Aqueous Extraction
Product No. 8288
The Raptor reader allows the user to simply add sample and walk
away - the instrument controls the volume, temperature and
timing of your assay. The system allows you to run three samples
at one time.
T2
/H
T2
T2
/HT
2
© Neogen Corporation, 2018. Neogen, Raptor and Reveal are registered trademarks of Neogen Corporation, Lansing, Michigan, USA. FD1225-0818
Reveal® Q+ MAX for T2/HT2
Aqueous Extraction
Note: Please read kit instructions completely before performing test.
1. Prepare by entering the QR code 2. Obtain a representative sample. 3. Add contents of 1 MAX 1 packet to 4. Add 50 mL distilled or deionized
into the Raptor® reader Grind and weigh out a 10 gram the sample water to the sample
sample
5. Shake vigorously for 3 minutes, or 6. Allow to settle, 7. Add 100 µl sample extract to the red 8. Add 1500 µl of sample diluent to
blend for 1 minute then filter collecting dilution cup and mix up and down 5 the red dilution cup and mix by
a minimum 3 mL filtrate times pipetting up and down 5 times
9. Insert strip into the Raptor 10. Insert cartidge into the Raptor 11. Add 400 µL sample from the red 12. Results shown on the Raptor reader
cartridge reader dilution cup to the back of the *Samples greater than 50 ppb must
Raptor cartridge. Reader will start be diluted & re-run.
automatically