HARMONIC ACE® Curved Shears for da Vinci® Systems

Assembly
Note: Read and follow the Instructions for Use prior to assembly and use.

Step 1 Step 2 Step 3


The hand piece and Insert screw together. Turn the Insert Slide the torque wrench over the closed jaws and onto the Remove the torque wrench. Slide the Insert with jaws closed
clockwise until it is finger tight. wrench flats. Hold only the hand piece and turn the wrench through the opening in the housing.
clockwise to tighten until you hear one “click,” which indicates
sufficient torque has been applied to secure the Insert.

Step 4


For 8 mm instruments: (a) Hand tighten the gray shaft, while holding the instrument housing and hand piece together until finger-
tight. (b) Slide the torque wrench over the jaws onto the wrench flats and turn the wrench in a clockwise direction until you hear one “click.”
See Side 2 for
For 5 mm instruments: Slide the torque wrench over the jaws onto the wrench flats. While firmly pushing the wrench and hand piece
together, turn the wrench in a clockwise direction until you hear one “click.”
Generator Self Test,
Note: If you do not hear the click after several turns of the wrench, pull the Insert partially out of the housing, rotate the Insert 180 degrees, and re-engage into the housing. HARMONIC ACE Components,
8 mm Instrument (a) 8 mm (b) 5 mm Instrument Product Comparison
and Contact Information

© 2010 Intuitive Surgical. All rights reserved. Intuitive®, Intuitive Surgical®, da Vinci®,
da Vinci® S™, da Vinci® Si™, InSite®, TilePro™, Single-Site™, and EndoWrist® are trade-
marks or registered trademarks of Intuitive Surgical. All other product names are
trademarks or registered trademarks of their respective holders.
Note: The housing and the hand Ethicon Endo-Surgery DSL #: 10-1438
piece must be held together to Note: The housing and the hand piece must be held to-
minimize the gap. gether to minimize the gap.

SIDE 1
Slide the wrench off the Insert, over the closed jaws, and remove it completely. Confirm that the Insert is securely attached to the housing by Customer Service: 1.800.876.1310
trying to slide the Insert out while holding the housing. www.intuitivesurgical.com
PN 550999-02 Rev. A
Generator Self-Test
Note: Do not turn the generator power on before the hand piece and instrument are connected to the generator. Read and follow the Instructions
for Use prior to assembly and use.

Step 1
➥ See Side 1 for
Assembly Instructions
SIDE 2
Plug the hand piece into the appropriate generator port. Turn on the generator by pressing the power button on HARMONIC ACE® Curved Shears Components
the front panel.

Step 2 5 mm Reusable 8 mm Reusable HARMONIC ACE® Curved


Instrument Housing Instrument Housing Shears Disposable Insert
When assembled, the jaws of the device close automatically. The jaws must be open during the generator self-test. (PN – 400274 / 420274) (PN – 400275 / 420275) (PN – 400272)
An error may occur if the jaws are closed during the generator self-test.
To open the jaws: Hold the instrument housing with one hand, and with the other hand push only the
plastic ring of the Insert toward the housing without touching the hand piece. Holding the hand piece while
pushing the Insert can cause the self-test to fail. Compatible Ethicon Endo-Surgery Components
With the jaws open and away from any obstruction, activate the HARMONIC System by pressing the foot pedal.
“TEST IN PROGRESS” will appear on the generator’s graphic display, and audible tones will indicate the test is in
progress. If no errors are identified, the device is ready for use.
HARMONIC Generator 300 HARMONIC Footswitch HARMONIC Torque Wrench HARMONIC Hand Piece
Product Code - GEN04 Product Code - FSW01 N/A Product Code - HP054

HARMONIC vs. HARMONIC ACE®


Note: HARMONIC Curved Shears components are not compatible or interchange-
able with the new HARMONIC ACE® Curved Shears components.

HARMONIC Curved Shears HARMONIC ACE® Curved Shears


Brand HARMONIC HARMONIC ACE®
Name Curved Shears Curved Shears
Insert Color Black Gray
Branding on INTUITIVE® HARMONIC ACE® (Side 1)
Insert (on both sides) INTUITIVE® (Side 2)
Contact Information Torque Wrench TLB01 1 Gray Torque Wrench 2
If you experience problems with a disposable HARMONIC De- If you experience problems with the reusable HARMONIC Hand piece, 1 – Not included with disposable insert (PN 400169)
vice that interfaces with the da Vinci® system, please contact: generator or footswitch, please contact: 2 – Included with disposable insert (PN 400272)
Intuitive Surgical Product Support Ethicon Endo-Surgery Customer Support Center
1.800.876.1310 1.800.USE.ENDO (1.800.873.3636)