.821.20.20)
For Assistance
Audio Audio
Functioning Functioning
Check Audio
Continue OR Set-up Ensure Microphone is On and Continue OR Set-up
Step One Power On Vision Cart and
and Homing Check Volume on Touchscreen and Homing
Volume on Surgeon Console
and Touchpad
and Vision Cart
Audio
Functioning
Check Audio on
Vision Cart and Surgeon Continue OR Set-up
Console. Adjust Volume if and Homing
Needed
Call 1-800-876-1310
(Int: +41.21.821.20.20)
For Assistance
Patient Cart
Powers On
Check AC Power Connection
Step One Plug into New Outlet
for Patient Cart
Patient Cart
Press Patient Cart Powers on Continue OR Set-up,
Step Five Power Button Start-up, and Homing
Call 1-800-876-1310
(Int: +41.21.821.20.20)
For Assistance
No
da Vinci Standard:
Replace Sterile Adapter
Step Two
da Vinci S and da Vinci Si:
Replace Drape with
Integrated Sterile Adapter
No
Call 1-800-876-1310
(Int: +41.21.821.20.20)
For Assistance
EndoWrist Instrument
Does Not Engage
Instrument
Engages
Remove and Reseat Instrument Arm
Step One Sterile Adapter and Continue
Endowrist Instrument
Instrument Does
Not Engage
Instrument
Engages
Remove EndoWrist Instrument
Step Two and Reinsert ANOTHER EndoWrist Continue
Instrument on the
Instrument Arm
Instrument Does
Not Engage
da Vinci Standard:
Replace Sterile Adapter Instrument
and Reinsert Instrument Engages
Instrument Does
Not Engage
Call 1-800-876-1310
(Int: +41.21.821.20.20)
For Assistance
Call 1-800-876-1310
(Int: +41.21.821.20.20)
For Assistance
Instrument Responds
Does the Generator to Foot Pedal
Activate When the Foot Continue
Pedal is pressed?
Instrument Responds
to Foot Pedal
Power On
Step Four Generator
Continue
Instrument Responds
1. Power Off to External Foot Pedal
Generator
Call 1-800-876-1310
(Int: +41.21.821.20.20)
For Assistance
For technical information, including full cautions and warnings on using the da Vinci System, please refer to the System User Manual. Read all instruc-
tions carefully. Failure to properly follow instructions, notes, cautions, warnings and danger messages associated with this equipment may lead to
serious injury or complications for the patient.
While clinical studies support the use of the da Vinci Surgical System as an effective tool for minimally invasive surgery, outcomes cannot be guaran-
teed, as surgery is patient and procedure specific.
© 2013 Intuitive Surgical. All rights reserved. Intuitive®, Intuitive Surgical®, da Vinci®, da Vinci® S™, InSite®, TilePro™ and EndoWrist® are trademarks or registered
trademarks of Intuitive Surgical. All other product names are trademarks or registered trademarks of their respective holders. PN 550866-03 Rev. B 12/2013
Troubleshooting Flow Chart
for Improving Vision during
Missing Image (One or Both Eyes)
Check Default Settings On CCUs Note: Please refer to the “Default Camera Control Unit Settings” tab for
instructions on restoring CCUs to their default settings (da Vinci Standard
and da Vinci S ONLY). Refer to Restore Factory Settings for da Vinci Si
Re-do Black Balance and/or
default video settings.
White Balance
(Refer Flowchart for Black or Note: Black Balance is for da Vinci Standard and da Vinci S only.
White Balance)
All Systems:
Check Image Good
Quality in the Continue
Surgeon Console
Bad
Ensure Camera Cable Connections are Tightly
Step Three connected to CCU and Camera Head
Yes
Replace Cable
Bad
Change Camera Head
Call 1.800.876.1310
(Intl: +41.21.821.20.20)
For Camera Head Repair
or Replacement
Note: Please refer to the “Default Camera Control Unit Settings” tab for
Check Default Settings on CCUs instructions on restoring CCUs to their default settings (da Vinci Standard
and da Vinci S ONLY). Refer to Restore Factory Settings for da Vinci Si
default video settings.
Check Image
Good
Quality in the
Continue
Surgeon Console
Dark Image
or Poor Color Call 1.800.876.1310
For Assistance Changing (Int: +41.21.821.20.20)
da Vinci Standard and da Vinci S:
Replace Light Guide Cable the Camera Head or Cable For Assistance
Step Three da Vinci Si: Replace Camera Head
or Camera Cables Dark Image or Poor Color
Good
Check Image Quality in the Continue
Surgeon Console
Flickering Image
Check and reseat Vision Cable Connections Including 1. Turn off Surgeon Console 2. Reseat Vision
Step Two Camera Cable at CCU Cables 3. Turn On Surgeon Console
Step Two
Flickering Image
1. Power On Surgeon Console and for da Vinci Standard
Step Three and da Vinci S, the Vision Cart Good
2. da Vinci Standard and da Vinci S: Ensure CCU’s are on Check Image Quality in the Continue
Surgeon Console
No No Call 1.800.876.1310
Does Video Noise Happen only Flickering Moved to other eye?
Step Four When Cautery is Applied?
