Candela Gentlelase Mini Error 5.2 - Syneron-Candela GentleLase Mini - MedWrench
Candela Gentlelase Mini Error 5.2 - Syneron-Candela GentleLase Mini - MedWrench
Forums (https://
www.medwrench.com/forums/ Symptoms
equipment/10070/syneron-
- Machine stops working during this error
candela-gentlelase-mini )
- Screen backlight doesn't work.
Documents
Videos
ChristineMae
News Reply from ChristineMae Mon Dec 19 2016
Hey,
Check that your CPU Board is receiving the signal from your
fiber optic cable. for your EOA issue,
Christine Ricoy
C.L.R. T.
Report Post
avatar placemark
Reply from LaserTech Wed Dec 28 2016
I don't think your refering to the right machine this doesnt have
LaserTech a lvps just a hvps. Thanks for the response but does someone
have a less generic answer?
Report Post
ChristineMae
Reply from ChristineMae Thu Dec 29 2016
Also if you have replaced the HVPS and still have this problem I
would look at the CPU or the connections from the HVPS to the
CPU.
If you have the time you may call our office and 3219843177
and I can try to assist you in troublshooting the unit.
Christine Ricoy
C.L.R. T.
Report Post
alsurabi
Reply from alsurabi Thu Sep 19 2019
Report Post