(Int: +41.21.821.20.20)
For Assistance
Yes Yes
Remove ESU from Vision Cart Replace Camera Cable Step Four
Good Good
Check Image Quality in the Check Image Quality in the Continue
Continue Surgeon Console
Surgeon Console
Flickering Image 4
Blurry Image
Blurry Image
Blurry Image
Call 1.800.876.1310
(Int: +41.21.821.20.20)
For Assistance
Blurry Image 5
Black Balance (Panasonic SD Vision Systems ONLY)
Step One Set CCUs to DEFAULT Settings Note: Please refer to the “Default Camera Control Unit
Settings” tab for instructions on restoring CCUs to their
default settings (da Vinci Standard and da Vinci S ONLY).
Set Illuminator to 0%
Yes
Cover Endoscope Tip Completely
Check Image
Bad
Step Three Quality on
Surgeon Console
Good
Step One
Set CCUs to DEFAULT Setting Note: Please refer to the “Default Set Illuminator to 100%
Camera Control Unit Settings” tab for
instructions on restoring CCUs to their
default settings (da Vinci Standard
Set Illuminator to 100% and da Vinci S ONLY). Refer to Restore Restore Factory Settings
Factory Settings for da Vinci Si default
video settings.
Place Sterile White Paper 4” (10 cm) from Endoscope Tip
Ensure Image on Vision Cart Monitor is Completely White Place Sterile White Paper 4” (10 cm) from Endoscope Tip
Good Second Fail: Place Sterile White Paper 4” (10 cm) from Endoscope Tip
LEDs Blink
Continue by Performing 3D Calibration and Stay Lit
Pass
Redo White Balance
Fail
White Balance 7
Unable to Focus
Unable to Focus
Unable
Able to Focus
to Focus
Unable to Focus
Check Image
Call 1.800.876.1310
Step Five Quality in the
(Intl: +41.21.821.20.20)
Surgeon console Unable to Focus for Assistance
Able to Focus
Continue
Unable to Focus 8
Default CCU Settings (da Vinci Standard and da Vinci S ONLY) WHITE
BALANCE GAIN
Panasonic® SD DC POWER GAIN SELECTOR KNOBS SCENE ELC LEVEL
Leave Power On
ON
OFF
GAIN
DC POWER PAGE ITEM
L
GAIN GAIN PANASONIC PANASONIC
DC POWER PAGE ITEM DC POWER PAGE ITEM
HIGH ATW
KSJDBCK KSJDBCK
HIGH
LOW
ATW
AWC L GP-US742
HIGH
LOW
ATW
AWC R GP-US742
OFF MANU OFF MANU
LOW AWC
ON ON
GP-US742
(FREEZE) (AWC) (ABC) (E-ZOOM) (FREEZE) (AWC) (ABC) (E-ZOOM)
CAM R B 1 ON CAM R B 1 ON
OFF OFF
OFF MANU
BAR BAR
ON
2 OFF 2 OFF
NTSC
GAIN SCENE ELC LEVEL CAMERA NTSC
GAIN SCENE ELC LEVEL CAMERA
(FREEZE) (AWC) (ABC) (E-ZOOM)
CAM R B 1 ON
L R
L R
OFF
BAR 2 OFF
GAIN SCENE ELC
FOCUS FOCUS
NTSC LEVEL CAMERA
IN OUT
DC POWER 1
SCENE
2 3 LOW
GAIN
HIGH
FREEZE
MENU SEL GP-US942
DC POWER 1
SCENE
2 3 LOW
GAIN
HIGH
FREEZE
MENU SEL GP-US942
DC POWER 1 2 3 LOW HIGH
ON ON
FREEZE
GP-US942
BRIGHTNESS BRIGHTNESS
OFF OFF
SCENE GAIN
SEL
EXPAND EXPAND
BAR
AWC
D RANGE
L BAR
AWC
D RANGE
R MENU
CAMERA CAMERA
ON
BRIGHTNESS
OFF EXPAND
BAR
AWC
D RANGE
L
FOCUS FOCUS
IN OUT
• SCENE: SCENE 1
• D-RANGE: OFF
• GAIN: OFF
Ikegami® HD AUTO
IRIS PAINT SENSITIVITY
SHUTTER
Leave Power On
Ikegami
L IRIS VIDEO LEVELS
WARN SCENE FILE
180º
AUTO
SHUTTER 1 2 3 4
Ikegami
L IRIS
POWER
WARN
MENU R PAINT R SENSITIVITY
POWER
L
MENU
POWER
MENU
FOCUS FOCUS
IN OUT
2
2
2
L
L
2
R
L R
R L
L
2
2
Default Camera Control Unit Settings (da Vinci Standard and da Vinci S ONLY) 9
Restore Factory Settings (da Vinci Si only)
For technical information, including full cautions and warnings on using the da Vinci System, please refer to the System User Manual. Read all instruc-
tions carefully. Failure to properly follow instructions, notes, cautions, warnings and danger messages associated with this equipment may lead to
serious injury or complications for the patient.
While clinical studies support the use of the da Vinci Surgical System as an effective tool for minimally invasive surgery, outcomes cannot be guaran-
teed, as surgery is patient and procedure specific.
© 2013 Intuitive Surgical. All rights reserved. Intuitive®, Intuitive Surgical®, da Vinci®, da Vinci® S™, da Vinci® Si™, InSite®, TilePro™ and EndoWrist® are trademarks or
registered trademarks of Intuitive Surgical. All other product names are trademarks or registered trademarks of their respective holders.
PN 550866-03 Rev. B 12/2013